Skip to main content
Top
Published in: The International Journal of Cardiovascular Imaging 5/2012

01-06-2012 | Original paper

Quantitative myocardial perfusion measurement using CT Perfusion: a validation study in a porcine model of reperfused acute myocardial infarction

Authors: Aaron So, Jiang Hsieh, Jian-Ying Li, Jennifer Hadway, Hua-Fu Kong, Ting-Yim Lee

Published in: The International Journal of Cardiovascular Imaging | Issue 5/2012

Login to get access

Abstract

We validated a CT Perfusion technique with beam hardening (BH) correction for quantitative measurement of myocardial blood flow (MBF). Acute myocardial infarction (AMI) was created in four pigs by occluding the distal LAD for 1 h followed by reperfusion. MBF was measured from dynamic contrast enhanced CT (DCE-CT) scanning of the heart, with correction of cardiac motion and BH, before ischemic insult and on day 7, 10 and 14 post. On day 14 post, radiolabeled microspheres were injected to measure MBF and the results were compared with those measured by CT Perfusion. Excised hearts were stained with 2,3,5-triphenyltetrazolium chloride (TTC) to determine the relationship between MBF measured by CT Perfusion and myocardial viability. MBF measured by CT Perfusion was strongly correlated with that by microspheres over a wide range of MBF values (R = 0.81, from 25 to 225 ml min−1 100 g−1). While MBF in the LAD territory decreased significantly from 98.4 ± 2.5 ml min−1 100 g−1 at baseline to 32.2 ± 9.1 ml min−1 100 g−1, P < 0.05 at day 7 and to 49.4 ± 9.3 ml min−1 100 g−1, P < 0.05 at day 14, the decrease in remote myocardium (LCx territory) from baseline (103.9 ± 1.9 ml min−1 100 g−1) was minimal throughout the study (90.6 ± 5.1 ml min−1 100 g−1 on day 14 post, P > 0.05). TTC staining confirmed incomplete infarction in the LAD territory and no infarction in the LCx territory. Microvascular obstruction in infarcted tissue resulted in no-reflow and hence persistently low MBF in the reperfused LAD territory which contained a mixture of viable and non-viable tissue. CT Perfusion measurement of MBF was accurate and correlated well with histology and microspheres measurements.
Literature
1.
go back to reference Miller JM, Rochitte CE, Dewey M et al (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359(22):2324–2336PubMedCrossRef Miller JM, Rochitte CE, Dewey M et al (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359(22):2324–2336PubMedCrossRef
2.
go back to reference George RT, Silva C, Cordeiro MA et al (2006) Multidetector computed tomography myocardial perfusion imaging during adenosine stress. J Am Coll Cardiol 48(1):153–160PubMedCrossRef George RT, Silva C, Cordeiro MA et al (2006) Multidetector computed tomography myocardial perfusion imaging during adenosine stress. J Am Coll Cardiol 48(1):153–160PubMedCrossRef
3.
go back to reference George RT, Jerosch-Herold M, Silva C et al (2007) Quantification of myocardial perfusion using dynamic 64-detector computed tomography. Invest Radiol 42(12):815–822PubMedCrossRef George RT, Jerosch-Herold M, Silva C et al (2007) Quantification of myocardial perfusion using dynamic 64-detector computed tomography. Invest Radiol 42(12):815–822PubMedCrossRef
4.
go back to reference Blankstein R, Shturman LD, Rogers IS et al (2009) Adenosine-induced stress myocardial perfusion imaging using dual-source cardiac computed tomography. J Am Coll Cardiol 54(12):1072–1084PubMedCrossRef Blankstein R, Shturman LD, Rogers IS et al (2009) Adenosine-induced stress myocardial perfusion imaging using dual-source cardiac computed tomography. J Am Coll Cardiol 54(12):1072–1084PubMedCrossRef
5.
go back to reference Bastarrika G, Ramos-Duran L, Rosenblum MA et al (2010) Adenosine-stress dynamic myocardial CT perfusion imaging: initial clinical experience. Invest Radiol 45(6):306–313PubMed Bastarrika G, Ramos-Duran L, Rosenblum MA et al (2010) Adenosine-stress dynamic myocardial CT perfusion imaging: initial clinical experience. Invest Radiol 45(6):306–313PubMed
