Open Access 01-10-2020 | Computed Tomography | Computed Tomography

Image quality evaluation of dual-layer spectral CT in comparison to single-layer CT in a reduced-dose setting

Authors: Thuy Duong Do, Stephan Rheinheimer, Hans-Ulrich Kauczor, Wolfram Stiller, Tim Weber, Stephan Skornitzke

Published in: European Radiology | Issue 10/2020

Login to get access

Abstract

Objectives

To quantitatively and qualitatively evaluate image quality in dual-layer CT (DLCT) compared to single-layer CT (SLCT) in the thorax, abdomen, and pelvis in a reduced-dose setting.

Methods

Intraindividual, retrospective comparisons were performed in 25 patients who received at least one acquisition of all three acquisition protocols SLCTlow (100 kVp), DLCThigh (120 kVp), and DLCTlow (120 kVp), all covering the venous-phase thorax, abdomen, and pelvis with matched CTDIvol between SLCTlow and DLCTlow. Reconstruction parameters were identical between all scans. Image quality was assessed quantitatively at 10 measurement locations in the thorax, abdomen, and pelvis by two independent observers, and subjectively with an intraindividual forced choice test between the three acquisitions. Dose-length product (DLP) and CTDIvol were extracted for dose comparison.

Results

Despite matched CTDIvol in acquisition protocols, CTDIvol and DLP were lower for SLCTlow compared to DLCTlow and DLCThigh (DLP 408.58, 444.68, 647.08 mGy·cm, respectively; p < 0.0004), as automated tube current modulation for DLCTlow reached the lower limit in the thorax (mean 66.1 mAs vs limit 65 mAs). Noise and CNR were comparable between SLCTlow and DLCTlow (p values, 0.29–0.51 and 0.05–0.20), but CT numbers were significantly higher for organs and vessels in the upper abdomen for SLCTlow compared to DLCTlow. DLCThigh had significantly better image quality (Noise and CNR). Subjective image quality was superior for DLCThigh, but no difference was found between SLCTlow and DLCTlow.

Conclusions

DLCTlow showed comparable image quality to SLCTlow, with the additional possibility of spectral post-processing. Further dose reduction seems possible by decreasing the lower limit of the tube current for the thorax.

