Published in: European Radiology 6/2019

01-06-2019 | Magnetic Resonance

Risk assessment of copper-containing contraceptives: the impact for women with implanted intrauterine devices during clinical MRI and CT examinations

Authors: Wiebke Neumann, Tanja Uhrig, Matthias Malzacher, Verena Kossmann, Lothar R. Schad, Frank G. Zoellner

Published in: European Radiology | Issue 6/2019

Login to get access

Abstract

Objectives

To assess the risks for implant users with copper-containing intrauterine devices (IUDs) during MR and CT examinations.

Methods

A tissue-mimicking phantom suitable for all experiments within this study was developed. Seven different types of copper IUDs were evaluated. Heating and dislocation of each IUD were investigated at two clinically relevant positions in 1.5 T and 3 T MR scanners. Artifacts in the field of view caused by each tested IUD were determined for clinical MR and CT imaging.

Results

No significant heating of any tested IUD was detected during MR measurements. The temperature increase was less than 0.6 K for all IUDs. Neither angular deflection nor translation of any IUD was detected. Artifacts in MR images were limited to the very vicinity of the IUDs except for one IUD containing a steel-visualizing element. Streaking artifacts in CT were severe (up to 75.5%) in the slices including the IUD.

Conclusion

No significant risk possibly harming the patient was determined during this phantom study, deeming MR examinations safe for women with an implanted copper IUD. Image quality was more impaired for CT than for MR imaging and needs careful consideration during diagnosis.

