Axillary vein puncture using fluoroscopic landmarks: a safe and effective approach for implantable cardioverter defibrillator leads

Authors: Federico Migliore, Mariachiara Siciliano, Manuel De Lazzari, Sonia Ferretto, Chiara Dalla Valle, Alessandro Zorzi, Domenico Corrado, Sabino Iliceto, Emanuele Bertaglia

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 3/2015

Login to get access

Abstract

Purpose

Axillary vein puncture is an effective method for pacemaker lead insertion with less complications compared with subclavian vein puncture; however, there are limited data on implantable cardioverter defibrillator (ICD) implantation with this technique. We reported our experience with a blind axillary vein puncture using fluoroscopic landmarks consisting of the outer edge of the first rib and the body surface of the second rib for ICD lead implantation.

Methods

The study population included 103 consecutive patients (mean age 59 ± 9 years) referred for ICD implantation using axillary vein puncture without contrast venography. An 18-gauge needle was advanced toward the outer edge of the fist rib below the clavicle or the body surface of the second rib. If the vein was not entered, the needle was withdrawn and the puncture was repeated with slight variations of needle direction for a maximum of four times, then contrast-guided vein puncture was performed.

Results

The total implanted leads were 152 including 103 right ventricular leads, 35 right atrial leads, and 14 left ventricular epicardial leads. Blind axillary vein puncture was successful obtained in 96 (93.2 %) patients. The rate of success was higher using the body surface of the second rib compared with the outer edge of the first rib (88.7 vs. 100 %; p = 0.04).Contrast venography was required in seven (6.8 %) patients because of vein course abnormality (n = 5) or vasospasm (n = 2). No acute complications or device-related complications were recorded during a mean follow-up of 12 ± 5 months.

