Published in: Clinical Research in Cardiology 9/2014

01-09-2014 | Original Paper

Impact of respiration gating on image integration guided atrial fibrillation ablation

Authors: Emin Evren Özcan, Gabor Szeplaki, Tamas Tahin, Istvan Osztheimer, Szabolcs Szilagyi, Astrid Apor, Pal Maurovich Horvath, Hajnalka Vago, Béla Merkely, Laszlo Geller

Published in: Clinical Research in Cardiology | Issue 9/2014

Login to get access

Abstract

Background

Radiofrequency (RF) catheter ablation guided by electroanatomic mapping systems is an effective therapy for atrial fibrillation. However, it may be affected by respiration movements. The aim of this study was to determine the impact of respiratory gating on procedural parameters in patients undergoing catheter ablation of atrial fibrillation (AF).

Methods and results

One-hundred forty consecutive patients undergoing pulmonary vein isolation were admitted to study. Respiratory gating module (AccuResp algorithm, Carto3, Biosense Webster) was enabled in 70 patients and disabled in 70 patients during procedures. Successful pulmonary vein isolation and sinus rhythm were obtained in all patients. A significant reduction in total procedure times [median 77, interquartile range (IQR 66–95) min vs median 82 (IQR 72–104) min, p < 0.05] and fluoroscopy times [median 14 (IQR 9–17) min vs median 16 (IQR 12–22) min, p < 0.05] were observed in the respiratory gated group. Although ablation times (duration between the first and last ablation) were significantly shorter in respiratory gated group [median 37 (IQR 32–53) min vs median 48 (IQR 39–65) min, p < 0.05], total RF application durations were not different between two groups [median 1,554 (IQR 1,213–2,196) s vs median 1,802 (IQR 1,344–2,448) s, p = 0.11]. Difference in electroanatomical map reconstruction times was not significant [median 14 (IQR 12–16) min in gated group vs median 13 (IQR 10–18) min in nongated group, p = 0.19].