6.
go back to reference Mahnken AH, Klotz E, Pietsch H et al (2010) Quantitative whole heart stress perfusion CT imaging as noninvasive assessment of hemodynamics in coronary artery stenosis: preliminary animal experience. Invest Radiol 45(6):298–305PubMed Mahnken AH, Klotz E, Pietsch H et al (2010) Quantitative whole heart stress perfusion CT imaging as noninvasive assessment of hemodynamics in coronary artery stenosis: preliminary animal experience. Invest Radiol 45(6):298–305PubMed
7.
go back to reference Ho KT, Chua KC, Klotz E et al (2010) Stress and rest dynamic myocardial perfusion imaging by evaluation of complete time-attenuation curves with dual-source CT. JACC Cardiovasc Imaging 3(8):811–820PubMedCrossRef Ho KT, Chua KC, Klotz E et al (2010) Stress and rest dynamic myocardial perfusion imaging by evaluation of complete time-attenuation curves with dual-source CT. JACC Cardiovasc Imaging 3(8):811–820PubMedCrossRef
8.
go back to reference Lee TY (2002) Functional CT: physiological models. Trends Biotechnol 20(Suppl):S3–S10CrossRef Lee TY (2002) Functional CT: physiological models. Trends Biotechnol 20(Suppl):S3–S10CrossRef
9.
go back to reference Lee TY, Purdie TG, Stewart E (2003) CT imaging of angiogenesis. Q J Nucl Med 47(3):171–187PubMed Lee TY, Purdie TG, Stewart E (2003) CT imaging of angiogenesis. Q J Nucl Med 47(3):171–187PubMed
10.
go back to reference So A, Hsieh J, Li JY et al (2009) Beam hardening correction in CT myocardial perfusion measurement. Phys Med Biol 54(10):3031–3050PubMedCrossRef So A, Hsieh J, Li JY et al (2009) Beam hardening correction in CT myocardial perfusion measurement. Phys Med Biol 54(10):3031–3050PubMedCrossRef
11.
go back to reference So A, Adam V, Acharya K et al (2003) Manual and semi-automatic segmentation versus retrospective ECG gating for correction of cardiac motion. SPIE Med Imaging Proc 5031:493–503 So A, Adam V, Acharya K et al (2003) Manual and semi-automatic segmentation versus retrospective ECG gating for correction of cardiac motion. SPIE Med Imaging Proc 5031:493–503
12.
go back to reference Cenic A, Nabavi DG, Craen RA et al (2000) A CT method to measure hemodynamics in brain tumors: validation and application of cerebral blood flow maps. AJNR Am J Neuroradiol 21(3):462–470PubMed Cenic A, Nabavi DG, Craen RA et al (2000) A CT method to measure hemodynamics in brain tumors: validation and application of cerebral blood flow maps. AJNR Am J Neuroradiol 21(3):462–470PubMed
13.
go back to reference Murphy BD, Fox AJ, Lee DH et al (2008) White matter thresholds for ischemic penumbra and infarct core in patients with acute stroke: CT perfusion study. Radiology 247(3):818–825PubMedCrossRef Murphy BD, Fox AJ, Lee DH et al (2008) White matter thresholds for ischemic penumbra and infarct core in patients with acute stroke: CT perfusion study. Radiology 247(3):818–825PubMedCrossRef
14.
go back to reference Meier P, Zierler KL (1954) On the theory of the indicator-dilution method for measurement of blood flow and volume. J Appl Physiol 6(12):731–744PubMed Meier P, Zierler KL (1954) On the theory of the indicator-dilution method for measurement of blood flow and volume. J Appl Physiol 6(12):731–744PubMed
15.
go back to reference Bassingthwaighte JB, Knopp TJ, Anderson DU (1970) Flow estimation by indicator dilution (bolus injection). Circ Res 27(2):277–291PubMedCrossRef Bassingthwaighte JB, Knopp TJ, Anderson DU (1970) Flow estimation by indicator dilution (bolus injection). Circ Res 27(2):277–291PubMedCrossRef
16.
go back to reference Johnson JA, Wilson TA (1966) A model for capillary exchange. Am J Physiol 210(6):1299–1303PubMed Johnson JA, Wilson TA (1966) A model for capillary exchange. Am J Physiol 210(6):1299–1303PubMed
17.
