Published in: Clinical Oral Investigations 3/2009

01-09-2009 | Original Article

Temperature rise and degree of photopolymerization conversion of nanocomposites and conventional dental composites

Authors: Mohammad Atai, Fariba Motevasselian

Published in: Clinical Oral Investigations | Issue 3/2009

Login to get access

Abstract

The aim of the study was to investigate the temperature rise of a nanocomposite and a conventional hybrid dental composite during photopolymerization when cured with halogen curing lamp (QHT) and light-emitting diode (LED). Temperature rise during photopolymerization of two commercially available composites (Filtek Supreme® and TetricCeram®) were measured using a K-type thermocouple and a digital thermometer. Different curing modes were utilized to cure the composites: a high-intensity QHT unit (Optilux 501) in two different modes (standard and ramp), a low-intensity QHT unit (Coltolux 50), and an LED unit (Ultralume-2). Total temperature rise, polymerization reaction exotherm, and irradiation-induced temperature rise of the composites were determined. Degree of conversion of the specimens was measured using FTIR spectroscopy. The results revealed that the Filtek Supreme® nanocomposite showed lower temperature rise and degree of conversion in comparison with the hybrid composite (p < 0.05). It was also found that the LED curing unit induced considerable total and irradiation temperature rise without any improvement in the degree of conversion. Ramp curing mode showed lower temperature rise and delayed gel point and was found to be more effective than QHT standard mode and LED units. Although it is claimed that the LED curing units exhibit lower temperature rise during the photopolymerization, the present study showed that the curing units have no advantage over the conventional QHT units regarding the temperature rise and degree of polymerization conversion.
Literature
1.
go back to reference Althoff O, Hartung M (2000) Advances in light curing. Am J Dent 13(special issue):77D–81DPubMed Althoff O, Hartung M (2000) Advances in light curing. Am J Dent 13(special issue):77D–81DPubMed
2.
go back to reference Altintas SH, Yondem I, Tak O, Usumez A (2008) Temperature rise during polymerization of three different provisional materials. Clin Oral Investig 12(3):283–286PubMedCrossRef Altintas SH, Yondem I, Tak O, Usumez A (2008) Temperature rise during polymerization of three different provisional materials. Clin Oral Investig 12(3):283–286PubMedCrossRef
3.
go back to reference Andrzejewska E (2001) Photopolymerization kinetics of multifunctional monomers. Prog Polym Sci 26(4):605–665CrossRef Andrzejewska E (2001) Photopolymerization kinetics of multifunctional monomers. Prog Polym Sci 26(4):605–665CrossRef
4.
go back to reference Anseth KS, Wang CM, Bowman CN (1994) Kinetic evidence of reaction diffusion during the polymerization of multi (meth) acrylate monomers. Macromolecules 27:650–655CrossRef Anseth KS, Wang CM, Bowman CN (1994) Kinetic evidence of reaction diffusion during the polymerization of multi (meth) acrylate monomers. Macromolecules 27:650–655CrossRef
5.
go back to reference Asmussen E, Peutzfeldt A (2005) Temperature rise induced by some light emitting diode and quartz–tungsten–halogen curing units. Eur J Oral Sci 113(1):96–98PubMedCrossRef Asmussen E, Peutzfeldt A (2005) Temperature rise induced by some light emitting diode and quartz–tungsten–halogen curing units. Eur J Oral Sci 113(1):96–98PubMedCrossRef
6.
go back to reference Asmussen E, Peutzfeldt A (2003) Light-emitting diode curing: Influence on selected properties of resin composites. Quintessence International 34(1):71–75PubMed Asmussen E, Peutzfeldt A (2003) Light-emitting diode curing: Influence on selected properties of resin composites. Quintessence International 34(1):71–75PubMed
7.
go back to reference Atai M, Watts DC, Atai Z (2005) Shrinkage strain-rates of dental resin-monomer and composite systems. Biomaterials 26:5015–5020PubMedCrossRef Atai M, Watts DC, Atai Z (2005) Shrinkage strain-rates of dental resin-monomer and composite systems. Biomaterials 26:5015–5020PubMedCrossRef
8.
