Published in: Clinical Oral Investigations 1/2014

01-01-2014 | Original Article

Comparison of intraoral radiography and cone-beam computed tomography for the detection of horizontal root fractures: an in vitro study

Authors: Hakan Avsever, Kaan Gunduz, Kaan Orhan, İsmail Uzun, Bilal Ozmen, Erol Egrioglu, Muhammed Midilli

Published in: Clinical Oral Investigations | Issue 1/2014

Login to get access

Abstract

Objectives

This study aimed to compare the diagnostic accuracy of two different cone-beam computed tomography (CBCT) units with several intraoral radiography techniques for detecting horizontal root fractures.

Methods

The study material comprised 82 extracted human maxillary incisors without root fractures that had not undergone any root canal treatment. Root fractures were created in the horizontal plane in 31 teeth by a mechanical force using a hammer with the tooth placed on a soft foundation as described in a previous study. The teeth were divided into two groups: a control group with no fractures and a test group with fractures. These were randomized to the empty maxillary anterior sockets of a dry human maxilla. Each tooth was imaged at various vertical angles using each of the following modalities: a 3D Accuitomo 170 CBCT, a NewTom 3G CBCT, a VistaScan PSP, a CCD sensor, and conventional film. Specificity and sensitivity for assessing horizontal root fracture by each radiographic technique were calculated. Chi-square statistics were used to evaluate differences between modalities. Kappa statistics assessed the agreement between observers. Results were considered significant at P < 0.05.

Results

The kappa values for inter-observer agreement between observers ranged between 0.88 and 0.98 for the 3D Accuitomo 170, 0.82 and 0.91 for the NewTom 3G, and 0.61 and 0.72 for the different types of intraoral images. The diagnostic accuracy for detecting fracture lines in 3D Accuitomo 170 (0.93) was significantly higher than NewTom 3G (0.87), VistaScan (0.71), CCD (0.70), and CF (0.68).

Conclusions

3D Accuitomo 170 has the highest sensitivity and diagnostic accuracy for detecting horizontal root fracture among the 5 radiographic modalities examined. CBCT should be considered as the most reliable imaging modality of choice for the diagnosis of horizontal root fracture.

