Fixation of mesh to the peritoneum using a fibrin glue: investigations with a biomechanical model and an experimental laparoscopic porcine model

Authors: C. Schug-Pass, H. Lippert, F. Köckerling

Published in: Surgical Endoscopy | Issue 12/2009

Login to get access

Abstract

Background

In recent years, the use of fibrin glue has become an established practice in several areas of surgical treatment. For example, fibrin glue is used increasingly as an alternative method for mesh fixation in hernia surgery, significantly helping to reduce the incidence of chronic pain. The experiments in this study were aimed at elucidating the extent to which tack- or suture-based permanent fixation can be replaced by fixation with fibrin glue for laparoscopic intraperitoneal repair of abdominal wall hernias.

Methods

In an initial series of experiments conducted with a biomechanical model, the strength of the fibrin glue for fixation of lightweight mesh (TiMesh light) to muscle tissue was compared with its strength of fixation to the peritoneum. In a second series of experiments, mesh was laparoscopically implanted in an established porcine model. Fibrin glue was used for mesh fixation in six animals. Laparoscopic exploration and explantation of the meshes were conducted after 4 months. Planimetric analysis was performed to investigate adhesion and shrinkage of the mesh surface.

Results

The strength of fibrin glue for fixation of mesh to the peritoneum was significantly less than for its fixation to muscle tissue (11.86 N vs. 47.88 N; p = 0.001). Three of the implanted meshes were not completely integrated, and two of these were dislocated. On the average, adhesions were seen on 16% of the mesh surfaces. The mesh shrinkage rate was 24.2%.

