Published in: Journal of Clinical Monitoring and Computing 2/2012

01-04-2012

Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin

Authors: Douglas A. Colquhoun, Katherine T. Forkin, Marcel E. Durieux, Robert H. Thiele

Published in: Journal of Clinical Monitoring and Computing | Issue 2/2012

Login to get access

Abstract

The decision to administer blood products is complex and multifactorial. Accurate assessment of the concentration of hemoglobin [Hgb] is a key component of this evaluation. Recently a noninvasive method of continuously measuring hemoglobin (SpHb) has become available with multi-wavelength Pulse CO-Oximetry. The accuracy of this device is well documented, but the trending ability of this monitor has not been previously described. Twenty patients undergoing major thoracic and lumbar spine surgery were recruited. All patients received radial arterial lines. On the contralateral index finger, a R1 25 sensor (Rev E) was applied and connected to a Radical-7 Pulse CO-Oximeter (both Masimo Corp, Irvine, CA). Blood samples were drawn intermittently at the anesthesia provider’s discretion and were analyzed by the operating room satellite laboratory CO-Oximeter. The value of Hgb and SpHb at that time point was compared. Trend analysis was performed by the four quadrant plot technique, testing directionality of change, and Critchley’s polar plot method testing both directionality and magnitude of the change in values. Eighty-eight samples recorded at times of sufficient signal quality were available for analysis. Four quadrant plot analysis revealed 94% of data within the quadrants associated with the correct direction change, and 90% of data points lay within the analysis bounds proposed by Critchley. Pulse CO-Oximetry offers an acceptable trend monitor in patients undergoing major spine surgery. Future work should explore the ability of this device to detect large changes in hemoglobin, as well as its applicability in additional surgical and non-surgical patient populations.
Literature
1.
go back to reference Hahn RG, Li Y, Zdolsek J. Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics. Acta Anaesthesiol Scand. 2010;54(10):1233–40.PubMedCrossRef Hahn RG, Li Y, Zdolsek J. Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics. Acta Anaesthesiol Scand. 2010;54(10):1233–40.PubMedCrossRef
2.
go back to reference Macknet MR, Allard M, Applegate RL, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111(6):1424–6.PubMedCrossRef Macknet MR, Allard M, Applegate RL, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111(6):1424–6.PubMedCrossRef
3.
go back to reference Causey MW, Miller S, Foster A, et al. Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg. 2011;201(5):592–8.PubMedCrossRef Causey MW, Miller S, Foster A, et al. Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg. 2011;201(5):592–8.PubMedCrossRef
4.
go back to reference Frasca D, Dahyot-Fizelier C, Catherine K, et al. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39(10):2277–82.PubMedCrossRef Frasca D, Dahyot-Fizelier C, Catherine K, et al. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39(10):2277–82.PubMedCrossRef
5.
go back to reference Gayat E, Bodin A, Sportiello C, et al. Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med. 2011;57(4):330–3.PubMedCrossRef Gayat E, Bodin A, Sportiello C, et al. Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med. 2011;57(4):330–3.PubMedCrossRef
6.
go back to reference Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.PubMedCrossRef Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.PubMedCrossRef
7.
go back to reference Ruppel GL, Wilson HA, Gall VK, Hempkens JA. Multi-wavelength pulse oximeter is not suitable for adjusting D(LCO) measurements. Respir Care. 2011;56(8):1115–21.PubMedCrossRef Ruppel GL, Wilson HA, Gall VK, Hempkens JA. Multi-wavelength pulse oximeter is not suitable for adjusting D(LCO) measurements. Respir Care. 2011;56(8):1115–21.PubMedCrossRef
8.
go back to reference Berkow L, Rotolo S, Mirski E. Continuous Noninvasive Hemoglobin Monitoring During Complex Spine Surgery. Analg: Anesth; 2011. Berkow L, Rotolo S, Mirski E. Continuous Noninvasive Hemoglobin Monitoring During Complex Spine Surgery. Analg: Anesth; 2011.
9.
go back to reference Nguyen B-V, Vincent J-L, Nowak E, et al. The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011;113(5):1052–7.PubMedCrossRef Nguyen B-V, Vincent J-L, Nowak E, et al. The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011;113(5):1052–7.PubMedCrossRef
10.
go back to reference Morey TE, Gravenstein N, Rice MJ. Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011;113(1):89–91.PubMedCrossRef Morey TE, Gravenstein N, Rice MJ. Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011;113(1):89–91.PubMedCrossRef
11.
go back to reference Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82.PubMedCrossRef Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82.PubMedCrossRef
12.
go back to reference Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8(2):135–60.PubMedCrossRef Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8(2):135–60.PubMedCrossRef
13.
go back to reference Critchley LA, Lee A, Ho AM-H. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111(5):1180–92.PubMedCrossRef Critchley LA, Lee A, Ho AM-H. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111(5):1180–92.PubMedCrossRef
14.
go back to reference Perrino AC, Harris SN, Luther MA. Intraoperative determination of cardiac output using multiplane transesophageal echocardiography: a comparison to thermodilution. Anesthesiology. 1998;89(2):350–7.PubMedCrossRef Perrino AC, Harris SN, Luther MA. Intraoperative determination of cardiac output using multiplane transesophageal echocardiography: a comparison to thermodilution. Anesthesiology. 1998;89(2):350–7.PubMedCrossRef
15.
go back to reference Perrino AC, O’Connor T, Luther M. Transtracheal Doppler cardiac output monitoring: comparison to thermodilution during noncardiac surgery. Anesth Analg. 1994;78(6):1060–6.PubMedCrossRef Perrino AC, O’Connor T, Luther M. Transtracheal Doppler cardiac output monitoring: comparison to thermodilution during noncardiac surgery. Anesth Analg. 1994;78(6):1060–6.PubMedCrossRef
16.
go back to reference Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25(3):536–46.PubMedCrossRef Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25(3):536–46.PubMedCrossRef
18.
go back to reference Cannesson M, Le Manach Y, Hofer CK, et al. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a “gray zone” approach. Anesthesiology. 2011;115(2):231–41.PubMedCrossRef Cannesson M, Le Manach Y, Hofer CK, et al. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a “gray zone” approach. Anesthesiology. 2011;115(2):231–41.PubMedCrossRef
Metadata
Title
Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin
Authors
Douglas A. Colquhoun
Katherine T. Forkin
Marcel E. Durieux
Robert H. Thiele
Publication date
01-04-2012
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 2/2012
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-012-9335-3

Other articles of this Issue 2/2012

Journal of Clinical Monitoring and Computing 2/2012 Go to the issue