Skip to main content
Top
Published in: European Archives of Oto-Rhino-Laryngology 3/2012

01-03-2012 | Laryngology

Reliability and efficacy of a new CO2 laser hollow fiber: a prospective study of 39 patients

Authors: Marc Remacle, Andrea Ricci-Maccarini, Nayla Matar, Georges Lawson, Flavio Pieri, Vincent Bachy, Marie-Cécile Nollevaux

Published in: European Archives of Oto-Rhino-Laryngology | Issue 3/2012

Login to get access

Abstract

We present the first series of patients treated by transoral laser surgery (TLS) using the new AcuPulse 40WG CO2 laser with the FiberLase flexible waveguide (CO2 LWG) (Lumenis, Santa Clara, CA) with the objective to test its reliability and efficacy. Patients older than 18 years, with oral, pharyngo-laryngeal or tracheal benign or premalignant lesions were enrolled after signing an informed consent. This prospective study was conducted between October 2010 and May 2011 in two tertiary care university hospitals. Thirty-nine patients were enrolled in the study. The mean age was 47.9 years (range 18–86 years). There were 21 women and 18 men. Thirteen patients had hypertrophy of lymphoid tissue (palatine and or lingual), nine patients had granulomas, four patients had an exudative glottic lesion, three patients had severe dysplasia (glottic and supraglottic), three patients had leukoplakia, two patients had glottal cysts, two patients had laryngeal papilloma, two patients had bilateral paralysis of the vocal folds and one patient suffered from spasmodic dysphonia. Eighty-two percent of the procedures were performed under general anesthesia with laryngo-tracheal intubation. The CO2 fiber passed through a handpiece was used with a microscope in the majority of the procedures. The laser delivery mode parameter used was: SuperPulse or Continuous Wave. Power levels were 3–15 Watts (W), continuous delivery. Each procedure utilized one CO2 fiber which performed adequately throughout the procedure. No complications were noted with the use of this technology. A bipolar cautery was needed to control bleeding in eight procedures; all these procedures were tonsillectomies. The CO2 LWG is a safe and reliable tool for TLS. It is durable enough to last through the entire surgical procedure without the need for replacement. Its use must be tailored depending on the type and location of the lesion, the CO2 lasers tissue effects as well as the surgeon’s experience.
Literature
1.
go back to reference Strong MS, Jako GJ (1972) Laser surgery in the larynx: early clinical experience with continuous CO2 laser. Ann Otol Rhinol Laryngol 81:791PubMed Strong MS, Jako GJ (1972) Laser surgery in the larynx: early clinical experience with continuous CO2 laser. Ann Otol Rhinol Laryngol 81:791PubMed
2.
go back to reference Remacle M, Hassan F, Cohen D, Lawson G, Delos M (2005) New computer-guided scanner for improving CO2 laser assisted microincision. Eur Arch Otorhinolaryngol 262:113–119PubMedCrossRef Remacle M, Hassan F, Cohen D, Lawson G, Delos M (2005) New computer-guided scanner for improving CO2 laser assisted microincision. Eur Arch Otorhinolaryngol 262:113–119PubMedCrossRef
3.
go back to reference Ayari-Khalfallah S, Fuchsmann C, Froehlich P (2008) Thulium laser in airway diseases in children. Curr Opin Otolaryngol Head Neck Surg 16:55–59PubMedCrossRef Ayari-Khalfallah S, Fuchsmann C, Froehlich P (2008) Thulium laser in airway diseases in children. Curr Opin Otolaryngol Head Neck Surg 16:55–59PubMedCrossRef
4.