Clinical relevance

CBCT imaging offers the clear advantage over conventional imaging that traumatized teeth can be visualized in all three dimensions—especially the oro-facial dimension
Literature
1.
go back to reference Bornstein MM, Wölner-Hanssen AB, Sendi P, von Arx T (2009) Comparison of intraoral radiography and limited cone beam computed tomography for the assessment of root-fractured permanent teeth. Dent Traumatol 25:571–577PubMedCrossRef Bornstein MM, Wölner-Hanssen AB, Sendi P, von Arx T (2009) Comparison of intraoral radiography and limited cone beam computed tomography for the assessment of root-fractured permanent teeth. Dent Traumatol 25:571–577PubMedCrossRef
2.
go back to reference Caliskan MK, Pehlivan Y (1996) Prognosis of root-fractured permanent incisors. Endod Dent Traumatol 12:129–136PubMedCrossRef Caliskan MK, Pehlivan Y (1996) Prognosis of root-fractured permanent incisors. Endod Dent Traumatol 12:129–136PubMedCrossRef
3.
go back to reference Kamburoğlu K, Ilker Cebeci AR, Gröndahl HG (2009) Effectiveness of limited cone-beam computed tomography in the detection of horizontal root fracture. Dent Traumatol 25:256–261PubMedCrossRef Kamburoğlu K, Ilker Cebeci AR, Gröndahl HG (2009) Effectiveness of limited cone-beam computed tomography in the detection of horizontal root fracture. Dent Traumatol 25:256–261PubMedCrossRef
4.
go back to reference Hovland EJ (1992) Horizontal root fractures. Treatment and repair. Dent Clin North Am 36:509–525PubMed Hovland EJ (1992) Horizontal root fractures. Treatment and repair. Dent Clin North Am 36:509–525PubMed
5.
go back to reference Poi WR, Manfrin TM, Holland R, Sonoda CK (2002) Repair characteristics of horizontal root fracture: a case report. Dent Traumatol 18:98–102PubMedCrossRef Poi WR, Manfrin TM, Holland R, Sonoda CK (2002) Repair characteristics of horizontal root fracture: a case report. Dent Traumatol 18:98–102PubMedCrossRef
6.
go back to reference Andreasen FM, Andreasen JO, Bayer T (1989) Prognosis of root fractured permanent incisors—prediction of healing modalities. Endod Dent Traumatol 5:11–22PubMedCrossRef Andreasen FM, Andreasen JO, Bayer T (1989) Prognosis of root fractured permanent incisors—prediction of healing modalities. Endod Dent Traumatol 5:11–22PubMedCrossRef
7.
go back to reference Cvek M, Andreasen JO, Borum MK (2001) Healing of 208 intraalveolar root-fractures in patients aged 7–17 years. Dent Traumatol 17:53–62PubMedCrossRef Cvek M, Andreasen JO, Borum MK (2001) Healing of 208 intraalveolar root-fractures in patients aged 7–17 years. Dent Traumatol 17:53–62PubMedCrossRef
8.
go back to reference Cvek M, Tsilingaridis G, Andreasen JO (2008) Survival of 534 incisors after intra-alveolar root fracture in patients aged 7–17 years. Dent Traumatol 24:379–387PubMedCrossRef Cvek M, Tsilingaridis G, Andreasen JO (2008) Survival of 534 incisors after intra-alveolar root fracture in patients aged 7–17 years. Dent Traumatol 24:379–387PubMedCrossRef
9.
go back to reference Kositbowornchai S, Nuansakul R, Sikram S, Sinahawattana S, Saengmontri S (2001) Root fracture detection: a comparison of direct digital radiography with conventional radiography. Dentomaxillofac Radiol 30:106–109PubMedCrossRef Kositbowornchai S, Nuansakul R, Sikram S, Sinahawattana S, Saengmontri S (2001) Root fracture detection: a comparison of direct digital radiography with conventional radiography. Dentomaxillofac Radiol 30:106–109PubMedCrossRef
10.
go back to reference Borelli P, Alibrandi P (1999) Unusual horizontal and vertical root fractures of maxillary molars: an 11-year follow-up. J Endod 25:136–139PubMedCrossRef Borelli P, Alibrandi P (1999) Unusual horizontal and vertical root fractures of maxillary molars: an 11-year follow-up. J Endod 25:136–139PubMedCrossRef
11.
go back to reference Flores MT, Andersson L, Andreasen JO, Bakland LK, Malmgren B, Barnett F et al (2007) Guidelines for the management of traumatic dental injuries. I Fractures and luxations of permanent teeth. Dent Traumatol 23:66–71PubMedCrossRef Flores MT, Andersson L, Andreasen JO, Bakland LK, Malmgren B, Barnett F et al (2007) Guidelines for the management of traumatic dental injuries. I Fractures and luxations of permanent teeth. Dent Traumatol 23:66–71PubMedCrossRef
12.
go back to reference Andreasen FM, Andreasen JO (1994) Root fractures. In: Andreasen JO, Andreasen FM (eds) Textbook and color atlas of traumatic injuries to the teeth, 3rd edn. Munsksgaard, Copenhagen, pp 279–311 Andreasen FM, Andreasen JO (1994) Root fractures. In: Andreasen JO, Andreasen FM (eds) Textbook and color atlas of traumatic injuries to the teeth, 3rd edn. Munsksgaard, Copenhagen, pp 279–311
