A novel device designed to improve the operability of energy devices with foot pedals in endoscopic surgery: the Foot-Site Monitor

Authors: Akihiro Kondo, Yuji Nishizawa, Yasumasa Horikiri, Hiroki Amemori, Yuichiro Tsukada, Takeshi Sasaki, Toshikazu Kawai, Hiroyuki Daiko, Masaaki Ito

Published in: Surgery Today | Issue 11/2019

Login to get access

Abstract

Purpose

We developed the “Foot-Site Monitor” (FSM) to permit easy identification of multiple foot pedals of energy devices in endoscopic surgery. The purpose of this study was to evaluate the effectiveness of the use of the FSM.

Methods

Using a training box, 20 surgeons performed a task consisting of pressing foot pedals with and without the FSM. The time from showing a color on a PC monitor to pressing the foot pedal that corresponded to this color (defined as the completion time), and the distance of the movement of the tip of the surgeon’s forceps from the beginning to the end of the task (defined as shaking of the forceps) were measured.

Results

The use of the FSM was associated with significantly reduced shaking of the forceps (1.95 vs. 2.47 mm; p = 0.014), and a tendency toward a shorter completion time (1.39 vs. 1.51 s; p = 0.053).

Conclusions

The use of the FSM in endoscopic surgery contributes to reduced shaking of the forceps and may shorten the operative time.
Literature
1.
go back to reference Kamiyoshihara M, Igai H, Ibe T, Kawatani N, Ohtaki Y, Shimizu K, et al. Right superior mediastinal lymph node dissection in thoracoscopic surgery using a bipolar sealing device. Innovations. 2013;8:258–63.CrossRefPubMed Kamiyoshihara M, Igai H, Ibe T, Kawatani N, Ohtaki Y, Shimizu K, et al. Right superior mediastinal lymph node dissection in thoracoscopic surgery using a bipolar sealing device. Innovations. 2013;8:258–63.CrossRefPubMed
2.
go back to reference Dural C, Akyuz M, Yazici P, Aksoy E, Aucejo F, Quintini C, et al. Safety and efficacy of a new bipolar energy device for parenchymal dissection in laparoscopic liver resection. Surg Laparosc Endosc Percutan Tech. 2016;26:21–4.CrossRefPubMed Dural C, Akyuz M, Yazici P, Aksoy E, Aucejo F, Quintini C, et al. Safety and efficacy of a new bipolar energy device for parenchymal dissection in laparoscopic liver resection. Surg Laparosc Endosc Percutan Tech. 2016;26:21–4.CrossRefPubMed
3.
go back to reference Weyhe D, Uslae VN, Tabriz N, Burkowski I, Heinzel R, Muller A, et al. Experience and dissection device are more relevant than patient-related factors for operation time in laparoscopic sigmoid resection-a retrospective 8-year observational study. Int J Colorectal Dis. 2017;32:1703–10.CrossRefPubMedPubMedCentral Weyhe D, Uslae VN, Tabriz N, Burkowski I, Heinzel R, Muller A, et al. Experience and dissection device are more relevant than patient-related factors for operation time in laparoscopic sigmoid resection-a retrospective 8-year observational study. Int J Colorectal Dis. 2017;32:1703–10.CrossRefPubMedPubMedCentral
4.
go back to reference Choussein S, Srouji SS, Farland LV, Gargiulo AR. Flexible carbon dioxide laser fiber versus ultrasonic scalpel in robot-assisted laparoscopic myomectomy. J Minim Invasive Gynecol. 2015;22:1183–90.CrossRefPubMed Choussein S, Srouji SS, Farland LV, Gargiulo AR. Flexible carbon dioxide laser fiber versus ultrasonic scalpel in robot-assisted laparoscopic myomectomy. J Minim Invasive Gynecol. 2015;22:1183–90.CrossRefPubMed
5.
go back to reference Gotohda N, Yamanaka T, Saiura A, Uesaka K, Hashimoto M, Konishi M, et al. Impact of energy devices during liver parenchymal transection: a multicenter randomized controlled trial. World J Surg. 2015;39:1543–9.CrossRefPubMed Gotohda N, Yamanaka T, Saiura A, Uesaka K, Hashimoto M, Konishi M, et al. Impact of energy devices during liver parenchymal transection: a multicenter randomized controlled trial. World J Surg. 2015;39:1543–9.CrossRefPubMed
6.
