Evaluation of a novel permanent capped helical coil fastener in a porcine model of laparoscopic ventral hernia repair

Authors: Arnab Majumder, Mojtaba Fayezizadeh, William W. Hope, Yuri W. Novitsky

Published in: Surgical Endoscopy | Issue 12/2016

Login to get access

Abstract

Background

Existing permanent helical coil fasteners, although commonly employed for mesh fixation during laparoscopic hernia repair, are associated with peritoneal tissue attachment formation and resultant visceral complications. We evaluated attachment formation, fastener engagement, and mesh/tissue integration associated with laparoscopic fixation using a novel permanent capped helical coil fastener (HC-Capped) compared to permanent non-capped helical coil fasteners (HC-Non-Capped) in a porcine model.

Methods

Twelve female pigs underwent bilateral laparoscopic intraperitoneal fixation of Composix™ L/P Mesh (10 × 15 cm oval) with HC-Capped or HC-Non-Capped fasteners. Thirty-two fasteners were used to secure each mesh utilizing a “double-crown” technique. Laparoscopy at 30 days was used to evaluate the presence and area coverage of attachments (Diamond Score) and percentage of engaged fasteners. At 90 days, following necropsy, each mesh was evaluated for the presence, percentage, and tenacity (Butler Score) of attachments and fastener engagement. Samples were biomechanically evaluated to assess tissue integration via T-peel testing.

Results

HC-Capped fasteners demonstrated a significantly lower attachment area score compared to the HC-Non-Capped group at 30 days (0.92 ± 0.26 vs. 2.50 ± 0.29/3.00, p = 0.002) and 90 days (0.60 ± 0.22 vs. 2.08 ± 0.29/3.00, p = 0.004). At 90 days, the HC-Capped group evidenced significantly lower attachment tenacity score (1.00 ± 0.37 vs. 2.75 ± 0.33/4.00, p = 0.013). Furthermore, at 30 and 90 days, a significantly greater percentage of HC-Capped fasteners remained properly engaged (30 days: 99.7 % vs. 86.5 %, p < 0.001 and 90 days: 99.4 % vs. 74.5 %, p = 0.001). T-peel biomechanical testing demonstrated significantly greater mesh/tissue integration for HC-Capped group (2.16 ± 0.24 vs. 1.16 ± 0.29 N/cm, p = 0.038).

