Feasibility of using multivector impedance to monitor pulmonary congestion in heart failure patients

Authors: Philip F. Binkley, James G. Porterfield, Linda M. Porterfield, Scott L. Beau, Raffaele Corbisiero, G. Stephen Greer, Charles J. Love, Melanie Turkel, Anders Bjorling, Fujian Qu, Taraneh G. Farazi

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2012

Login to get access

Abstract

Purpose

Pulmonary edema (PE) is associated with fluid accumulation in the lungs. Device-based impedance measurements have been used to detect fluid overload prior to hospitalization. However, studies have reported a high false positive rate (FPR). The objective of this study was to develop and test a new multivector impedance-based algorithm that reliably tracks PE clinical events.

Methods

We enrolled patients with implanted CRT-Ds in 23 US centers within 2 weeks of device implant. Six-vector impedance data was collected automatically by the CRT-Ds every 30 min during emergency department visits/hospitalizations and every 2 h at all other times. Detection algorithms for cardiac resynchronization therapy defibrillator (CRT-D) and implantable cardiac defibrillator (ICD) devices were developed using those impedance vectors that would be available in corresponding devices and retrospectively evaluated.

Results

There were 75 patients (69 % male), mean age 66 ± 12 years, with a LVEF of 23 ± 6 % and QRS of 149 ± 25 ms. Twenty-one major clinical events occurred over 8.2 ± 2.6 months of follow-up time. CRT-D vector combinations resulted in a sensitivity of 71.4 % (95 % confidence interval 47.8–88.7) and a FPR of 0.56 (0.30–0.94) false positives per patient-year (FPs/pt-yr); ICD vector combinations resulted in a sensitivity of 61.9 % (38.4–81.9) and a FPR of 0.63 (0.36–0.90) FPs/pt-yr. In comparison, the single-vector RVCoil-Can implementation of this algorithm resulted in a sensitivity of 57.1 % (34.0–78.2) and a FPR of 0.74 (0.44–1.12) FPs/pt-yr.

