01-02-2016 | Original Article

Validity of non-invasive point-of-care hemoglobin estimation in healthy and sick children—a method comparison study

Authors: Aditya Bhat, Amit Upadhyay, Vijay Jaiswal, Deepak Chawla, Dharamveer Singh, Mithilesh Kumar, C. P. Yadav

Published in: European Journal of Pediatrics | Issue 2/2016

Login to get access

Abstract

This study was conducted at a tertiary care center in northern India to evaluate the validity of non-invasive transcutaneous hemoglobin estimation in healthy and sick children in comparison to hemoglobin estimation by traditional lab method. A method comparison study was conducted including 150 subjects. Enrolled patients included 80 neonates with average age of 3.9 ± 2.1 days, and 70 children with average age of 5.8 ± 2 years. Each population (newborn and children) comprised of almost equal numbers of healthy and critically ill patients with shock. Hemoglobin (Hb) was estimated on enrolment by transcutaneous spectrophotometry (SpHb) and traditional automated lab analyzer (Hb-Lab). Difference between Hb levels by the two methods (called bias) was measured and analyzed using Bland-Altman method. Out of 148 data pairs analyzed, bias between SpHb and Hb-Lab was −1.52 ± 1.91 g/dl (mean ± SD). SpHb showed excellent positive correlation with Hb-Lab (r = 0.94 (p < 0.001)) and good visual agreement on Bland-Altman plots. Bias was higher in sick subjects with shock as compared to healthy ones in both neonatal and pediatric population (−2.31 ± 2.21 g/dl versus −0.77 ± 1.2 g/dl, respectively).
Conclusions: SpHb showed good accuracy and correlated well with lab estimated Hb levels in healthy children. However, in children with impaired peripheral perfusion, its diagnostic accuracy was inadequate to justify routine use for quantification of severity of anemia and making transfusion decisions solely on non-invasive estimation of hemoglobin.
What is Known:
Non-invasive hemoglobin estimation is a relatively new and novel method which has given mixed results regarding its potential efficacy in adults.
There is limited data regarding usefulness and accuracy of non-invasive Hb estimation by SpHb in sick neonates and children.
What is New:
Non-invasive Hb estimation by SpHb monitor is reasonably accurate in healthy neonates and children.
It can be used in critically ill children and neonates, but in conjunction with lab confirmation of Hb values.
Literature
1.
go back to reference Amano I, Murakami A (2013) Use of non invasive total hemoglobin measurement as a screening tool for anemia in children. Pediatr Int 55(6):803–805CrossRefPubMed Amano I, Murakami A (2013) Use of non invasive total hemoglobin measurement as a screening tool for anemia in children. Pediatr Int 55(6):803–805CrossRefPubMed
2.
go back to reference Applegate RL, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW (2012) Evaluation of pulse co-oximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology 116:65–72CrossRefPubMed Applegate RL, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW (2012) Evaluation of pulse co-oximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology 116:65–72CrossRefPubMed
3.
go back to reference Awada WN, Mohmoued MF, Radwan TM, Hussien GZ, Elkady HW (2015) Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery: a prospective cohort study. J Clin Monit Comput: 1–8 Awada WN, Mohmoued MF, Radwan TM, Hussien GZ, Elkady HW (2015) Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery: a prospective cohort study. J Clin Monit Comput: 1–8
4.
go back to reference Berkow L, Rotolo S, Mirski E (2011) Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg 113(6):1396–402CrossRefPubMed Berkow L, Rotolo S, Mirski E (2011) Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg 113(6):1396–402CrossRefPubMed
5.
go back to reference Butwick A, Hilton G, Carvalho B (2012) Non-invasive hemoglobin measurement in patients undergoing elective Caesarean section. Br J Anaesth 108:271–277CrossRefPubMed Butwick A, Hilton G, Carvalho B (2012) Non-invasive hemoglobin measurement in patients undergoing elective Caesarean section. Br J Anaesth 108:271–277CrossRefPubMed