Conclusions

In a porcine model, HC-Capped fasteners demonstrated significantly less attachment coverage and tenacity in the early postoperative period. Furthermore, the HC-Capped cohort evidenced significantly greater percentage of properly engaged fasteners and greater mesh/tissue integration. Data suggest that shielding exposed fastener points on the visceral mesh surface with polymer caps may reduce attachment formation and aid in mesh fixation and integration.
Literature
1.
go back to reference Leblanc KA, Booth WV (1993) Laparoscopic repair of incisional abdominal hernias using expanded polytetrafluoroethylene: preliminary findings. Surg Laparosc Endosc 3:39–41PubMed Leblanc KA, Booth WV (1993) Laparoscopic repair of incisional abdominal hernias using expanded polytetrafluoroethylene: preliminary findings. Surg Laparosc Endosc 3:39–41PubMed
2.
go back to reference Forbes SS, Eskicioglu C, Mcleod RS, Okrainec A (2009) Meta-analysis of randomized controlled trials comparing open and laparoscopic ventral and incisional hernia repair with mesh. Br J Surg 96:851–858CrossRefPubMed Forbes SS, Eskicioglu C, Mcleod RS, Okrainec A (2009) Meta-analysis of randomized controlled trials comparing open and laparoscopic ventral and incisional hernia repair with mesh. Br J Surg 96:851–858CrossRefPubMed
3.
go back to reference Goodney PP, Birkmeyer CM, Birkmeyer JD (2002) Short-term outcomes of laparoscopic and open ventral hernia repair: a meta-analysis. Arch Surg 137:1161–1165CrossRefPubMed Goodney PP, Birkmeyer CM, Birkmeyer JD (2002) Short-term outcomes of laparoscopic and open ventral hernia repair: a meta-analysis. Arch Surg 137:1161–1165CrossRefPubMed
4.
go back to reference Wassenaar EB, Raymakers JT, Rakic S (2008) Impact of the mesh fixation technique on operation time in laparoscopic repair of ventral hernias. Hernia 12:23–25CrossRefPubMed Wassenaar EB, Raymakers JT, Rakic S (2008) Impact of the mesh fixation technique on operation time in laparoscopic repair of ventral hernias. Hernia 12:23–25CrossRefPubMed
5.
go back to reference Reynvoet E, Deschepper E, Rogiers X, Troisi R, Berrevoet F (2014) Laparoscopic ventral hernia repair: is there an optimal mesh fixation technique? A systematic review. Langenbecks Arch Surg 399:55–63CrossRefPubMed Reynvoet E, Deschepper E, Rogiers X, Troisi R, Berrevoet F (2014) Laparoscopic ventral hernia repair: is there an optimal mesh fixation technique? A systematic review. Langenbecks Arch Surg 399:55–63CrossRefPubMed
6.
go back to reference Schreinemacher MH, Van Barneveld KW, Peeters E, Miserez M, Gijbels MJ, Greve JW, Bouvy ND (2014) Adhesions to sutures, tackers, and glue for intraperitoneal mesh fixation: an experimental study. Hernia 18:865–872CrossRefPubMed Schreinemacher MH, Van Barneveld KW, Peeters E, Miserez M, Gijbels MJ, Greve JW, Bouvy ND (2014) Adhesions to sutures, tackers, and glue for intraperitoneal mesh fixation: an experimental study. Hernia 18:865–872CrossRefPubMed
7.
go back to reference Joels CS, Matthews BD, Kercher KW, Austin C, Norton HJ, Williams TC, Heniford BT (2005) Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture. Surg Endosc 19:780–785CrossRefPubMed Joels CS, Matthews BD, Kercher KW, Austin C, Norton HJ, Williams TC, Heniford BT (2005) Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture. Surg Endosc 19:780–785CrossRefPubMed
8.
go back to reference Wassenaar EB, Schoenmaeckers EJ, Raymakers JT, Rakic S (2010) Subsequent abdominal surgery after laparoscopic ventral and incisional hernia repair with an expanded polytetrafluoroethylene mesh: a single institution experience with 72 reoperations. Hernia 14:137–142CrossRefPubMed Wassenaar EB, Schoenmaeckers EJ, Raymakers JT, Rakic S (2010) Subsequent abdominal surgery after laparoscopic ventral and incisional hernia repair with an expanded polytetrafluoroethylene mesh: a single institution experience with 72 reoperations. Hernia 14:137–142CrossRefPubMed
9.
go back to reference Byrd JF, Agee N, Swan RZ, Lau KN, Heath JJ, Mckillop IH, Sindram D, Martinie JB, Iannitti DA (2011) Evaluation of absorbable and permanent mesh fixation devices: adhesion formation and mechanical strength. Hernia 15:553–558CrossRefPubMed Byrd JF, Agee N, Swan RZ, Lau KN, Heath JJ, Mckillop IH, Sindram D, Martinie JB, Iannitti DA (2011) Evaluation of absorbable and permanent mesh fixation devices: adhesion formation and mechanical strength. Hernia 15:553–558CrossRefPubMed
10.
go back to reference Schreinemacher MH, Van Barneveld KW, Dikmans RE, Gijbels MJ, Greve JW, Bouvy ND (2013) Coated meshes for hernia repair provide comparable intraperitoneal adhesion prevention. Surg Endosc 27:4202–4209CrossRefPubMed Schreinemacher MH, Van Barneveld KW, Dikmans RE, Gijbels MJ, Greve JW, Bouvy ND (2013) Coated meshes for hernia repair provide comparable intraperitoneal adhesion prevention. Surg Endosc 27:4202–4209CrossRefPubMed
11.
go back to reference Deeken CR, Faucher KM, Matthews BD (2012) A review of the composition, characteristics, and effectiveness of barrier mesh prostheses utilized for laparoscopic ventral hernia repair. Surg Endosc 26:566–575CrossRefPubMed Deeken CR, Faucher KM, Matthews BD (2012) A review of the composition, characteristics, and effectiveness of barrier mesh prostheses utilized for laparoscopic ventral hernia repair. Surg Endosc 26:566–575CrossRefPubMed