Conclusions

Axillary vein access using fluoroscopic landmarks, especially the body surface of the second rib, is an effective approach for ICD implantation and offers the potential to avoid complications usually observed with traditional subclavian vein approach
Literature
1.
go back to reference Littleford, P. O., Parsonnet, V., & Spector, S. D. (1979). Method for the rapid and atraumatic insertion of permanent endocardial pacemaker electrodes through the subclavian vein. American Journal of Cardiology, 43, 980–982.PubMedCrossRef Littleford, P. O., Parsonnet, V., & Spector, S. D. (1979). Method for the rapid and atraumatic insertion of permanent endocardial pacemaker electrodes through the subclavian vein. American Journal of Cardiology, 43, 980–982.PubMedCrossRef
2.
go back to reference Aggarwal, R. K., Connelly, D. T., Ray, S. G., Ball, J., & Charles, R. G. (1995). Early complications of permanent pacemaker implantation. British Heart Journal, 73, 571–575.PubMedCentralPubMedCrossRef Aggarwal, R. K., Connelly, D. T., Ray, S. G., Ball, J., & Charles, R. G. (1995). Early complications of permanent pacemaker implantation. British Heart Journal, 73, 571–575.PubMedCentralPubMedCrossRef
3.
go back to reference Chauhan, A., Grace, A. A., & Newell, S. A. (1994). Early complications after dual chamber versus single chamber pacemaker implantation. PACE, 17(Pt. II), 2012–2013.PubMedCrossRef Chauhan, A., Grace, A. A., & Newell, S. A. (1994). Early complications after dual chamber versus single chamber pacemaker implantation. PACE, 17(Pt. II), 2012–2013.PubMedCrossRef
4.
go back to reference Kazama, S., Nishiyama, K., & Machii, M. (1993). Long term follow up of ventricular endocardial pacing leads. Japanese Heart Journal, 34, 193–200.PubMedCrossRef Kazama, S., Nishiyama, K., & Machii, M. (1993). Long term follow up of ventricular endocardial pacing leads. Japanese Heart Journal, 34, 193–200.PubMedCrossRef
5.
go back to reference Alt, E., Volker, R., & Blomer, H. (1987). Lead fracture in pacemaker patients. Thoracic and Cardiovascular Surgeon, 35, 101–104.PubMedCrossRef Alt, E., Volker, R., & Blomer, H. (1987). Lead fracture in pacemaker patients. Thoracic and Cardiovascular Surgeon, 35, 101–104.PubMedCrossRef
6.
go back to reference Antonelli, D., Rosenfeld, T., Freedberg, N. A., Palma, E., Gross, J. N., & Furman, S. (1998). Insulation lead failure: is it a matter of insulation coating, venous approach, or both? Pacing and Clinical Electrophysiology, 21, 418–421.PubMedCrossRef Antonelli, D., Rosenfeld, T., Freedberg, N. A., Palma, E., Gross, J. N., & Furman, S. (1998). Insulation lead failure: is it a matter of insulation coating, venous approach, or both? Pacing and Clinical Electrophysiology, 21, 418–421.PubMedCrossRef
7.
go back to reference Magney, J. E., Flynn, D. M., Parsons, J. A., Staplin, D. H., Chin-Purcell, M. V., Milstein, S., & Hunter, D. W. (1993). Anatomical mechanisms explaining damage to pacemaker leads, defibrillator leads, and failure of central venous catheters adjacent to the sternoclavicular joint. PACE, 16, 445.PubMedCrossRef Magney, J. E., Flynn, D. M., Parsons, J. A., Staplin, D. H., Chin-Purcell, M. V., Milstein, S., & Hunter, D. W. (1993). Anatomical mechanisms explaining damage to pacemaker leads, defibrillator leads, and failure of central venous catheters adjacent to the sternoclavicular joint. PACE, 16, 445.PubMedCrossRef
8.
go back to reference Femenia, F., Diez, J. C., Arce, M., & Baranchuk, A. (2011). Subclavian crush syndrome: a cause of pacemaker lead fracture. Cardiovascular Journal of Africa, 22, 201–202.PubMedCrossRef Femenia, F., Diez, J. C., Arce, M., & Baranchuk, A. (2011). Subclavian crush syndrome: a cause of pacemaker lead fracture. Cardiovascular Journal of Africa, 22, 201–202.PubMedCrossRef
9.
go back to reference Roelke, M., O’Nunain, S. S., & Osswald, S. (1995). Subclavian crush syndrome complicating transvenous cardioverter defibrillator systems. Pacing and Clinical Electrophysiology, 18, 973–979.PubMedCrossRef Roelke, M., O’Nunain, S. S., & Osswald, S. (1995). Subclavian crush syndrome complicating transvenous cardioverter defibrillator systems. Pacing and Clinical Electrophysiology, 18, 973–979.PubMedCrossRef
10.
go back to reference Migliore F, Curnis A, Bertaglia E. Axillary vein technique for pacemaker and implantable defibrillator leads implantation: a safe and alternative approach? J Cardiovasc Med 2014; [Epub ahead of print]. Migliore F, Curnis A, Bertaglia E. Axillary vein technique for pacemaker and implantable defibrillator leads implantation: a safe and alternative approach? J Cardiovasc Med 2014; [Epub ahead of print].