go back to reference Atai M, Watts DC (2006) A new kinetic model for the photopolymerization shrinkage-strain of dental composites and resin-monomers. Dent Mater 22(8):785–791PubMedCrossRef Atai M, Watts DC (2006) A new kinetic model for the photopolymerization shrinkage-strain of dental composites and resin-monomers. Dent Mater 22(8):785–791PubMedCrossRef
9.
go back to reference Baldissara P, Catapano S, Scotti R (1997) Clinical and histological evaluation of thermal injury thresholds in human teeth: a preliminary study. J Oral Rehabil 24(11):791–801PubMedCrossRef Baldissara P, Catapano S, Scotti R (1997) Clinical and histological evaluation of thermal injury thresholds in human teeth: a preliminary study. J Oral Rehabil 24(11):791–801PubMedCrossRef
10.
go back to reference Chung KH, Greener EH (1990) Correlation between degree of conversion, filler concentration and mechanical properties of posterior composite resins. J Oral Rehabil 17(5):487–494PubMedCrossRef Chung KH, Greener EH (1990) Correlation between degree of conversion, filler concentration and mechanical properties of posterior composite resins. J Oral Rehabil 17(5):487–494PubMedCrossRef
11.
go back to reference Davidson CL, Feilzer AJ (1997) Polymerization shrinkage and polymerization shrinkage stress in polymer-based restoratives. J Dent 25(6):435–440PubMedCrossRef Davidson CL, Feilzer AJ (1997) Polymerization shrinkage and polymerization shrinkage stress in polymer-based restoratives. J Dent 25(6):435–440PubMedCrossRef
12.
go back to reference Dunn WJ, Bush AC (2002) A comparison of polymerization by light-emitting diode and halogen-based light-curing units. JADA 133(3):335–341PubMed Dunn WJ, Bush AC (2002) A comparison of polymerization by light-emitting diode and halogen-based light-curing units. JADA 133(3):335–341PubMed
13.
go back to reference Goodis HE, White JM, Andrews J, Watanabe LG (1989) Measurement of temperature generated by visible-light-cure lamps in an in vitro model. Dent Mater 5(4):230–234PubMedCrossRef Goodis HE, White JM, Andrews J, Watanabe LG (1989) Measurement of temperature generated by visible-light-cure lamps in an in vitro model. Dent Mater 5(4):230–234PubMedCrossRef
14.
go back to reference Halvorson RH, Erickson RL, Davidson CL (2003) The effect of filler and silane content on conversion of resin-based composite. Dent Mater 19(4):327–333PubMedCrossRef Halvorson RH, Erickson RL, Davidson CL (2003) The effect of filler and silane content on conversion of resin-based composite. Dent Mater 19(4):327–333PubMedCrossRef
15.
go back to reference Hannig M, Bott B (1999) In-vitro pulp chamber temperature rise during composite resin polymerization with various light-curing sources. Dent Mater 15(4):275–281PubMedCrossRef Hannig M, Bott B (1999) In-vitro pulp chamber temperature rise during composite resin polymerization with various light-curing sources. Dent Mater 15(4):275–281PubMedCrossRef
16.
go back to reference Hofmann N, Hugo B, Klaiber B (2002) Effect of irradiation type (LED or QTH) on photo-activated composite shrinkage strain kinetics, temperature rise, and hardness. Eur J Oral Sci 110(6):471–479PubMedCrossRef Hofmann N, Hugo B, Klaiber B (2002) Effect of irradiation type (LED or QTH) on photo-activated composite shrinkage strain kinetics, temperature rise, and hardness. Eur J Oral Sci 110(6):471–479PubMedCrossRef
17.
go back to reference Hussey DL, Biagioni PA, Lamey P-J (1995) Thermographic measurement of temperature change during resin composite polymerization in vivo. J Dent 23(5):267–271PubMedCrossRef Hussey DL, Biagioni PA, Lamey P-J (1995) Thermographic measurement of temperature change during resin composite polymerization in vivo. J Dent 23(5):267–271PubMedCrossRef
18.
go back to reference Ilie N, Kunzelmann K-H, Visvanathan A, Hickel R (2005) Curing behavior of a nanocomposite as a function of polymerization procedure. Dent Mater J 24(4):469–477PubMed Ilie N, Kunzelmann K-H, Visvanathan A, Hickel R (2005) Curing behavior of a nanocomposite as a function of polymerization procedure. Dent Mater J 24(4):469–477PubMed
19.