go back to reference Richter S, Kollmar O, Schuld J, Moussavian MR, Igna D, Schilling MK, Chirurgische Arbeitsgemeinschaft OP-Technik und OP-Strukturen (CAOP) of the Deutsche Gesellschaft fur Allgemein- und Viszeralchirurgie. Randomized clinical trial of efficacy and costs of three dissection devices in liver resection. Br J Surg. 2009;96:593–601.CrossRefPubMed Richter S, Kollmar O, Schuld J, Moussavian MR, Igna D, Schilling MK, Chirurgische Arbeitsgemeinschaft OP-Technik und OP-Strukturen (CAOP) of the Deutsche Gesellschaft fur Allgemein- und Viszeralchirurgie. Randomized clinical trial of efficacy and costs of three dissection devices in liver resection. Br J Surg. 2009;96:593–601.CrossRefPubMed
7.
go back to reference Ikeda M, Hasegawa K, Seno K, Imamura H, Beck Y, Sugawara Y, et al. The vessel sealing system (LigaSure) in hepatic resection: a randomized controlled trial. Ann Surg. 2009;250:199–203.CrossRefPubMed Ikeda M, Hasegawa K, Seno K, Imamura H, Beck Y, Sugawara Y, et al. The vessel sealing system (LigaSure) in hepatic resection: a randomized controlled trial. Ann Surg. 2009;250:199–203.CrossRefPubMed
8.
go back to reference Targarona EM, Balague C, Marin J, Neto RB, Martinez C, Garriga J, et al. Energy sources for laparoscopic colectomy: a prospective randomized comparison of conventional electrosurgery, bipolar computer-controlled electrosurgery and ultrasonic dissection. Operative outcome and costs analysis. Surg Innov. 2005;12:339–44.CrossRefPubMed Targarona EM, Balague C, Marin J, Neto RB, Martinez C, Garriga J, et al. Energy sources for laparoscopic colectomy: a prospective randomized comparison of conventional electrosurgery, bipolar computer-controlled electrosurgery and ultrasonic dissection. Operative outcome and costs analysis. Surg Innov. 2005;12:339–44.CrossRefPubMed
9.
go back to reference Kossi J, Luostarinen M, Kontula I, Laato M. Laparoscopic sigmoid and rectal resection using an electrothermal bipolar vessel sealing device. J Laparoendosc Adv Surg Tech A. 2007;17:719–22.CrossRefPubMed Kossi J, Luostarinen M, Kontula I, Laato M. Laparoscopic sigmoid and rectal resection using an electrothermal bipolar vessel sealing device. J Laparoendosc Adv Surg Tech A. 2007;17:719–22.CrossRefPubMed
10.
go back to reference Milsom J, Trencheva K, Menette S, Pavoor R, Shukla P, Ma J, et al. Evaluation of the safety, efficacy, and versatility of a new surgical device (THUNDERBEAT) in comparison with Harmonic ACE, LigaSure V, and EnSeal devices in a porcine model. J Laparoendosc Adv Surg Tech A. 2012;22:378–86.CrossRefPubMed Milsom J, Trencheva K, Menette S, Pavoor R, Shukla P, Ma J, et al. Evaluation of the safety, efficacy, and versatility of a new surgical device (THUNDERBEAT) in comparison with Harmonic ACE, LigaSure V, and EnSeal devices in a porcine model. J Laparoendosc Adv Surg Tech A. 2012;22:378–86.CrossRefPubMed
11.
go back to reference Garfinkle R, Boutros M, Hippalgaonkar N, Maimon G, da Silva G, Potenti F, et al. Electrothermal bipolar vessel ligation improves operative time during laparoscopic total proctocolectomy: a large single-center experience. Surg Endosc. 2016;30:2840–7.CrossRefPubMed Garfinkle R, Boutros M, Hippalgaonkar N, Maimon G, da Silva G, Potenti F, et al. Electrothermal bipolar vessel ligation improves operative time during laparoscopic total proctocolectomy: a large single-center experience. Surg Endosc. 2016;30:2840–7.CrossRefPubMed
12.
go back to reference Kim YN, Yoo YC, Guner A, Cho I, Kwon IG, Kim YN, et al. Comparison of perioperative surgical outcomes between a bipolar device and an ultrasonic device during laparoscopic gastrectomy for gastric cancer. Surg Endosc. 2015;29:589–95.CrossRefPubMed Kim YN, Yoo YC, Guner A, Cho I, Kwon IG, Kim YN, et al. Comparison of perioperative surgical outcomes between a bipolar device and an ultrasonic device during laparoscopic gastrectomy for gastric cancer. Surg Endosc. 2015;29:589–95.CrossRefPubMed
13.
go back to reference Vizzielli G, Conte C, Romano M, Fagotti A, Constantini B, Lodoli C, et al. Clinical impact of a surgical energy device in advanced ovarian cancer surgery including bowel resection. In Vivo. 2018;32:359–64.PubMedPubMedCentral Vizzielli G, Conte C, Romano M, Fagotti A, Constantini B, Lodoli C, et al. Clinical impact of a surgical energy device in advanced ovarian cancer surgery including bowel resection. In Vivo. 2018;32:359–64.PubMedPubMedCentral