11.
go back to reference Sharma, G., Senguttuvan, N. B., Thachil, A., Leong, D., Naik, N., Yadav, R., & Juneja, R. (2012). A comparison of lead placement through the subclavian vein technique with fluoroscopy-guided axillary vein technique for permanent pacemaker insertion. Canadian Journal of Cardiology, 28, 542–546.PubMedCrossRef Sharma, G., Senguttuvan, N. B., Thachil, A., Leong, D., Naik, N., Yadav, R., & Juneja, R. (2012). A comparison of lead placement through the subclavian vein technique with fluoroscopy-guided axillary vein technique for permanent pacemaker insertion. Canadian Journal of Cardiology, 28, 542–546.PubMedCrossRef
12.
go back to reference Silvetti, M. S., Placidi, S., Palmieri, R., Righi, D., Rava, L., & Drago, F. (2013). Percutaneous axillary vein approach in pediatric pacing: comparison with subclavian vein approach. Pacing Clini Electrophysiol, 36, 1550–1557.CrossRef Silvetti, M. S., Placidi, S., Palmieri, R., Righi, D., Rava, L., & Drago, F. (2013). Percutaneous axillary vein approach in pediatric pacing: comparison with subclavian vein approach. Pacing Clini Electrophysiol, 36, 1550–1557.CrossRef
13.
go back to reference Yamada, T., Ando, K., Goya, M., Yasumoto, H., & Nobuyoshi, M. (2006). Long term lead survival after two types of vein puncture [abstract]. Heart Rhythm, 3, 141.CrossRef Yamada, T., Ando, K., Goya, M., Yasumoto, H., & Nobuyoshi, M. (2006). Long term lead survival after two types of vein puncture [abstract]. Heart Rhythm, 3, 141.CrossRef
14.
go back to reference Ki Hun, K., Kee-Joon, C., & Kyyoung-Min, P. (2013). Long term efficacy of axillary vein approach compared with subclavian vein approach for permanent pacemaker implantation [abstract]. American College of Cardiology, 61. Ki Hun, K., Kee-Joon, C., & Kyyoung-Min, P. (2013). Long term efficacy of axillary vein approach compared with subclavian vein approach for permanent pacemaker implantation [abstract]. American College of Cardiology, 61.
15.
go back to reference Byrd, C. (1993). Clinical experience with the extrathoracic introducer insertion technique. Pacing and Clinical Electrophysiology, 16, 1781–1784.PubMedCrossRef Byrd, C. (1993). Clinical experience with the extrathoracic introducer insertion technique. Pacing and Clinical Electrophysiology, 16, 1781–1784.PubMedCrossRef
16.
go back to reference Gardini, A., & Benedini, G. (1998). Blind extrathoracic subclavian venipuncture for pacemaker implant: a 3-year experience in 250 patients. Pacing and Clinical Electrophysiology, 21, 2304–2308.PubMedCrossRef Gardini, A., & Benedini, G. (1998). Blind extrathoracic subclavian venipuncture for pacemaker implant: a 3-year experience in 250 patients. Pacing and Clinical Electrophysiology, 21, 2304–2308.PubMedCrossRef
17.
go back to reference Magney, J., Staplin, D., Flynn, D., & Hunter, D. (1993). A new approach to percutaneous subclavian venipuncture to avoid lead fracture or central venous catheter occlusion. Pacing and Clinical Electrophysiology, 16, 2133–2142.PubMedCrossRef Magney, J., Staplin, D., Flynn, D., & Hunter, D. (1993). A new approach to percutaneous subclavian venipuncture to avoid lead fracture or central venous catheter occlusion. Pacing and Clinical Electrophysiology, 16, 2133–2142.PubMedCrossRef
18.
go back to reference Burri, H., Sunthorn, H., Dorsaz, P., & Shah, D. (2005). Prospective study of axillary vein puncture with or without contrast venography for pacemaker and defibrillator lead implantation. Pacing and Clinical Electrophysiology, 28(Suppl 1), S280–S283.PubMedCrossRef Burri, H., Sunthorn, H., Dorsaz, P., & Shah, D. (2005). Prospective study of axillary vein puncture with or without contrast venography for pacemaker and defibrillator lead implantation. Pacing and Clinical Electrophysiology, 28(Suppl 1), S280–S283.PubMedCrossRef
19.
go back to reference Calkins, H., Ramza, B. M., Brinker, J., Atiga, W., Donahue, K., Nsah, E., & Taylor, E. (2001). Prospective randomized comparison of the safety and effectiveness of placement of endocardial pacemaker and defibrillator leads using the extrathoracic subclavian vein guided by contrast venography versus the cephalic approach. Pacing and Clinical Electrophysiology, 24, 456–464.PubMedCrossRef Calkins, H., Ramza, B. M., Brinker, J., Atiga, W., Donahue, K., Nsah, E., & Taylor, E. (2001). Prospective randomized comparison of the safety and effectiveness of placement of endocardial pacemaker and defibrillator leads using the extrathoracic subclavian vein guided by contrast venography versus the cephalic approach. Pacing and Clinical Electrophysiology, 24, 456–464.PubMedCrossRef