go back to reference Jandt KD, Mills RW, Blackwell GB, Ashworth SH (2000) Depth of cure and compressive strength of dental composites cured with blue light emitting diodes (LEDs). Dent Mater 16(1):41–47PubMedCrossRef Jandt KD, Mills RW, Blackwell GB, Ashworth SH (2000) Depth of cure and compressive strength of dental composites cured with blue light emitting diodes (LEDs). Dent Mater 16(1):41–47PubMedCrossRef
20.
go back to reference Knežević A, Tarle Z, Meniga A, Šutalo J, Pichler G, Ristić M (2001) Degree of conversion and temperature rise during polymerization of composite resin samples with blue diodes. J Oral Rehabil 28(6):586–591PubMedCrossRef Knežević A, Tarle Z, Meniga A, Šutalo J, Pichler G, Ristić M (2001) Degree of conversion and temperature rise during polymerization of composite resin samples with blue diodes. J Oral Rehabil 28(6):586–591PubMedCrossRef
21.
go back to reference Leonard DL, Charlton DG, Roberts HW, Cohen ME (2002) Polymerization efficiency of LED curing lights. J Esthet Restor Dent 14(5):286–295PubMedCrossRef Leonard DL, Charlton DG, Roberts HW, Cohen ME (2002) Polymerization efficiency of LED curing lights. J Esthet Restor Dent 14(5):286–295PubMedCrossRef
22.
go back to reference Lloyd CH, Joshi A, McGlynn E (1986) Temperature rises produced by light sources and composites during curing. Dent Mater 2(4):170–174PubMedCrossRef Lloyd CH, Joshi A, McGlynn E (1986) Temperature rises produced by light sources and composites during curing. Dent Mater 2(4):170–174PubMedCrossRef
23.
go back to reference Masutani S, Setcos JC, Schnell RJ, Phillips RW (1988) Temperature rise during polymerization of visible light-activated composite resins. Dent Mater 4(4):174–178PubMedCrossRef Masutani S, Setcos JC, Schnell RJ, Phillips RW (1988) Temperature rise during polymerization of visible light-activated composite resins. Dent Mater 4(4):174–178PubMedCrossRef
24.
go back to reference Meyer GR, Ernst CP, Willershausen B (2002) Decrease in power output of new light-emitting diode (LED) curing devices with increasing distance to filling surface. J Adhes Dent 4(3):197–204PubMed Meyer GR, Ernst CP, Willershausen B (2002) Decrease in power output of new light-emitting diode (LED) curing devices with increasing distance to filling surface. J Adhes Dent 4(3):197–204PubMed
25.
go back to reference Mills RW, Jandt KD, Ashworth SH (1999) Dental composite depth of cure with halogen and blue light emitting diode technology. Br Dent J 186(8):388–391PubMed Mills RW, Jandt KD, Ashworth SH (1999) Dental composite depth of cure with halogen and blue light emitting diode technology. Br Dent J 186(8):388–391PubMed
26.
go back to reference Mills RW, Uhl A, Jandt KD (2002) Optical power outputs, spectra and dental composite depths of cure, obtained with blue light emitting diode (LED) and halogen light curing units (LCUs). Br Dent J 193(8):459–463PubMedCrossRef Mills RW, Uhl A, Jandt KD (2002) Optical power outputs, spectra and dental composite depths of cure, obtained with blue light emitting diode (LED) and halogen light curing units (LCUs). Br Dent J 193(8):459–463PubMedCrossRef
27.
go back to reference Murchison DF, Moore BK (1992) Influence of curing time and distance on microhardness of eight light-cured liners. Oper Dent 17(4):135–141PubMed Murchison DF, Moore BK (1992) Influence of curing time and distance on microhardness of eight light-cured liners. Oper Dent 17(4):135–141PubMed
28.
go back to reference Nomoto R, McCabe JF, Hirano S (2004) Comparison of halogen, plasma and LED curing units. Oper Dent 29(3):287–294PubMed Nomoto R, McCabe JF, Hirano S (2004) Comparison of halogen, plasma and LED curing units. Oper Dent 29(3):287–294PubMed
29.
go back to reference Nomura Y, Teshima W, Tanaka N, Yoshida Y, Nahara Y, Okazaki M (2002) Thermal analysis of dental resins cured with blue light-emitting diodes (LEDs). J Biomed Mater Res 63(2):209–213PubMedCrossRef Nomura Y, Teshima W, Tanaka N, Yoshida Y, Nahara Y, Okazaki M (2002) Thermal analysis of dental resins cured with blue light-emitting diodes (LEDs). J Biomed Mater Res 63(2):209–213PubMedCrossRef
30.
go back to reference Porko C, Hietala EL (2001) Pulpal temperature change with visible light-curing. Oper Dent 26:181–185 Porko C, Hietala EL (2001) Pulpal temperature change with visible light-curing. Oper Dent 26:181–185
