Abstract

Purpose

In this study, we aimed to report our preliminary experience regarding the impact of the novel short-tip (ST) cryoballoon (CB) on procedural efficacy and signal quality during pulmonary vein (PV) isolation for both paroxysmal and persistent atrial fibrillation.

Methods

Between March, 2015, and August, 2015, we enrolled a total of 64 patients (47 patients male, 73 %) with a mean age of 60 ± 11 years. In the study population, 31 patients (48 %) underwent PVI using Advance (ADV) CB and 33 (52 %) patients with ST CB. In all patients, a 28-mm balloon was used.

Results

Acute procedural success rates were 100 % for the entire study population. A statistically insignificant increase in the percentage of PV signal recordings was observed with ST CB in all PVs compared to ADV CB [88 vs. 81 % for left superior PV (LSPV), 82 vs. 78 % for left inferior PV (LIPV), 85 vs. 84 % for right superior PV (RSPV), 82 vs. 71 % for right inferior PV (RIPV), p < 0.05]. Additionally, the difference in minimum temperature reached during the procedure per PV was not statistically significant between ST CB and ADV CB except the LIPV (LSPV −44.2 ± 5.9 vs. −45.6 ± 5.3 °C, p = 0.970; LIPV −38.7 ± 4.6 °C vs −44.6 ± 6.8 °C, p < 0.001; RSPV −45.6 ± 7.4 °C vs.−47.2 ± 6.1 °C, p = 0.168; RIPV −41.4 ± 5.1 °C vs.−43.7 ± 6.3 °C, p = 0.360). Time to isolation for each PV was similar between ST and ADV CB (p > 0.05).