go back to reference St Lawrence KS, Lee TY (1998) An adiabatic approximation to the tissue homogeneity model for water exchange in the brain: I. Theoretical derivation. J Cereb Blood Flow Metab 18(12):1365–1377PubMedCrossRef St Lawrence KS, Lee TY (1998) An adiabatic approximation to the tissue homogeneity model for water exchange in the brain: I. Theoretical derivation. J Cereb Blood Flow Metab 18(12):1365–1377PubMedCrossRef
18.
go back to reference Alexander CS, Liu SM (1976) Effect of hypercapnia and hypocapnia on myocardial blood flow and performance in anaesthetized dogs. Cardiovasc Res 10(3):341–348PubMedCrossRef Alexander CS, Liu SM (1976) Effect of hypercapnia and hypocapnia on myocardial blood flow and performance in anaesthetized dogs. Cardiovasc Res 10(3):341–348PubMedCrossRef
19.
go back to reference Eliades D, Weiss HR (1986) Effect of hypercapnia on coronary circulation. Cardiovasc Res 20(2):127–133PubMedCrossRef Eliades D, Weiss HR (1986) Effect of hypercapnia on coronary circulation. Cardiovasc Res 20(2):127–133PubMedCrossRef
20.
go back to reference Heymann MA, Payne BD, Hoffman JI et al (1977) Blood flow measurements with radionuclide-labeled particles. Prog Cardiovasc Dis 20(1):55–79PubMedCrossRef Heymann MA, Payne BD, Hoffman JI et al (1977) Blood flow measurements with radionuclide-labeled particles. Prog Cardiovasc Dis 20(1):55–79PubMedCrossRef
21.
go back to reference Cerqueira MD, Weissman NJ, Dilsizian V et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the cardiac imaging committee of the council on clinical cardiology of the American heart association. Circulation 105(4):539–542PubMedCrossRef Cerqueira MD, Weissman NJ, Dilsizian V et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the cardiac imaging committee of the council on clinical cardiology of the American heart association. Circulation 105(4):539–542PubMedCrossRef
22.
go back to reference Geleijnse ML, Fioretti PM, Roelandt JR (1997) Methodology, feasibility, safety and diagnostic accuracy of dobutamine stress echocardiography. J Am Coll Cardiol 30(3):595–606PubMedCrossRef Geleijnse ML, Fioretti PM, Roelandt JR (1997) Methodology, feasibility, safety and diagnostic accuracy of dobutamine stress echocardiography. J Am Coll Cardiol 30(3):595–606PubMedCrossRef
23.
go back to reference Ambrosio G, Weisman HF, Mannisi JA et al (1989) Progressive impairment of regional myocardial perfusion after initial restoration of postischemic blood flow. Circulation 80(6):1846–1861PubMedCrossRef Ambrosio G, Weisman HF, Mannisi JA et al (1989) Progressive impairment of regional myocardial perfusion after initial restoration of postischemic blood flow. Circulation 80(6):1846–1861PubMedCrossRef
24.
go back to reference Kloner RA, Ganote CE, Jennings RB (1974) The “no-reflow” phenomenon after temporary coronary occlusion in the dog. J Clin Invest 54(6):1496–1508PubMedCrossRef Kloner RA, Ganote CE, Jennings RB (1974) The “no-reflow” phenomenon after temporary coronary occlusion in the dog. J Clin Invest 54(6):1496–1508PubMedCrossRef
25.
go back to reference Kloner RA, Rude RE, Carlson N et al (1980) Ultrastructural evidence of microvascular damage and myocardial cell injury after coronary artery occlusion: which comes first? Circulation 62(5):945–952PubMedCrossRef Kloner RA, Rude RE, Carlson N et al (1980) Ultrastructural evidence of microvascular damage and myocardial cell injury after coronary artery occlusion: which comes first? Circulation 62(5):945–952PubMedCrossRef
26.
go back to reference Manciet LH, Poole DC, McDonagh PF et al (1994) Microvascular compression during myocardial ischemia: mechanistic basis for no-reflow phenomenon. Am J Physiol 266(4 Pt 2):H1541–H1550PubMed Manciet LH, Poole DC, McDonagh PF et al (1994) Microvascular compression during myocardial ischemia: mechanistic basis for no-reflow phenomenon. Am J Physiol 266(4 Pt 2):H1541–H1550PubMed
27.
