Skip to main content
Top
Published in: European Radiology 5/2020

01-05-2020 | Computed Tomography | Breast

Comparison of propagation-based CT using synchrotron radiation and conventional cone-beam CT for breast imaging

Authors: Seyedamir Tavakoli Taba, Patrycja Baran, Yakov I. Nesterets, Serena Pacile, Susanne Wienbeck, Christian Dullin, Konstantin Pavlov, Anton Maksimenko, Darren Lockie, Sheridan C. Mayo, Harry M. Quiney, Diego Dreossi, Fulvia Arfelli, Giuliana Tromba, Sarah Lewis, Timur E. Gureyev, Patrick C. Brennan

Published in: European Radiology | Issue 5/2020

Login to get access

Abstract

Objectives

To evaluate and compare the image quality of propagation-based phase-contrast computed tomography (PB-CT) using synchrotron radiation and conventional cone-beam breast computed tomography (CBBCT) based on various radiological image quality criteria.

Methods

Eight excised breast tissue samples of various sizes and containing different lesion types were scanned using PB-CT at a synchrotron facility and using CBBCT at a university-affiliated breast imaging centre. PB-CT scans were performed at two different mean glandular dose (MGD) levels: standard (5.8 mGy) and low (1.5 mGy), for comparison with CBBCT scans at the standard MGD (5.8 mGy). Image quality assessment was carried out using six quality criteria and six independent medical imaging experts in a reading room with mammography workstations. The interobserver agreement between readers was evaluated using intraclass correlation coefficient (ICC), and image quality was compared between the two breast imaging modalities using the area under the visual grading characteristic curve (AUCVGC).

Results

Interobserver agreement between the readers showed moderate reliability for five image criteria (ICC: ranging from 0.488 to 0.633) and low reliability for one criterion (image noise) (ICC 0.307). For five image quality criteria (overall quality, perceptible contrast, lesion sharpness, normal tissue interfaces, and calcification visibility), both standard-dose PB-CT images (AUCVGC 0.958 to 1, p ≤ .05) and low dose PB-CT images (AUCVGC 0.785 to 0.834, p ≤ .05) were of significantly higher image quality than standard-dose CBBCT images.

Conclusions

Synchrotron-based PB-CT can achieve a significantly higher radiological image quality at a substantially lower radiation dose compared with conventional CBBCT.