Conclusions

This multivector impedance algorithm was effective in tracking PE clinical events in this patient population. Additional studies are needed to prospectively evaluate the performance of this algorithm in a larger population.
Literature
1.
go back to reference Lloyd-Jones, D., Adams, R. J., Brown, T. M., Carnethon, M., Dai, S., De, S. G., et al. (2010). Heart disease and stroke statistics—2010 update: a report from the American Heart Association. Circulation, 121(7), e46–e215.PubMedCrossRef Lloyd-Jones, D., Adams, R. J., Brown, T. M., Carnethon, M., Dai, S., De, S. G., et al. (2010). Heart disease and stroke statistics—2010 update: a report from the American Heart Association. Circulation, 121(7), e46–e215.PubMedCrossRef
2.
go back to reference Keenan, P. S., Normand, S. L., Lin, Z., Drye, E. E., Bhat, K. R., Ross, J. S., et al. (2008). An administrative claims measure suitable for profiling hospital performance on the basis of 30-day all-cause readmission rates among patients with heart failure. Circulation Cardiovascular Quality and Outcomes, 1(1), 29–37.PubMedCrossRef Keenan, P. S., Normand, S. L., Lin, Z., Drye, E. E., Bhat, K. R., Ross, J. S., et al. (2008). An administrative claims measure suitable for profiling hospital performance on the basis of 30-day all-cause readmission rates among patients with heart failure. Circulation Cardiovascular Quality and Outcomes, 1(1), 29–37.PubMedCrossRef
3.
go back to reference Jencks, S. F., Williams, M. V., & Coleman, E. A. (2009). Rehospitalizations among patients in the Medicare fee-for-service program. The New England Journal of Medicine, 360(14), 1418–1428.PubMedCrossRef Jencks, S. F., Williams, M. V., & Coleman, E. A. (2009). Rehospitalizations among patients in the Medicare fee-for-service program. The New England Journal of Medicine, 360(14), 1418–1428.PubMedCrossRef
4.
go back to reference Ross, J. S., Chen, J., Lin, Z., Bueno, H., Curtis, J. P., Keenan, P. S., et al. (2010). Recent national trends in readmission rates after heart failure hospitalization. Circulation Heart Failure, 3(1), 97–103.PubMedCrossRef Ross, J. S., Chen, J., Lin, Z., Bueno, H., Curtis, J. P., Keenan, P. S., et al. (2010). Recent national trends in readmission rates after heart failure hospitalization. Circulation Heart Failure, 3(1), 97–103.PubMedCrossRef
5.
go back to reference Levy, D., Kenchaiah, S., Larson, M. G., Benjamin, E. J., Kupka, M. J., Ho, K. K., et al. (2002). Long-term trends in the incidence of and survival with heart failure. The New England Journal of Medicine, 347(18), 1397–1402.PubMedCrossRef Levy, D., Kenchaiah, S., Larson, M. G., Benjamin, E. J., Kupka, M. J., Ho, K. K., et al. (2002). Long-term trends in the incidence of and survival with heart failure. The New England Journal of Medicine, 347(18), 1397–1402.PubMedCrossRef
6.
go back to reference Yu, C. M., Wang, L., Chau, E., Chan, R. H., Kong, S. L., Tang, M. O., et al. (2005). Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation, 112(6), 841–848.PubMedCrossRef Yu, C. M., Wang, L., Chau, E., Chan, R. H., Kong, S. L., Tang, M. O., et al. (2005). Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation, 112(6), 841–848.PubMedCrossRef
7.
go back to reference Bourge, R. C., Abraham, W. T., Adamson, P. B., Aaron, M. F., Aranda, J. M., Jr., Magalski, A., et al. (2008). Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: the COMPASS-HF study. Journal of the American College of Cardiology, 51(11), 1073–1079.PubMedCrossRef Bourge, R. C., Abraham, W. T., Adamson, P. B., Aaron, M. F., Aranda, J. M., Jr., Magalski, A., et al. (2008). Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: the COMPASS-HF study. Journal of the American College of Cardiology, 51(11), 1073–1079.PubMedCrossRef
8.
go back to reference Ritzema, J., Troughton, R., Melton, I., Crozier, I., Doughty, R., Krum, H., et al. (2010). Physician-directed patient self-management of left atrial pressure in advanced chronic heart failure. Circulation, 121(9), 1086–1095.PubMedCrossRef Ritzema, J., Troughton, R., Melton, I., Crozier, I., Doughty, R., Krum, H., et al. (2010). Physician-directed patient self-management of left atrial pressure in advanced chronic heart failure. Circulation, 121(9), 1086–1095.PubMedCrossRef
9.
go back to reference Abraham, W. T., Adamson, P. B., Bourge, R. C., Aaron, M. F., Costanzo, M. R., Stevenson, L. W., et al. (2011). Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. Lancet, 377(9766), 658–666.PubMedCrossRef Abraham, W. T., Adamson, P. B., Bourge, R. C., Aaron, M. F., Costanzo, M. R., Stevenson, L. W., et al. (2011). Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. Lancet, 377(9766), 658–666.PubMedCrossRef
10.
go back to reference Wang, L., Lahtinen, S., Lentz, L., Rakow, N., Kaszas, C., Ruetz, L., et al. (2005). Feasibility of using an implantable system to measure thoracic congestion in an ambulatory chronic heart failure canine model. Pacing and Clinical Electrophysiology, 28(5), 404–411.PubMedCrossRef Wang, L., Lahtinen, S., Lentz, L., Rakow, N., Kaszas, C., Ruetz, L., et al. (2005). Feasibility of using an implantable system to measure thoracic congestion in an ambulatory chronic heart failure canine model. Pacing and Clinical Electrophysiology, 28(5), 404–411.PubMedCrossRef