14.
go back to reference Stotz L, Joukhadar R, Hamza A, Thangarajah F, Bardens D, Juhasz-Boss I, et al. Instrument usage in laparoscopic gynecologic surgery: a prospective clinical trial. Arch Gynecol Obstet. 2018;298:773–9.CrossRefPubMed Stotz L, Joukhadar R, Hamza A, Thangarajah F, Bardens D, Juhasz-Boss I, et al. Instrument usage in laparoscopic gynecologic surgery: a prospective clinical trial. Arch Gynecol Obstet. 2018;298:773–9.CrossRefPubMed
16.
go back to reference Matsunaga R, Nishizawa Y, Saito N, Kobayashi A, Ohdaira T, Ito M. Quantitative evaluation of 3D imaging in laparoscopic surgery. Surg Today. 2017;47:440–4.CrossRefPubMed Matsunaga R, Nishizawa Y, Saito N, Kobayashi A, Ohdaira T, Ito M. Quantitative evaluation of 3D imaging in laparoscopic surgery. Surg Today. 2017;47:440–4.CrossRefPubMed
17.
go back to reference Rashid TG, Kini M, Ind TE. Comparing the learning curve for robotically assisted and straight stick laparoscopic procedures in surgical novices. Int J Med Robot. 2010;6:306–10.CrossRefPubMed Rashid TG, Kini M, Ind TE. Comparing the learning curve for robotically assisted and straight stick laparoscopic procedures in surgical novices. Int J Med Robot. 2010;6:306–10.CrossRefPubMed
18.
go back to reference Sackier J, Wang Y. Robotically assisted laparoscopic surgery: From concept to development. Surg Endosc. 1994;8:63–6.CrossRefPubMed Sackier J, Wang Y. Robotically assisted laparoscopic surgery: From concept to development. Surg Endosc. 1994;8:63–6.CrossRefPubMed
19.
go back to reference Buess G, Arezzo A, Schurr M, Ulmer F, Fisher H, Gumb L, et al. A new remote-controlled endoscope positioning system for endoscopic solo surgery: the FIPS endoarm. Surg Endosc. 2000;14:395–9.CrossRefPubMed Buess G, Arezzo A, Schurr M, Ulmer F, Fisher H, Gumb L, et al. A new remote-controlled endoscope positioning system for endoscopic solo surgery: the FIPS endoarm. Surg Endosc. 2000;14:395–9.CrossRefPubMed
20.
go back to reference Allaf M, Jackman S, Schulam P, Cadeddu J, Lee B, Moore R, et al. Laparoscopic visual field: voice vs foot pedal interfaces for control of the AESOP robot. Surg Endosc. 1998;12:1415–8.CrossRefPubMed Allaf M, Jackman S, Schulam P, Cadeddu J, Lee B, Moore R, et al. Laparoscopic visual field: voice vs foot pedal interfaces for control of the AESOP robot. Surg Endosc. 1998;12:1415–8.CrossRefPubMed
21.
go back to reference Vara-Thorbeck C, Munoz V, Toscano R, Gomez J, Fernandez J, Felices M, et al. A new robotic endoscope manipulator: a prelim trial to evaluate the performance of a voice-operated industrial robot and a human assistant in several simulated and real endoscopic operation. Surg Endosc. 2001;15:924–7.CrossRefPubMed Vara-Thorbeck C, Munoz V, Toscano R, Gomez J, Fernandez J, Felices M, et al. A new robotic endoscope manipulator: a prelim trial to evaluate the performance of a voice-operated industrial robot and a human assistant in several simulated and real endoscopic operation. Surg Endosc. 2001;15:924–7.CrossRefPubMed
22.
go back to reference van der Voort M, Heijnsdijk EA, Gouma DJ. Bowel injury as a complication of laparoscopy. Br J Surg. 2004;91:1253–8.CrossRefPubMed van der Voort M, Heijnsdijk EA, Gouma DJ. Bowel injury as a complication of laparoscopy. Br J Surg. 2004;91:1253–8.CrossRefPubMed
Metadata
Title
A novel device designed to improve the operability of energy devices with foot pedals in endoscopic surgery: the Foot-Site Monitor
Authors
Akihiro Kondo
Yuji Nishizawa
Yasumasa Horikiri
Hiroki Amemori
Yuichiro Tsukada
Takeshi Sasaki
Toshikazu Kawai
Hiroyuki Daiko
Masaaki Ito
Publication date
01-11-2019
Publisher
Springer Singapore
Published in
Surgery Today / Issue 11/2019
Print ISSN: 0941-1291
Electronic ISSN: 1436-2813
DOI
https://doi.org/10.1007/s00595-019-01824-7

Other articles of this Issue 11/2019

Surgery Today 11/2019 Go to the issue