Skip to main content
Top
Published in: European Journal of Orthopaedic Surgery & Traumatology 3/2019

01-04-2019 | Up-to date Review and Case Report • HIP¨- ARTHROPLASTY

Modular femoral neck failure after revision of a total hip arthroplasty: a finite element analysis

Authors: Jan Zajc, Jožef Predan, Nenad Gubeljak, Andrej Moličnik, Samo K. Fokter

Published in: European Journal of Orthopaedic Surgery & Traumatology | Issue 3/2019

Login to get access

Abstract

The authors report on a case of modular femoral neck fracture which appeared 21 months after revision of acetabular component. The revision surgery was performed 8 years after the primary total hip arthroplasty due to aseptic loosening of the acetabular component. During acetabular revision, the primary implanted short (S, − 3.5 mm) femoral head was also exchanged with extra-long (XL, + 7.0 mm) femoral head fitting the modular femoral neck with a longer lever arm. Numerical analysis has shown that this has resulted in a 19.9% increase in tensile stress at the neck–stem coupling during normal walking cycle. This could result in microcrack initiation and propagation and finally lead to modular neck failure of the otherwise well-fixed stem. Surgeons should avoid excessive loading of the exchangeable neck (dual-modular) femoral stem designs as the stem–neck couplings are subject to corrosion and are not as reliable as monoblock stems.
Literature
1.
go back to reference Traina F, De Clerico M, Biondi F, Pilla F, Tassinari E, Toni A (2009) Sex differences in hip morphology: is stem modularity effective for total hip replacement? J Bone Jt Surg Am 91(Suppl 6):121–128CrossRef Traina F, De Clerico M, Biondi F, Pilla F, Tassinari E, Toni A (2009) Sex differences in hip morphology: is stem modularity effective for total hip replacement? J Bone Jt Surg Am 91(Suppl 6):121–128CrossRef
2.
go back to reference Wright C, Sporer S, Urban R, Jacobs J (2010) Fracture of a modular femoral neck after total hip arthroplasty. J Bone Jt Surg Am 92(Suppl 6):1518–1521CrossRef Wright C, Sporer S, Urban R, Jacobs J (2010) Fracture of a modular femoral neck after total hip arthroplasty. J Bone Jt Surg Am 92(Suppl 6):1518–1521CrossRef
3.
go back to reference Grupp TM, Weik T, Bloemer W, Knaebel HP (2010) Modular titanium alloy neck adapter failures in hip replacement–failure mode analysis and influence of implant material. BMC Musculoskelet Disord 11:3CrossRefPubMedPubMedCentral Grupp TM, Weik T, Bloemer W, Knaebel HP (2010) Modular titanium alloy neck adapter failures in hip replacement–failure mode analysis and influence of implant material. BMC Musculoskelet Disord 11:3CrossRefPubMedPubMedCentral
4.
go back to reference Fokter SK, Rudolf R, Moličnik A (2016) Titanium alloy femoral neck fracture–clinical and metallurgical analysis in 6 cases. Acta Orthop 87(2):197–202PubMed Fokter SK, Rudolf R, Moličnik A (2016) Titanium alloy femoral neck fracture–clinical and metallurgical analysis in 6 cases. Acta Orthop 87(2):197–202PubMed
5.
go back to reference Weiser M, Chen D (2016) Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls. Curr Rev Musculoskelet Med 9(1):75–83CrossRefPubMedPubMedCentral Weiser M, Chen D (2016) Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls. Curr Rev Musculoskelet Med 9(1):75–83CrossRefPubMedPubMedCentral
6.
go back to reference Naal F, Impellizzeri F, Leunig M (2008) Which is the best activity rating scale for patients undergoing total joint arthroplasty? Clin Orthop Relat Res 467(4):958–965CrossRefPubMedPubMedCentral Naal F, Impellizzeri F, Leunig M (2008) Which is the best activity rating scale for patients undergoing total joint arthroplasty? Clin Orthop Relat Res 467(4):958–965CrossRefPubMedPubMedCentral