Key Points

Clinical use of reduced-dose DLCT is feasible despite the required higher tube potential.
DLCT with reduced dose shows comparable objective and subjective image quality to reduced-dose SLCT.
Further dose reduction in the thorax might be possible by adjusting mAs thresholds.
Literature
1.
go back to reference Hickethier T, Byrtus J, Hauger M et al (2018) Utilization of virtual mono-energetic images (MonoE) derived from a dual-layer spectral detector CT (SDCT) for the assessment of abdominal arteries in venous contrast phase scans. Eur J Radiol 99:28–33CrossRef Hickethier T, Byrtus J, Hauger M et al (2018) Utilization of virtual mono-energetic images (MonoE) derived from a dual-layer spectral detector CT (SDCT) for the assessment of abdominal arteries in venous contrast phase scans. Eur J Radiol 99:28–33CrossRef
2.
go back to reference Doerner J, Luetkens JA, Iuga AI et al (2018) Poly-energetic and virtual mono-energetic images from a novel dual-layer spectral detector CT: optimization of window settings is crucial to improve subjective image quality in abdominal CT angiographies. Abdom Radiol (NY) 43:742–750CrossRef Doerner J, Luetkens JA, Iuga AI et al (2018) Poly-energetic and virtual mono-energetic images from a novel dual-layer spectral detector CT: optimization of window settings is crucial to improve subjective image quality in abdominal CT angiographies. Abdom Radiol (NY) 43:742–750CrossRef
3.
go back to reference Doerner J, Hauger M, Hickethier T et al (2017) Image quality evaluation of dual-layer spectral detector CT of the chest and comparison with conventional CT imaging. Eur J Radiol 93:52–58CrossRef Doerner J, Hauger M, Hickethier T et al (2017) Image quality evaluation of dual-layer spectral detector CT of the chest and comparison with conventional CT imaging. Eur J Radiol 93:52–58CrossRef
4.
go back to reference Borggrefe J, Neuhaus VF, Le Blanc M et al (2019) Accuracy of iodine density thresholds for the separation of vertebral bone metastases from healthy-appearing trabecular bone in spectral detector computed tomography. Eur Radiol 29:3253–3261CrossRef Borggrefe J, Neuhaus VF, Le Blanc M et al (2019) Accuracy of iodine density thresholds for the separation of vertebral bone metastases from healthy-appearing trabecular bone in spectral detector computed tomography. Eur Radiol 29:3253–3261CrossRef
5.
go back to reference Lennartz S, Laukamp KR, Neuhaus V et al (2018) Dual-layer detector CT of the head: initial experience in visualization of intracranial hemorrhage and hypodense brain lesions using virtual monoenergetic images. Eur J Radiol 108:177–183CrossRef Lennartz S, Laukamp KR, Neuhaus V et al (2018) Dual-layer detector CT of the head: initial experience in visualization of intracranial hemorrhage and hypodense brain lesions using virtual monoenergetic images. Eur J Radiol 108:177–183CrossRef
6.
go back to reference Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435CrossRef Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435CrossRef
7.
go back to reference Mettler FA Jr, Bhargavan M, Faulkner K et al (2009) Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources--1950-2007. Radiology 253:520–531CrossRef Mettler FA Jr, Bhargavan M, Faulkner K et al (2009) Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources--1950-2007. Radiology 253:520–531CrossRef
9.
go back to reference Garcia-Garrigos E, Arenas-Jimenez JJ, Sanchez-Paya J (2018) Best protocol for combined contrast-enhanced thoracic and abdominal CT for lung cancer: a single-institution randomized crossover clinical trial. AJR Am J Roentgenol 210:1226–1234CrossRef Garcia-Garrigos E, Arenas-Jimenez JJ, Sanchez-Paya J (2018) Best protocol for combined contrast-enhanced thoracic and abdominal CT for lung cancer: a single-institution randomized crossover clinical trial. AJR Am J Roentgenol 210:1226–1234CrossRef
10.
go back to reference Pahn G, Skornitzke S, Schlemmer HP, Kauczor HU, Stiller W (2016) Toward standardized quantitative image quality (IQ) assessment in computed tomography (CT): a comprehensive framework for automated and comparative IQ analysis based on ICRU report 87. Phys Med 32:104–115CrossRef Pahn G, Skornitzke S, Schlemmer HP, Kauczor HU, Stiller W (2016) Toward standardized quantitative image quality (IQ) assessment in computed tomography (CT): a comprehensive framework for automated and comparative IQ analysis based on ICRU report 87. Phys Med 32:104–115CrossRef
11.
go back to reference Gur D, Rubin DA, Kart BH et al (1997) Forced choice and ordinal discrete rating assessment of image quality: a comparison. J Digit Imaging 10:103–107CrossRef Gur D, Rubin DA, Kart BH et al (1997) Forced choice and ordinal discrete rating assessment of image quality: a comparison. J Digit Imaging 10:103–107CrossRef
13.
go back to reference Brink JA, Morin RL (2012) Size-specific dose estimation for CT: how should it be used and what does it mean? Radiology 265:666–668CrossRef Brink JA, Morin RL (2012) Size-specific dose estimation for CT: how should it be used and what does it mean? Radiology 265:666–668CrossRef
14.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174 Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174
15.
go back to reference van Ommen F, de Jong H, Dankbaar JW, Bennink E, Leiner T, Schilham AMR (2019) Dose of CT protocols acquired in clinical routine using a dual-layer detector CT scanner: a preliminary report. Eur J Radiol 112:65–71CrossRef van Ommen F, de Jong H, Dankbaar JW, Bennink E, Leiner T, Schilham AMR (2019) Dose of CT protocols acquired in clinical routine using a dual-layer detector CT scanner: a preliminary report. Eur J Radiol 112:65–71CrossRef
16.
