Skip to main content
Top
Published in: Surgical Endoscopy 5/2015

01-05-2015

A prospective trial evaluating the clinical performance of a novel surgical energy device in laparoscopic colon surgery

Authors: J. W. Milsom, K. Trencheva, T. Sonoda, G. Nandakumar, P. J. Shukla, S. Lee

Published in: Surgical Endoscopy | Issue 5/2015

Login to get access

Abstract

Introduction

The aim of this prospective human trial was to evaluate the clinical performance of a novel THUNDERBEAT (TB) energy device in laparoscopic colon surgery. This study reports the first human trial in USA with this combined energy device in colon surgery.

Methods and procedures

This is a prospective pilot study with 30 subjects undergoing left or right laparoscopic colon resection for neoplasm in a single institution. All soft tissue dissections and all vessel ligations were performed using TB. No other energy device was used within the abdomen. Recorded end-points were dissection time (from the start of colon mobilization to specimen removal), surgical procedure time, the number of times TB taken out of the abdominal cavity, intraoperative complications (bleeding at the time of mesenteric dissection or vessel ligation, thermal injury during surgery, injury of other organs), technical device problems, postoperative complications (bleeding, delayed thermal injuries, other complications within 30 days), length of hospital stay, and mortality.

Results

Thirty subjects (15 males) were enrolled in the study with median age and range 68.5 (21–86) and BMI kg/m2 25.5 (20–35). Twelve subjects underwent right and 18 left laparoscopic hemicolectomy. The mean surgical procedure time was 163 ± 86 min and for dissection using TB device 80.6 ± 35 min. Major vessel ligation was successful in all subjects. The median number of TB applications to seal inferior mesenteric artery was 3 (2–8). TB was taken out of the abdominal cavity during dissection for tip cleaning a medium number of two times/per case. No intraoperative or postoperative complications (bleeding, thermal injuries, etc.) related to use of TB were noted.

Conclusions

The TB device demonstrated efficient and successful performance at tissue dissection and vessel ligation in left and right colectomies. TB technology can be employed in complex abdominal surgery and may save time through faster dissection but comparative studies with other energy devices are needed to confirm this.
Literature
1.
go back to reference Milsom J, Trencheva K, Monette S, Pavoor R, Shukla P, Sonoda T (2012) Evaluation of the safety, efficacy, & versatility of a new surgical energy device (THUNDERBEAT) in comparison with harmonic ACE, ligasure V and enseal devices in a porcine model. J Laparoendosc Adv Surg Tech A 22(4):378–386CrossRefPubMed Milsom J, Trencheva K, Monette S, Pavoor R, Shukla P, Sonoda T (2012) Evaluation of the safety, efficacy, & versatility of a new surgical energy device (THUNDERBEAT) in comparison with harmonic ACE, ligasure V and enseal devices in a porcine model. J Laparoendosc Adv Surg Tech A 22(4):378–386CrossRefPubMed
2.
go back to reference Seehofer D, Mogl M, Boas-Knoop S, Unger J, Schirmeier A, Chopra S, Eurich D (2012) Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5 mm sealing and cutting instruments. Surg Endosc 26(9):2541–2549. doi:10.1007/s00464-012-2229-0 CrossRefPubMedCentralPubMed Seehofer D, Mogl M, Boas-Knoop S, Unger J, Schirmeier A, Chopra S, Eurich D (2012) Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5 mm sealing and cutting instruments. Surg Endosc 26(9):2541–2549. doi:10.​1007/​s00464-012-2229-0 CrossRefPubMedCentralPubMed
3.
go back to reference Fagotti A, Vizzielli G, Fanfani F, Gallotta V, Rossitto C, Costantini B, Gueli-Alletti S, Avenia N, Iodice R, Scambia G (2013) A randomized study comparing the use of Thunderbeat technology vs. standard electrosurgery during laparoscopic radical hysterectomy and pelvic lymphadenectomy for gynecological cancer. J Minim Invasive Gynecol S1553-4650(13):01432–01435 Fagotti A, Vizzielli G, Fanfani F, Gallotta V, Rossitto C, Costantini B, Gueli-Alletti S, Avenia N, Iodice R, Scambia G (2013) A randomized study comparing the use of Thunderbeat technology vs. standard electrosurgery during laparoscopic radical hysterectomy and pelvic lymphadenectomy for gynecological cancer. J Minim Invasive Gynecol S1553-4650(13):01432–01435
4.
go back to reference Newcomb WL, Hope WW, Schmelzer TM et al (2009) Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices. Surg Endosc 23(1):90–96CrossRefPubMed Newcomb WL, Hope WW, Schmelzer TM et al (2009) Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices. Surg Endosc 23(1):90–96CrossRefPubMed
Metadata
Title
A prospective trial evaluating the clinical performance of a novel surgical energy device in laparoscopic colon surgery
Authors
J. W. Milsom
K. Trencheva
T. Sonoda
G. Nandakumar
P. J. Shukla
S. Lee
Publication date
01-05-2015
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 5/2015
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-014-3783-4

Other articles of this Issue 5/2015

Surgical Endoscopy 5/2015 Go to the issue