12.
go back to reference Zeng Q, Yu Z, You J, Zhang Q (2007) Efficacy and safety of Seprafilm for preventing postoperative abdominal adhesion: systematic review and meta-analysis. World J Surg 31:2125–2131 (discussion 2132) CrossRefPubMed Zeng Q, Yu Z, You J, Zhang Q (2007) Efficacy and safety of Seprafilm for preventing postoperative abdominal adhesion: systematic review and meta-analysis. World J Surg 31:2125–2131 (discussion 2132) CrossRefPubMed
13.
go back to reference Inoue M, Uchida K, Otake K, Nagano Y, Ide S, Hashimoto K, Matsushita K, Koike Y, Mohri Y, Kusunoki M (2013) Efficacy of Seprafilm for preventing adhesive bowel obstruction and cost-benefit analysis in pediatric patients undergoing laparotomy. J Pediatr Surg 48:1528–1534CrossRefPubMed Inoue M, Uchida K, Otake K, Nagano Y, Ide S, Hashimoto K, Matsushita K, Koike Y, Mohri Y, Kusunoki M (2013) Efficacy of Seprafilm for preventing adhesive bowel obstruction and cost-benefit analysis in pediatric patients undergoing laparotomy. J Pediatr Surg 48:1528–1534CrossRefPubMed
14.
go back to reference Li X, He J, Bian W, Li Z, Zhang W, Li D, Snedeker JG (2014) A novel silk-based artificial ligament and tricalcium phosphate/polyether ether ketone anchor for anterior cruciate ligament reconstruction—safety and efficacy in a porcine model. Acta Biomater 10:3696–3704CrossRefPubMed Li X, He J, Bian W, Li Z, Zhang W, Li D, Snedeker JG (2014) A novel silk-based artificial ligament and tricalcium phosphate/polyether ether ketone anchor for anterior cruciate ligament reconstruction—safety and efficacy in a porcine model. Acta Biomater 10:3696–3704CrossRefPubMed
15.
go back to reference Abdullah MR, Goharian A, Abdul Kadir MR, Wahit MU (2015) Biomechanical and bioactivity concepts of polyetheretherketone composites for use in orthopedic implants—a review. J Biomed Mater Res A. doi:10.1002/jbm.a.35480 PubMed Abdullah MR, Goharian A, Abdul Kadir MR, Wahit MU (2015) Biomechanical and bioactivity concepts of polyetheretherketone composites for use in orthopedic implants—a review. J Biomed Mater Res A. doi:10.​1002/​jbm.​a.​35480 PubMed
17.
go back to reference Rivard CH, Rhalmi S, Coillard C (2002) In vivo biocompatibility testing of peek polymer for a spinal implant system: a study in rabbits. J Biomed Mater Res 62:488–498CrossRefPubMed Rivard CH, Rhalmi S, Coillard C (2002) In vivo biocompatibility testing of peek polymer for a spinal implant system: a study in rabbits. J Biomed Mater Res 62:488–498CrossRefPubMed
18.
go back to reference Hallab NJ, Bao QB, Brown T (2013) Assessment of epidural versus intradiscal biocompatibility of PEEK implant debris: an in vivo rabbit model. Eur Spine J 22:2740–2751CrossRefPubMedPubMedCentral Hallab NJ, Bao QB, Brown T (2013) Assessment of epidural versus intradiscal biocompatibility of PEEK implant debris: an in vivo rabbit model. Eur Spine J 22:2740–2751CrossRefPubMedPubMedCentral
19.
go back to reference Hollinsky C, Kolbe T, Walter I, Joachim A, Sandberg S, Koch T, Rulicke T, Tuchmann A (2010) Tensile strength and adhesion formation of mesh fixation systems used in laparoscopic incisional hernia repair. Surg Endosc 24:1318–1324CrossRefPubMed Hollinsky C, Kolbe T, Walter I, Joachim A, Sandberg S, Koch T, Rulicke T, Tuchmann A (2010) Tensile strength and adhesion formation of mesh fixation systems used in laparoscopic incisional hernia repair. Surg Endosc 24:1318–1324CrossRefPubMed
20.
go back to reference Peach G, Tan LC (2008) Small bowel obstruction and perforation due to a displaced spiral tacker: a rare complication of laparoscopic inguinal hernia repair. Hernia 12:303–305CrossRefPubMed Peach G, Tan LC (2008) Small bowel obstruction and perforation due to a displaced spiral tacker: a rare complication of laparoscopic inguinal hernia repair. Hernia 12:303–305CrossRefPubMed
21.
go back to reference Ladurner R, Mussack T (2004) Small bowel perforation due to protruding spiral tackers: a rare complication in laparoscopic incisional hernia repair. Surg Endosc 18:1001PubMed Ladurner R, Mussack T (2004) Small bowel perforation due to protruding spiral tackers: a rare complication in laparoscopic incisional hernia repair. Surg Endosc 18:1001PubMed
22.
go back to reference Fitzgerald HL, Orenstein SB, Novitsky YW (2010) Small bowel obstruction owing to displaced spiral tack after laparoscopic TAPP inguinal hernia repair. Surg Laparosc Endosc Percutan Tech 20:e132–e135CrossRefPubMed Fitzgerald HL, Orenstein SB, Novitsky YW (2010) Small bowel obstruction owing to displaced spiral tack after laparoscopic TAPP inguinal hernia repair. Surg Laparosc Endosc Percutan Tech 20:e132–e135CrossRefPubMed
Metadata
Title
Evaluation of a novel permanent capped helical coil fastener in a porcine model of laparoscopic ventral hernia repair
Authors
Arnab Majumder
Mojtaba Fayezizadeh
William W. Hope
Yuri W. Novitsky
Publication date
01-12-2016
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 12/2016
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-016-4874-1

Other articles of this Issue 12/2016

Surgical Endoscopy 12/2016 Go to the issue