Published in: The International Journal of Cardiovascular Imaging 2/2014

01-02-2014 | Original Paper

Comparison of longitudinal geometric measurement in human coronary arteries between frequency-domain optical coherence tomography and intravascular ultrasound

Authors: Yong Liu, Kunihiro Shimamura, Takashi Kubo, Atsuhi Tanaka, Hironori Kitabata, Yasushi Ino, Takashi Tanimoto, Yasutsugu Shiono, Makoto Orii, Takashi Yamano, Tomoyuki Yamaguchi, Kumiko Hirata, Toshio Imanishi, Takashi Akasaka

Published in: The International Journal of Cardiovascular Imaging | Issue 2/2014

Login to get access

Abstract

Previous studies have demonstrated the higher accuracy of frequency-domain optical coherence tomography (FD-OCT) for quantitative measurements in comparison with intravascular ultrasound (IVUS). However, those analyses were based on the cross-sectional images. The aim of this study was to assess the accuracy of FD-OCT for longitudinal geometric measurements of coronary arteries in comparison with IVUS. Between October 2011 and March 2012, we performed prospective FD-OCT and IVUS examinations in consecutive 77 patients who underwent percutaneous coronary intervention with single stent. Regression analysis and Bland–Altman analysis revealed an excellent correlation between the FD-OCT-measured stent lengths and IVUS-measured stent lengths (r = 0.986, p < 0.001; mean difference = −0.51 mm). There was an excellent agreement between the actual stent lengths and the FD-OCT-measured stent lengths (r = 0.993, p < 0.001) as well as between the actual stent lengths and the IVUS-measured stent lengths (r = 0.981, p < 0.001). The difference between the actual stent lengths and the FD-OCT-measured stent lengths was significantly smaller than that between the actual stent lengths and the IVUS-measured stent lengths (0.15 ± 0.68 vs. 0.70 ± 1.15 mm, p < 0.001). Both FD-OCT (mean difference = −0.04 and −0.04 mm, respectively) and IVUS (mean difference = −0.06 and −0.06 mm, respectively) showed an excellent intra-observer and inter-observer reproducibility for the stent length measurements. In conclusion, FD-OCT provides accurate longitudinal measurement with excellent intra-observer and inter-observer reproducibility. FD-OCT might be a reliable technique for longitudinal geometric measurement in human coronary arteries.
Literature
1.
go back to reference Kwon SW, Kim BK, Kim TH, Kim JS, Ko YG, Choi D, Jang Y, Hong MK (2011) Qualitative assessment of neointimal tissue after drug-eluting stent implantation: comparison between follow-up optical coherence tomography and intravascular ultrasound. Am Heart J 161(2):367–372PubMedCrossRef Kwon SW, Kim BK, Kim TH, Kim JS, Ko YG, Choi D, Jang Y, Hong MK (2011) Qualitative assessment of neointimal tissue after drug-eluting stent implantation: comparison between follow-up optical coherence tomography and intravascular ultrasound. Am Heart J 161(2):367–372PubMedCrossRef
2.
go back to reference Diletti R, Garcia–Garcia HM, Gomez-Lara J et al (2011) Assessment of coronary atherosclerosis progression and regression at bifurcations using combined IVUS and OCT. JACC Cardiovasc Imaging 4(7):774–780PubMedCrossRef Diletti R, Garcia–Garcia HM, Gomez-Lara J et al (2011) Assessment of coronary atherosclerosis progression and regression at bifurcations using combined IVUS and OCT. JACC Cardiovasc Imaging 4(7):774–780PubMedCrossRef
3.
go back to reference Kubo T, Imanishi T, Takarada S et al (2007) Assessment of culprit lesion morphology in acute myocardial infarction: ability of optical coherence tomography compared with intravascular ultrasound and coronary angioscopy. J Am Coll Cardiol 50(10):933–939PubMedCrossRef Kubo T, Imanishi T, Takarada S et al (2007) Assessment of culprit lesion morphology in acute myocardial infarction: ability of optical coherence tomography compared with intravascular ultrasound and coronary angioscopy. J Am Coll Cardiol 50(10):933–939PubMedCrossRef
4.
