01-09-2012 | REVIEW

Subcutaneous chronic implantable defibrillation systems in humans

Authors: Riccardo Cappato, Warren M. Smith, Margaret A. Hood, Ian G. Crozier, Luc Jordaens, Stefan G. Spitzer, Andrey V. Ardashev, Lucas Boersma, Pierpaolo Lupo, Andrew A. Grace, Gust H. Bardy

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 3/2012

Login to get access

Abstract

The recent introduction of subcutaneous implantable cardioverter defibrillator (S-ICD) has raised attention about the potential of this technology for clinical use in daily clinical practice. We review the methods and results of the four studies conducted in humans for approval of this innovative technology for daily practice. Two studies using a temporary S-ICD system (acute human studies) were conducted to search for an appropriate lead configuration and energy requirements. For this purpose, 4 S-ICD configurations were tested in 78 patients at the time of transvenous (TV)-ICD implantation. The optimal configuration was tested in 49 more patients to comparatively assess the subcutaneous defibrillation threshold (S-DFT) versus the standard TV-ICD. Long-term implants were evaluated in 55 patients using an implanted system (chronic human study). The acute humans studies led to an optimal S-ICD configuration comprising a parasternal electrode and left anterolateral thoracic pulse generator. Both configurations successfully terminated 98% of induced ventricular fibrillation (VF), but significantly higher energy levels were required with S-ICD than with TV-ICD systems (36.6 ± 19.8 J vs. 11.1 ± 8.5 J). In the chronic study, all 137 VF episodes induced at time of implant were detected with a 98% conversion rate. Two pocket infections and four lead revisions were required during 10 ± 1 months of follow-up. During this period, survival was 98%, and 12 spontaneous ventricular tachyarrhythmias were detected and treated by the device. These data show that the S-ICD systems here consistently detected and converted VF induced at time of implant as well as sustained ventricular tachyarrhythmias occurring during follow-up (248).
Literature
1.
go back to reference The AVID Investigators. (1997). A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. The New England Journal of Medicine, 337, 1576–1583.CrossRef The AVID Investigators. (1997). A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. The New England Journal of Medicine, 337, 1576–1583.CrossRef
2.
go back to reference Connolly, S. J., Hallstrom, A. P., Cappato, R., Schron, E. B., Kuck, K. H., Zipes, D. P., Greene, H. L., Boczor, S., Domanski, M., Follmann, D., Gent, M., & Roberts, R. S. (2000). on behalf of the investigators of the AVID, CASH and CIDS studies. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. European Heart Journal, 21, 2071–2078.PubMedCrossRef Connolly, S. J., Hallstrom, A. P., Cappato, R., Schron, E. B., Kuck, K. H., Zipes, D. P., Greene, H. L., Boczor, S., Domanski, M., Follmann, D., Gent, M., & Roberts, R. S. (2000). on behalf of the investigators of the AVID, CASH and CIDS studies. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. European Heart Journal, 21, 2071–2078.PubMedCrossRef
3.
go back to reference Moss, A. J., Hall, W. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Klein, H., Levine, J. H., Saksena, S., Waldo, A. L., Wilber, D., Brown, M. W., & Heo, M. (1996). for the Multicenter Automatic Defibrillator Implantation Trial Investigators. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. The New England Journal of Medicine, 335, 1933–1940.PubMedCrossRef Moss, A. J., Hall, W. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Klein, H., Levine, J. H., Saksena, S., Waldo, A. L., Wilber, D., Brown, M. W., & Heo, M. (1996). for the Multicenter Automatic Defibrillator Implantation Trial Investigators. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. The New England Journal of Medicine, 335, 1933–1940.PubMedCrossRef
4.
go back to reference Moss, A. J., Zareba, W., Hall, W. J., Klein, H., Wilber, D. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Brown, M. W., & Andrews, M. L. (2002). for the Multicenter Automatic Difibrillator Implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. The New England Journal of Medicine, 346, 877–883.PubMedCrossRef Moss, A. J., Zareba, W., Hall, W. J., Klein, H., Wilber, D. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Brown, M. W., & Andrews, M. L. (2002). for the Multicenter Automatic Difibrillator Implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. The New England Journal of Medicine, 346, 877–883.PubMedCrossRef
5.
