Impedance-pH monitoring on medications does not reliably confirm the presence of gastroesophageal reflux disease in patients referred for antireflux surgery

Authors: Marc A. Ward, Christy M. Dunst, Ezra N. Teitelbaum, Valerie J. Halpin, Kevin M. Reavis, Lee L. Swanström, Steven R. DeMeester

Published in: Surgical Endoscopy | Issue 2/2018

Login to get access

Abstract

Introduction

The gold standard for the objective diagnosis of gastroesophageal reflux disease (GERD) is ambulatory-pH monitoring off medications. Increasingly, impedance-pH (MII-pH) monitoring on medications is being used to evaluate refractory symptoms. The aim of this study was to determine whether an MII-pH test on medications can reliably detect the presence of GERD.

Methods

Patients referred for persistent reflux symptoms despite pH confirmed adequate acid suppression (DeMeester score ≤14.7) were reviewed retrospectively. All patients who originally had MII-pH testing on medications were re-evaluated with an off medication Bravo-pH study. Acid exposure results (defined by off medication Bravo) were compared to the original on medication MII-pH.

Results

There were 49 patients who met study criteria (median age 51). Patients had normal acid exposure during their MII-pH test on medications (average DMS 4.35). Impedance was abnormal (normal ≤47) in 25 of the 49 patients (51%). On subsequent Bravo-pH off medications, 37 patients (75.7%) showed increased esophageal acid exposure (average DMS 36.4). Bravo-pH testing was abnormal in 84% of patients with abnormal MII testing and in 67% with normal MII testing. The sensitivity and specificity of an abnormal MII-pH on medications for increased esophageal acid exposure off medications was 56.8 and 66.7%, respectively. The positive predictive value of confirming GERD from an abnormal MII-pH on medications is 84%, while the negative predictive value is 33.3%. A receiver operating characteristic (ROC) curve was generated and the area under the curve was 0.71, indicating that MII-pH on medications is a fair test (0.7–0.8) in diagnosing pathologic GERD.