20.
go back to reference Rodorf, R., Poggio, L., Savastano, S., Vicentini, A., Petracci, B., Chieffo, E., & Klersy, C. (2013). Failure of implantable cardioverter-defibrillator leads: a matter of lead size? Heart Rhythm, 10, 184–190.CrossRef Rodorf, R., Poggio, L., Savastano, S., Vicentini, A., Petracci, B., Chieffo, E., & Klersy, C. (2013). Failure of implantable cardioverter-defibrillator leads: a matter of lead size? Heart Rhythm, 10, 184–190.CrossRef
21.
go back to reference Dora, S., Kumar, V., Bhat, A., & Tharakan, J. (2003). Venogram-guided extrathoracic subclavian vein puncture. Indian Heart Journal, 55, 637–640.PubMed Dora, S., Kumar, V., Bhat, A., & Tharakan, J. (2003). Venogram-guided extrathoracic subclavian vein puncture. Indian Heart Journal, 55, 637–640.PubMed
22.
go back to reference Nash, A., Burrell, C., Ring, N., & Marshall, A. (1998). Evaluation of an ultrasonically guided vein puncture technique for the placement of permanent pacing electrodes. Pacing and Clinical Electrophysiology, 21, 452–455.PubMedCrossRef Nash, A., Burrell, C., Ring, N., & Marshall, A. (1998). Evaluation of an ultrasonically guided vein puncture technique for the placement of permanent pacing electrodes. Pacing and Clinical Electrophysiology, 21, 452–455.PubMedCrossRef
24.
go back to reference Antonelli, D., Feldman, A., Freedberg, N. A., & Turgeman, Y. (2013). Axillary vein puncture without contrast venography for pacemaker and defibrillator leads implantation. Pacing and Clinical Electrophysiology, 36, 1107–1110.PubMedCrossRef Antonelli, D., Feldman, A., Freedberg, N. A., & Turgeman, Y. (2013). Axillary vein puncture without contrast venography for pacemaker and defibrillator leads implantation. Pacing and Clinical Electrophysiology, 36, 1107–1110.PubMedCrossRef
25.
go back to reference Jiang, M., Gong, X., Zhou, S., Pu, J., Mao, J., & He, B. (2013). A comparison of steep and shallow needle trajectories in blind axillary vein puncture. Pacing and Clinical Electrophysiology, 36, 1150–1155.PubMedCrossRef Jiang, M., Gong, X., Zhou, S., Pu, J., Mao, J., & He, B. (2013). A comparison of steep and shallow needle trajectories in blind axillary vein puncture. Pacing and Clinical Electrophysiology, 36, 1150–1155.PubMedCrossRef
26.
go back to reference Ramza, B. M., Rosenthal, L., Hui, R., Nsah, E., Savader, S., Lawrence, J. H., Tomaselli, G., et al. (1997). Safety and effectiveness of placement of pacemaker and defibrillator leads in the axillary vein guided by contrast venography. American Journal of Cardiology, 80, 892–896.PubMedCrossRef Ramza, B. M., Rosenthal, L., Hui, R., Nsah, E., Savader, S., Lawrence, J. H., Tomaselli, G., et al. (1997). Safety and effectiveness of placement of pacemaker and defibrillator leads in the axillary vein guided by contrast venography. American Journal of Cardiology, 80, 892–896.PubMedCrossRef
27.
go back to reference Krahn, A. D., Morissette, J., Lahm, R., Haddad, T., Baxter, W. W., McVenes, R., Crystal, E., et al. (2014). Radiographic predictors of lead conductor fracture. Circ Arrhythm Electrophysiol, 7, 1070–1077. Krahn, A. D., Morissette, J., Lahm, R., Haddad, T., Baxter, W. W., McVenes, R., Crystal, E., et al. (2014). Radiographic predictors of lead conductor fracture. Circ Arrhythm Electrophysiol, 7, 1070–1077.
28.
go back to reference Kim, K. H., Park, K. M., Nam, G. B., Kim, D. K., Oh, M., & Choi, H. (2014). Comparison of the axillary venous approach and subclavian venous approach for efficacy of permanent implantation. 8-year follow-up results. Circulation Journal, 78, 865–871.PubMedCrossRef Kim, K. H., Park, K. M., Nam, G. B., Kim, D. K., Oh, M., & Choi, H. (2014). Comparison of the axillary venous approach and subclavian venous approach for efficacy of permanent implantation. 8-year follow-up results. Circulation Journal, 78, 865–871.PubMedCrossRef
Metadata
Title
Axillary vein puncture using fluoroscopic landmarks: a safe and effective approach for implantable cardioverter defibrillator leads
Authors
Federico Migliore
Mariachiara Siciliano
Manuel De Lazzari
Sonia Ferretto
Chiara Dalla Valle
Alessandro Zorzi
Domenico Corrado
Sabino Iliceto
Emanuele Bertaglia
Publication date
01-09-2015
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 3/2015
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-015-0011-7

Other articles of this Issue 3/2015

Journal of Interventional Cardiac Electrophysiology 3/2015 Go to the issue