7.
go back to reference Bergmann G, Graichen F, Rohlmann A (1993) Hip joint loading during walking and running, measured in two patients. J Biomech 26(8):969–990CrossRefPubMed Bergmann G, Graichen F, Rohlmann A (1993) Hip joint loading during walking and running, measured in two patients. J Biomech 26(8):969–990CrossRefPubMed
8.
go back to reference Fokter S, Moličnik A, Kavalar R, Pelicon P, Rudolf R, Gubeljak N (2017) Why do some titanium-alloy total hip arthroplasty modular necks fail? J Mech Behav Biomed Mater 69:107–114CrossRefPubMed Fokter S, Moličnik A, Kavalar R, Pelicon P, Rudolf R, Gubeljak N (2017) Why do some titanium-alloy total hip arthroplasty modular necks fail? J Mech Behav Biomed Mater 69:107–114CrossRefPubMed
9.
go back to reference Wright Medical Technology, Inc. (2010) Masters Hip Post Training Document, Amsterdam (July 2010) Wright Medical Technology, Inc. (2010) Masters Hip Post Training Document, Amsterdam (July 2010)
10.
go back to reference Massin P, Geais L, Astoin E, Simondi M, Lavaste F (2000) The anatomic basis for the concept of lateralized femoral stems. J Arthroplasty 15(1):93–101CrossRefPubMed Massin P, Geais L, Astoin E, Simondi M, Lavaste F (2000) The anatomic basis for the concept of lateralized femoral stems. J Arthroplasty 15(1):93–101CrossRefPubMed
11.
go back to reference Duwelius P, Burkhart B, Carnahan C, Branam G, Ko L, Wu Y et al (2013) Modular versus nonmodular neck femoral implants in primary total hip arthroplasty: which is better? Clin Orthop Relat Res 472(4):1240–1245CrossRefPubMedPubMedCentral Duwelius P, Burkhart B, Carnahan C, Branam G, Ko L, Wu Y et al (2013) Modular versus nonmodular neck femoral implants in primary total hip arthroplasty: which is better? Clin Orthop Relat Res 472(4):1240–1245CrossRefPubMedPubMedCentral
12.
go back to reference Atwood S, Patten E, Bozic K, Pruitt L, Ries M (2010) Corrosion-induced fracture of a double-modular hip prosthesis. J Bone Jt Surg Am 92(6):1522–1525CrossRef Atwood S, Patten E, Bozic K, Pruitt L, Ries M (2010) Corrosion-induced fracture of a double-modular hip prosthesis. J Bone Jt Surg Am 92(6):1522–1525CrossRef
13.
go back to reference Wilson D, Dunbar M, Amirault J, Farhat Z (2010) Early failure of a modular femoral neck total hip arthroplasty component. J Bone Jt Surg Am 92(6):1514–1517CrossRef Wilson D, Dunbar M, Amirault J, Farhat Z (2010) Early failure of a modular femoral neck total hip arthroplasty component. J Bone Jt Surg Am 92(6):1514–1517CrossRef
14.
go back to reference Dangles C, Altstetter C (2010) Failure of the modular neck in a total hip arthroplasty. J Arthroplasty 25(7):1169.e5–1169.e7CrossRef Dangles C, Altstetter C (2010) Failure of the modular neck in a total hip arthroplasty. J Arthroplasty 25(7):1169.e5–1169.e7CrossRef
15.
go back to reference Skendzel J, Blaha J, Urquhart A (2011) Total hip arthroplasty modular neck failure. J Arthroplasty 26(2):338.e1–338.e4CrossRef Skendzel J, Blaha J, Urquhart A (2011) Total hip arthroplasty modular neck failure. J Arthroplasty 26(2):338.e1–338.e4CrossRef
16.
go back to reference Fitch D, Ancarani C, Bordini B (2015) Long-term survivorship and complication rate comparison of a cementless modular stem and cementless fixed neck stems for primary total hip replacement. Int Orthop 39(9):1827–1832CrossRefPubMed Fitch D, Ancarani C, Bordini B (2015) Long-term survivorship and complication rate comparison of a cementless modular stem and cementless fixed neck stems for primary total hip replacement. Int Orthop 39(9):1827–1832CrossRefPubMed
