Published in: European Radiology 5/2021

01-05-2021 | Magnetic Resonance Imaging | Magnetic Resonance

Determination of skeletal tumor extent: is an isotropic T1-weighted 3D sequence adequate?

Authors: Rodrigo Luna, Jan Fritz, Filippo del Grande, Shivani Ahlawat, Laura M. Fayad

Published in: European Radiology | Issue 5/2021

Login to get access

Abstract

Objectives

To test the hypothesis that an accelerated, T1-weighted 3D CAIPIRINHA SPACE sequence with isotropic voxel size offers a similar performance to conventional T1-weighted 2D TSE (turbo spin echo) for the evaluation of bone tumor extent and characteristics.

Methods

Thirty-four patients who underwent 3-T MRI with 3DT1 (CAIPIRINHA SPACE TSE) and 2DT1 (TSE) were included. Sequence acquisition time was reported. Two radiologists independently evaluated each technique for tumor location, size/length, tumor-to-joint distance, signal intensity, margin/extraosseous extension, and signal-to-noise (SNR) and contrast-to-noise (CNR) ratios.

Results

Tumors were located in long (20/36, 55.5%) and pelvic (16/36, 44.4%) bones. 3DT1 sequence required an average acquisition time of 235 s (± 42 s, range 156–372), while two plane 2DT1 sequences combined (coronal and axial) had an average acquisition time of 381 s (± 73 s, range 312–523). There was no difference in the measurements of tumor length and tumor-to-joint distance (p = 0.95) between 3DT1 and 2DT1 images. Tumors were hypointense (17/36, 47.2% vs 17/36, 47.2%), isointense (12/36, 33.3% vs 12/36, 33.3%), or hyperintense (7/36, 19.4% vs 7/36, 19.4%) on 3DT1 vs 2DT1, respectively. Assessment of tumor margins and extraosseous extension was similar, and there was no difference in tumor SNR or CNR (p > 0.05).

Conclusions

An accelerated 3D CAIPIRINHA SPACE T1 sequence provides comparable assessments of intramedullary bone tumor extent and similar tumor characteristics to conventional 2DT1 MRI. For the assessment of bone tumors, the isotropic volume acquisition and multiplanar reformation capability of the 3DT1 datasets can obviate the need for 2DT1 acquisitions in multiple planes.