go back to reference Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R., Domanski, M., Troutman, C., Anderson, J., Johnson, G., McNulty, S. E., Clapp-Channing, N., Davidson-Ray, L. D., Fraulo, E. S., Fishbein, D. P., & Luceri, R. M. (2005). Ip JH for the SCD-HeFT Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. The New England Journal of Medicine, 352, 225–237.PubMedCrossRef Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R., Domanski, M., Troutman, C., Anderson, J., Johnson, G., McNulty, S. E., Clapp-Channing, N., Davidson-Ray, L. D., Fraulo, E. S., Fishbein, D. P., & Luceri, R. M. (2005). Ip JH for the SCD-HeFT Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. The New England Journal of Medicine, 352, 225–237.PubMedCrossRef
6.
go back to reference Mirowski, M., Reid, P. R., Mower, M. M., Walkins, L. Jr., Platia, E. V., Griffith, L. S., & Juanteguv, J. Ml. (1984). The automatic implantable cardioverter-defibrillator. Pacing and Clinical Electrophysiology, 7, 534–540.PubMedCrossRef Mirowski, M., Reid, P. R., Mower, M. M., Walkins, L. Jr., Platia, E. V., Griffith, L. S., & Juanteguv, J. Ml. (1984). The automatic implantable cardioverter-defibrillator. Pacing and Clinical Electrophysiology, 7, 534–540.PubMedCrossRef
7.
go back to reference Brodman, R., Fisher, J. D., Furman, S., Johnston, D. R., Kim, S. G., Matos, J. A., & Waspe, L. E. (1984). Implantation of automatic cardioverter defibrillators via median sternotomy. Pacing and Clinical Electrophysiology, 7, 1363–1369.PubMedCrossRef Brodman, R., Fisher, J. D., Furman, S., Johnston, D. R., Kim, S. G., Matos, J. A., & Waspe, L. E. (1984). Implantation of automatic cardioverter defibrillators via median sternotomy. Pacing and Clinical Electrophysiology, 7, 1363–1369.PubMedCrossRef
8.
go back to reference Echt, D. S., Armstrong, K., Schmidt, P., Oyer, P. E., Stinson, E. B., & Winckle, R. A. (1985). Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator. Circulation, 71, 289–296.PubMedCrossRef Echt, D. S., Armstrong, K., Schmidt, P., Oyer, P. E., Stinson, E. B., & Winckle, R. A. (1985). Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator. Circulation, 71, 289–296.PubMedCrossRef
9.
go back to reference Ellenbogen, K. A., Hellkamp, A. S., Wilkoff, B. L., Camunas, J. L., Love, J. C., Hadjis, T. A., Lee, K. L., & Lamas, G. A. (2003). Complications arising after implantation of DDD pacemakers: the MOST experience. The American Journal of Cardiology, 92(6), 740–741.PubMedCrossRef Ellenbogen, K. A., Hellkamp, A. S., Wilkoff, B. L., Camunas, J. L., Love, J. C., Hadjis, T. A., Lee, K. L., & Lamas, G. A. (2003). Complications arising after implantation of DDD pacemakers: the MOST experience. The American Journal of Cardiology, 92(6), 740–741.PubMedCrossRef
10.
go back to reference Parsonnet, V., Bernstein, A. D., & Lindsay, B. (1989). Pacemaker-implantation complication rates: an analysis of some contributing factors. Journal of the American College of Cardiology, 13, 917–921.PubMedCrossRef Parsonnet, V., Bernstein, A. D., & Lindsay, B. (1989). Pacemaker-implantation complication rates: an analysis of some contributing factors. Journal of the American College of Cardiology, 13, 917–921.PubMedCrossRef
11.