11.
go back to reference Khoury, D. S., Naware, M., Siou, J., Blomqvist, A., Mathuria, N. S., Wang, J., et al. (2009). Ambulatory monitoring of congestive heart failure by multiple bioelectric impedance vectors. Journal of the American College of Cardiology, 53(12), 1075–1081.PubMedCrossRef Khoury, D. S., Naware, M., Siou, J., Blomqvist, A., Mathuria, N. S., Wang, J., et al. (2009). Ambulatory monitoring of congestive heart failure by multiple bioelectric impedance vectors. Journal of the American College of Cardiology, 53(12), 1075–1081.PubMedCrossRef
12.
go back to reference Conraads, V. M., Tavazzi, L., Santini, M., Oliva, F., Gerritse, B., Yu, C. M., et al. (2011). Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial. European Heart Journal, 32(18), 2266–2273.PubMedCrossRef Conraads, V. M., Tavazzi, L., Santini, M., Oliva, F., Gerritse, B., Yu, C. M., et al. (2011). Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial. European Heart Journal, 32(18), 2266–2273.PubMedCrossRef
13.
go back to reference Abraham, W. T., Compton, S., Haas, G., Foreman, B., Canby, R. C., Fishel, R., et al. (2011). Intrathoracic impedance vs daily weight monitoring for predicting worsening heart failure events: results of the fluid accumulation status trial (FAST). Congestive Heart Failure, 17(2), 51–55.PubMedCrossRef Abraham, W. T., Compton, S., Haas, G., Foreman, B., Canby, R. C., Fishel, R., et al. (2011). Intrathoracic impedance vs daily weight monitoring for predicting worsening heart failure events: results of the fluid accumulation status trial (FAST). Congestive Heart Failure, 17(2), 51–55.PubMedCrossRef
14.
go back to reference Vollmann, D., Nagele, H., Schauerte, P., Wiegand, U., Butter, C., Zanotto, G., et al. (2007). Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure. European Heart Journal, 28(15), 1835–1840.PubMedCrossRef Vollmann, D., Nagele, H., Schauerte, P., Wiegand, U., Butter, C., Zanotto, G., et al. (2007). Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure. European Heart Journal, 28(15), 1835–1840.PubMedCrossRef
15.
go back to reference Ypenburg, C., Bax, J. J., van der Wall, E. E., Schalij, M. J., & van Erven, L. (2007). Intrathoracic impedance monitoring to predict decompensated heart failure. The American Journal of Cardiology, 99(4), 554–557.PubMedCrossRef Ypenburg, C., Bax, J. J., van der Wall, E. E., Schalij, M. J., & van Erven, L. (2007). Intrathoracic impedance monitoring to predict decompensated heart failure. The American Journal of Cardiology, 99(4), 554–557.PubMedCrossRef
16.
go back to reference Catanzariti, D., Lunati, M., Landolina, M., Zanotto, G., Lonardi, G., Iacopino, S., et al. (2009). Monitoring intrathoracic impedance with an implantable defibrillator reduces hospitalizations in patients with heart failure. Pacing and Clinical Electrophysiology, 32(3), 363–370.PubMedCrossRef Catanzariti, D., Lunati, M., Landolina, M., Zanotto, G., Lonardi, G., Iacopino, S., et al. (2009). Monitoring intrathoracic impedance with an implantable defibrillator reduces hospitalizations in patients with heart failure. Pacing and Clinical Electrophysiology, 32(3), 363–370.PubMedCrossRef
17.
go back to reference Wang, L. (2007). Fundamentals of intrathoracic impedance monitoring in heart failure. The American Journal of Cardiology, 99(10A), 3G–10G.PubMedCrossRef Wang, L. (2007). Fundamentals of intrathoracic impedance monitoring in heart failure. The American Journal of Cardiology, 99(10A), 3G–10G.PubMedCrossRef
18.
go back to reference Kramer, D. B., & Maisel, W. H. (2010). An unusual cause of abnormal intrathoracic impedance in a patient with arrhythmogenic right ventricular cardiomyopathy. Pacing and Clinical Electrophysiology, 34(7), e60–3.PubMedCrossRef Kramer, D. B., & Maisel, W. H. (2010). An unusual cause of abnormal intrathoracic impedance in a patient with arrhythmogenic right ventricular cardiomyopathy. Pacing and Clinical Electrophysiology, 34(7), e60–3.PubMedCrossRef
19.
go back to reference Ware, L. B., & Matthay, M. A. (2005). Clinical practice. Acute pulmonary edema. The New England Journal of Medicine, 353(26), 2788–2796.PubMedCrossRef Ware, L. B., & Matthay, M. A. (2005). Clinical practice. Acute pulmonary edema. The New England Journal of Medicine, 353(26), 2788–2796.PubMedCrossRef
Metadata
Title
Feasibility of using multivector impedance to monitor pulmonary congestion in heart failure patients
Authors
Philip F. Binkley
James G. Porterfield
Linda M. Porterfield
Scott L. Beau
Raffaele Corbisiero
G. Stephen Greer
Charles J. Love
Melanie Turkel
Anders Bjorling
Fujian Qu
Taraneh G. Farazi
Publication date
01-11-2012
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2012
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-012-9693-2

Other articles of this Issue 2/2012

Journal of Interventional Cardiac Electrophysiology 2/2012 Go to the issue