Key Points

• PB-CT using synchrotron radiation results in higher image quality than conventional CBBCT for breast imaging.
• PB-CT using synchrotron radiation requires a lower radiation dose than conventional CBBCT for breast imaging.
• PB-CT can help clinicians diagnose patients with breast cancer.
Literature
1.
go back to reference Australian Institute of Health and Welfare (2017) Australian Cancer Incidence and Mortality (ACIM) books. AIHW, Canberra Australian Institute of Health and Welfare (2017) Australian Cancer Incidence and Mortality (ACIM) books. AIHW, Canberra
2.
go back to reference Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A (2018) Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 68:394–424PubMedCrossRef Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A (2018) Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 68:394–424PubMedCrossRef
3.
go back to reference Tavakoli Taba S, Brennan PC, Lewis S (2019) Dynamics of breast imaging research: a global scoping review and Sino-Australian comparison case study. PLoS One 14:e0210256PubMedPubMedCentralCrossRef Tavakoli Taba S, Brennan PC, Lewis S (2019) Dynamics of breast imaging research: a global scoping review and Sino-Australian comparison case study. PLoS One 14:e0210256PubMedPubMedCentralCrossRef
4.
go back to reference Independent UK Panel on Breast Cancer Screening (2012) The benefits and harms of breast cancer screening: an independent review. Lancet 380:1778–1786 Independent UK Panel on Breast Cancer Screening (2012) The benefits and harms of breast cancer screening: an independent review. Lancet 380:1778–1786
5.
go back to reference Ciatto S, Houssami N, Bernardi D et al (2013) Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14:583–589PubMedCrossRef Ciatto S, Houssami N, Bernardi D et al (2013) Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14:583–589PubMedCrossRef
6.
go back to reference Pisano ED, Gatsonis C, Hendrick E et al (2005) Diagnostic performance of digital versus film mammography for breast-cancer screening. N Engl J Med 353:1773–1783PubMedCrossRef Pisano ED, Gatsonis C, Hendrick E et al (2005) Diagnostic performance of digital versus film mammography for breast-cancer screening. N Engl J Med 353:1773–1783PubMedCrossRef
7.
go back to reference Mandelson MT, Oestreicher N, Porter PL et al (2000) Breast density as a predictor of mammographic detection: comparison of interval- and screen-detected cancers. J Natl Cancer Inst 92:1081–1087PubMedCrossRef Mandelson MT, Oestreicher N, Porter PL et al (2000) Breast density as a predictor of mammographic detection: comparison of interval- and screen-detected cancers. J Natl Cancer Inst 92:1081–1087PubMedCrossRef
9.
go back to reference Australian Institute of Health and Welfare (2018) BreastScreen Australia monitoring report 2018. AIHW, Canberra Australian Institute of Health and Welfare (2018) BreastScreen Australia monitoring report 2018. AIHW, Canberra
10.
go back to reference Alakhras M, Bourne R, Rickard M, Ng KH, Pietrzyk M, Brennan PC (2013) Digital tomosynthesis: a new future for breast imaging? Clin Radiol 68:e225–e236PubMedCrossRef Alakhras M, Bourne R, Rickard M, Ng KH, Pietrzyk M, Brennan PC (2013) Digital tomosynthesis: a new future for breast imaging? Clin Radiol 68:e225–e236PubMedCrossRef
11.
go back to reference Skaane P, Bandos AI, Gullien R et al (2013) Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56PubMedCrossRef Skaane P, Bandos AI, Gullien R et al (2013) Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56PubMedCrossRef
12.
go back to reference Durand MA, Haas BM, Yao X et al (2015) Early clinical experience with digital breast tomosynthesis for screening mammography. Radiology 274:85–92PubMedCrossRef Durand MA, Haas BM, Yao X et al (2015) Early clinical experience with digital breast tomosynthesis for screening mammography. Radiology 274:85–92PubMedCrossRef
13.
go back to reference Taba ST, Gureyev TE, Alakhras M, Lewis S, Lockie D, Brennan PC (2018) X-ray phase-contrast technology in breast imaging: principles, options, and clinical application. AJR Am J Roentgenol 211:133–145PubMedCrossRef Taba ST, Gureyev TE, Alakhras M, Lewis S, Lockie D, Brennan PC (2018) X-ray phase-contrast technology in breast imaging: principles, options, and clinical application. AJR Am J Roentgenol 211:133–145PubMedCrossRef