Conclusion

Respiratory gating significantly improves fluoroscopy and ablation times during electroanatomic mapping guided AF ablation. Respiratory gated maps may provide uninterrupted continuous ablation applications. Furthermore, using automatic respiratory gating module does not prolong mapping times.
Literature
1.
go back to reference Calkins H, Kuck KH, Cappato R et al (2012) 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace 14:528–606PubMedCrossRef Calkins H, Kuck KH, Cappato R et al (2012) 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace 14:528–606PubMedCrossRef
2.
go back to reference Camm AJ, Lip GY, De Caterina R et al (2012) 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation. Developed with the special contribution of the European Heart Rhythm Association. Eur Heart J 33:2719–2747PubMedCrossRef Camm AJ, Lip GY, De Caterina R et al (2012) 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation. Developed with the special contribution of the European Heart Rhythm Association. Eur Heart J 33:2719–2747PubMedCrossRef
3.
go back to reference Said SM, Braun-Dullaeus RC (2011) Comment on the European guidelines for the management of atrial fibrillation. Clin Res Cardiol 100:543–544PubMed Said SM, Braun-Dullaeus RC (2011) Comment on the European guidelines for the management of atrial fibrillation. Clin Res Cardiol 100:543–544PubMed
4.
go back to reference Chun KR, Schmidt B, Kuck KH et al (2013) Catheter ablation of atrial fibrillation in the young: insights from the German Ablation Registry. Clin Res Cardiol 102:459–468PubMed Chun KR, Schmidt B, Kuck KH et al (2013) Catheter ablation of atrial fibrillation in the young: insights from the German Ablation Registry. Clin Res Cardiol 102:459–468PubMed
5.
go back to reference Richter B, Gwechenberger M, Socas A et al (2012) Markers of oxidative stress after ablation of atrial fibrillation are associated with inflammation, delivered radiofrequency energy and early recurrence of atrial fibrillation. Clin Res Cardiol 101:217–225PubMed Richter B, Gwechenberger M, Socas A et al (2012) Markers of oxidative stress after ablation of atrial fibrillation are associated with inflammation, delivered radiofrequency energy and early recurrence of atrial fibrillation. Clin Res Cardiol 101:217–225PubMed
6.
go back to reference Noseworthy PA, Malchano ZJ, Ahmed J, Holmvang G, Ruskin JN, Reddy VY (2005) The impact of respiration on left atrial and pulmonary venous anatomy: implications for image-guided intervention. Heart Rhythm 2:1173–1178PubMedCrossRef Noseworthy PA, Malchano ZJ, Ahmed J, Holmvang G, Ruskin JN, Reddy VY (2005) The impact of respiration on left atrial and pulmonary venous anatomy: implications for image-guided intervention. Heart Rhythm 2:1173–1178PubMedCrossRef
7.
go back to reference Klemm HU, Steven D, Johnsen C et al (2007) Catheter motion during atrial ablation due to the beating heart and respiration: impact on accuracy and spatial referencing in three-dimensional mapping. Heart Rhythm 4:587–592PubMedCrossRef Klemm HU, Steven D, Johnsen C et al (2007) Catheter motion during atrial ablation due to the beating heart and respiration: impact on accuracy and spatial referencing in three-dimensional mapping. Heart Rhythm 4:587–592PubMedCrossRef
8.
go back to reference Beinart R, Kabra R, Heist KE et al (2011) Respiratory compensation improves the accuracy of electroanatomic mapping of the left atrium and pulmonary veins during atrial fibrillation ablation. J Interv Card Electrophysiol 32:105–110PubMedCrossRef Beinart R, Kabra R, Heist KE et al (2011) Respiratory compensation improves the accuracy of electroanatomic mapping of the left atrium and pulmonary veins during atrial fibrillation ablation. J Interv Card Electrophysiol 32:105–110PubMedCrossRef
9.
go back to reference Ector J, De Buck S, Loeckx D et al (2008) Changes in left atrial anatomy due to respiration: impact on three-dimensional image integration during atrial fibrillation ablation. J Cardiovasc Electrophysiol 19:828–834PubMedCrossRef Ector J, De Buck S, Loeckx D et al (2008) Changes in left atrial anatomy due to respiration: impact on three-dimensional image integration during atrial fibrillation ablation. J Cardiovasc Electrophysiol 19:828–834PubMedCrossRef
10.
go back to reference Rotter M, Takahashi Y, Sanders P et al (2005) Reduction of fluoroscopy exposure and procedure duration during ablation of atrial fibrillation using a novel anatomical navigation system. Eur Heart J 26:1415–1421PubMedCrossRef Rotter M, Takahashi Y, Sanders P et al (2005) Reduction of fluoroscopy exposure and procedure duration during ablation of atrial fibrillation using a novel anatomical navigation system. Eur Heart J 26:1415–1421PubMedCrossRef
11.
go back to reference Bertaglia E, Bella PD, Tondo C et al (2009) Image integration increases efficacy of paroxysmal atrial fibrillation catheter ablation: results from the CartoMerge Italian Registry. Europace 11:1004–1010PubMedCrossRef Bertaglia E, Bella PD, Tondo C et al (2009) Image integration increases efficacy of paroxysmal atrial fibrillation catheter ablation: results from the CartoMerge Italian Registry. Europace 11:1004–1010PubMedCrossRef
12.
go back to reference Sra J, Ratnakumar S (2008) Cardiac image registration of the left atrium and pulmonary veins. Heart Rhythm 5:609–617PubMedCrossRef Sra J, Ratnakumar S (2008) Cardiac image registration of the left atrium and pulmonary veins. Heart Rhythm 5:609–617PubMedCrossRef