go back to reference Tobin, K., Stewart, J., Westveer, D., & Frumin, H. (2000). Acute complications of permanent pacemaker implantation: their financial implications and relation to volume and operator experience. The American Journal of Cardiology, 85, 774–776.PubMedCrossRef Tobin, K., Stewart, J., Westveer, D., & Frumin, H. (2000). Acute complications of permanent pacemaker implantation: their financial implications and relation to volume and operator experience. The American Journal of Cardiology, 85, 774–776.PubMedCrossRef
12.
go back to reference Trigano, A. J., Taramasco, V., Paganelli, F., Gerard, R., & Levy, S. (1996). Incidence of perforation and other mechanical complications during dual active fixation. PACE, 19, 1828–1831.PubMedCrossRef Trigano, A. J., Taramasco, V., Paganelli, F., Gerard, R., & Levy, S. (1996). Incidence of perforation and other mechanical complications during dual active fixation. PACE, 19, 1828–1831.PubMedCrossRef
13.
go back to reference Kai, H., Honma, T., & Imaizumi, T. (2001). Superior vena cava syndrome caused by multiple pacing leads. Heart, 86, 80.PubMedCrossRef Kai, H., Honma, T., & Imaizumi, T. (2001). Superior vena cava syndrome caused by multiple pacing leads. Heart, 86, 80.PubMedCrossRef
14.
go back to reference Chua, J. D., Wilkoff, B. L., Lee, I., Juratli, N., Longworth, D. L., & Gordon, S. M. (2000). Diagnosis and management of infections involving implantable electrophysiologic cardiac devices. Ann Int Med, 133, 604–608.PubMed Chua, J. D., Wilkoff, B. L., Lee, I., Juratli, N., Longworth, D. L., & Gordon, S. M. (2000). Diagnosis and management of infections involving implantable electrophysiologic cardiac devices. Ann Int Med, 133, 604–608.PubMed
15.
go back to reference Grimm, W., Flores, B. F., & Marchlinski, F. E. (1993). Complications of implantable cardioverter defibrillator therapy: follow-up of 241 patients. PACE, 16, 218–222.PubMedCrossRef Grimm, W., Flores, B. F., & Marchlinski, F. E. (1993). Complications of implantable cardioverter defibrillator therapy: follow-up of 241 patients. PACE, 16, 218–222.PubMedCrossRef
16.
go back to reference Pavia, S., & Wilkoff, B. (2001). The management of surgical complications of pacemakers and implantable cardioverter-defibrillators. Current Opinion in Cardiology, 16, 66–71.PubMedCrossRef Pavia, S., & Wilkoff, B. (2001). The management of surgical complications of pacemakers and implantable cardioverter-defibrillators. Current Opinion in Cardiology, 16, 66–71.PubMedCrossRef
17.
go back to reference Dorwarth, U., Frey, B., Dugas, M., Matis, T., Fiek, M., Schmoeckel, M., Remp, T., Durchlaub, I., Gerth, A., Steinbeck, G., & Hoffmann, E. (2003). Transvenous defibrillation leads: high incidence of failure during long-term follow-up. Journal Cardiovasc Electrophys, 14, 38–43.CrossRef Dorwarth, U., Frey, B., Dugas, M., Matis, T., Fiek, M., Schmoeckel, M., Remp, T., Durchlaub, I., Gerth, A., Steinbeck, G., & Hoffmann, E. (2003). Transvenous defibrillation leads: high incidence of failure during long-term follow-up. Journal Cardiovasc Electrophys, 14, 38–43.CrossRef
18.
go back to reference Ellenbogen, K. A., Wood, M. A., Shepard, R. K., Clemo, H. F., Vaughn, T., Holloman, K., Dow, M., Leffler, J., Abeyratne, A., & Verness, D. (2003). Detection and management of an implantable cardioverter defibrillator lead failure incidence and clinical implications. Journal of the American College of Cardiology, 41, 73–80.PubMedCrossRef Ellenbogen, K. A., Wood, M. A., Shepard, R. K., Clemo, H. F., Vaughn, T., Holloman, K., Dow, M., Leffler, J., Abeyratne, A., & Verness, D. (2003). Detection and management of an implantable cardioverter defibrillator lead failure incidence and clinical implications. Journal of the American College of Cardiology, 41, 73–80.PubMedCrossRef
19.
