Skip to main content
Top
Published in: European Journal of Pediatrics 7/2017

Open Access 01-07-2017 | Original Article

Comparison of four different intraosseous access devices during simulated pediatric resuscitation. A randomized crossover manikin trial

Authors: Karol Bielski, Lukasz Szarpak, Jacek Smereka, Jerzy R. Ladny, Steve Leung, Kurt Ruetzler

Published in: European Journal of Pediatrics | Issue 7/2017

Login to get access

Abstract

The aim of the study was to compare the success rate, procedure time, and user satisfaction of pediatric NIO™ compared to Pediatric BIG®, EZ-IO®, and Jamshidi intraosseous access devices. This was a randomized, crossover manikin trial with 87 paramedics. The correct location of intraosseous access when using NIO, BIG, EZ-IO, and Jamshidi was varied and was respectively 100, 90, 90, and 90%. The time required to obtain intravascular access (time T1) in the case of NIO, BIG, EZ-IO, and Jamshidi was varied and amounted to 9 s [IQR, 8–12] for NIO, 12 s [IQR, 9–16] for BIG, 13.5 s [IQR, 11–17] for the EZ-IO, and 15 s [IQR, 13–19] for Jamshidi. The paramedics evaluated each device on the subjective ease with which they performed the procedures. The intraosseous device, which proved the easiest to use was NIO, which in the case of CPR received a median rating of 1.5 (IQR, 0.5–1.5) points.
Conclusion: Our study found that NIO® is superior to BIG®, EZ-IO®, and Jamshidi. NIO® achieved the highest first attempt success rate. NIO® also required the least time to insert and easiest to operate even by novice users. Further study is needed to test our findings in cadavers or human subjects. Based on our findings, NIO® is a promising intraosseous device for use in pediatric resuscitation.
What is Known:
Venous access in acutely ill pediatric patients, such as those undergoing cardiopulmonary resuscitation, is needed for prompt administration of drugs and fluids.
Intraosseous access is recommended by American Heart Association and European Resuscitation council if vascular access is not readily obtainable to prevent delay in treatment.
What is New:
This simulated pediatric resuscitation compared performance of four commercially available pediatric intraosseous devices in a manikin model.
NIO® outperformed BIG®, EZ-IO®, and Jamshidi in first attempt success rates and time of procedure among novice users.
Literature
1.
go back to reference Andersen LW, Berg KM, Saindon BZ, Massaro JM, Raymond TT, Berg RA, Nadkarni VM, Donnino MW, American Heart Association Get With the Guidelines–Resuscitation Investigators (2015) Time to epinephrine and survival after pediatric in-hospital cardiac arrest. JAMA 314(8):802–810. doi:10.1001/jama.2015.9678 CrossRefPubMed Andersen LW, Berg KM, Saindon BZ, Massaro JM, Raymond TT, Berg RA, Nadkarni VM, Donnino MW, American Heart Association Get With the Guidelines–Resuscitation Investigators (2015) Time to epinephrine and survival after pediatric in-hospital cardiac arrest. JAMA 314(8):802–810. doi:10.​1001/​jama.​2015.​9678 CrossRefPubMed
2.
go back to reference Black KJ, Pusic MV, Harmidy D, McGillivray D (2005) Pediatric intravenous insertion in the emergency department: bevel up or bevel down? Pediatr Emerg Care 21(11):707–711CrossRefPubMed Black KJ, Pusic MV, Harmidy D, McGillivray D (2005) Pediatric intravenous insertion in the emergency department: bevel up or bevel down? Pediatr Emerg Care 21(11):707–711CrossRefPubMed
4.
go back to reference de Caen AR, Berg MD, Chameides L, Gooden CK, Hickey RW, Scott HF, Sutton RM, Tijssen JA, Topjian A, van der Jagt ÉW, Schexnayder SM, Samson RA (2015) Part 12: pediatric advanced life support: 2015 American Heart Association Guidelines update for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 132(18 Suppl 2):S526–S542. doi:10.1161/CIR.0000000000000266 CrossRefPubMed de Caen AR, Berg MD, Chameides L, Gooden CK, Hickey RW, Scott HF, Sutton RM, Tijssen JA, Topjian A, van der Jagt ÉW, Schexnayder SM, Samson RA (2015) Part 12: pediatric advanced life support: 2015 American Heart Association Guidelines update for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 132(18 Suppl 2):S526–S542. doi:10.​1161/​CIR.​0000000000000266​ CrossRefPubMed
10.
12.