Key Points

• Risk assessment of copper-containing IUDs with regard to heating, dislocation, and artifacts during MR and CT imaging.
• Neither significant heating nor dislocation was determined in MR; image quality was more impaired for CT than for MR imaging and needs careful consideration during diagnosis.
• The tested IUDs pose no additional risks for implant users during MR and CT examinations.
Literature
1.
go back to reference Kavanaugh ML, Jerman J (2018) Contraceptive method use in the United States: trends and characteristics between 2008, 2012 and 2014. Contraception 97:14–21CrossRefPubMed Kavanaugh ML, Jerman J (2018) Contraceptive method use in the United States: trends and characteristics between 2008, 2012 and 2014. Contraception 97:14–21CrossRefPubMed
2.
go back to reference United Nations Population Fund (2016) TCu380A intrauterine contraceptive device (IUD) WHO/UNFPA technical specification and prequalification guidance. United Nations Population Fund, NY, USA United Nations Population Fund (2016) TCu380A intrauterine contraceptive device (IUD) WHO/UNFPA technical specification and prequalification guidance. United Nations Population Fund, NY, USA
3.
go back to reference World Health Organization (ed) (2009) Medical eligibility criteria for contraceptive use – 4th ed. World Health Organization, Geneva World Health Organization (ed) (2009) Medical eligibility criteria for contraceptive use – 4th ed. World Health Organization, Geneva
5.
go back to reference Dietrich O, Reiser MF, Schoenberg SO (2008) Artifacts in 3-T MRI: physical background and reduction strategies. Eur J Radiol 65:29–35CrossRefPubMed Dietrich O, Reiser MF, Schoenberg SO (2008) Artifacts in 3-T MRI: physical background and reduction strategies. Eur J Radiol 65:29–35CrossRefPubMed
6.
go back to reference Guerin B, Serano P, Iacono MI et al (2018) Realistic modeling of deep brain stimulation implants for electromagnetic MRI safety studies. Phys Med Biol 63:095015CrossRefPubMedPubMedCentral Guerin B, Serano P, Iacono MI et al (2018) Realistic modeling of deep brain stimulation implants for electromagnetic MRI safety studies. Phys Med Biol 63:095015CrossRefPubMedPubMedCentral
7.
go back to reference Golestanirad L, Angelone LM, Iacono MI, Katnani H, Wald LL, Bonmassar G (2017) Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: a simulation study of the effect of extracranial loops. Magn Reson Med 78:1558–1565CrossRefPubMed Golestanirad L, Angelone LM, Iacono MI, Katnani H, Wald LL, Bonmassar G (2017) Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: a simulation study of the effect of extracranial loops. Magn Reson Med 78:1558–1565CrossRefPubMed
8.
go back to reference Noureddine Y, Kraff O, Ladd ME et al (2018) In vitro and in silico assessment of RF-induced heating around intracranial aneurysm clips at 7 Tesla. Magn Reson Med 79:568–581CrossRefPubMed Noureddine Y, Kraff O, Ladd ME et al (2018) In vitro and in silico assessment of RF-induced heating around intracranial aneurysm clips at 7 Tesla. Magn Reson Med 79:568–581CrossRefPubMed
10.
go back to reference Winter L, Oberacker E, Özerdem C et al (2015) On the RF heating of coronary stents at 7.0 Tesla MRI. Magn Reson Med 74:999–1010CrossRefPubMed Winter L, Oberacker E, Özerdem C et al (2015) On the RF heating of coronary stents at 7.0 Tesla MRI. Magn Reson Med 74:999–1010CrossRefPubMed
11.
go back to reference Tse ZT, Elhawary H, Montesinos CA, Rea M, Young I, Lampérth M (2011) Testing MR image artifacts generated by engineering materials. Concepts Magn Reson 39B:109–117CrossRef Tse ZT, Elhawary H, Montesinos CA, Rea M, Young I, Lampérth M (2011) Testing MR image artifacts generated by engineering materials. Concepts Magn Reson 39B:109–117CrossRef
12.
go back to reference Santoro D, Winter L, Müller A et al (2012) Detailing radio frequency heating induced by coronary stents: a 7.0 Tesla magnetic resonance study. PLoS One 7:e49963CrossRefPubMedPubMedCentral Santoro D, Winter L, Müller A et al (2012) Detailing radio frequency heating induced by coronary stents: a 7.0 Tesla magnetic resonance study. PLoS One 7:e49963CrossRefPubMedPubMedCentral
13.
go back to reference Elhawary H, Zivanovic A, Davies B, Lamperth M (2006) A review of magnetic resonance imaging compatible manipulators in surgery. Proc Inst Mech Eng H 220:413–424CrossRefPubMed Elhawary H, Zivanovic A, Davies B, Lamperth M (2006) A review of magnetic resonance imaging compatible manipulators in surgery. Proc Inst Mech Eng H 220:413–424CrossRefPubMed
14.
go back to reference Christoforou EG, Tsekos NV, Özcan A (2006) Design and testing of a robotic system for MR image-guided interventions. J Intell Robot Syst 47:175–196CrossRef Christoforou EG, Tsekos NV, Özcan A (2006) Design and testing of a robotic system for MR image-guided interventions. J Intell Robot Syst 47:175–196CrossRef
15.
go back to reference Tse ZT, Janssen H, Hamed A, Ristic M, Young I, Lamperth M (2009) Magnetic resonance elastography hardware design: a survey. Proc Inst Mech Eng H 223:497–514CrossRefPubMed Tse ZT, Janssen H, Hamed A, Ristic M, Young I, Lamperth M (2009) Magnetic resonance elastography hardware design: a survey. Proc Inst Mech Eng H 223:497–514CrossRefPubMed
16.
go back to reference Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691CrossRef Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691CrossRef
17.
go back to reference Lee MJ, Kim S, Lee SA et al (2007) Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27:791–803CrossRef Lee MJ, Kim S, Lee SA et al (2007) Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27:791–803CrossRef