6.
go back to reference Causey MW, Miller S, Foster A, Beekley A, Zenger D, Martin M (2011) Validation of non invasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg 201:592–598CrossRefPubMed Causey MW, Miller S, Foster A, Beekley A, Zenger D, Martin M (2011) Validation of non invasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg 201:592–598CrossRefPubMed
7.
go back to reference Colquhoun DA, Forkin KT, Durieux ME, Thiele RH (2012) Ability of the Masimo pulse co-oximeter to detect changes in hemoglobin. J Clin Monit Comput 26:69–73CrossRefPubMed Colquhoun DA, Forkin KT, Durieux ME, Thiele RH (2012) Ability of the Masimo pulse co-oximeter to detect changes in hemoglobin. J Clin Monit Comput 26:69–73CrossRefPubMed
8.
go back to reference Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O (2011) Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med 39(10):2277–2282CrossRefPubMed Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O (2011) Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med 39(10):2277–2282CrossRefPubMed
9.
go back to reference García-Soler P., Camacho-Alonso J.M. Noninvasive Monitoring of Hemoglobin Concentration in Pediatric Critical Patients. 24th Annual Meeting of the European Society of Paediatric and Neonatal Intensive Care. June 12–15, Rotterdam, Netherlands García-Soler P., Camacho-Alonso J.M. Noninvasive Monitoring of Hemoglobin Concentration in Pediatric Critical Patients. 24th Annual Meeting of the European Society of Paediatric and Neonatal Intensive Care. June 12–15, Rotterdam, Netherlands
10.
go back to reference Gayat E, Aulagnier J, Matthieu E, Boisson M, Fischler M (2012) Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies. PloS One 7(1):e30065PubMedCentralCrossRefPubMed Gayat E, Aulagnier J, Matthieu E, Boisson M, Fischler M (2012) Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies. PloS One 7(1):e30065PubMedCentralCrossRefPubMed
11.
go back to reference Gayat E, Bodin A, Sportiello C, Boisson M, Dreyfus J-F, Mathieu E et al (2011) Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med 57(4):330–333CrossRefPubMed Gayat E, Bodin A, Sportiello C, Boisson M, Dreyfus J-F, Mathieu E et al (2011) Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med 57(4):330–333CrossRefPubMed
12.
go back to reference Gibson B, Todd A, Roberts I, Pamphilon D, Rodeck C, Bolton-Maggs P (2004) Transfusion guidelines for neonates and older children. Br J Haematol 124(4):433CrossRefPubMed Gibson B, Todd A, Roberts I, Pamphilon D, Rodeck C, Bolton-Maggs P (2004) Transfusion guidelines for neonates and older children. Br J Haematol 124(4):433CrossRefPubMed
13.
go back to reference Goldstein B, Giroir B, Randolph A (2005) International pediatric sepsis consensus conference: definitions for sepsis and organ dysfunction in pediatrics. Pediatr Crit Care Med 6(1):2–8CrossRefPubMed Goldstein B, Giroir B, Randolph A (2005) International pediatric sepsis consensus conference: definitions for sepsis and organ dysfunction in pediatrics. Pediatr Crit Care Med 6(1):2–8CrossRefPubMed
14.
go back to reference Istaphanous GK, Wheeler DS, Lisco SJ, Shander A (2011) Red blood cell transfusion in critically ill children: a narrative review. Pediatr Crit Care Med J 12(2):174–183CrossRef Istaphanous GK, Wheeler DS, Lisco SJ, Shander A (2011) Red blood cell transfusion in critically ill children: a narrative review. Pediatr Crit Care Med J 12(2):174–183CrossRef
15.
go back to reference Jung YH, Lee J, Kim H-S, Shin SH, Sohn JA, Kim E-K (2013) The efficacy of noninvasive hemoglobin measurement by pulse co-oximetry in neonates. Pediatr Crit Care Med J 14(1):70–73CrossRef Jung YH, Lee J, Kim H-S, Shin SH, Sohn JA, Kim E-K (2013) The efficacy of noninvasive hemoglobin measurement by pulse co-oximetry in neonates. Pediatr Crit Care Med J 14(1):70–73CrossRef
17.
go back to reference Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B (2011) Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology 115(3):548–554CrossRefPubMed Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B (2011) Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology 115(3):548–554CrossRefPubMed