Conclusion

Compared to the gold standard, MII-pH on medications does not reliably confirm the presence of GERD. Excellent outcomes with antireflux surgery are dependent on the presence of GERD; thus, patients should not be offered antireflux surgery until GERD is confirmed with pH testing off medications.
Literature
2.
go back to reference Locke GR, Talley NJ, Fett SL et al (1997) Prevalence and clinical spectrum of gastroesophageal reflux: a population-based study in Olmsted County, Minnesota. Gastroenterology 112:1448–1456CrossRefPubMed Locke GR, Talley NJ, Fett SL et al (1997) Prevalence and clinical spectrum of gastroesophageal reflux: a population-based study in Olmsted County, Minnesota. Gastroenterology 112:1448–1456CrossRefPubMed
3.
go back to reference Anderson WD 3rd, Strayer SM, Mull SR (2015) Common questions about the management of gastroesophageal reflux disease. Am Fam Physician 91(10):692–697PubMed Anderson WD 3rd, Strayer SM, Mull SR (2015) Common questions about the management of gastroesophageal reflux disease. Am Fam Physician 91(10):692–697PubMed
4.
go back to reference Kahrilas PJ, Shaheen NJ, Vaezi MF (2008) American Gastroenterological Association Institute technical review on the management of gastroesophageal reflux disease. Gastroenterology 135:1383–1413CrossRefPubMed Kahrilas PJ, Shaheen NJ, Vaezi MF (2008) American Gastroenterological Association Institute technical review on the management of gastroesophageal reflux disease. Gastroenterology 135:1383–1413CrossRefPubMed
5.
go back to reference Kenneth R, DeVault DOC (2005) Updated guidelines for the diagnosis and treatment of gastroesophageal reflux disease. Am J Gastroenterol 100:190–200CrossRef Kenneth R, DeVault DOC (2005) Updated guidelines for the diagnosis and treatment of gastroesophageal reflux disease. Am J Gastroenterol 100:190–200CrossRef
6.
go back to reference Tutuian R, Castell D (2006) Review article: complete gastro-oesophageal reflux monitoring—combined pH and impedance. Aliment Pharmacol Ther 24(Suppl 2):27–37CrossRefPubMed Tutuian R, Castell D (2006) Review article: complete gastro-oesophageal reflux monitoring—combined pH and impedance. Aliment Pharmacol Ther 24(Suppl 2):27–37CrossRefPubMed
7.
go back to reference Agrawal A, Castell D (2008) Clinical importance of impedance measurements. J Clin Gastroenterol 42:579–583CrossRefPubMed Agrawal A, Castell D (2008) Clinical importance of impedance measurements. J Clin Gastroenterol 42:579–583CrossRefPubMed
8.
go back to reference Park W, Vaezi M (2005) Esophageal impedance recording: clinical utility and limitations. Curr Gastroenterol Rep 7:182–189CrossRefPubMed Park W, Vaezi M (2005) Esophageal impedance recording: clinical utility and limitations. Curr Gastroenterol Rep 7:182–189CrossRefPubMed
9.
go back to reference Sifrim D, Blondeau K (2006) Technology insight: the role of impedancetesting for esophageal disorders. Nat Clin Pract Gastroenterol Hepatol 3:210–219CrossRefPubMed Sifrim D, Blondeau K (2006) Technology insight: the role of impedancetesting for esophageal disorders. Nat Clin Pract Gastroenterol Hepatol 3:210–219CrossRefPubMed
10.
go back to reference Peters JH, DeMeester TR, Crookes P, Oberg S, de Vos Shoop M, Hagen JA, Bremner CG (1998) The treatment of gastroesophageal reflux disease with laparoscopic Nissen fundoplication: prospective evaluation of 100 patients with “typical” symptoms. Ann Surg 228(1):40–50CrossRefPubMedPubMedCentral Peters JH, DeMeester TR, Crookes P, Oberg S, de Vos Shoop M, Hagen JA, Bremner CG (1998) The treatment of gastroesophageal reflux disease with laparoscopic Nissen fundoplication: prospective evaluation of 100 patients with “typical” symptoms. Ann Surg 228(1):40–50CrossRefPubMedPubMedCentral
11.
go back to reference Tutuian R, Mainie I, Agrawal A et al (2006) Normal values for ambulatory 24-h combined impedance-pH monitoring on acid suppressive therapy. Gastroenterology 130(Suppl 2):A171 Tutuian R, Mainie I, Agrawal A et al (2006) Normal values for ambulatory 24-h combined impedance-pH monitoring on acid suppressive therapy. Gastroenterology 130(Suppl 2):A171
12.
go back to reference Pandolfino J, Richter J, Ours T et al (2003) Ambulatory esophageal pH monitoring using a wireless system. Am J Gastroenterol 98:740–749CrossRefPubMed Pandolfino J, Richter J, Ours T et al (2003) Ambulatory esophageal pH monitoring using a wireless system. Am J Gastroenterol 98:740–749CrossRefPubMed
13.