17.
go back to reference Pour A, Borden R, Murayama T, Groll-Brown M, Blaha J (2015) High risk of failure with bimodular femoral components in THA. Clin Orthop Relat Res 474(1):146–153CrossRefPubMedPubMedCentral Pour A, Borden R, Murayama T, Groll-Brown M, Blaha J (2015) High risk of failure with bimodular femoral components in THA. Clin Orthop Relat Res 474(1):146–153CrossRefPubMedPubMedCentral
18.
go back to reference Gofton W, Illical E, Feibel R, Kim P, Beaulé P (2017) A single-center experience with a titanium modular neck total hip arthroplasty. J Arthroplasty 32(8):2450–2456CrossRefPubMed Gofton W, Illical E, Feibel R, Kim P, Beaulé P (2017) A single-center experience with a titanium modular neck total hip arthroplasty. J Arthroplasty 32(8):2450–2456CrossRefPubMed
19.
go back to reference Fokter SK, Levašič V, Kovač S (2017) The inovation trap: modular neck in total hip arthroplasty. Zdrav Vest 86:115–126 Fokter SK, Levašič V, Kovač S (2017) The inovation trap: modular neck in total hip arthroplasty. Zdrav Vest 86:115–126
20.
go back to reference Bernstein D, Meftah M, Paranilam J, Incavo S (2016) Eighty-six percent failure rate of a modular-neck femoral stem design at 3 to 5 years. J Bone Jt Surg 98(12):e49CrossRef Bernstein D, Meftah M, Paranilam J, Incavo S (2016) Eighty-six percent failure rate of a modular-neck femoral stem design at 3 to 5 years. J Bone Jt Surg 98(12):e49CrossRef
21.
go back to reference Pelayo-de-Tomás J, Rodrigo-Pérez J, Novoa-Parra C, Lizaur-Utrilla A, Morales-Suárez-Varela M, Blas-Dobón J (2017) Cementless modular neck stems: are they a safe option in primary total hip arthroplasty? Eur J Orthop Surg Traumatol 28(3):463–469CrossRefPubMed Pelayo-de-Tomás J, Rodrigo-Pérez J, Novoa-Parra C, Lizaur-Utrilla A, Morales-Suárez-Varela M, Blas-Dobón J (2017) Cementless modular neck stems: are they a safe option in primary total hip arthroplasty? Eur J Orthop Surg Traumatol 28(3):463–469CrossRefPubMed
22.
go back to reference Colas S, Allalou A, Poichotte A, Piriou P, Dray-Spira R, Zureik M (2017) Exchangeable femoral neck (dual-modular) THA prostheses have poorer survivorship than other designs: a nationwide cohort of 324,108 patients. Clin Orthop Relat Res 475(8):2046–2059CrossRefPubMedPubMedCentral Colas S, Allalou A, Poichotte A, Piriou P, Dray-Spira R, Zureik M (2017) Exchangeable femoral neck (dual-modular) THA prostheses have poorer survivorship than other designs: a nationwide cohort of 324,108 patients. Clin Orthop Relat Res 475(8):2046–2059CrossRefPubMedPubMedCentral
23.
go back to reference Mumme T, Müller-Rath R, Jakobi N, Weißkopf M, Dott W, Marx R et al (2004) In vitro serum levels of metal ions released from orthopaedic implants. Eur J Orthop Surg Traumatol 15(2):83–89CrossRef Mumme T, Müller-Rath R, Jakobi N, Weißkopf M, Dott W, Marx R et al (2004) In vitro serum levels of metal ions released from orthopaedic implants. Eur J Orthop Surg Traumatol 15(2):83–89CrossRef
24.
go back to reference Fokter SK, Repše-Fokter A, Takač I (2009) Case report: femoral neuropathy secondary to total hip arthroplasty wear debris. Clin Orthop Relat Res 467(11):3032–3035CrossRefPubMedPubMedCentral Fokter SK, Repše-Fokter A, Takač I (2009) Case report: femoral neuropathy secondary to total hip arthroplasty wear debris. Clin Orthop Relat Res 467(11):3032–3035CrossRefPubMedPubMedCentral
25.