go back to reference Kleemann, T., Becker, T., Doenges, K., Vater, M., Senges, J., Schneider, S., Saggau, W., Weisse, U., & Seidl, K. (2007). Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of > 10years. Circulation, 115, 2474–2480.PubMedCrossRef Kleemann, T., Becker, T., Doenges, K., Vater, M., Senges, J., Schneider, S., Saggau, W., Weisse, U., & Seidl, K. (2007). Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of > 10years. Circulation, 115, 2474–2480.PubMedCrossRef
20.
go back to reference Tung, R., Zimetbaum, P., & Josephson, M. E. (2008). A critical appraisal of implantable cardioverter- defibrillator therapy for the prevention of sudden cardiac death. JACC, 52, 1111–1121.PubMed Tung, R., Zimetbaum, P., & Josephson, M. E. (2008). A critical appraisal of implantable cardioverter- defibrillator therapy for the prevention of sudden cardiac death. JACC, 52, 1111–1121.PubMed
21.
go back to reference Maisel, W. H. (2007). Transvenous implantable cardioverter defibrillator leads: the weakest link. Circulation, 115, 2461–2463.PubMedCrossRef Maisel, W. H. (2007). Transvenous implantable cardioverter defibrillator leads: the weakest link. Circulation, 115, 2461–2463.PubMedCrossRef
22.
go back to reference Kitamura, S., Satomi, K., Kurita, T., Shimizu, W., Suyama, K., Aihara, N., Niwaya, K., Kobayashi, J., & Kamakura, S. (2006). Long-term follow-up of transvenous defibrillation leads: high incidence of fracture in coaxial polyurethane lead. Circulation Journal, 70, 273–277.PubMedCrossRef Kitamura, S., Satomi, K., Kurita, T., Shimizu, W., Suyama, K., Aihara, N., Niwaya, K., Kobayashi, J., & Kamakura, S. (2006). Long-term follow-up of transvenous defibrillation leads: high incidence of fracture in coaxial polyurethane lead. Circulation Journal, 70, 273–277.PubMedCrossRef
25.
go back to reference Jones, G. K., et al. (1995). Mechanical complications after implantation of multiple-lead nonthoracotomy defibrillator systems: implications for management and future system design. American Heart Journal, 130, 327–333.PubMedCrossRef Jones, G. K., et al. (1995). Mechanical complications after implantation of multiple-lead nonthoracotomy defibrillator systems: implications for management and future system design. American Heart Journal, 130, 327–333.PubMedCrossRef
26.
go back to reference Bardy, G. H., et al. (1993). Implantable transvenous cardioverter-defibrillators. Circulation, 87, 1152–1168.PubMedCrossRef Bardy, G. H., et al. (1993). Implantable transvenous cardioverter-defibrillators. Circulation, 87, 1152–1168.PubMedCrossRef
27.
go back to reference Kron, J. (2001). Lead- and device-related complications in the antiarrhythmics versus implantable defibrillators trial. American Heart Journal, 141, 92–98.PubMedCrossRef Kron, J. (2001). Lead- and device-related complications in the antiarrhythmics versus implantable defibrillators trial. American Heart Journal, 141, 92–98.PubMedCrossRef
28.
go back to reference Maisel, W. H., & Kramer, D. B. (2008). Implantable cardioverter-defibrillator lead performance. Circulation, 117, 2721–2723.PubMedCrossRef Maisel, W. H., & Kramer, D. B. (2008). Implantable cardioverter-defibrillator lead performance. Circulation, 117, 2721–2723.PubMedCrossRef
29.
go back to reference Cappato, R., & Bardy, G. H. (2005). Sub-cutaneous electrical defibrillation in canines. Journal of the American College of Cardiology, 45 Suppl, 106A. Abstract. Cappato, R., & Bardy, G. H. (2005). Sub-cutaneous electrical defibrillation in canines. Journal of the American College of Cardiology, 45 Suppl, 106A. Abstract.
