Published in: Clinical Oral Investigations 8/2018

Open Access 01-11-2018 | Original Article

CAD/CAM produces dentures with improved fit

Authors: Otto Steinmassl, Herbert Dumfahrt, Ingrid Grunert, Patricia-Anca Steinmassl

Published in: Clinical Oral Investigations | Issue 8/2018

Login to get access

Abstract

Objectives

Resin polymerisation shrinkage reduces the congruence of the denture base with denture-bearing tissues and thereby decreases the retention of conventionally fabricated dentures. CAD/CAM denture manufacturing is a subtractive process, and polymerisation shrinkage is not an issue anymore. Therefore, CAD/CAM dentures are assumed to show a higher denture base congruence than conventionally fabricated dentures. It has been the aim of this study to test this hypothesis.

Materials and methods

CAD/CAM dentures provided by four different manufacturers (AvaDent, Merz Dental, Whole You, Wieland/Ivoclar) were generated from ten different master casts. Ten conventional dentures (pack and press, long-term heat polymerisation) made from the same master casts served as control group. The master casts and all denture bases were scanned and matched digitally. The absolute incongruences were measured using a 2-mm mesh.

Results

Conventionally fabricated dentures showed a mean deviation of 0.105 mm, SD = 0.019 from the master cast. All CAD/CAM dentures showed lower mean incongruences. From all CAD/CAM dentures, AvaDent Digital Dentures showed the highest congruence with the master cast surface with a mean deviation of 0.058 mm, SD = 0.005. Wieland Digital Dentures showed a mean deviation of 0.068 mm, SD = 0.005, Whole You Nexteeth prostheses showed a mean deviation of 0.074 mm, SD = 0.011 and Baltic Denture System prostheses showed a mean deviation of 0.086 mm, SD = 0.012.

Conclusions

CAD/CAM produces dentures with better fit than conventional dentures.

Clinical Relevance

The present study explains the clinically observed enhanced retention and lower traumatic ulcer-frequency in CAD/CAM dentures.
Literature
4.
go back to reference Darvell BW, Clark RK (2000) The physical mechanisms of complete denture retention. Br Dent J 189(5):248–252PubMed Darvell BW, Clark RK (2000) The physical mechanisms of complete denture retention. Br Dent J 189(5):248–252PubMed
6.
go back to reference Murray MD, Darvell BW (1989) Reappraisal of the physics of denture retention. Int J Prosthodont 2(3):234–242PubMed Murray MD, Darvell BW (1989) Reappraisal of the physics of denture retention. Int J Prosthodont 2(3):234–242PubMed
12.
go back to reference Kattadiyil MT, Goodacre CJ, Baba NZ (2013) CAD/CAM complete dentures: a review of two commercial fabrication systems. J Calif Dent Assoc 41(6):407–416PubMed Kattadiyil MT, Goodacre CJ, Baba NZ (2013) CAD/CAM complete dentures: a review of two commercial fabrication systems. J Calif Dent Assoc 41(6):407–416PubMed
18.
go back to reference Ender A, Mehl A (2013) Influence of scanning strategies on the accuracy of digital intraoral scanning systems. Int J Comput Dent 16(1):11–21PubMed Ender A, Mehl A (2013) Influence of scanning strategies on the accuracy of digital intraoral scanning systems. Int J Comput Dent 16(1):11–21PubMed
Metadata
Title
CAD/CAM produces dentures with improved fit
Authors
Otto Steinmassl
Herbert Dumfahrt
Ingrid Grunert
Patricia-Anca Steinmassl
Publication date
01-11-2018
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 8/2018
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-018-2369-2

Other articles of this Issue 8/2018

Clinical Oral Investigations 8/2018 Go to the issue