13.
go back to reference de Ruvo E, Dottori S, Sciarra L et al (2013) Impact of respiration on electroanatomical mapping of the right atrium: implication for cavotricuspid isthmus ablation. J Interv Card Electrophysiol 36:33–40PubMedCrossRef de Ruvo E, Dottori S, Sciarra L et al (2013) Impact of respiration on electroanatomical mapping of the right atrium: implication for cavotricuspid isthmus ablation. J Interv Card Electrophysiol 36:33–40PubMedCrossRef
14.
go back to reference Kumar S, Morton JB, Halloran K et al (2012) Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmias. Heart Rhythm 9:1041–1047PubMedCrossRef Kumar S, Morton JB, Halloran K et al (2012) Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmias. Heart Rhythm 9:1041–1047PubMedCrossRef
15.
go back to reference Wakili R, Clauss S, Schmidt V et al (2014) Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation. Clin Res Cardiol 103:97–106PubMed Wakili R, Clauss S, Schmidt V et al (2014) Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation. Clin Res Cardiol 103:97–106PubMed
16.
go back to reference Efstathopoulos EP, Katritsis DG, Kottou S et al (2006) Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis. Europace 8:443–448PubMedCrossRef Efstathopoulos EP, Katritsis DG, Kottou S et al (2006) Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis. Europace 8:443–448PubMedCrossRef
17.
go back to reference McFadden SL, Mooney RB, Shepherd PH (2002) X-ray dose and associated risks from radiofrequency catheter ablation procedures. Br J Radiol 75:253–265PubMedCrossRef McFadden SL, Mooney RB, Shepherd PH (2002) X-ray dose and associated risks from radiofrequency catheter ablation procedures. Br J Radiol 75:253–265PubMedCrossRef
18.
go back to reference Pantos I, Koukorava C, Nirgianaki E et al (2012) Radiation exposure of the operator during cardiac catheter ablation procedures. Radiat Prot Dosimetry 150:306–311PubMedCrossRef Pantos I, Koukorava C, Nirgianaki E et al (2012) Radiation exposure of the operator during cardiac catheter ablation procedures. Radiat Prot Dosimetry 150:306–311PubMedCrossRef
19.
go back to reference Lindsay BD, Eichling JO, Ambos HD, Cain ME (1992) Radiation exposure to patients and medical personnel during radiofrequency catheter ablation for supraventricular tachycardia. Am J Cardiol 70(2):218–223PubMedCrossRef Lindsay BD, Eichling JO, Ambos HD, Cain ME (1992) Radiation exposure to patients and medical personnel during radiofrequency catheter ablation for supraventricular tachycardia. Am J Cardiol 70(2):218–223PubMedCrossRef
20.
go back to reference Lickfett L, Mahesh M, Vasamreddy C et al (2004) Radiation exposure during catheter ablation of atrial fibrillation. Circulation 110:3003–3010PubMedCrossRef Lickfett L, Mahesh M, Vasamreddy C et al (2004) Radiation exposure during catheter ablation of atrial fibrillation. Circulation 110:3003–3010PubMedCrossRef
21.
go back to reference Macle L, Weerasooriya R, Jais P et al (2003) Radiation exposure during radiofrequency catheter ablation for atrial fibrillation. Pacing Clin Electrophysiol 26:288–291PubMedCrossRef Macle L, Weerasooriya R, Jais P et al (2003) Radiation exposure during radiofrequency catheter ablation for atrial fibrillation. Pacing Clin Electrophysiol 26:288–291PubMedCrossRef
22.
go back to reference Estner HL, Deisenhofer I, Luik A et al (2006) Electrical isolation of pulmonary veins in patients with atrial fibrillation: reduction of fluoroscopy exposure and procedure duration by the use of a non-fluoroscopic navigation system (NavX). Europace 8:583–587PubMedCrossRef Estner HL, Deisenhofer I, Luik A et al (2006) Electrical isolation of pulmonary veins in patients with atrial fibrillation: reduction of fluoroscopy exposure and procedure duration by the use of a non-fluoroscopic navigation system (NavX). Europace 8:583–587PubMedCrossRef
23.
go back to reference Sporton SC, Earley MJ, Nathan AW, Schilling RJ (2004) Electroanatomic versus fluoroscopic mapping for catheter ablation procedures: a prospective randomized study. J Cardiovasc Electrophysiol 15:310–315PubMedCrossRef Sporton SC, Earley MJ, Nathan AW, Schilling RJ (2004) Electroanatomic versus fluoroscopic mapping for catheter ablation procedures: a prospective randomized study. J Cardiovasc Electrophysiol 15:310–315PubMedCrossRef
24.
go back to reference Stabile G, Scaglione M, del Greco M et al (2012) Reduced fluoroscopy exposure during ablation of atrial fibrillation using a novel electroanatomical navigation system: a multicentre experience. Europace 14:60–65PubMedCrossRef Stabile G, Scaglione M, del Greco M et al (2012) Reduced fluoroscopy exposure during ablation of atrial fibrillation using a novel electroanatomical navigation system: a multicentre experience. Europace 14:60–65PubMedCrossRef
Metadata
Title
Impact of respiration gating on image integration guided atrial fibrillation ablation
Authors
Emin Evren Özcan
Gabor Szeplaki
Tamas Tahin
Istvan Osztheimer
Szabolcs Szilagyi
Astrid Apor
Pal Maurovich Horvath
Hajnalka Vago
Béla Merkely
Laszlo Geller
Publication date
01-09-2014
Publisher
Springer Berlin Heidelberg
Published in
Clinical Research in Cardiology / Issue 9/2014
Print ISSN: 1861-0684
Electronic ISSN: 1861-0692
DOI
https://doi.org/10.1007/s00392-014-0707-1

Other articles of this Issue 9/2014

Clinical Research in Cardiology 9/2014 Go to the issue