18.
go back to reference Macknet MR, Allard M, Applegate RL 2nd, Rook J (2010) The accuracy of non invasive and continuous total hemoglobin measurement by pulse co-oximetry in human subjects undergoing hemodilution. Anesth Analg 111(6):142–146CrossRef Macknet MR, Allard M, Applegate RL 2nd, Rook J (2010) The accuracy of non invasive and continuous total hemoglobin measurement by pulse co-oximetry in human subjects undergoing hemodilution. Anesth Analg 111(6):142–146CrossRef
19.
go back to reference Miller RD, Ward TA, Shiboski SC, Cohen NH (2011) A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg 112(4):858–863CrossRefPubMed Miller RD, Ward TA, Shiboski SC, Cohen NH (2011) A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg 112(4):858–863CrossRefPubMed
20.
go back to reference Nguyen BV, Vincent JL, Nowak E, Coat M, Paleiron N, Gouny P, Ould-Ahmed M, Guillouet M, Arvieux CC, Gueret G (2011) The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg 113:1052–1057CrossRefPubMed Nguyen BV, Vincent JL, Nowak E, Coat M, Paleiron N, Gouny P, Ould-Ahmed M, Guillouet M, Arvieux CC, Gueret G (2011) The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg 113:1052–1057CrossRefPubMed
22.
go back to reference Oski FA, Naiman JL (1972) Hematologic problems in the newborn, 2nd edn. WB Saunders, Philadelphia, p 13 Oski FA, Naiman JL (1972) Hematologic problems in the newborn, 2nd edn. WB Saunders, Philadelphia, p 13
23.
go back to reference Phillips MR, Khoury AL, Bortsov AV, Marzinsky A, Short KA, Cairns BA et al (2015) A noninvasive hemoglobin monitor in the pediatric intensive care unit. J Surg Res 195(1):257–262CrossRefPubMed Phillips MR, Khoury AL, Bortsov AV, Marzinsky A, Short KA, Cairns BA et al (2015) A noninvasive hemoglobin monitor in the pediatric intensive care unit. J Surg Res 195(1):257–262CrossRefPubMed
24.
go back to reference International Committee for Standardization in Haematology; Expert Panel on Hemoglobinometry (1987) Recommendations for reference method for hemoglobinometry in human blood (ICSH standard 1986) and specifications for international haemiglobincyanide reference preparation (3rd edition). Clin Lab Haematol 9(1):739 International Committee for Standardization in Haematology; Expert Panel on Hemoglobinometry (1987) Recommendations for reference method for hemoglobinometry in human blood (ICSH standard 1986) and specifications for international haemiglobincyanide reference preparation (3rd edition). Clin Lab Haematol 9(1):739
25.
go back to reference Rice MJ, Gravenstein N, Morey TE (2013) Noninvasive hemoglobin monitoring: how accurate is enough? Anesth Analg 117(4):902–907CrossRefPubMed Rice MJ, Gravenstein N, Morey TE (2013) Noninvasive hemoglobin monitoring: how accurate is enough? Anesth Analg 117(4):902–907CrossRefPubMed
26.
go back to reference Shah N, Osea EA, Martinez GJ (2013) Accuracy of noninvasive hemoglobin and invasive point of-care hemoglobin testing compared with a laboratory analyzer. Int J Lab Hematol 36(1):56–61PubMedCentralCrossRefPubMed Shah N, Osea EA, Martinez GJ (2013) Accuracy of noninvasive hemoglobin and invasive point of-care hemoglobin testing compared with a laboratory analyzer. Int J Lab Hematol 36(1):56–61PubMedCentralCrossRefPubMed
27.
go back to reference Von Schweinitz B, De Lorenzo R, Cuenca P, Anschutz R, Allen P (2015) Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill? Intern Emerg Med 10(1):55–61CrossRef Von Schweinitz B, De Lorenzo R, Cuenca P, Anschutz R, Allen P (2015) Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill? Intern Emerg Med 10(1):55–61CrossRef
Metadata
Title
Validity of non-invasive point-of-care hemoglobin estimation in healthy and sick children—a method comparison study
Authors
Aditya Bhat
Amit Upadhyay
Vijay Jaiswal
Deepak Chawla
Dharamveer Singh
Mithilesh Kumar
C. P. Yadav
Publication date
01-02-2016
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Pediatrics / Issue 2/2016
Print ISSN: 0340-6199
Electronic ISSN: 1432-1076
DOI
https://doi.org/10.1007/s00431-015-2602-9

Other articles of this Issue 2/2016

European Journal of Pediatrics 2/2016 Go to the issue