Key Points

• 3DT1 offers an equivalent performance to 2DT1 for the assessment of bone tumor characteristics, with faster and higher resolution capability, obviating the need for acquiring 2DT1 in multiple planes.
• There was no difference in the measurements of tumor length and tumor-to-joint distance obtained on 3DT1 and 2DT1 images.
• There was no difference in signal-to-noise ratio (SNR) or contrast-to-noise ratio (CNR) measures between 3DT1 and 2DT1.
Literature
1.
go back to reference Notohamiprodjo M, Kuschel B, Horng A et al (2012) 3D-MRI of the ankle with optimized 3D-SPACE. Invest Radiol 47:231–239CrossRef Notohamiprodjo M, Kuschel B, Horng A et al (2012) 3D-MRI of the ankle with optimized 3D-SPACE. Invest Radiol 47:231–239CrossRef
2.
go back to reference Yoon MA, Hong SJ, Lee KC, Lee CH (2019) Contrast-enhanced magnetic resonance imaging of pelvic bone metastases at 3.0 T: comparison between 3-dimensional T1-weighted CAIPIRINHA-VIBE sequence and 2-dimensional T1-weighted turbo spin-echo sequence. J Comput Assist Tomogr 43:46–50CrossRef Yoon MA, Hong SJ, Lee KC, Lee CH (2019) Contrast-enhanced magnetic resonance imaging of pelvic bone metastases at 3.0 T: comparison between 3-dimensional T1-weighted CAIPIRINHA-VIBE sequence and 2-dimensional T1-weighted turbo spin-echo sequence. J Comput Assist Tomogr 43:46–50CrossRef
3.
go back to reference Shakoor D, Guermazi A, Kijowski R et al (2019) Cruciate ligament injuries of the knee: a meta-analysis of the diagnostic performance of 3D MRI. J Magn Reson Imaging 50:1545–1560CrossRef Shakoor D, Guermazi A, Kijowski R et al (2019) Cruciate ligament injuries of the knee: a meta-analysis of the diagnostic performance of 3D MRI. J Magn Reson Imaging 50:1545–1560CrossRef
4.
go back to reference Del Grande F, Subhawong T, Weber K, Aro M, Mugera C, Fayad LM (2014) Detection of soft-tissue sarcoma recurrence: added value of functional MR imaging techniques at 3.0 T. Radiology 271:499–511CrossRef Del Grande F, Subhawong T, Weber K, Aro M, Mugera C, Fayad LM (2014) Detection of soft-tissue sarcoma recurrence: added value of functional MR imaging techniques at 3.0 T. Radiology 271:499–511CrossRef
5.
go back to reference Van Dyck P, Gielen JL, Vanhoenacker FM et al (2012) Diagnostic performance of 3D SPACE for comprehensive knee joint assessment at 3 T. Insights Imaging 3:603–610CrossRef Van Dyck P, Gielen JL, Vanhoenacker FM et al (2012) Diagnostic performance of 3D SPACE for comprehensive knee joint assessment at 3 T. Insights Imaging 3:603–610CrossRef
6.
go back to reference Jung JY, Yoon YC, Kim HR, Choe BK, Wang JH, Jung JY (2013) Knee derangements: comparison of isotropic 3D fast spin-echo, isotropic 3D balanced fast field-echo, and conventional 2D fast spin-echo MR imaging. Radiology 268:802–813CrossRef Jung JY, Yoon YC, Kim HR, Choe BK, Wang JH, Jung JY (2013) Knee derangements: comparison of isotropic 3D fast spin-echo, isotropic 3D balanced fast field-echo, and conventional 2D fast spin-echo MR imaging. Radiology 268:802–813CrossRef
7.
go back to reference Kijowski R, Davis KW, Woods MA et al (2009) Knee joint: comprehensive assessment with 3D isotropic resolution fast spin-echo MR imaging--diagnostic performance compared with that of conventional MR imaging at 3.0 T. Radiology 252:486–495CrossRef Kijowski R, Davis KW, Woods MA et al (2009) Knee joint: comprehensive assessment with 3D isotropic resolution fast spin-echo MR imaging--diagnostic performance compared with that of conventional MR imaging at 3.0 T. Radiology 252:486–495CrossRef
8.