31.
go back to reference Poulos JG, Styner DL (1997) Curing lights: changes in intensity output with use over time. Gen Dent 45(1):70–73PubMed Poulos JG, Styner DL (1997) Curing lights: changes in intensity output with use over time. Gen Dent 45(1):70–73PubMed
32.
go back to reference Schneider LF, Consani S, Correr–Sobrinho L, Correr AB, Sinhoreti MA (2006) Halogen and LED light curing of composite: temperature increase and Knoop hardness. Clin Oral Investig 10(1):66–71PubMedCrossRef Schneider LF, Consani S, Correr–Sobrinho L, Correr AB, Sinhoreti MA (2006) Halogen and LED light curing of composite: temperature increase and Knoop hardness. Clin Oral Investig 10(1):66–71PubMedCrossRef
33.
go back to reference Shortall AC, Harrington E (1998) Temperature rise during polymerization of light-activated resin composites. J Oral Rehabil 25(12):908–913PubMedCrossRef Shortall AC, Harrington E (1998) Temperature rise during polymerization of light-activated resin composites. J Oral Rehabil 25(12):908–913PubMedCrossRef
34.
go back to reference Smail SRJ, Patterson CJW, Mclundie AC, Strang R (1988) In vitro temperature rises during visible-light curing of a lining material and a posterior composite. J Oral Rehabil 15(4):361–366PubMedCrossRef Smail SRJ, Patterson CJW, Mclundie AC, Strang R (1988) In vitro temperature rises during visible-light curing of a lining material and a posterior composite. J Oral Rehabil 15(4):361–366PubMedCrossRef
35.
go back to reference Soh MS, Yap AU, Yu T, Shen ZX (2004) Analysis of the degree of conversion of LED and halogen lights using micro-Raman spectroscopy. Oper Dent 29(5):571–577PubMed Soh MS, Yap AU, Yu T, Shen ZX (2004) Analysis of the degree of conversion of LED and halogen lights using micro-Raman spectroscopy. Oper Dent 29(5):571–577PubMed
36.
go back to reference Tarle Z, Meniga A, Knezevic A, Sutalo J, Ristic M, Pichler G (2002) Composite conversion and temperature rise using a conventional, plasma arc, and an experimental blue LED curing unit. J Oral Rehabil 29(7):662–667PubMedCrossRef Tarle Z, Meniga A, Knezevic A, Sutalo J, Ristic M, Pichler G (2002) Composite conversion and temperature rise using a conventional, plasma arc, and an experimental blue LED curing unit. J Oral Rehabil 29(7):662–667PubMedCrossRef
37.
go back to reference Uhl A, Mills RW, Jandt KD (2003) Polymerization and light-induced heat of dental composites cured with LED and halogen technology. Biomaterials 24(10):1809–1820PubMedCrossRef Uhl A, Mills RW, Jandt KD (2003) Polymerization and light-induced heat of dental composites cured with LED and halogen technology. Biomaterials 24(10):1809–1820PubMedCrossRef
38.
go back to reference Vandewalle KS, Roberts HW, Tiba A, Charlton DG (2005) Thermal emission and curing efficiency of LED and halogen curing lights. Oper Dent 30(2):257–264PubMed Vandewalle KS, Roberts HW, Tiba A, Charlton DG (2005) Thermal emission and curing efficiency of LED and halogen curing lights. Oper Dent 30(2):257–264PubMed
39.
go back to reference Watts DC (2005) Reaction kinetics and mechanics in photo-polymerised networks. Dent Mater 21(1):27–35PubMedCrossRef Watts DC (2005) Reaction kinetics and mechanics in photo-polymerised networks. Dent Mater 21(1):27–35PubMedCrossRef
40.
go back to reference Yap AU, Soh MS (2003) Thermal emission by different light-curing units. Oper Dent 28(3):260–266PubMed Yap AU, Soh MS (2003) Thermal emission by different light-curing units. Oper Dent 28(3):260–266PubMed
41.
go back to reference Zach L, Cohen G (1965) Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol 19:515–530PubMedCrossRef Zach L, Cohen G (1965) Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol 19:515–530PubMedCrossRef
Metadata
Title
Temperature rise and degree of photopolymerization conversion of nanocomposites and conventional dental composites
Authors
Mohammad Atai
Fariba Motevasselian
Publication date
01-09-2009
Publisher
Springer-Verlag
Published in
Clinical Oral Investigations / Issue 3/2009
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-008-0236-2

Other articles of this Issue 3/2009

Clinical Oral Investigations 3/2009 Go to the issue