Conclusion

Mesh fixation alone to the undamaged peritoneum in the intraperitoneal region cannot be recommended because of the risk for dislocation. Additional fixation using sutures, tacks, or both is needed until the mesh material is completely integrated.
Literature
3.
go back to reference Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233:18–25CrossRefPubMed Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233:18–25CrossRefPubMed
4.
go back to reference Katkhouda N (2004) A new technique for laparoscopic hernia repair using fibrin sealant. Surg Technol Int 12:120–126PubMed Katkhouda N (2004) A new technique for laparoscopic hernia repair using fibrin sealant. Surg Technol Int 12:120–126PubMed
5.
go back to reference Olmi S, Addis A, Domeneghini C, Scaini A, Croce E (2007) Experimental comparison of type of tissucol dilution and composite mesh (Parietex) for laparoscopic repair of groin and abdominal hernia: observational study conducted in a university laboratory. Hernia 11:211–215CrossRefPubMed Olmi S, Addis A, Domeneghini C, Scaini A, Croce E (2007) Experimental comparison of type of tissucol dilution and composite mesh (Parietex) for laparoscopic repair of groin and abdominal hernia: observational study conducted in a university laboratory. Hernia 11:211–215CrossRefPubMed
6.
go back to reference ten Hallers EJ, Jansen JA, Marres HA, Rakhorst G, Verkerke GJ (2007) Histological assessment of titanium and polypropylene fiber mesh implantation with and without fibrin tissue glue. J Biomed Mater Res A 80:372–380PubMed ten Hallers EJ, Jansen JA, Marres HA, Rakhorst G, Verkerke GJ (2007) Histological assessment of titanium and polypropylene fiber mesh implantation with and without fibrin tissue glue. J Biomed Mater Res A 80:372–380PubMed
7.
go back to reference Topart P, Vandenbroucke F, Lozach P (2005) Tisseel versus tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias: a retrospective analysis. Surg Endosc 19:724–727CrossRefPubMed Topart P, Vandenbroucke F, Lozach P (2005) Tisseel versus tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias: a retrospective analysis. Surg Endosc 19:724–727CrossRefPubMed
8.
go back to reference Schwab R, Willms A, Kröger A, Becker HP (2006) Less chronic pain following mesh fixation using a fibrin sealant in TEP inguinal hernia repair. Hernia 10:272–277CrossRefPubMed Schwab R, Willms A, Kröger A, Becker HP (2006) Less chronic pain following mesh fixation using a fibrin sealant in TEP inguinal hernia repair. Hernia 10:272–277CrossRefPubMed
9.
go back to reference Lovisetto F, Zonta S, Rota E, Mazzilli M, Bardone M, Bottero L, Faillance G, Longoni M (2007) Use of human fibrin glue (Tissucol) versus staples for mesh fixation in laparoscopic transabdominal preperitoneal hernioplasty. Ann Surg 245:222–231CrossRefPubMed Lovisetto F, Zonta S, Rota E, Mazzilli M, Bardone M, Bottero L, Faillance G, Longoni M (2007) Use of human fibrin glue (Tissucol) versus staples for mesh fixation in laparoscopic transabdominal preperitoneal hernioplasty. Ann Surg 245:222–231CrossRefPubMed
10.
go back to reference Olmi S, Scaini A, Erba L, Guaglio M, Croce E (2007) Quantification of pain in laparoscopic transabdominal preperitoneal (TAPP) inguinal hernioplasty idendifies marked differences between prosthesis fixation systems. Surgery 142:40–46CrossRefPubMed Olmi S, Scaini A, Erba L, Guaglio M, Croce E (2007) Quantification of pain in laparoscopic transabdominal preperitoneal (TAPP) inguinal hernioplasty idendifies marked differences between prosthesis fixation systems. Surgery 142:40–46CrossRefPubMed
11.
go back to reference Nienhuis S, Staal E, Strobbe L, Rosman C, Groenewoud H, Bleichrodt R (2007) Chronic pain after mesh repair of inguinal hernia: a systematic review. Am J Surg 194:394–400CrossRef Nienhuis S, Staal E, Strobbe L, Rosman C, Groenewoud H, Bleichrodt R (2007) Chronic pain after mesh repair of inguinal hernia: a systematic review. Am J Surg 194:394–400CrossRef
12.
go back to reference Morris-Stiff G, Hughes LE (1998) The outcomes of nonabsorbable mesh placed within the abdominal cavity: literature review and clinical experience. J Am Coll Surg 186:352–357CrossRefPubMed Morris-Stiff G, Hughes LE (1998) The outcomes of nonabsorbable mesh placed within the abdominal cavity: literature review and clinical experience. J Am Coll Surg 186:352–357CrossRefPubMed