14.
go back to reference Dullum JR, Lewis EC, Mayer JA (2000) Rates and correlates of discomfort associated with mammography. Radiology 214:547–552PubMedCrossRef Dullum JR, Lewis EC, Mayer JA (2000) Rates and correlates of discomfort associated with mammography. Radiology 214:547–552PubMedCrossRef
15.
go back to reference Wienbeck S, Lotz J, Fischer U (2017) Review of clinical studies and first clinical experiences with a commercially available cone-beam breast CT in Europe. Clin Imaging 42:50–59PubMedCrossRef Wienbeck S, Lotz J, Fischer U (2017) Review of clinical studies and first clinical experiences with a commercially available cone-beam breast CT in Europe. Clin Imaging 42:50–59PubMedCrossRef
16.
go back to reference Uhlig J, Uhlig A, Biggemann L, Fischer U, Lotz J, Wienbeck S (2019) Diagnostic accuracy of cone-beam breast computed tomography: a systematic review and diagnostic meta-analysis. Eur Radiol 29:1194–1202PubMedCrossRef Uhlig J, Uhlig A, Biggemann L, Fischer U, Lotz J, Wienbeck S (2019) Diagnostic accuracy of cone-beam breast computed tomography: a systematic review and diagnostic meta-analysis. Eur Radiol 29:1194–1202PubMedCrossRef
17.
go back to reference Seifert P, Conover D, Zhang Y et al (2014) Evaluation of malignant breast lesions in the diagnostic setting with cone beam breast computed tomography (breast CT): feasibility study. Breast J 20:364–374PubMedCrossRef Seifert P, Conover D, Zhang Y et al (2014) Evaluation of malignant breast lesions in the diagnostic setting with cone beam breast computed tomography (breast CT): feasibility study. Breast J 20:364–374PubMedCrossRef
18.
go back to reference He N, Wu YP, Kong Y et al (2016) The utility of breast cone-beam computed tomography, ultrasound, and digital mammography for detecting malignant breast tumors: a prospective study with 212 patients. Eur J Radiol 85:392–403PubMedCrossRef He N, Wu YP, Kong Y et al (2016) The utility of breast cone-beam computed tomography, ultrasound, and digital mammography for detecting malignant breast tumors: a prospective study with 212 patients. Eur J Radiol 85:392–403PubMedCrossRef
19.
20.
go back to reference O'Connell A, Conover DL, Zhang Y et al (2010) Cone-beam CT for breast imaging: radiation dose, breast coverage, and image quality. AJR Am J Roentgenol 195:496–509PubMedCrossRef O'Connell A, Conover DL, Zhang Y et al (2010) Cone-beam CT for breast imaging: radiation dose, breast coverage, and image quality. AJR Am J Roentgenol 195:496–509PubMedCrossRef
21.
go back to reference Wilkins S, Gureyev T, Gao D, Pogany A, Stevenson A (1996) Phase-contrast imaging using polychromatic hard X-rays. Nature 384:335–338CrossRef Wilkins S, Gureyev T, Gao D, Pogany A, Stevenson A (1996) Phase-contrast imaging using polychromatic hard X-rays. Nature 384:335–338CrossRef
22.
go back to reference Arfelli F, Assante M, Bonvicini V et al (1998) Low-dose phase contrast x-ray medical imaging. Phys Med Biol 43:2845PubMedCrossRef Arfelli F, Assante M, Bonvicini V et al (1998) Low-dose phase contrast x-ray medical imaging. Phys Med Biol 43:2845PubMedCrossRef
24.
go back to reference Nesterets YI, Gureyev TE (2014) Noise propagation in x-ray phase-contrast imaging and computed tomography. J Phys D Appl Phys 47:105402CrossRef Nesterets YI, Gureyev TE (2014) Noise propagation in x-ray phase-contrast imaging and computed tomography. J Phys D Appl Phys 47:105402CrossRef
26.
go back to reference Longo R, Arfelli F, Bellazzini R et al (2016) Towards breast tomography with synchrotron radiation at Elettra: first images. Phys Med Biol 61:1634–1649PubMedCrossRef Longo R, Arfelli F, Bellazzini R et al (2016) Towards breast tomography with synchrotron radiation at Elettra: first images. Phys Med Biol 61:1634–1649PubMedCrossRef
27.