go back to reference Bolli R, Triana JF, Jeroudi MO (1990) Prolonged impairment of coronary vasodilation after reversible ischemia. Evidence for microvascular “stunning”. Circ Res 67(2):332–343PubMedCrossRef Bolli R, Triana JF, Jeroudi MO (1990) Prolonged impairment of coronary vasodilation after reversible ischemia. Evidence for microvascular “stunning”. Circ Res 67(2):332–343PubMedCrossRef
28.
go back to reference Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301(5):500–507PubMedCrossRef Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301(5):500–507PubMedCrossRef
29.
go back to reference Thibault JB, Sauer KD, Bouman CA et al (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34(11):4526–4544PubMedCrossRef Thibault JB, Sauer KD, Bouman CA et al (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34(11):4526–4544PubMedCrossRef
30.
go back to reference Alvarez RE, Macovski A (1976) Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol 21(5):733–744PubMedCrossRef Alvarez RE, Macovski A (1976) Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol 21(5):733–744PubMedCrossRef
31.
go back to reference Kalender WA, Perman WH, Vetter JR et al (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys 13(3):334–339PubMedCrossRef Kalender WA, Perman WH, Vetter JR et al (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys 13(3):334–339PubMedCrossRef
32.
go back to reference Francone M, Carbone I, Danti M et al (2006) ECG-gated multi-detector row spiral CT in the assessment of myocardial infarction: correlation with non-invasive angiographic findings. Eur Radiol 16(1):15–24PubMedCrossRef Francone M, Carbone I, Danti M et al (2006) ECG-gated multi-detector row spiral CT in the assessment of myocardial infarction: correlation with non-invasive angiographic findings. Eur Radiol 16(1):15–24PubMedCrossRef
33.
go back to reference Bassingthwaighte JB, Wang CY, Chan IS (1989) Blood-tissue exchange via transport and transformation by capillary endothelial cells. Circ Res 65(4):997–1020PubMedCrossRef Bassingthwaighte JB, Wang CY, Chan IS (1989) Blood-tissue exchange via transport and transformation by capillary endothelial cells. Circ Res 65(4):997–1020PubMedCrossRef
35.
go back to reference Schmitt M, Horstick G, Petersen SE et al (2005) Quantification of resting myocardial blood flow in a pig model of acute ischemia based on first-pass MRI. Magn Reson Med 53(5):1223–1227PubMedCrossRef Schmitt M, Horstick G, Petersen SE et al (2005) Quantification of resting myocardial blood flow in a pig model of acute ischemia based on first-pass MRI. Magn Reson Med 53(5):1223–1227PubMedCrossRef
36.
go back to reference Kroll K, Wilke N, Jerosch-Herold M et al (1996) Modeling regional myocardial flows from residue functions of an intravascular indicator. Am J Physiol 271(4 Pt 2):H1643–H1655PubMed Kroll K, Wilke N, Jerosch-Herold M et al (1996) Modeling regional myocardial flows from residue functions of an intravascular indicator. Am J Physiol 271(4 Pt 2):H1643–H1655PubMed
37.
go back to reference Jerosch-Herold M, Wilke N, Stillman AE (1998) Magnetic resonance quantification of the myocardial perfusion reserve with a fermi function model for constrained deconvolution. Med Phys 25(1):73–84PubMedCrossRef Jerosch-Herold M, Wilke N, Stillman AE (1998) Magnetic resonance quantification of the myocardial perfusion reserve with a fermi function model for constrained deconvolution. Med Phys 25(1):73–84PubMedCrossRef
Metadata
Title
Quantitative myocardial perfusion measurement using CT Perfusion: a validation study in a porcine model of reperfused acute myocardial infarction
Authors
Aaron So
Jiang Hsieh
Jian-Ying Li
Jennifer Hadway
Hua-Fu Kong
Ting-Yim Lee
Publication date
01-06-2012
Publisher
Springer Netherlands
Published in
The International Journal of Cardiovascular Imaging / Issue 5/2012
Print ISSN: 1569-5794
Electronic ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-011-9927-x

Other articles of this Issue 5/2012

The International Journal of Cardiovascular Imaging 5/2012 Go to the issue