13.
go back to reference Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72:75–80PubMed Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72:75–80PubMed
14.
15.
go back to reference White SC, Pharoah MJ (2008) The evolution and application of dental maxillofacial imaging modalities. Dent Clin North Am 52:689–705PubMedCrossRef White SC, Pharoah MJ (2008) The evolution and application of dental maxillofacial imaging modalities. Dent Clin North Am 52:689–705PubMedCrossRef
16.
go back to reference Patel S, Dawood A, Whaites E, Pitt Ford T (2009) New dimensions in endodontic imaging: part 1. Conventional and alternative radiographic systems. Int Endod J 42:447–462PubMedCrossRef Patel S, Dawood A, Whaites E, Pitt Ford T (2009) New dimensions in endodontic imaging: part 1. Conventional and alternative radiographic systems. Int Endod J 42:447–462PubMedCrossRef
17.
go back to reference Gunduz K, Avsever H, Orhan K, Çelenk P, Ozmen B et al (2013) Comparison of intraoral radiography and cone-beam computed tomography for the detection of vertical root fractures: an in vitro study. Oral Radiology 1:6–12CrossRef Gunduz K, Avsever H, Orhan K, Çelenk P, Ozmen B et al (2013) Comparison of intraoral radiography and cone-beam computed tomography for the detection of vertical root fractures: an in vitro study. Oral Radiology 1:6–12CrossRef
18.
go back to reference Tanimoto H, Arai Y (2009) The effect of voxel size on image reconstruction in cone-beam computed tomography. Oral Radiol 25:149–153CrossRef Tanimoto H, Arai Y (2009) The effect of voxel size on image reconstruction in cone-beam computed tomography. Oral Radiol 25:149–153CrossRef
19.
go back to reference Wenzel A, Kirkevang LL (2005) High resolution charge-coupled device sensor vs medium resolution photostimulable phosphor plate digital receptors for detection of root fractures in vitro. Dent Traumatol 21:32–36PubMedCrossRef Wenzel A, Kirkevang LL (2005) High resolution charge-coupled device sensor vs medium resolution photostimulable phosphor plate digital receptors for detection of root fractures in vitro. Dent Traumatol 21:32–36PubMedCrossRef
20.
go back to reference Majorana A, Pasini S, Bardellini E, Keller E (2002) Clinical and epidemiological study of traumatic root fractures. Dent Traumatol 18:77–80PubMedCrossRef Majorana A, Pasini S, Bardellini E, Keller E (2002) Clinical and epidemiological study of traumatic root fractures. Dent Traumatol 18:77–80PubMedCrossRef
21.
go back to reference Hashimoto K, Kawashima S, Kameoka S, Akiyama Y, Honjoya T, Ejima K et al (2005) High resolution charge-coupled device sensor vs. medium resolution photostimulable phosphor plate digital receptors for detection of root fractures in vitro. Dent Traumatol 21:32–36CrossRef Hashimoto K, Kawashima S, Kameoka S, Akiyama Y, Honjoya T, Ejima K et al (2005) High resolution charge-coupled device sensor vs. medium resolution photostimulable phosphor plate digital receptors for detection of root fractures in vitro. Dent Traumatol 21:32–36CrossRef
22.
go back to reference Kamburoğlu K, Murat S, Yüksel SP, Cebeci AR, Horasan S (2010) Detection of vertical root fracture using cone-beam computerized tomography: an in vitro assessment. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 109:e74–e81PubMedCrossRef Kamburoğlu K, Murat S, Yüksel SP, Cebeci AR, Horasan S (2010) Detection of vertical root fracture using cone-beam computerized tomography: an in vitro assessment. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 109:e74–e81PubMedCrossRef
23.
go back to reference Costa FF, Gaia BF, Umetsubo OS, Cavalcanti MG (2011) Detection of horizontal root fracture with small-volume cone-beam computed tomography in the presence and absence of intracanal metallic post. J Endod 37:1456–1459PubMedCrossRef Costa FF, Gaia BF, Umetsubo OS, Cavalcanti MG (2011) Detection of horizontal root fracture with small-volume cone-beam computed tomography in the presence and absence of intracanal metallic post. J Endod 37:1456–1459PubMedCrossRef
24.