go back to reference Food and Drug Administration, Department of Health and Human Services. 510(k) clearance: K050541 Omniguide BeamPath CO2 Mark I Laser Beam Delivery System, 2005 Food and Drug Administration, Department of Health and Human Services. 510(k) clearance: K050541 Omniguide BeamPath CO2 Mark I Laser Beam Delivery System, 2005
5.
go back to reference Torres D, Weisberg O, Shapira G, Anastassiou Ch, Temelkuran B, Shurgalin M, Jacobs SA, Ahmad RU, Wang T, Kolodny U, Shapshay SM, Wang Z, Devaiah AK, Upadhyay UD, Koufman JA (2005) Omniguide photonic bandgap fibers for flexible delivery of CO2 laser energy for laryngeal and airway surgery. Proc SPIE 5686:310–321CrossRef Torres D, Weisberg O, Shapira G, Anastassiou Ch, Temelkuran B, Shurgalin M, Jacobs SA, Ahmad RU, Wang T, Kolodny U, Shapshay SM, Wang Z, Devaiah AK, Upadhyay UD, Koufman JA (2005) Omniguide photonic bandgap fibers for flexible delivery of CO2 laser energy for laryngeal and airway surgery. Proc SPIE 5686:310–321CrossRef
6.
go back to reference Shurgalin M, Anastassiou Ch (2008) A new modality for minimally invasive CO2 laser surgery: flexible hollow-core photonic bandgap fibers. Biomed Instrum Technol 42:318–325 Shurgalin M, Anastassiou Ch (2008) A new modality for minimally invasive CO2 laser surgery: flexible hollow-core photonic bandgap fibers. Biomed Instrum Technol 42:318–325
7.
go back to reference Ossoff RH (1989) Laser safety in otolaryngology- head and neck surgery: anesthetic and educational considerations for laryngeal surgery. Laryngoscope 99:1–26PubMedCrossRef Ossoff RH (1989) Laser safety in otolaryngology- head and neck surgery: anesthetic and educational considerations for laryngeal surgery. Laryngoscope 99:1–26PubMedCrossRef
8.
go back to reference Koufman JA, Rees KJ, Frazier WD, Wright SC, Kilpatrick, Halum SL, Postma GN (2007) Office-based laryngeal laser surgery: a review of 443 cases using three wavelengths. Otolaryngol Head Neck Surg 137:146–151PubMedCrossRef Koufman JA, Rees KJ, Frazier WD, Wright SC, Kilpatrick, Halum SL, Postma GN (2007) Office-based laryngeal laser surgery: a review of 443 cases using three wavelengths. Otolaryngol Head Neck Surg 137:146–151PubMedCrossRef
9.
go back to reference Zeitels SM, Akst LM, Burns JA, Hillman RE, Broadhurst MS, Anderson RR (2006) Office-based 532-nm pulsed KTP laser treatment of glottal papillomatosis and dysplasia. Ann Otol Rhinol Laryngol 115(9):679–685PubMed Zeitels SM, Akst LM, Burns JA, Hillman RE, Broadhurst MS, Anderson RR (2006) Office-based 532-nm pulsed KTP laser treatment of glottal papillomatosis and dysplasia. Ann Otol Rhinol Laryngol 115(9):679–685PubMed
10.
go back to reference Remacle M, Delos M, Lawson G, Jamart J (2002) Accuracy of histological examination following endoscopic CO2 laser assisted laryngectomy. Otorhinolaryngol Nova 12:16–20CrossRef Remacle M, Delos M, Lawson G, Jamart J (2002) Accuracy of histological examination following endoscopic CO2 laser assisted laryngectomy. Otorhinolaryngol Nova 12:16–20CrossRef
Metadata
Title
Reliability and efficacy of a new CO2 laser hollow fiber: a prospective study of 39 patients
Authors
Marc Remacle
Andrea Ricci-Maccarini
Nayla Matar
Georges Lawson
Flavio Pieri
Vincent Bachy
Marie-Cécile Nollevaux
Publication date
01-03-2012
Publisher
Springer-Verlag
Published in
European Archives of Oto-Rhino-Laryngology / Issue 3/2012
Print ISSN: 0937-4477
Electronic ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-011-1822-5

Other articles of this Issue 3/2012

European Archives of Oto-Rhino-Laryngology 3/2012 Go to the issue