13.
go back to reference Rudmik LR, Schieman C, Dixon E, Debru E (2006) Laparoscopic incisional hernia repair: a review of the literature. Hernia 10:110–119CrossRefPubMed Rudmik LR, Schieman C, Dixon E, Debru E (2006) Laparoscopic incisional hernia repair: a review of the literature. Hernia 10:110–119CrossRefPubMed
14.
go back to reference Heniford BT, Park A, Ramshaw BJ, Voeller G (2003) Laparoscopic repair of ventral hernias: nine years experience with 850 consecutive hernias. Ann Surg 238:391–400PubMed Heniford BT, Park A, Ramshaw BJ, Voeller G (2003) Laparoscopic repair of ventral hernias: nine years experience with 850 consecutive hernias. Ann Surg 238:391–400PubMed
15.
go back to reference Berger D, Bientzle M, Muller A (2002) Postoperative complications after laparoscopic incisional hernia repair. Surg Endosc 16:1720–1723CrossRefPubMed Berger D, Bientzle M, Muller A (2002) Postoperative complications after laparoscopic incisional hernia repair. Surg Endosc 16:1720–1723CrossRefPubMed
16.
go back to reference Bellows CF, Berger DH (2006) Infiltration of suture sites with local anesthesia for management of pain following laparoscopic ventral hernia repairs: a prospective randomized trial. JSLS 10:345–350PubMed Bellows CF, Berger DH (2006) Infiltration of suture sites with local anesthesia for management of pain following laparoscopic ventral hernia repairs: a prospective randomized trial. JSLS 10:345–350PubMed
17.
go back to reference Le Blanc KA (2007) Laparoscopic incisional hernia repair: are transfascial sutures necessary? A review of the literature. Surg Endosc 21:508–513CrossRef Le Blanc KA (2007) Laparoscopic incisional hernia repair: are transfascial sutures necessary? A review of the literature. Surg Endosc 21:508–513CrossRef
18.
go back to reference van’t Riet M, de Vos van Steenwijk PJ, Kleinrensink GJ, Steyerberg EW, Bonjer HJ (2002) Tensile strength of mesh fixation methods in laparoscopic incisional hernia repair. Surg Endosc 16:1713–1716CrossRefPubMed van’t Riet M, de Vos van Steenwijk PJ, Kleinrensink GJ, Steyerberg EW, Bonjer HJ (2002) Tensile strength of mesh fixation methods in laparoscopic incisional hernia repair. Surg Endosc 16:1713–1716CrossRefPubMed
19.
go back to reference Winslow ER, Diaz S, Desai K, Meininger T, Soper NJ, Klingensmith ME (2004) Laparoscopic incisional hernia repair in a porcine model: what do transfixation sutures add? Surg Endosc 18:529–535CrossRefPubMed Winslow ER, Diaz S, Desai K, Meininger T, Soper NJ, Klingensmith ME (2004) Laparoscopic incisional hernia repair in a porcine model: what do transfixation sutures add? Surg Endosc 18:529–535CrossRefPubMed
20.
go back to reference Joels CS, Matthews BD, Kercher KW, Austin C, Norton HJ, Williams TC, Heniford BT (2005) Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture. Surg Endosc 19:780–785CrossRefPubMed Joels CS, Matthews BD, Kercher KW, Austin C, Norton HJ, Williams TC, Heniford BT (2005) Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture. Surg Endosc 19:780–785CrossRefPubMed
21.
go back to reference Petter-Puchner AH, Fortelny R, Mittermayr R, Öhlinger W, Redl H (2005) Fibrin sealing versus stapling of hernia meshes in an onlay model in the rat. Hernia 9:322–329CrossRefPubMed Petter-Puchner AH, Fortelny R, Mittermayr R, Öhlinger W, Redl H (2005) Fibrin sealing versus stapling of hernia meshes in an onlay model in the rat. Hernia 9:322–329CrossRefPubMed
22.
go back to reference Olmi S, Scaini A, Erba L, Bertolini A, Croce E (2007) Laparoscopic repair of inguinal hernias using an intraperitoneal onlay mesh technique and a Parietex composite mesh fixed with fibrin glue (Tissucol): personal technique and preliminary results. Surg Endosc 21:1961–1964CrossRefPubMed Olmi S, Scaini A, Erba L, Bertolini A, Croce E (2007) Laparoscopic repair of inguinal hernias using an intraperitoneal onlay mesh technique and a Parietex composite mesh fixed with fibrin glue (Tissucol): personal technique and preliminary results. Surg Endosc 21:1961–1964CrossRefPubMed
23.