go back to reference Subhas N, Kao A, Freire M, Polster JM, Obuchowski NA, Winalski CS (2011) MRI of the knee ligaments and menisci: comparison of isotropic-resolution 3D and conventional 2D fast spin-echo sequences at 3 T. AJR Am J Roentgenol 197:442–450CrossRef Subhas N, Kao A, Freire M, Polster JM, Obuchowski NA, Winalski CS (2011) MRI of the knee ligaments and menisci: comparison of isotropic-resolution 3D and conventional 2D fast spin-echo sequences at 3 T. AJR Am J Roentgenol 197:442–450CrossRef
9.
go back to reference Lee S, Jee WH, Jung JY, Lee SY, Ryu KS, Ha KY (2015) MRI of the lumbar spine: comparison of 3D isotropic turbo spin-echo SPACE sequence versus conventional 2D sequences at 3.0 T. Acta Radiol 56:174–181CrossRef Lee S, Jee WH, Jung JY, Lee SY, Ryu KS, Ha KY (2015) MRI of the lumbar spine: comparison of 3D isotropic turbo spin-echo SPACE sequence versus conventional 2D sequences at 3.0 T. Acta Radiol 56:174–181CrossRef
10.
go back to reference Fritz J, Raithel E, Thawait GK, Gilson W, Papp DF (2016) Six-fold acceleration of high-spatial resolution 3D SPACE MRI of the knee through incoherent k-space undersampling and iterative reconstruction-first experience. Invest Radiol 51:400–409CrossRef Fritz J, Raithel E, Thawait GK, Gilson W, Papp DF (2016) Six-fold acceleration of high-spatial resolution 3D SPACE MRI of the knee through incoherent k-space undersampling and iterative reconstruction-first experience. Invest Radiol 51:400–409CrossRef
11.
go back to reference Fritz J, Fritz B, Thawait GG, Meyer H, Gilson WD, Raithel E (2016) Three-dimensional CAIPIRINHA SPACE TSE for 5-minute high-resolution MRI of the knee. Invest Radiol 51:609–617CrossRef Fritz J, Fritz B, Thawait GG, Meyer H, Gilson WD, Raithel E (2016) Three-dimensional CAIPIRINHA SPACE TSE for 5-minute high-resolution MRI of the knee. Invest Radiol 51:609–617CrossRef
12.
go back to reference Naraghi A, White LM (2012) Three-dimensional MRI of the musculoskeletal system. AJR Am J Roentgenol 199:W283–W293CrossRef Naraghi A, White LM (2012) Three-dimensional MRI of the musculoskeletal system. AJR Am J Roentgenol 199:W283–W293CrossRef
13.
go back to reference Ahlawat S, Morris C, Fayad LM (2016) Three-dimensional volumetric MRI with isotropic resolution: improved speed of acquisition, spatial resolution and assessment of lesion conspicuity in patients with recurrent soft tissue sarcoma. Skeletal Radiol 45:645–652CrossRef Ahlawat S, Morris C, Fayad LM (2016) Three-dimensional volumetric MRI with isotropic resolution: improved speed of acquisition, spatial resolution and assessment of lesion conspicuity in patients with recurrent soft tissue sarcoma. Skeletal Radiol 45:645–652CrossRef
14.
go back to reference Putta T, Gibikote S, Madhuri V, Walter N (2016) Accuracy of various MRI sequences in determining the tumour margin in musculoskeletal tumours. Pol J Radiol 81:540–548PubMedPubMedCentral Putta T, Gibikote S, Madhuri V, Walter N (2016) Accuracy of various MRI sequences in determining the tumour margin in musculoskeletal tumours. Pol J Radiol 81:540–548PubMedPubMedCentral
15.
go back to reference Vogler JB 3rd, Murphy WA (1988) Bone marrow imaging. Radiology 168:679–693CrossRef Vogler JB 3rd, Murphy WA (1988) Bone marrow imaging. Radiology 168:679–693CrossRef
16.
go back to reference Shiga NT, Del Grande F, Lardo O, Fayad LM (2013) Imaging of primary bone tumors: determination of tumor extent by non-contrast sequences. Pediatr Radiol 43:1017–1023CrossRef Shiga NT, Del Grande F, Lardo O, Fayad LM (2013) Imaging of primary bone tumors: determination of tumor extent by non-contrast sequences. Pediatr Radiol 43:1017–1023CrossRef
17.
go back to reference Hwang S, Panicek DM (2007) Magnetic resonance imaging of bone marrow in oncology, Part 1. Skeletal Radiol 36:913–920CrossRef Hwang S, Panicek DM (2007) Magnetic resonance imaging of bone marrow in oncology, Part 1. Skeletal Radiol 36:913–920CrossRef