go back to reference Haneder S, Siedek F, Doerner J et al (2018) Thoracic-abdominal imaging with a novel dual-layer spectral detector CT: intra-individual comparison of image quality and radiation dose with 128-row single-energy acquisition. Acta Radiol 59:1458–1465CrossRef Haneder S, Siedek F, Doerner J et al (2018) Thoracic-abdominal imaging with a novel dual-layer spectral detector CT: intra-individual comparison of image quality and radiation dose with 128-row single-energy acquisition. Acta Radiol 59:1458–1465CrossRef
17.
go back to reference Nagayama Y, Nakaura T, Oda S et al (2018) Dual-layer detector CT of chest, abdomen, and pelvis with a one-third iodine dose: image quality, radiation dose, and optimal monoenergetic settings. Clin Radiol 73:1058 e1021–1058 e1029 Nagayama Y, Nakaura T, Oda S et al (2018) Dual-layer detector CT of chest, abdomen, and pelvis with a one-third iodine dose: image quality, radiation dose, and optimal monoenergetic settings. Clin Radiol 73:1058 e1021–1058 e1029
18.
go back to reference van Ommen F, Bennink E, Vlassenbroek A et al (2018) Image quality of conventional images of dual-layer SPECTRAL CT: a phantom study. Med Phys 45:3031–3042CrossRef van Ommen F, Bennink E, Vlassenbroek A et al (2018) Image quality of conventional images of dual-layer SPECTRAL CT: a phantom study. Med Phys 45:3031–3042CrossRef
19.
go back to reference Schenzle JC, Sommer WH, Neumaier K et al (2010) Dual energy CT of the chest: how about the dose? Invest Radiol 45:347–353CrossRef Schenzle JC, Sommer WH, Neumaier K et al (2010) Dual energy CT of the chest: how about the dose? Invest Radiol 45:347–353CrossRef
20.
go back to reference Uhrig M, Simons D, Kachelriess M, Pisana F, Kuchenbecker S, Schlemmer HP (2016) Advanced abdominal imaging with dual energy CT is feasible without increasing radiation dose. Cancer Imaging 16:15CrossRef Uhrig M, Simons D, Kachelriess M, Pisana F, Kuchenbecker S, Schlemmer HP (2016) Advanced abdominal imaging with dual energy CT is feasible without increasing radiation dose. Cancer Imaging 16:15CrossRef
21.
go back to reference Siegel MJ, Curtis WA, Ramirez-Giraldo JC (2016) Effects of dual-energy technique on radiation exposure and image quality in pediatric body CT. AJR Am J Roentgenol 207:826–835CrossRef Siegel MJ, Curtis WA, Ramirez-Giraldo JC (2016) Effects of dual-energy technique on radiation exposure and image quality in pediatric body CT. AJR Am J Roentgenol 207:826–835CrossRef
22.
go back to reference Singh R, Sharma A, McDermott S et al (2019) Comparison of image quality and radiation doses between rapid kV-switching and dual-source DECT techniques in the chest. Eur J Radiol 119:108639CrossRef Singh R, Sharma A, McDermott S et al (2019) Comparison of image quality and radiation doses between rapid kV-switching and dual-source DECT techniques in the chest. Eur J Radiol 119:108639CrossRef
23.
go back to reference Johnson TR (2012) Dual-energy CT: general principles. AJR Am J Roentgenol 199:S3–S8CrossRef Johnson TR (2012) Dual-energy CT: general principles. AJR Am J Roentgenol 199:S3–S8CrossRef
24.
go back to reference Boland GW, O'Malley ME, Saez M, Fernandez-del-Castillo C, Warshaw AL, Mueller PR (1999) Pancreatic-phase versus portal vein-phase helical CT of the pancreas: optimal temporal window for evaluation of pancreatic adenocarcinoma. AJR Am J Roentgenol 172:605–608CrossRef Boland GW, O'Malley ME, Saez M, Fernandez-del-Castillo C, Warshaw AL, Mueller PR (1999) Pancreatic-phase versus portal vein-phase helical CT of the pancreas: optimal temporal window for evaluation of pancreatic adenocarcinoma. AJR Am J Roentgenol 172:605–608CrossRef
25.
go back to reference Bankier AA, Kressel HY (2012) Through the looking glass revisited: the need for more meaning and less drama in the reporting of dose and dose reduction in CT. Radiology 265:4–8CrossRef Bankier AA, Kressel HY (2012) Through the looking glass revisited: the need for more meaning and less drama in the reporting of dose and dose reduction in CT. Radiology 265:4–8CrossRef
26.
go back to reference Alkadhi H, Schindera ST (2011) State of the art low-dose CT angiography of the body. Eur J Radiol 80:36–40CrossRef Alkadhi H, Schindera ST (2011) State of the art low-dose CT angiography of the body. Eur J Radiol 80:36–40CrossRef
27.
go back to reference Zopfs D, Lennartz S, Laukamp K et al (2018) Improved depiction of atherosclerotic carotid artery stenosis in virtual monoenergetic reconstructions of venous phase dual-layer computed tomography in comparison to polyenergetic reconstructions. Eur J Radiol 100:36–42CrossRef Zopfs D, Lennartz S, Laukamp K et al (2018) Improved depiction of atherosclerotic carotid artery stenosis in virtual monoenergetic reconstructions of venous phase dual-layer computed tomography in comparison to polyenergetic reconstructions. Eur J Radiol 100:36–42CrossRef
28.
go back to reference Tabari A, Ramandeep S, Khera RD et al (2019) Can fully iterative reconstruction technique enable routine abdominal CT at less than 1 mSv? Eur J Radiol Open 6:225–230CrossRef Tabari A, Ramandeep S, Khera RD et al (2019) Can fully iterative reconstruction technique enable routine abdominal CT at less than 1 mSv? Eur J Radiol Open 6:225–230CrossRef
Metadata
Title
Image quality evaluation of dual-layer spectral CT in comparison to single-layer CT in a reduced-dose setting
Authors
Thuy Duong Do
Stephan Rheinheimer
Hans-Ulrich Kauczor
Wolfram Stiller
Tim Weber
Stephan Skornitzke
Publication date
01-10-2020
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 10/2020
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-06894-7

Other articles of this Issue 10/2020

European Radiology 10/2020 Go to the issue