go back to reference Lau H (2005) Fibrin sealant versus mechanical stapling for mesh fixation during endoscopic extraperitoneal inguinal hernioplasty. Ann Surg 242:670–675CrossRefPubMed Lau H (2005) Fibrin sealant versus mechanical stapling for mesh fixation during endoscopic extraperitoneal inguinal hernioplasty. Ann Surg 242:670–675CrossRefPubMed
24.
go back to reference Conze J, Rosch R, Klinge U, Weiss C, Anurov M, Titkowa S, Oettinger A, Schumpelick V (2004) Polypropylene in the intraabdominal position: influence of pore size and surface area. Hernia 8:365–372CrossRefPubMed Conze J, Rosch R, Klinge U, Weiss C, Anurov M, Titkowa S, Oettinger A, Schumpelick V (2004) Polypropylene in the intraabdominal position: influence of pore size and surface area. Hernia 8:365–372CrossRefPubMed
25.
go back to reference Schug-Paß C, Tamme C, Tannapfel A, Köckerling F (2006) A lightweight polypropylene mesh (TiMesh) for laparoscopic intraperitoneal repair of abdominal wall hernias: comparison of biocompatibilty with the DualMesh in an experimental study using the porcine model. Surg Endosc 20:402–409CrossRefPubMed Schug-Paß C, Tamme C, Tannapfel A, Köckerling F (2006) A lightweight polypropylene mesh (TiMesh) for laparoscopic intraperitoneal repair of abdominal wall hernias: comparison of biocompatibilty with the DualMesh in an experimental study using the porcine model. Surg Endosc 20:402–409CrossRefPubMed
26.
go back to reference Schug-Paß C, Sommerer F, Tannapfel A, Lippert H, Köckerling F (2008) Does the additional application of a polylactide film (SurgiWrap®) to a lightweight mesh (TiMesh light®) reduce adhesions in laparoscopic intraperitoneal implantation? Experimental results obtained in a porcine model. Surg Endosc 22:2433–2439CrossRefPubMed Schug-Paß C, Sommerer F, Tannapfel A, Lippert H, Köckerling F (2008) Does the additional application of a polylactide film (SurgiWrap®) to a lightweight mesh (TiMesh light®) reduce adhesions in laparoscopic intraperitoneal implantation? Experimental results obtained in a porcine model. Surg Endosc 22:2433–2439CrossRefPubMed
27.
go back to reference de Virgilio C, Elbassir M, Hidalgo A, Schaber B, French S, Amin S, Stabile BE (1999) Fibrin glue reduces the severity of intraabdominal adhesions in a rat model. Am J Surg 178:577–580CrossRefPubMed de Virgilio C, Elbassir M, Hidalgo A, Schaber B, French S, Amin S, Stabile BE (1999) Fibrin glue reduces the severity of intraabdominal adhesions in a rat model. Am J Surg 178:577–580CrossRefPubMed
28.
go back to reference Toosie K, Gallego K, Stabile BE, Schaber B, French S, de Virgilio C (2000) Fibrin glue reduces intraabdominal adhesions to synthetic mesh in a rat ventral hernia model. Am Surg 66:41–45PubMed Toosie K, Gallego K, Stabile BE, Schaber B, French S, de Virgilio C (2000) Fibrin glue reduces intraabdominal adhesions to synthetic mesh in a rat ventral hernia model. Am Surg 66:41–45PubMed
29.
go back to reference Martin-Cartes JA, Morales-Conde S, Suarez-Grau JM, Bustos-Jimenez M, Cadet-Dussort H, Socas-Macias M, Lopez-Bernal F, Alamo-Martinez JM, Tutosaus-Gomez JD, Moralez-Mendez S (2006) Prevention of peritoneal adhesions to intraperitoneal prostheses: an experimental study in pigs. Cir Esp 80:214–219CrossRefPubMed Martin-Cartes JA, Morales-Conde S, Suarez-Grau JM, Bustos-Jimenez M, Cadet-Dussort H, Socas-Macias M, Lopez-Bernal F, Alamo-Martinez JM, Tutosaus-Gomez JD, Moralez-Mendez S (2006) Prevention of peritoneal adhesions to intraperitoneal prostheses: an experimental study in pigs. Cir Esp 80:214–219CrossRefPubMed
30.
go back to reference Saed GM, Kruger M, Diamond MP (2004) Expression of transforming growth factor-beta and extracellular matrix by human peritoneal mesothelial cells and by fibroblasts from normal peritoneum and adhesions: effect of Tisseel. Wound Repair Regen 12:557–564CrossRefPubMed Saed GM, Kruger M, Diamond MP (2004) Expression of transforming growth factor-beta and extracellular matrix by human peritoneal mesothelial cells and by fibroblasts from normal peritoneum and adhesions: effect of Tisseel. Wound Repair Regen 12:557–564CrossRefPubMed
31.
go back to reference Klinge U, Conze J, Limberg W, Brücker C, Ottinger AP, Schumpelick V (1996) Pathophysiology of the abdominal wall. Chirurg 67:229–233PubMed Klinge U, Conze J, Limberg W, Brücker C, Ottinger AP, Schumpelick V (1996) Pathophysiology of the abdominal wall. Chirurg 67:229–233PubMed