go back to reference Suzuki Y, Ikeno F, Koizumi T, Tio F, Yeung AC, Yock PG, Fitzgerald PJ, Fearon WF (2008) In vivo comparison between optical coherence tomography and intravascular ultrasound for detecting small degrees of in-stent neointima after stent implantation. JACC Cardiovasc Interv 1(2):168–173PubMedCrossRef Suzuki Y, Ikeno F, Koizumi T, Tio F, Yeung AC, Yock PG, Fitzgerald PJ, Fearon WF (2008) In vivo comparison between optical coherence tomography and intravascular ultrasound for detecting small degrees of in-stent neointima after stent implantation. JACC Cardiovasc Interv 1(2):168–173PubMedCrossRef
5.
go back to reference Tearney GJ, Regar E, Akasaka T et al (2012) Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation. J Am Coll Cardiol 59(12):1058–1072PubMedCrossRef Tearney GJ, Regar E, Akasaka T et al (2012) Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation. J Am Coll Cardiol 59(12):1058–1072PubMedCrossRef
6.
go back to reference Liu L, Gardecki JA, Nadkarni SK, Toussaint JD, Yagi Y, Bouma BE, Tearney GJ (2011) Imaging the subcellular structure of human coronary atherosclerosis using micro-optical coherence tomography. Nat Med 17(8):1010–1014PubMedCentralPubMedCrossRef Liu L, Gardecki JA, Nadkarni SK, Toussaint JD, Yagi Y, Bouma BE, Tearney GJ (2011) Imaging the subcellular structure of human coronary atherosclerosis using micro-optical coherence tomography. Nat Med 17(8):1010–1014PubMedCentralPubMedCrossRef
7.
go back to reference Kim JS, Jang IK, Fan C et al (2009) Evaluation in 3 months duration of neointimal coverage after zotarolimus-eluting stent implantation by optical coherence tomography: the ENDEAVOR OCT trial. JACC Cardiovasc Interv 2(12):1240–1247PubMedCrossRef Kim JS, Jang IK, Fan C et al (2009) Evaluation in 3 months duration of neointimal coverage after zotarolimus-eluting stent implantation by optical coherence tomography: the ENDEAVOR OCT trial. JACC Cardiovasc Interv 2(12):1240–1247PubMedCrossRef
8.
go back to reference Yoon JH, Di Vito L, Moses JW, Fearon WF, Yeung AC, Zhang S, Bezerra HG, Costa MA, Jang IK (2012) Feasibility and safety of the second-generation, frequency domain optical coherence tomography (FD-OCT): a multicenter study. J Invasive Cardiol 24(5):206–209PubMed Yoon JH, Di Vito L, Moses JW, Fearon WF, Yeung AC, Zhang S, Bezerra HG, Costa MA, Jang IK (2012) Feasibility and safety of the second-generation, frequency domain optical coherence tomography (FD-OCT): a multicenter study. J Invasive Cardiol 24(5):206–209PubMed
9.
go back to reference Fedele S, Biondi-Zoccai G, Kwiatkowski P et al (2012) Reproducibility of coronary optical coherence tomography for lumen and length measurements in humans (The CLI-VAR [Centro per la Lotta contro l’Infarto-VARiability] study). Am J Cardiol 110(8):1106–1112PubMedCrossRef Fedele S, Biondi-Zoccai G, Kwiatkowski P et al (2012) Reproducibility of coronary optical coherence tomography for lumen and length measurements in humans (The CLI-VAR [Centro per la Lotta contro l’Infarto-VARiability] study). Am J Cardiol 110(8):1106–1112PubMedCrossRef
10.
go back to reference Takarada S, Imanishi T, Liu Y et al (2010) Advantage of next-generation frequency-domain optical coherence tomography compared with conventional time-domain system in the assessment of coronary lesion. Catheter Cardiovasc Interv 75(2):202–206PubMedCrossRef Takarada S, Imanishi T, Liu Y et al (2010) Advantage of next-generation frequency-domain optical coherence tomography compared with conventional time-domain system in the assessment of coronary lesion. Catheter Cardiovasc Interv 75(2):202–206PubMedCrossRef
11.
go back to reference Gonzalo N, Tearney GJ, Serruys PW et al (2010) Second-generation optical coherence tomography in clinical practice. High-speed data acquisition is highly reproducible in patients undergoing percutaneous coronary intervention. Rev Esp Cardiol 63(8):893–903PubMedCrossRef Gonzalo N, Tearney GJ, Serruys PW et al (2010) Second-generation optical coherence tomography in clinical practice. High-speed data acquisition is highly reproducible in patients undergoing percutaneous coronary intervention. Rev Esp Cardiol 63(8):893–903PubMedCrossRef
12.
go back to reference Alfonso F, Dutary J, Paulo M et al (2012) Combined use of optical coherence tomography and intravascular ultrasound imaging in patients undergoing coronary interventions for stent thrombosis. Heart 98(16):1213–1220PubMedCrossRef Alfonso F, Dutary J, Paulo M et al (2012) Combined use of optical coherence tomography and intravascular ultrasound imaging in patients undergoing coronary interventions for stent thrombosis. Heart 98(16):1213–1220PubMedCrossRef