Conclusion

Our preliminary findings indicated similar acute procedural success for the novel ST CB compared with second generation ADV CB. The increase in the ratio of PV signal recordings obtained during the PVI using the novel ST CB was not significant.
Literature
1.
go back to reference Metzner A, Reissmann B, Rausch P, Mathew S, Wohlmuth P, Tilz R, et al. One-year clinical outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon. Circ Arrhythm Electrophysiol. 2014;7:288–92.CrossRefPubMed Metzner A, Reissmann B, Rausch P, Mathew S, Wohlmuth P, Tilz R, et al. One-year clinical outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon. Circ Arrhythm Electrophysiol. 2014;7:288–92.CrossRefPubMed
2.
go back to reference Aytemir K, Gurses KM, Yalcin MU, Kocyigit D, Dural M, Evranos B, et al. Safety and efficacy outcomes in patients undergoing pulmonary vein isolation with second-generation cryoballoondagger. Europace. 2015;17:379–87.CrossRefPubMed Aytemir K, Gurses KM, Yalcin MU, Kocyigit D, Dural M, Evranos B, et al. Safety and efficacy outcomes in patients undergoing pulmonary vein isolation with second-generation cryoballoondagger. Europace. 2015;17:379–87.CrossRefPubMed
3.
go back to reference Ciconte G, Ottaviano L, de Asmundis C, Baltogiannis G, Conte G, Sieira J, et al. Pulmonary vein isolation as index procedure for persistent atrial fibrillation: one-year clinical outcome after ablation using the second-generation cryoballoon. Heart Rhythm. 2015;12:60–6.CrossRefPubMed Ciconte G, Ottaviano L, de Asmundis C, Baltogiannis G, Conte G, Sieira J, et al. Pulmonary vein isolation as index procedure for persistent atrial fibrillation: one-year clinical outcome after ablation using the second-generation cryoballoon. Heart Rhythm. 2015;12:60–6.CrossRefPubMed
4.
go back to reference Koektuerk B, Yorgun H, Hengeoz O, Turan CH, Dahmen A, Yang A, et al. Cryoballoon ablation for pulmonary vein isolation in patients with persistent atrial fibrillation: one-year outcome using second generation cryoballoon. Circ Arrhythm Electrophysiol. 2015;8:1073–9.CrossRefPubMed Koektuerk B, Yorgun H, Hengeoz O, Turan CH, Dahmen A, Yang A, et al. Cryoballoon ablation for pulmonary vein isolation in patients with persistent atrial fibrillation: one-year outcome using second generation cryoballoon. Circ Arrhythm Electrophysiol. 2015;8:1073–9.CrossRefPubMed
5.
go back to reference Furnkranz A, Bordignon S, Dugo D, Perotta L, Gunawardene M, Schulte-Hahn B, et al. Improved 1-year clinical success rate of pulmonary vein isolation with the second-generation cryoballoon in patients with paroxysmal atrial fibrillation. J Cardiovasc Electrophysiol. 2014;25:840–4.CrossRefPubMed Furnkranz A, Bordignon S, Dugo D, Perotta L, Gunawardene M, Schulte-Hahn B, et al. Improved 1-year clinical success rate of pulmonary vein isolation with the second-generation cryoballoon in patients with paroxysmal atrial fibrillation. J Cardiovasc Electrophysiol. 2014;25:840–4.CrossRefPubMed
6.
go back to reference Chun KR, Furnkranz A, Metzner A, Schmidt B, Tilz R, Zerm T, et al. Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. J Cardiovasc Electrophysiol. 2009;20:1203–10.CrossRefPubMed Chun KR, Furnkranz A, Metzner A, Schmidt B, Tilz R, Zerm T, et al. Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. J Cardiovasc Electrophysiol. 2009;20:1203–10.CrossRefPubMed
7.
go back to reference Kuhne M, Knecht S, Altmann D, Ammann P, Schaer B, Osswald S, et al. Validation of a novel spiral mapping catheter for real-time recordings from the pulmonary veins during cryoballoon ablation of atrial fibrillation. Heart Rhythm. 2013;10:241–6.CrossRefPubMed Kuhne M, Knecht S, Altmann D, Ammann P, Schaer B, Osswald S, et al. Validation of a novel spiral mapping catheter for real-time recordings from the pulmonary veins during cryoballoon ablation of atrial fibrillation. Heart Rhythm. 2013;10:241–6.CrossRefPubMed
8.
go back to reference Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation—developed with the special contribution of the European Heart Rhythm Association. Europace. 2012;14:1385–413.CrossRefPubMed Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation—developed with the special contribution of the European Heart Rhythm Association. Europace. 2012;14:1385–413.CrossRefPubMed
9.
go back to reference Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998;339:659–66.CrossRefPubMed Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998;339:659–66.CrossRefPubMed
10.