go back to reference Longo R, Arfelli F, Bonazza D et al (2019) Advancements towards the implementation of clinical phase-contrast breast computed tomography at Elettra. J Synchrotron Radiat 26:1343–1353PubMedCrossRef Longo R, Arfelli F, Bonazza D et al (2019) Advancements towards the implementation of clinical phase-contrast breast computed tomography at Elettra. J Synchrotron Radiat 26:1343–1353PubMedCrossRef
28.
go back to reference Tavakoli Taba S, Baran P, Lewis S et al (2019) Toward improving breast cancer imaging: radiological assessment of propagation-based phase-contrast CT technology. Acad Radiol 26:e79–e89PubMedCrossRef Tavakoli Taba S, Baran P, Lewis S et al (2019) Toward improving breast cancer imaging: radiological assessment of propagation-based phase-contrast CT technology. Acad Radiol 26:e79–e89PubMedCrossRef
29.
go back to reference Baran P, Pacile S, Nesterets YI et al (2017) Optimization of propagation-based x-ray phase-contrast tomography for breast cancer imaging. Phys Med Biol 62:2315–2332PubMedCrossRef Baran P, Pacile S, Nesterets YI et al (2017) Optimization of propagation-based x-ray phase-contrast tomography for breast cancer imaging. Phys Med Biol 62:2315–2332PubMedCrossRef
30.
go back to reference Gureyev T, Mayo S, Nesterets YI et al (2014) Investigation of the imaging quality of synchrotron-based phase-contrast mammographic tomography. J Phys D Appl Phys 47:365401CrossRef Gureyev T, Mayo S, Nesterets YI et al (2014) Investigation of the imaging quality of synchrotron-based phase-contrast mammographic tomography. J Phys D Appl Phys 47:365401CrossRef
31.
go back to reference Pacile S, Brun F, Dullin C et al (2015) Clinical application of low-dose phase contrast breast CT: methods for the optimization of the reconstruction workflow. Biomed Opt Express 6:3099–3112PubMedPubMedCentralCrossRef Pacile S, Brun F, Dullin C et al (2015) Clinical application of low-dose phase contrast breast CT: methods for the optimization of the reconstruction workflow. Biomed Opt Express 6:3099–3112PubMedPubMedCentralCrossRef
32.
go back to reference Nesterets YI, Gureyev TE, Mayo SC et al (2015) A feasibility study of X-ray phase-contrast mammographic tomography at the Imaging and Medical beamline of the Australian Synchrotron. J Synchrotron Radiat 22:1509–1523PubMedCrossRef Nesterets YI, Gureyev TE, Mayo SC et al (2015) A feasibility study of X-ray phase-contrast mammographic tomography at the Imaging and Medical beamline of the Australian Synchrotron. J Synchrotron Radiat 22:1509–1523PubMedCrossRef
33.
go back to reference Brombal L, Donato S, Dreossi D et al (2018) Phase-contrast breast CT: the effect of propagation distance. Phys Med Biol 63:24 nt03CrossRef Brombal L, Donato S, Dreossi D et al (2018) Phase-contrast breast CT: the effect of propagation distance. Phys Med Biol 63:24 nt03CrossRef
34.
go back to reference Brombal L, Golosio B, Arfelli F et al (2019) Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction. J Med Imaging (Bellingham) 6:031402 Brombal L, Golosio B, Arfelli F et al (2019) Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction. J Med Imaging (Bellingham) 6:031402
35.
go back to reference Boss A (2018) Editorial comment: cone-beam and phase contrast CT: new horizons in breast imaging? Eur Radiol 28:3729–3730PubMedCrossRef Boss A (2018) Editorial comment: cone-beam and phase contrast CT: new horizons in breast imaging? Eur Radiol 28:3729–3730PubMedCrossRef
37.
go back to reference Castelli E, Tonutti M, Arfelli F et al (2011) Mammography with synchrotron radiation: first clinical experience with phase-detection technique. Radiology 259:684–694PubMedCrossRef Castelli E, Tonutti M, Arfelli F et al (2011) Mammography with synchrotron radiation: first clinical experience with phase-detection technique. Radiology 259:684–694PubMedCrossRef
38.
go back to reference Fedon C, Longo F, Mettivier G, Longo R (2015) GEANT4 for breast dosimetry: parameters optimization study. Phys Med Biol 60:N311–N323PubMedCrossRef Fedon C, Longo F, Mettivier G, Longo R (2015) GEANT4 for breast dosimetry: parameters optimization study. Phys Med Biol 60:N311–N323PubMedCrossRef
39.