go back to reference Maconochie IK, Bingham R, Eich C, López-Herce J, Rodríguez-Núñez A, Rajka T, Van de Voorde P, Zideman DA, Biarent D, Paediatric life support section Collaborators (2015) European Resuscitation Council Guidelines for Resuscitation 2015: section 6. Paediatric life support. Resuscitation 95:223–248. doi:10.1016/j.resuscitation.2015.07.028 CrossRefPubMed Maconochie IK, Bingham R, Eich C, López-Herce J, Rodríguez-Núñez A, Rajka T, Van de Voorde P, Zideman DA, Biarent D, Paediatric life support section Collaborators (2015) European Resuscitation Council Guidelines for Resuscitation 2015: section 6. Paediatric life support. Resuscitation 95:223–248. doi:10.​1016/​j.​resuscitation.​2015.​07.​028 CrossRefPubMed
13.
14.
go back to reference Reigart JR, Chamberlain KH, Eldridge D, O'Brien ES, Freeland KD, Larsen P, Goff D, Hartzog TH (2012) Peripheral intravenous access in pediatric inpatients. Clin Pediatr (Phila) 51(5):468–472CrossRef Reigart JR, Chamberlain KH, Eldridge D, O'Brien ES, Freeland KD, Larsen P, Goff D, Hartzog TH (2012) Peripheral intravenous access in pediatric inpatients. Clin Pediatr (Phila) 51(5):468–472CrossRef
15.
go back to reference Ross EM, Mapp J, Kharod C, Wampler DA, Velasquez C, Miramontes DA (2016) Time to epinephrine in out-of-hospital cardiac arrest: a retrospective analysis of intraosseous versus intravenous access. Am J Disaster Med 11(2):119–123. doi:10.5055/ajdm.2016.0230 CrossRefPubMed Ross EM, Mapp J, Kharod C, Wampler DA, Velasquez C, Miramontes DA (2016) Time to epinephrine in out-of-hospital cardiac arrest: a retrospective analysis of intraosseous versus intravenous access. Am J Disaster Med 11(2):119–123. doi:10.​5055/​ajdm.​2016.​0230 CrossRefPubMed
19.
go back to reference Smereka J, Szarpak L, Smereka A, Leung S, Ruetzler K (2017) Evaluation of new two-thumb chest compression technique for infant CPR performed by novice physicians. A randomized, crossover, manikin trial. Am J Emerg Med 35(4):604–609. doi:10.1016/j.ajem.2016.12.045 CrossRefPubMed Smereka J, Szarpak L, Smereka A, Leung S, Ruetzler K (2017) Evaluation of new two-thumb chest compression technique for infant CPR performed by novice physicians. A randomized, crossover, manikin trial. Am J Emerg Med 35(4):604–609. doi:10.​1016/​j.​ajem.​2016.​12.​045 CrossRefPubMed
20.
21.
go back to reference Suyama J, Knutsen CC, Northington WE, Hahn M, Hostler D (2007) IO versus IV access while wearing personal protective equipment in a HazMat scenario. Prehosp Emerg Care 11(4):467–472CrossRefPubMed Suyama J, Knutsen CC, Northington WE, Hahn M, Hostler D (2007) IO versus IV access while wearing personal protective equipment in a HazMat scenario. Prehosp Emerg Care 11(4):467–472CrossRefPubMed
22.
go back to reference Szarpak L, Truszewski Z, Smereka J, Krajewski P, Fudalej M, Adamczyk P, Czyzewski L (2016) A randomized cadaver study comparing first-attempt success between tibial and humeral intraosseous insertions using NIO device by paramedics: a preliminary investigation. Medicine (Baltimore) 95(20):e3724. doi:10.1097/MD.0000000000003724 CrossRef Szarpak L, Truszewski Z, Smereka J, Krajewski P, Fudalej M, Adamczyk P, Czyzewski L (2016) A randomized cadaver study comparing first-attempt success between tibial and humeral intraosseous insertions using NIO device by paramedics: a preliminary investigation. Medicine (Baltimore) 95(20):e3724. doi:10.​1097/​MD.​0000000000003724​ CrossRef
23.
go back to reference Szarpak L, Truszewski Z, Smereka J, Czyzewski L (2016) Does the use of a chest compression system in children improve the effectiveness of chest compressions? A randomised crossover simulation pilot study. Kardiol Pol 74(12):1499–1504. doi:10.5603/KP.a2016.0107 CrossRefPubMed Szarpak L, Truszewski Z, Smereka J, Czyzewski L (2016) Does the use of a chest compression system in children improve the effectiveness of chest compressions? A randomised crossover simulation pilot study. Kardiol Pol 74(12):1499–1504. doi:10.​5603/​KP.​a2016.​0107 CrossRefPubMed
Metadata
Title
Comparison of four different intraosseous access devices during simulated pediatric resuscitation. A randomized crossover manikin trial
Authors
Karol Bielski
Lukasz Szarpak
Jacek Smereka
Jerzy R. Ladny
Steve Leung
Kurt Ruetzler
Publication date
01-07-2017
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Pediatrics / Issue 7/2017
Print ISSN: 0340-6199
Electronic ISSN: 1432-1076
DOI
https://doi.org/10.1007/s00431-017-2922-z

Other articles of this Issue 7/2017

European Journal of Pediatrics 7/2017 Go to the issue