18.
go back to reference Fayad LM, Jacobs MA, Wang X, Carrino JA, Bluemke DA (2012) Musculoskeletal tumors: how to use anatomic, functional, and metabolic MR techniques. Radiology 265:340–356CrossRef Fayad LM, Jacobs MA, Wang X, Carrino JA, Bluemke DA (2012) Musculoskeletal tumors: how to use anatomic, functional, and metabolic MR techniques. Radiology 265:340–356CrossRef
19.
go back to reference Lecouvet FE, Pasoglou V, Van Nieuwenhove S et al (2020) Shortening the acquisition time of whole-body MRI: 3D T1 gradient echo Dixon vs fast spin echo for metastatic screening in prostate cancer. Eur Radiol 30:3083–3093CrossRef Lecouvet FE, Pasoglou V, Van Nieuwenhove S et al (2020) Shortening the acquisition time of whole-body MRI: 3D T1 gradient echo Dixon vs fast spin echo for metastatic screening in prostate cancer. Eur Radiol 30:3083–3093CrossRef
20.
go back to reference Kalia V, Fritz B, Johnson R, Gilson WD, Raithel E, Fritz J (2017) CAIPIRINHA accelerated SPACE enables 10-min isotropic 3D TSE MRI of the ankle for optimized visualization of curved and oblique ligaments and tendons. Eur Radiol 27:3652–3661CrossRef Kalia V, Fritz B, Johnson R, Gilson WD, Raithel E, Fritz J (2017) CAIPIRINHA accelerated SPACE enables 10-min isotropic 3D TSE MRI of the ankle for optimized visualization of curved and oblique ligaments and tendons. Eur Radiol 27:3652–3661CrossRef
21.
go back to reference Fritz J, Ahlawat S, Fritz B et al (2019) 10-min 3D turbo spin echo MRI of the knee in children: arthroscopy-validated accuracy for the diagnosis of internal derangement. J Magn Reson Imaging 49:e139–e151CrossRef Fritz J, Ahlawat S, Fritz B et al (2019) 10-min 3D turbo spin echo MRI of the knee in children: arthroscopy-validated accuracy for the diagnosis of internal derangement. J Magn Reson Imaging 49:e139–e151CrossRef
22.
go back to reference Fritz B, Bensler S, Thawait GK, Raithel E, Stern SE, Fritz J (2019) CAIPIRINHA-accelerated 10-min 3D TSE MRI of the ankle for the diagnosis of painful ankle conditions: performance evaluation in 70 patients. Eur Radiol 29:609–619CrossRef Fritz B, Bensler S, Thawait GK, Raithel E, Stern SE, Fritz J (2019) CAIPIRINHA-accelerated 10-min 3D TSE MRI of the ankle for the diagnosis of painful ankle conditions: performance evaluation in 70 patients. Eur Radiol 29:609–619CrossRef
23.
go back to reference Del Grande F, Delcogliano M, Guglielmi R et al (2018) Fully automated 10-minute 3D CAIPIRINHA SPACE TSE MRI of the knee in adults: a multicenter, multireader, multifield-strength validation study. Invest Radiol 53:689–697CrossRef Del Grande F, Delcogliano M, Guglielmi R et al (2018) Fully automated 10-minute 3D CAIPIRINHA SPACE TSE MRI of the knee in adults: a multicenter, multireader, multifield-strength validation study. Invest Radiol 53:689–697CrossRef
24.
go back to reference Zhai H, Lv Y, Kong X, Liu X, Liu D (2019) Magnetic resonance neurography appearance and diagnostic evaluation of peripheral nerve sheath tumors. Sci Rep 9:6939CrossRef Zhai H, Lv Y, Kong X, Liu X, Liu D (2019) Magnetic resonance neurography appearance and diagnostic evaluation of peripheral nerve sheath tumors. Sci Rep 9:6939CrossRef
25.
go back to reference Wolff SD, Balaban RS (1997) Assessing contrast on MR images. Radiology 202:25–29CrossRef Wolff SD, Balaban RS (1997) Assessing contrast on MR images. Radiology 202:25–29CrossRef
26.
go back to reference Ahmad S, Stevenson J, Mangham C, Cribb G, Cool P (2014) Accuracy of magnetic resonance imaging in planning the osseous resection margins of bony tumours in the proximal femur: based on coronal T1-weighted versus STIR images. Skeletal Radiol 43:1679–1686CrossRef Ahmad S, Stevenson J, Mangham C, Cribb G, Cool P (2014) Accuracy of magnetic resonance imaging in planning the osseous resection margins of bony tumours in the proximal femur: based on coronal T1-weighted versus STIR images. Skeletal Radiol 43:1679–1686CrossRef