go back to reference Mettivier G, Fedon C, Di Lillo F et al (2016) Glandular dose in breast computed tomography with synchrotron radiation. Phys Med Biol 61:569–587PubMedCrossRef Mettivier G, Fedon C, Di Lillo F et al (2016) Glandular dose in breast computed tomography with synchrotron radiation. Phys Med Biol 61:569–587PubMedCrossRef
40.
go back to reference Gureyev TE, Nesterets Y, Ternovski D et al (2011) Toolbox for advanced X-ray image processing. SPIE Optical Engineering + Applications. International Society for Optics and Photonics, pp 81410B-81410B-81414 Gureyev TE, Nesterets Y, Ternovski D et al (2011) Toolbox for advanced X-ray image processing. SPIE Optical Engineering + Applications. International Society for Optics and Photonics, pp 81410B-81410B-81414
41.
go back to reference Brun F, Pacilè S, Accardo A et al (2015) Enhanced and flexible software tools for X-ray computed tomography at the Italian synchrotron radiation facility Elettra. Fundam Inform 141:233–243CrossRef Brun F, Pacilè S, Accardo A et al (2015) Enhanced and flexible software tools for X-ray computed tomography at the Italian synchrotron radiation facility Elettra. Fundam Inform 141:233–243CrossRef
42.
go back to reference Paganin D, Mayo SC, Gureyev TE, Miller PR, Wilkins SW (2002) Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object. J Microsc 206:33–40PubMedCrossRef Paganin D, Mayo SC, Gureyev TE, Miller PR, Wilkins SW (2002) Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object. J Microsc 206:33–40PubMedCrossRef
43.
go back to reference Raupach R, Flohr T (2012) Performance evaluation of x-ray differential phase contrast computed tomography (PCT) with respect to medical imaging. Med Phys 39:4761–4774PubMedCrossRef Raupach R, Flohr T (2012) Performance evaluation of x-ray differential phase contrast computed tomography (PCT) with respect to medical imaging. Med Phys 39:4761–4774PubMedCrossRef
45.
go back to reference Myers GR, Thomas CDL, Paganin DM, Gureyev TE, Clement JG (2010) A general few-projection method for tomographic reconstruction of samples consisting of several distinct materials. Appl Phys Lett 96:021105CrossRef Myers GR, Thomas CDL, Paganin DM, Gureyev TE, Clement JG (2010) A general few-projection method for tomographic reconstruction of samples consisting of several distinct materials. Appl Phys Lett 96:021105CrossRef
46.
go back to reference Van der Sluis A, van der Vorst HA (1990) SIRT-and CG-type methods for the iterative solution of sparse linear least-squares problems. Linear Algebra Appl 130:257–303CrossRef Van der Sluis A, van der Vorst HA (1990) SIRT-and CG-type methods for the iterative solution of sparse linear least-squares problems. Linear Algebra Appl 130:257–303CrossRef
47.
go back to reference Båth M, Månsson LG (2007) Visual grading characteristics (VGC) analysis: a non-parametric rank-invariant statistical method for image quality evaluation. Br J Radiol 80:169–176PubMedCrossRef Båth M, Månsson LG (2007) Visual grading characteristics (VGC) analysis: a non-parametric rank-invariant statistical method for image quality evaluation. Br J Radiol 80:169–176PubMedCrossRef
48.
go back to reference Båth M, Hansson J (2016) VGC Analyzer – a software for statistical analysis of multiple-reader multiple-case visual grading characteristics (VGC) studies. Radiat Prot Dosimetry 169:46–53PubMedCrossRef Båth M, Hansson J (2016) VGC Analyzer – a software for statistical analysis of multiple-reader multiple-case visual grading characteristics (VGC) studies. Radiat Prot Dosimetry 169:46–53PubMedCrossRef
49.
go back to reference Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428PubMedCrossRef Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428PubMedCrossRef
50.
go back to reference Lai CJ, Shaw CC, Chen L et al (2007) Visibility of microcalcification in cone beam breast CT: effects of X-ray tube voltage and radiation dose. Med Phys 34:2995–3004PubMedCrossRef Lai CJ, Shaw CC, Chen L et al (2007) Visibility of microcalcification in cone beam breast CT: effects of X-ray tube voltage and radiation dose. Med Phys 34:2995–3004PubMedCrossRef
51.