go back to reference Costa FF, Gaia BF, Umetsubo OS, Pinheiro LR, Tortamano IP, Cavalcanti MG (2012) Use of large-volume cone-beam computed tomography in identification and localization of horizontal root fracture in the presence and absence of intracanal metallic post. J Endod 38:856–859. doi:10.1016/j.joen.2012.03.011 PubMedCrossRef Costa FF, Gaia BF, Umetsubo OS, Pinheiro LR, Tortamano IP, Cavalcanti MG (2012) Use of large-volume cone-beam computed tomography in identification and localization of horizontal root fracture in the presence and absence of intracanal metallic post. J Endod 38:856–859. doi:10.​1016/​j.​joen.​2012.​03.​011 PubMedCrossRef
26.
go back to reference Iikubo M, Kobayashi K, Mishima A, Shimoda S, Daimaruya T, Igarashi C, Imanaka M, Yuasa M, Sakamoto M, Sasano T (2009) Accuracy of intraoral radiography, multidetector helical CT, and limited cone-beam CT for the detection of horizontal tooth root fracture. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 108:e70–e74PubMedCrossRef Iikubo M, Kobayashi K, Mishima A, Shimoda S, Daimaruya T, Igarashi C, Imanaka M, Yuasa M, Sakamoto M, Sasano T (2009) Accuracy of intraoral radiography, multidetector helical CT, and limited cone-beam CT for the detection of horizontal tooth root fracture. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 108:e70–e74PubMedCrossRef
27.
go back to reference Al-Ekrish AA, Ekram MI, Al Faleh W, Al-Khader M, Al-Sadhan R (2012) The validity of different display monitors in the assessment of dental implant site dimensions in cone beam computed tomography images. Acta Odontol Scand. Nov 21. [Epub ahead of print] Al-Ekrish AA, Ekram MI, Al Faleh W, Al-Khader M, Al-Sadhan R (2012) The validity of different display monitors in the assessment of dental implant site dimensions in cone beam computed tomography images. Acta Odontol Scand. Nov 21. [Epub ahead of print]
28.
go back to reference Ilgüy M, Dinçer S, Ilgüy D, Bayirli G (2009) Detection of artificial occlusal caries in a phosphor imaging plate system with two types of LCD monitors versus three different films. J Digit Imaging 22:242–249PubMedCentralPubMedCrossRef Ilgüy M, Dinçer S, Ilgüy D, Bayirli G (2009) Detection of artificial occlusal caries in a phosphor imaging plate system with two types of LCD monitors versus three different films. J Digit Imaging 22:242–249PubMedCentralPubMedCrossRef
29.
go back to reference Kutcher M, Kalathingal S, Ludlow J, Abreu M, Platin E (2006) The effect of lighting conditions on caries interpretation with a laptop computer in a clinical setting. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 102:537–543PubMedCrossRef Kutcher M, Kalathingal S, Ludlow J, Abreu M, Platin E (2006) The effect of lighting conditions on caries interpretation with a laptop computer in a clinical setting. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 102:537–543PubMedCrossRef
30.
go back to reference Hellen-Halme K, Petersson A, Warfvinge G, Nilsson M (2008) Effect of ambient light and monitor brightness and contrast settings on the detection of approximal caries in digital radiographs: an in vitro study. Dentomaxillofac Radiol 37:380–384PubMedCrossRef Hellen-Halme K, Petersson A, Warfvinge G, Nilsson M (2008) Effect of ambient light and monitor brightness and contrast settings on the detection of approximal caries in digital radiographs: an in vitro study. Dentomaxillofac Radiol 37:380–384PubMedCrossRef
31.
go back to reference Lofthag-Hansen S, Thilander-Klang A, Ekestubbe A, Helmrot E, Grondahl K (2008) Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD. Dentomaxillofac Radiol 37:72–79PubMedCrossRef Lofthag-Hansen S, Thilander-Klang A, Ekestubbe A, Helmrot E, Grondahl K (2008) Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD. Dentomaxillofac Radiol 37:72–79PubMedCrossRef
32.
go back to reference Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol 32:229–234PubMedCrossRef Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol 32:229–234PubMedCrossRef
33.
go back to reference National Radiological Protection Board. Occupational, Public and Medical Exposure. Documents of the NRPB. Didcot, UK: NRPB, 1993; 4. National Radiological Protection Board. Occupational, Public and Medical Exposure. Documents of the NRPB. Didcot, UK: NRPB, 1993; 4.
Metadata
Title
Comparison of intraoral radiography and cone-beam computed tomography for the detection of horizontal root fractures: an in vitro study
Authors
Hakan Avsever
Kaan Gunduz
Kaan Orhan
İsmail Uzun
Bilal Ozmen
Erol Egrioglu
Muhammed Midilli
Publication date
01-01-2014
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 1/2014
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-013-0940-4

Other articles of this Issue 1/2014

Clinical Oral Investigations 1/2014 Go to the issue