go back to reference Chen SA, Hsieh MH, Tai CT, Tsai CF, Prakash VS, Yu WC, et al. Initiation of atrial fibrillation by ectopic beats originating from the pulmonary veins: electrophysiological characteristics, pharmacological responses, and effects of radiofrequency ablation. Circulation. 1999;100:1879–86.CrossRefPubMed Chen SA, Hsieh MH, Tai CT, Tsai CF, Prakash VS, Yu WC, et al. Initiation of atrial fibrillation by ectopic beats originating from the pulmonary veins: electrophysiological characteristics, pharmacological responses, and effects of radiofrequency ablation. Circulation. 1999;100:1879–86.CrossRefPubMed
11.
go back to reference Kuck KH, Furnkranz A. Cryoballoon ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2010;21:1427–31.CrossRefPubMed Kuck KH, Furnkranz A. Cryoballoon ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2010;21:1427–31.CrossRefPubMed
12.
go back to reference Verma A, Jiang CY, Betts TR, Chen J, Deisenhofer I, Mantovan R, et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372:1812–22.CrossRefPubMed Verma A, Jiang CY, Betts TR, Chen J, Deisenhofer I, Mantovan R, et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372:1812–22.CrossRefPubMed
13.
go back to reference Ho SY, Cabrera JA, Tran VH, Farre J, Anderson RH, Sanchez-Quintana D. Architecture of the pulmonary veins: relevance to radiofrequency ablation. Heart. 2001;86:265–70.CrossRefPubMedPubMedCentral Ho SY, Cabrera JA, Tran VH, Farre J, Anderson RH, Sanchez-Quintana D. Architecture of the pulmonary veins: relevance to radiofrequency ablation. Heart. 2001;86:265–70.CrossRefPubMedPubMedCentral
14.
go back to reference Saito T, Waki K, Becker AE. Left atrial myocardial extension onto pulmonary veins in humans: anatomic observations relevant for atrial arrhythmias. J Cardiovasc Electrophysiol. 2000;11:888–94.CrossRefPubMed Saito T, Waki K, Becker AE. Left atrial myocardial extension onto pulmonary veins in humans: anatomic observations relevant for atrial arrhythmias. J Cardiovasc Electrophysiol. 2000;11:888–94.CrossRefPubMed
15.
go back to reference Su W, Kowal R, Kowalski M, Metzner A, Svinarich JT, Wheelan K, et al. Best practice guide for cryoballoon ablation in atrial fibrillation: the compilation experience of more than 3000 procedures. Heart Rhythm. 2015;12:1658–66.CrossRefPubMed Su W, Kowal R, Kowalski M, Metzner A, Svinarich JT, Wheelan K, et al. Best practice guide for cryoballoon ablation in atrial fibrillation: the compilation experience of more than 3000 procedures. Heart Rhythm. 2015;12:1658–66.CrossRefPubMed
16.
go back to reference Furnkranz A, Koster I, Chun KR, Metzner A, Mathew S, Konstantinidou M, et al. Cryoballoon temperature predicts acute pulmonary vein isolation. Heart Rhythm. 2011;8:821–5.CrossRefPubMed Furnkranz A, Koster I, Chun KR, Metzner A, Mathew S, Konstantinidou M, et al. Cryoballoon temperature predicts acute pulmonary vein isolation. Heart Rhythm. 2011;8:821–5.CrossRefPubMed
17.
go back to reference Ciconte G, de Asmundis C, Sieira J, Conte G, Di Giovanni G, Mugnai G, et al. Single 3-minute freeze for second-generation cryoballoon ablation: one-year follow-up after pulmonary vein isolation. Heart Rhythm. 2015;12:673–80.CrossRefPubMed Ciconte G, de Asmundis C, Sieira J, Conte G, Di Giovanni G, Mugnai G, et al. Single 3-minute freeze for second-generation cryoballoon ablation: one-year follow-up after pulmonary vein isolation. Heart Rhythm. 2015;12:673–80.CrossRefPubMed
18.
go back to reference Metzner A, Rausch P, Lemes C, Reissmann B, Bardyszewski A, Tilz R, et al. The incidence of phrenic nerve injury during pulmonary vein isolation using the second-generation 28 mm cryoballoon. J Cardiovasc Electrophysiol. 2014;25:466–70.CrossRefPubMed Metzner A, Rausch P, Lemes C, Reissmann B, Bardyszewski A, Tilz R, et al. The incidence of phrenic nerve injury during pulmonary vein isolation using the second-generation 28 mm cryoballoon. J Cardiovasc Electrophysiol. 2014;25:466–70.CrossRefPubMed
19.
go back to reference Straube F, Dorwarth U, Schmidt M, Wankerl M, Ebersberger U, Hoffmann E. Comparison of the first and second cryoballoon: high-volume single-center safety and efficacy analysis. Circ Arrhythm Electrophysiol. 2014;7:293–9.CrossRefPubMed Straube F, Dorwarth U, Schmidt M, Wankerl M, Ebersberger U, Hoffmann E. Comparison of the first and second cryoballoon: high-volume single-center safety and efficacy analysis. Circ Arrhythm Electrophysiol. 2014;7:293–9.CrossRefPubMed
Metadata
Title
Cryoballoon ablation for pulmonary vein isolation in patients with atrial fibrillation: preliminary results using novel short-tip cryoballoon
Authors
Buelent Koektuerk
Hikmet Yorgun
Oezlem Koektuerk
Cem H. Turan
Muhammet Necati Murat Aksoy
Ramazan G. Turan
Eduard Gorr
Paul M. Bansmann
Christian Hoppe
Marc Horlitz
Publication date
01-10-2016
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2016
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-016-0134-5

Other articles of this Issue 1/2016

Journal of Interventional Cardiac Electrophysiology 1/2016 Go to the issue