go back to reference Rafferty EA, Durand MA, Conant EF et al (2016) Breast cancer screening using tomosynthesis and digital mammography in dense and nondense breasts. JAMA 315:1784–1786 Rafferty EA, Durand MA, Conant EF et al (2016) Breast cancer screening using tomosynthesis and digital mammography in dense and nondense breasts. JAMA 315:1784–1786
52.
go back to reference Wienbeck S, Fischer U, Luftner-Nagel S, Lotz J, Uhlig J (2018) Contrast-enhanced cone-beam breast-CT (CBBCT): clinical performance compared to mammography and MRI. Eur Radiol 28:3731–3741PubMedCrossRef Wienbeck S, Fischer U, Luftner-Nagel S, Lotz J, Uhlig J (2018) Contrast-enhanced cone-beam breast-CT (CBBCT): clinical performance compared to mammography and MRI. Eur Radiol 28:3731–3741PubMedCrossRef
53.
go back to reference Zhao Y, Brun E, Coan P et al (2012) High-resolution, low-dose phase contrast X-ray tomography for 3D diagnosis of human breast cancers. Proc Natl Acad Sci U S A 109:18290–18294PubMedPubMedCentralCrossRef Zhao Y, Brun E, Coan P et al (2012) High-resolution, low-dose phase contrast X-ray tomography for 3D diagnosis of human breast cancers. Proc Natl Acad Sci U S A 109:18290–18294PubMedPubMedCentralCrossRef
54.
go back to reference Longo R, Tonutti M, Rigon L et al (2014) Clinical study in phase-contrast mammography: image-quality analysis. Philos Trans A Math Phys Eng Sci 372:20130025 Longo R, Tonutti M, Rigon L et al (2014) Clinical study in phase-contrast mammography: image-quality analysis. Philos Trans A Math Phys Eng Sci 372:20130025
55.
go back to reference Lewis SJ, Gureyev TE, Baran P et al (2018) Towards clinic-friendly solutions for patient trials in breast cancer phase contrast imaging. Proc. SPIE 10718, 14th International Workshop on Breast Imaging (IWBI 2018), 107181P Lewis SJ, Gureyev TE, Baran P et al (2018) Towards clinic-friendly solutions for patient trials in breast cancer phase contrast imaging. Proc. SPIE 10718, 14th International Workshop on Breast Imaging (IWBI 2018), 107181P
56.
go back to reference Tang R, Xi Y, Chai WM et al (2011) Microbubble-based synchrotron radiation phase contrast imaging: basic study and angiography applications. Phys Med Biol 56:3503–3512PubMedCrossRef Tang R, Xi Y, Chai WM et al (2011) Microbubble-based synchrotron radiation phase contrast imaging: basic study and angiography applications. Phys Med Biol 56:3503–3512PubMedCrossRef
57.
go back to reference Millard TP, Endrizzi M, Everdell N et al (2015) Evaluation of microbubble contrast agents for dynamic imaging with x-ray phase contrast. Sci Rep 5:12509PubMedPubMedCentralCrossRef Millard TP, Endrizzi M, Everdell N et al (2015) Evaluation of microbubble contrast agents for dynamic imaging with x-ray phase contrast. Sci Rep 5:12509PubMedPubMedCentralCrossRef
58.
go back to reference Wu D, Wong MD, Yang K et al (2018) Using microbubble as contrast agent for high-energy X-ray in-line phase contrast imaging: demonstration and comparison study. IEEE Trans Biomed Eng 65:1117–1123PubMed Wu D, Wong MD, Yang K et al (2018) Using microbubble as contrast agent for high-energy X-ray in-line phase contrast imaging: demonstration and comparison study. IEEE Trans Biomed Eng 65:1117–1123PubMed
59.
go back to reference Lång K, Arboleda C, Forte S et al (2019) Microbubbles as a contrast agent in grating interferometry mammography: an ex vivo proof-of-mechanism study. Eur Radiol Exp 3:19PubMedPubMedCentralCrossRef Lång K, Arboleda C, Forte S et al (2019) Microbubbles as a contrast agent in grating interferometry mammography: an ex vivo proof-of-mechanism study. Eur Radiol Exp 3:19PubMedPubMedCentralCrossRef
Metadata
Title
Comparison of propagation-based CT using synchrotron radiation and conventional cone-beam CT for breast imaging
Authors
Seyedamir Tavakoli Taba
Patrycja Baran
Yakov I. Nesterets
Serena Pacile
Susanne Wienbeck
Christian Dullin
Konstantin Pavlov
Anton Maksimenko
Darren Lockie
Sheridan C. Mayo
Harry M. Quiney
Diego Dreossi
Fulvia Arfelli
Giuliana Tromba
Sarah Lewis
Timur E. Gureyev
Patrick C. Brennan
Publication date
01-05-2020
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 5/2020
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06567-0

Other articles of this Issue 5/2020

European Radiology 5/2020 Go to the issue