30.
go back to reference Bardy, G., Smith, W. M., Hood, M. A., Crozier, I. G., Melton, I. C., Jordaens, L., Theuns, D., Park, R., Wright, D. J., Connelly, D. T., Fynn, S. P., Murgatroyd, F. D., Sperzel, J., Neuzner, J., Spitzer, S., Ardashev, A. V., Oduro, A., Boersma, L., Maass, A. H., Van Gelder, I. C., Wilde, A. A., van Dessel, P. F., Knops, R. E., Barr, C. S., Lupo, P. P., Cappato, R., & Grace, A. A. (2010). An entirely subcutaneous implantable cardioverter defibrillator. New Engl Journal of Medicine, 308, 23–31. Bardy, G., Smith, W. M., Hood, M. A., Crozier, I. G., Melton, I. C., Jordaens, L., Theuns, D., Park, R., Wright, D. J., Connelly, D. T., Fynn, S. P., Murgatroyd, F. D., Sperzel, J., Neuzner, J., Spitzer, S., Ardashev, A. V., Oduro, A., Boersma, L., Maass, A. H., Van Gelder, I. C., Wilde, A. A., van Dessel, P. F., Knops, R. E., Barr, C. S., Lupo, P. P., Cappato, R., & Grace, A. A. (2010). An entirely subcutaneous implantable cardioverter defibrillator. New Engl Journal of Medicine, 308, 23–31.
31.
go back to reference Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., Hlatky, M. A., Newby, L. K., Page, R. L., Schoenfeld, M. H., Silka, M. J., Stevenson, L. W., & Sweeney, M. O. (2008). ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: executive summary. Heart Rhythm, 5, 935–955. Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., Hlatky, M. A., Newby, L. K., Page, R. L., Schoenfeld, M. H., Silka, M. J., Stevenson, L. W., & Sweeney, M. O. (2008). ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: executive summary. Heart Rhythm, 5, 935–955.
32.
go back to reference Gradaus, R., Hammel, D., Kotthoff, S., & Bocker, D. (2001). Nonthoracotomy implantable cardioverter defibrillator placement in children: use of subcutaneous array leads and abdominally placed implantable cardioverter defibrillator in children. Journal of Cardiovascular Electrophysiology, 12, 356–360.PubMedCrossRef Gradaus, R., Hammel, D., Kotthoff, S., & Bocker, D. (2001). Nonthoracotomy implantable cardioverter defibrillator placement in children: use of subcutaneous array leads and abdominally placed implantable cardioverter defibrillator in children. Journal of Cardiovascular Electrophysiology, 12, 356–360.PubMedCrossRef
33.
go back to reference Thogersen, A. M., Helvind, M., Jensen, T., Andersen, H., Jacobsen, J. R., & Chen, X. (2001). Implantable cardioverter defibrillator in a 4month old infant with cardiac arrest associated with a vascular heart tumor. PACE, 24, 1699–1700.PubMedCrossRef Thogersen, A. M., Helvind, M., Jensen, T., Andersen, H., Jacobsen, J. R., & Chen, X. (2001). Implantable cardioverter defibrillator in a 4month old infant with cardiac arrest associated with a vascular heart tumor. PACE, 24, 1699–1700.PubMedCrossRef
34.
go back to reference Alter, P., Waldhaus, S., Plachta, E., Moosdorf, R., & Grimm, W. (2005). Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing and Clinical Electrophysiology, 28, 926–932.PubMedCrossRef Alter, P., Waldhaus, S., Plachta, E., Moosdorf, R., & Grimm, W. (2005). Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing and Clinical Electrophysiology, 28, 926–932.PubMedCrossRef
35.