32.
go back to reference Scheidbach H, Tamme C, Tannapfel A, Lippert H, Köckerling F (2004) In vivo studies comparing the biocompatibility of various polypropylene meshes and their handling properties during endoscopic total extraperitoneal (TEP) patchplasty: an experimental study in pigs. Surg Endosc 18:211–220CrossRefPubMed Scheidbach H, Tamme C, Tannapfel A, Lippert H, Köckerling F (2004) In vivo studies comparing the biocompatibility of various polypropylene meshes and their handling properties during endoscopic total extraperitoneal (TEP) patchplasty: an experimental study in pigs. Surg Endosc 18:211–220CrossRefPubMed
33.
go back to reference Scheidbach H, Tannapfel A, Schmidt U, Lippert H, Köckerling F (2004) Influence of titanium coating on the biocompatibility of heavyweight polypropylene mesh: an animal experimental model. Eur Surg Res 36:313–317CrossRefPubMed Scheidbach H, Tannapfel A, Schmidt U, Lippert H, Köckerling F (2004) Influence of titanium coating on the biocompatibility of heavyweight polypropylene mesh: an animal experimental model. Eur Surg Res 36:313–317CrossRefPubMed
34.
go back to reference Schwab R, Schumacher O, Junge K, Binnebösel M, Klinge U, Becker HP, Schumpelick V (2008) Biomechanical analysis of mesh fixation in TAPP and TEP hernia repair. Surg Endosc 22:731–738CrossRefPubMed Schwab R, Schumacher O, Junge K, Binnebösel M, Klinge U, Becker HP, Schumpelick V (2008) Biomechanical analysis of mesh fixation in TAPP and TEP hernia repair. Surg Endosc 22:731–738CrossRefPubMed
35.
go back to reference Binnebösel M, Rosch R, Junge K, Flanagan TC, Schwab R, Schumpelick V, Klinge U (2007) Biomechanical analyses of overlap and mesh dislocation in an incisional hernia model in vitro. Surgery 142:365–371CrossRefPubMed Binnebösel M, Rosch R, Junge K, Flanagan TC, Schwab R, Schumpelick V, Klinge U (2007) Biomechanical analyses of overlap and mesh dislocation in an incisional hernia model in vitro. Surgery 142:365–371CrossRefPubMed
36.
go back to reference Knook MTT, van Rosmalen AC, Yoder BE, Kleinrensink GJ, Snijders CJ, Looman CWN, van Steesel CJ (2001) Optimal mesh size for endoscopic inguinal hernia repair. Surg Endosc 15:1471–1477CrossRefPubMed Knook MTT, van Rosmalen AC, Yoder BE, Kleinrensink GJ, Snijders CJ, Looman CWN, van Steesel CJ (2001) Optimal mesh size for endoscopic inguinal hernia repair. Surg Endosc 15:1471–1477CrossRefPubMed
37.
go back to reference Klinge U, Klosterhalfen B, Birkenhauer V, Junge K, Conze J, Schumpelick V (2002) Impact of polymer pore size on the interface scar formation in a rat model. J Surg Res 103:208–214CrossRefPubMed Klinge U, Klosterhalfen B, Birkenhauer V, Junge K, Conze J, Schumpelick V (2002) Impact of polymer pore size on the interface scar formation in a rat model. J Surg Res 103:208–214CrossRefPubMed
38.
go back to reference Garrard CL, Clemets RH, Nanney L, Davidson JM, Richards WO (1999) Adhesion formation is reduced after laparoscopic surgery. Surg Endosc 13:10–13CrossRefPubMed Garrard CL, Clemets RH, Nanney L, Davidson JM, Richards WO (1999) Adhesion formation is reduced after laparoscopic surgery. Surg Endosc 13:10–13CrossRefPubMed
39.
go back to reference Gutt CN, Oniu T, Mehrabi A, Büchler W (2004) Fewer adhesions induced by laparoscopic surgery? Surg Endosc 18:898–906CrossRefPubMed Gutt CN, Oniu T, Mehrabi A, Büchler W (2004) Fewer adhesions induced by laparoscopic surgery? Surg Endosc 18:898–906CrossRefPubMed
40.
go back to reference Duron JJ (2007) Postoperative intraperitoneal adhesion pathophysiology. Colorectal Dis 9(Suppl 2):14–24CrossRefPubMed Duron JJ (2007) Postoperative intraperitoneal adhesion pathophysiology. Colorectal Dis 9(Suppl 2):14–24CrossRefPubMed
Metadata
Title
Fixation of mesh to the peritoneum using a fibrin glue: investigations with a biomechanical model and an experimental laparoscopic porcine model
Authors
C. Schug-Pass
H. Lippert
F. Köckerling
Publication date
01-12-2009
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 12/2009
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-009-0509-0

Other articles of this Issue 12/2009

Surgical Endoscopy 12/2009 Go to the issue