27.
go back to reference Onikul E, Fletcher BD, Parham DM, Chen G (1996) Accuracy of MR imaging for estimating intraosseous extent of osteosarcoma. AJR Am J Roentgenol 167:1211–1215CrossRef Onikul E, Fletcher BD, Parham DM, Chen G (1996) Accuracy of MR imaging for estimating intraosseous extent of osteosarcoma. AJR Am J Roentgenol 167:1211–1215CrossRef
28.
go back to reference Thompson MJ, Shapton JC, Punt SE, Johnson CN, Conrad EU 3rd (2018) MRI identification of the osseous extent of pediatric bone sarcomas. Clin Orthop Relat Res 476:559–564CrossRef Thompson MJ, Shapton JC, Punt SE, Johnson CN, Conrad EU 3rd (2018) MRI identification of the osseous extent of pediatric bone sarcomas. Clin Orthop Relat Res 476:559–564CrossRef
29.
go back to reference Tokuda O, Hayashi N, Matsunaga N (2004) MRI of bone tumors: fast STIR imaging as a substitute for T1-weighted contrast-enhanced fat-suppressed spin-echo imaging. J Magn Reson Imaging 19:475–481CrossRef Tokuda O, Hayashi N, Matsunaga N (2004) MRI of bone tumors: fast STIR imaging as a substitute for T1-weighted contrast-enhanced fat-suppressed spin-echo imaging. J Magn Reson Imaging 19:475–481CrossRef
30.
go back to reference Bellanova L, Schubert T, Cartiaux O et al (2014) MRI-based assessment of safe margins in tumor surgery. Sarcoma 2014:686790CrossRef Bellanova L, Schubert T, Cartiaux O et al (2014) MRI-based assessment of safe margins in tumor surgery. Sarcoma 2014:686790CrossRef
31.
go back to reference Jin T, Deng ZP, Liu WF, Xu HR, Li Y, Niu XH (2017) Magnetic resonance imaging for the assessment of long bone tumors. Chin Med J (Engl) 130:2547–2550CrossRef Jin T, Deng ZP, Liu WF, Xu HR, Li Y, Niu XH (2017) Magnetic resonance imaging for the assessment of long bone tumors. Chin Med J (Engl) 130:2547–2550CrossRef
32.
go back to reference O'Flanagan SJ, Stack JP, McGee HM, Dervan P, Hurson B (1991) Imaging of intramedullary tumour spread in osteosarcoma. A comparison of techniques. J Bone Joint Surg Br 73:998–1001CrossRef O'Flanagan SJ, Stack JP, McGee HM, Dervan P, Hurson B (1991) Imaging of intramedullary tumour spread in osteosarcoma. A comparison of techniques. J Bone Joint Surg Br 73:998–1001CrossRef
33.
go back to reference Del Grande F, Tatizawa-Shiga N, Jalali Farahani S, Chalian M, Fayad LM (2014) Chemical shift imaging: preliminary experience as an alternative sequence for defining the extent of a bone tumor. Quant Imaging Med Surg 4:173–180PubMedPubMedCentral Del Grande F, Tatizawa-Shiga N, Jalali Farahani S, Chalian M, Fayad LM (2014) Chemical shift imaging: preliminary experience as an alternative sequence for defining the extent of a bone tumor. Quant Imaging Med Surg 4:173–180PubMedPubMedCentral
34.
go back to reference Grossman JW, De Smet AA, Shinki K (2009) Comparison of the accuracy rates of 3-T and 1.5-T MRI of the knee in the diagnosis of meniscal tear. AJR Am J Roentgenol 193:509–514CrossRef Grossman JW, De Smet AA, Shinki K (2009) Comparison of the accuracy rates of 3-T and 1.5-T MRI of the knee in the diagnosis of meniscal tear. AJR Am J Roentgenol 193:509–514CrossRef
Metadata
Title
Determination of skeletal tumor extent: is an isotropic T1-weighted 3D sequence adequate?
Authors
Rodrigo Luna
Jan Fritz
Filippo del Grande
Shivani Ahlawat
Laura M. Fayad
Publication date
01-05-2021
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 5/2021
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-07394-4

Other articles of this Issue 5/2021

European Radiology 5/2021 Go to the issue