go back to reference Khairy, P., Landzberg, M. J., Gatzoulis, M. A., Mercier, L. A., Fernandes, S. M., Côte, J. M., Lavoie, J. P., Fournier, A., Guerra, P. G., Frogoudaki, A., & Walsh, E. P. (2006). Dore A for the Epicardial Versus ENdocardial pacing and Thromboembolic events (EVENT) Investigators. Transvenous pacing leads and systemic thromboemboli in patients with intracardiac shunts: a multicenter study. Circulation, 113, 2391–2397.PubMedCrossRef Khairy, P., Landzberg, M. J., Gatzoulis, M. A., Mercier, L. A., Fernandes, S. M., Côte, J. M., Lavoie, J. P., Fournier, A., Guerra, P. G., Frogoudaki, A., & Walsh, E. P. (2006). Dore A for the Epicardial Versus ENdocardial pacing and Thromboembolic events (EVENT) Investigators. Transvenous pacing leads and systemic thromboemboli in patients with intracardiac shunts: a multicenter study. Circulation, 113, 2391–2397.PubMedCrossRef
36.
go back to reference Sohail, M. R., Uslan, D. Z., Khan, A. H., Friedman, P. A., Hayes, D. L., Wilson, W. R., Steckelberg, J. M., Stoner, S., & Baddour, L. M. (2007). Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. Journal of the American College of Cardiology, 49, 1851–1859.PubMedCrossRef Sohail, M. R., Uslan, D. Z., Khan, A. H., Friedman, P. A., Hayes, D. L., Wilson, W. R., Steckelberg, J. M., Stoner, S., & Baddour, L. M. (2007). Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. Journal of the American College of Cardiology, 49, 1851–1859.PubMedCrossRef
37.
go back to reference Danik, S. B., Mansour, M., Singh, J., Reddy, V. Y., Ellinor, P. T., Milan, D., & Heist, K. (2007). ‘Avila A, Ruskin JN, Mela T. Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator. Heart Rhythm, 4, 439–442.PubMedCrossRef Danik, S. B., Mansour, M., Singh, J., Reddy, V. Y., Ellinor, P. T., Milan, D., & Heist, K. (2007). ‘Avila A, Ruskin JN, Mela T. Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator. Heart Rhythm, 4, 439–442.PubMedCrossRef
38.
go back to reference Reynolds, M. R., Cohen, D. J., Kugelmass, A. D., Brown, P. P., Becker, E. R., Culler, S. D., & Simon, A. W. (2006). The frequency and incremental cost of major complications among medicare beneficiaries receiving implantable cardioverter-defibrillators. Journal of the American College of Cardiology, 47, 2493–2497.PubMedCrossRef Reynolds, M. R., Cohen, D. J., Kugelmass, A. D., Brown, P. P., Becker, E. R., Culler, S. D., & Simon, A. W. (2006). The frequency and incremental cost of major complications among medicare beneficiaries receiving implantable cardioverter-defibrillators. Journal of the American College of Cardiology, 47, 2493–2497.PubMedCrossRef
39.
go back to reference Byrd, C. L., Wilkoff, B. L., Love, C. J., et al. (1999). Intravascular extraction of problematic or infected permanent pacemaker leads: 1994–1996. U.S. Extraction Database, MED Institute. Pacing and Clinical Electrophysiology, 22, 1348–1357.PubMedCrossRef Byrd, C. L., Wilkoff, B. L., Love, C. J., et al. (1999). Intravascular extraction of problematic or infected permanent pacemaker leads: 1994–1996. U.S. Extraction Database, MED Institute. Pacing and Clinical Electrophysiology, 22, 1348–1357.PubMedCrossRef
Metadata
Title
Subcutaneous chronic implantable defibrillation systems in humans
Authors
Riccardo Cappato
Warren M. Smith
Margaret A. Hood
Ian G. Crozier
Luc Jordaens
Stefan G. Spitzer
Andrey V. Ardashev
Lucas Boersma
Pierpaolo Lupo
Andrew A. Grace
Gust H. Bardy
Publication date
01-09-2012
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 3/2012
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-012-9665-6

Other articles of this Issue 3/2012

Journal of Interventional Cardiac Electrophysiology 3/2012 Go to the issue