go back to reference Campos GM, Peters JH, DeMeester TR, Öberg S, Crookes PF, Tan S, DeMeester SR, Hagen JA, Bremner CG (1999) Multivariate analysis of factors predicting outcome after laparoscopic Nissen fundoplication. J Gastrointest Surg 3(3):292–300CrossRefPubMed Campos GM, Peters JH, DeMeester TR, Öberg S, Crookes PF, Tan S, DeMeester SR, Hagen JA, Bremner CG (1999) Multivariate analysis of factors predicting outcome after laparoscopic Nissen fundoplication. J Gastrointest Surg 3(3):292–300CrossRefPubMed
14.
go back to reference Hemmink GJ, Bredenoord AJ, Weusten BL, Monkelbaan JF, Timmer R, Smout AJ (2008) Esophageal pH-impedance monitoring in patients with therapy-resistant reflux symptoms:‘on’or ‘off’proton pump inhibitor? Am J Gastroenterol 103(10):2446–2453CrossRefPubMed Hemmink GJ, Bredenoord AJ, Weusten BL, Monkelbaan JF, Timmer R, Smout AJ (2008) Esophageal pH-impedance monitoring in patients with therapy-resistant reflux symptoms:‘on’or ‘off’proton pump inhibitor? Am J Gastroenterol 103(10):2446–2453CrossRefPubMed
15.
go back to reference Ayazi S, Lipham JC, Portale G, Peyre CG, Streets CG, Leers JM, DeMeester SR, Banki F, Chan LS, Hagen JA, DeMeester TR (2009) Bravo catheter-free pH monitoring: normal values, concordance, optimal diagnostic thresholds, and accuracy. Clin Gastroenterol Hepatol 7(1):60–67CrossRefPubMed Ayazi S, Lipham JC, Portale G, Peyre CG, Streets CG, Leers JM, DeMeester SR, Banki F, Chan LS, Hagen JA, DeMeester TR (2009) Bravo catheter-free pH monitoring: normal values, concordance, optimal diagnostic thresholds, and accuracy. Clin Gastroenterol Hepatol 7(1):60–67CrossRefPubMed
16.
go back to reference Zentilin P, Iiritano E, Dulbecco P, Bilardi C, Savarino E, De Conca S, Parodi A, Reglioni S, Vigneri S, Savarino V (2006) Normal values of 24-h ambulatory intraluminal impedance combined with pH-metry in subjects eating a Mediterranean diet. Digest Liver Dis 38(4):226–232CrossRef Zentilin P, Iiritano E, Dulbecco P, Bilardi C, Savarino E, De Conca S, Parodi A, Reglioni S, Vigneri S, Savarino V (2006) Normal values of 24-h ambulatory intraluminal impedance combined with pH-metry in subjects eating a Mediterranean diet. Digest Liver Dis 38(4):226–232CrossRef
17.
go back to reference Doulami G, Triantafyllou S, Natoudi M, Albanopoulos K, Menenakos E, Filis K, Zografos G, Theodorou D (2016) Normal values of 24H multichannel intraluminal impedance pH-metry in a Greek obese population based on montreal definition of Gerd. Obes Surg 26(1):126–131CrossRefPubMed Doulami G, Triantafyllou S, Natoudi M, Albanopoulos K, Menenakos E, Filis K, Zografos G, Theodorou D (2016) Normal values of 24H multichannel intraluminal impedance pH-metry in a Greek obese population based on montreal definition of Gerd. Obes Surg 26(1):126–131CrossRefPubMed
18.
go back to reference Moawad FJ, Betteridge JD, Boger JA, Cheng FK, Belle LS, Chen YJ, Maydonovitch CL, Wong RK (2013) Reflux episodes detected by impedance in patients on and off esomeprazole: a randomised double-blinded placebo-controlled crossover study. Aliment Pharmacol Ther 37(10):1011–1018CrossRefPubMed Moawad FJ, Betteridge JD, Boger JA, Cheng FK, Belle LS, Chen YJ, Maydonovitch CL, Wong RK (2013) Reflux episodes detected by impedance in patients on and off esomeprazole: a randomised double-blinded placebo-controlled crossover study. Aliment Pharmacol Ther 37(10):1011–1018CrossRefPubMed
19.
go back to reference Zerbib F, des Varannes S, Roman S et al (2005) Normal values and day-to-day variability of 24-h ambulatory oesophageal impedance-pH monitoring in a Belgian; French cohort of healthy subjects. Aliment Pharmacol Ther 22:1011–1021CrossRefPubMed Zerbib F, des Varannes S, Roman S et al (2005) Normal values and day-to-day variability of 24-h ambulatory oesophageal impedance-pH monitoring in a Belgian; French cohort of healthy subjects. Aliment Pharmacol Ther 22:1011–1021CrossRefPubMed
20.
go back to reference Xiao YL, Lin JK, Cheung TK, Wong NY, Yang L, Hung IF, Wong BC, Chen MH (2009) Normal values of 24-hour combined esophageal multichannel intraluminal impedance and pH monitoring in the Chinese population. Digestion 79(2):109–114CrossRefPubMed Xiao YL, Lin JK, Cheung TK, Wong NY, Yang L, Hung IF, Wong BC, Chen MH (2009) Normal values of 24-hour combined esophageal multichannel intraluminal impedance and pH monitoring in the Chinese population. Digestion 79(2):109–114CrossRefPubMed
Metadata
Title
Impedance-pH monitoring on medications does not reliably confirm the presence of gastroesophageal reflux disease in patients referred for antireflux surgery
Authors
Marc A. Ward
Christy M. Dunst
Ezra N. Teitelbaum
Valerie J. Halpin
Kevin M. Reavis
Lee L. Swanström
Steven R. DeMeester
Publication date
01-02-2018
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 2/2018
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-017-5759-7

Other articles of this Issue 2/2018

Surgical Endoscopy 2/2018 Go to the issue