go back to reference Lindgren J, Brismar B, Wikstrom A (2011) Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis. J Bone Jt Surg Br 93-B(10):1427–1430CrossRef Lindgren J, Brismar B, Wikstrom A (2011) Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis. J Bone Jt Surg Br 93-B(10):1427–1430CrossRef
26.
go back to reference Gill I, Webb J, Sloan K, Beaver R (2012) Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Jt Surg Br 94-B(7):895–900CrossRef Gill I, Webb J, Sloan K, Beaver R (2012) Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Jt Surg Br 94-B(7):895–900CrossRef
27.
go back to reference Cooper H, Urban R, Wixson R, Meneghini R, Jacobs J (2013) Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Jt Surg Am 95(10):865–872CrossRef Cooper H, Urban R, Wixson R, Meneghini R, Jacobs J (2013) Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Jt Surg Am 95(10):865–872CrossRef
28.
go back to reference Sharan D (1999) The problem of corrosion in orthopaedic implant materials. Orthop Update (India) 9(1):1–5 Sharan D (1999) The problem of corrosion in orthopaedic implant materials. Orthop Update (India) 9(1):1–5
30.
go back to reference Gilbert JL, Mali S, Urban RM, Silverton CD, Jacobs JJ (2011) In vivo oxide-induced stress corrosion cracking of Ti–6Al–4V in a neck-stem modular taper: Emergent behavior in a new mechanism of in vivo corrosion. J Biomed Mater Res B Appl Biomater 100B(2):584–594CrossRef Gilbert JL, Mali S, Urban RM, Silverton CD, Jacobs JJ (2011) In vivo oxide-induced stress corrosion cracking of Ti–6Al–4V in a neck-stem modular taper: Emergent behavior in a new mechanism of in vivo corrosion. J Biomed Mater Res B Appl Biomater 100B(2):584–594CrossRef
31.
go back to reference Silverton C, Jacobs J, Devitt J, Cooper H (2014) Midterm results of a femoral stem with a modular neck design: clinical outcomes and metal ion analysis. J Arthroplasty 29(9):1768–1773CrossRefPubMed Silverton C, Jacobs J, Devitt J, Cooper H (2014) Midterm results of a femoral stem with a modular neck design: clinical outcomes and metal ion analysis. J Arthroplasty 29(9):1768–1773CrossRefPubMed
32.
go back to reference Grupp T, Weik T, Bloemer W, Knaebel H (2010) Modular titanium alloy neck adapter failures in hip replacement: failure mode analysis and influence of implant material. BMC Musculoskelet Disord 11(1):3CrossRefPubMedPubMedCentral Grupp T, Weik T, Bloemer W, Knaebel H (2010) Modular titanium alloy neck adapter failures in hip replacement: failure mode analysis and influence of implant material. BMC Musculoskelet Disord 11(1):3CrossRefPubMedPubMedCentral
33.
go back to reference Mencière M, Amouyel T, Taviaux J, Bayle M, Laterza C, Mertl P (2014) Fracture of the cobalt–chromium modular femoral neck component in total hip arthroplasty. Orthop Traumatol Surg Res 100(5):565–568CrossRefPubMed Mencière M, Amouyel T, Taviaux J, Bayle M, Laterza C, Mertl P (2014) Fracture of the cobalt–chromium modular femoral neck component in total hip arthroplasty. Orthop Traumatol Surg Res 100(5):565–568CrossRefPubMed
Metadata
Title
Modular femoral neck failure after revision of a total hip arthroplasty: a finite element analysis
Authors
Jan Zajc
Jožef Predan
Nenad Gubeljak
Andrej Moličnik
Samo K. Fokter
Publication date
01-04-2019
Publisher
Springer Paris
Published in
European Journal of Orthopaedic Surgery & Traumatology / Issue 3/2019
Print ISSN: 1633-8065
Electronic ISSN: 1432-1068
DOI
https://doi.org/10.1007/s00590-018-2314-8

Other articles of this Issue 3/2019

European Journal of Orthopaedic Surgery & Traumatology 3/2019 Go to the issue