13.
go back to reference Habara M, Nasu K, Terashima M et al (2012) Impact of frequency-domain optical coherence tomography guidance for optimal coronary stent implantation in comparison with intravascular ultrasound guidance. Circ Cardiovasc Interv 5(2):193–201PubMedCrossRef Habara M, Nasu K, Terashima M et al (2012) Impact of frequency-domain optical coherence tomography guidance for optimal coronary stent implantation in comparison with intravascular ultrasound guidance. Circ Cardiovasc Interv 5(2):193–201PubMedCrossRef
14.
go back to reference Ramesh S, Papayannis A, Abdel-karim AR, Banerjee S, Brilakis E (2012) In vivo comparison of Fourier-domain optical coherence tomography and intravascular ultrasonography. J Invasive Cardiol 24(3):111–115PubMed Ramesh S, Papayannis A, Abdel-karim AR, Banerjee S, Brilakis E (2012) In vivo comparison of Fourier-domain optical coherence tomography and intravascular ultrasonography. J Invasive Cardiol 24(3):111–115PubMed
15.
go back to reference Bezerra HG, Attizzani GF, Sirbu V et al (2013) Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention. JACC Cardiovasc Interv 6(3):228–236PubMedCrossRef Bezerra HG, Attizzani GF, Sirbu V et al (2013) Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention. JACC Cardiovasc Interv 6(3):228–236PubMedCrossRef
16.
go back to reference Imola F, Mallus MT, Ramazzotti V, Manzoli A, Pappalardo A, Di Giorgio A, Albertucci M, Prati F (2010) Safety and feasibility of frequency domain optical coherence tomography to guide decision making in percutaneous coronary intervention. EuroIntervention 6(5):575–581PubMedCrossRef Imola F, Mallus MT, Ramazzotti V, Manzoli A, Pappalardo A, Di Giorgio A, Albertucci M, Prati F (2010) Safety and feasibility of frequency domain optical coherence tomography to guide decision making in percutaneous coronary intervention. EuroIntervention 6(5):575–581PubMedCrossRef
17.
go back to reference Tahara S, Bezerra HG, Baibars M, Kyono H, Wang W, Pokras S, Mehanna E, Petersen CL, Costa MA (2011) In vitro validation of new Fourier-domain optical coherence tomography. EuroIntervention 6(7):875–882PubMedCrossRef Tahara S, Bezerra HG, Baibars M, Kyono H, Wang W, Pokras S, Mehanna E, Petersen CL, Costa MA (2011) In vitro validation of new Fourier-domain optical coherence tomography. EuroIntervention 6(7):875–882PubMedCrossRef
18.
go back to reference Abnousi F, Waseda K, Kume T et al (2013) Variability in quantitative and qualitative analysis of intravascular ultrasound and frequency domain optical coherence tomography. Catheter Cardiovasc Interv 82(3):E192–E199PubMedCrossRef Abnousi F, Waseda K, Kume T et al (2013) Variability in quantitative and qualitative analysis of intravascular ultrasound and frequency domain optical coherence tomography. Catheter Cardiovasc Interv 82(3):E192–E199PubMedCrossRef
19.
go back to reference Jamil Z, Tearney G, Bruining N, Sihan K, van Soest G, Ligthart J, van Domburg R, Bouma B, Regar E (2013) Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection. Int J Cardiovasc Imaging 29(1):39–51PubMedCentralPubMedCrossRef Jamil Z, Tearney G, Bruining N, Sihan K, van Soest G, Ligthart J, van Domburg R, Bouma B, Regar E (2013) Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection. Int J Cardiovasc Imaging 29(1):39–51PubMedCentralPubMedCrossRef
Metadata
Title
Comparison of longitudinal geometric measurement in human coronary arteries between frequency-domain optical coherence tomography and intravascular ultrasound
Authors
Yong Liu
Kunihiro Shimamura
Takashi Kubo
Atsuhi Tanaka
Hironori Kitabata
Yasushi Ino
Takashi Tanimoto
Yasutsugu Shiono
Makoto Orii
Takashi Yamano
Tomoyuki Yamaguchi
Kumiko Hirata
Toshio Imanishi
Takashi Akasaka
Publication date
01-02-2014
Publisher
Springer Netherlands
Published in
The International Journal of Cardiovascular Imaging / Issue 2/2014
Print ISSN: 1569-5794
Electronic ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-013-0330-7

Other articles of this Issue 2/2014

The International Journal of Cardiovascular Imaging 2/2014 Go to the issue