18.
go back to reference Kalender WA (2011) Computer Tomography, 3rd edn. Publicis MCD Verlag, Germany Kalender WA (2011) Computer Tomography, 3rd edn. Publicis MCD Verlag, Germany
19.
20.
go back to reference Hess T, Stepanow B, Knopp MV (1996) Safety of intrauterine contraceptive devices during MR imaging. Eur Radiol 6:66–68CrossRefPubMed Hess T, Stepanow B, Knopp MV (1996) Safety of intrauterine contraceptive devices during MR imaging. Eur Radiol 6:66–68CrossRefPubMed
21.
go back to reference Pasquale SA, Russer TJ, Foldesy R, Mezrich RS (1997) Lack of interaction between magnetic resonance imaging and the copper-T380A IUD. Contraception 55:169–173CrossRefPubMed Pasquale SA, Russer TJ, Foldesy R, Mezrich RS (1997) Lack of interaction between magnetic resonance imaging and the copper-T380A IUD. Contraception 55:169–173CrossRefPubMed
22.
go back to reference Berger-Kulemann V, Einspieler H, Hachemian N et al (2013) Magnetic field interactions of copper-containing intrauterine devices in 3.0-Tesla magnetic resonance imaging: in vivo study. Korean J Radiol 14:416–422CrossRefPubMedPubMedCentral Berger-Kulemann V, Einspieler H, Hachemian N et al (2013) Magnetic field interactions of copper-containing intrauterine devices in 3.0-Tesla magnetic resonance imaging: in vivo study. Korean J Radiol 14:416–422CrossRefPubMedPubMedCentral
23.
go back to reference ASTM Standard F2119-07 (2007) Standard test method for evaluation of MR image artifacts from passive implants. ASTM International, West Conshohocken ASTM Standard F2119-07 (2007) Standard test method for evaluation of MR image artifacts from passive implants. ASTM International, West Conshohocken
24.
go back to reference ASTM Standard F2182-11a (2011) Standard test method measurements of radio frequency induced heating on or near passive implants during magnetic resonance imaging. ASTM International, West Conshohocken ASTM Standard F2182-11a (2011) Standard test method measurements of radio frequency induced heating on or near passive implants during magnetic resonance imaging. ASTM International, West Conshohocken
25.
go back to reference ASTM Standard F2213-06 (2006) Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance environment. ASTM International, West Conshohocken ASTM Standard F2213-06 (2006) Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance environment. ASTM International, West Conshohocken
26.
go back to reference EN60601 (2017) Medical electrical equipment - Part 2–33: particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis. Beuth Verlag, Berlin EN60601 (2017) Medical electrical equipment - Part 2–33: particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis. Beuth Verlag, Berlin
27.
go back to reference Neumann W, Lietzmann F, Schad LR, Zöllner FG (2017) Design of a multimodal (1H/23NaMR/CT) anthropomorphic thorax phantom. Z Med Phys 27:124–131 Neumann W, Lietzmann F, Schad LR, Zöllner FG (2017) Design of a multimodal (1H/23NaMR/CT) anthropomorphic thorax phantom. Z Med Phys 27:124–131
28.
go back to reference Kato H, Kuroda M, Yoshimura K et al (2005) Composition of MRI phantom equivalent to human tissues. Med Phys 32:3199–3208CrossRefPubMed Kato H, Kuroda M, Yoshimura K et al (2005) Composition of MRI phantom equivalent to human tissues. Med Phys 32:3199–3208CrossRefPubMed
29.
go back to reference Schuenke P, Koehler C, Korzowski A et al (2017) Adiabatically prepared spin-lock approach for T1ρ-based dynamic glucose enhanced MRI at ultrahigh fields. Magn Reson Med 78:215–225CrossRefPubMed Schuenke P, Koehler C, Korzowski A et al (2017) Adiabatically prepared spin-lock approach for T1ρ-based dynamic glucose enhanced MRI at ultrahigh fields. Magn Reson Med 78:215–225CrossRefPubMed
30.
go back to reference Nordbeck P, Fidler F, Weiss I et al (2008) Spatial distribution of RF-induced E-fields and implant heating in MRI. Magn Reson Med 60:312–319CrossRefPubMed Nordbeck P, Fidler F, Weiss I et al (2008) Spatial distribution of RF-induced E-fields and implant heating in MRI. Magn Reson Med 60:312–319CrossRefPubMed
32.
go back to reference Shellock FG (2002) Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system. J Magn Reson Imaging 16(6):721–732CrossRefPubMed Shellock FG (2002) Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system. J Magn Reson Imaging 16(6):721–732CrossRefPubMed
33.
go back to reference van der Schaaf I, van Leeuwen M, Vlassenbroek A, Velthuis B (2006) Minimizing clip artifacts in multi CT angiography of clipped patients. AJNR Am J Neuroradiol 27:60–66PubMed van der Schaaf I, van Leeuwen M, Vlassenbroek A, Velthuis B (2006) Minimizing clip artifacts in multi CT angiography of clipped patients. AJNR Am J Neuroradiol 27:60–66PubMed
34.
go back to reference Ulaby FT, Michielssen E, Ravaioli U (2015) Fundamentals of applied electromagnetics, global edition. Pearson Education Limited, Essex Ulaby FT, Michielssen E, Ravaioli U (2015) Fundamentals of applied electromagnetics, global edition. Pearson Education Limited, Essex
35.
go back to reference Yeung CJ, Susil RC, Atalar E (2002) RF safety of wires in interventional MRI: using a safety index. Magn Reson Med 47:187–193CrossRefPubMed Yeung CJ, Susil RC, Atalar E (2002) RF safety of wires in interventional MRI: using a safety index. Magn Reson Med 47:187–193CrossRefPubMed
36.
go back to reference Samoudi AM, Vermeeren G, Tanghe E, Van Holen R, Martens L, Josephs W (2016) Numerically simulated exposure of children and adults to pulsed gradient fields in MRI. J Magn Reson Imaging 44:1360–1367CrossRefPubMed Samoudi AM, Vermeeren G, Tanghe E, Van Holen R, Martens L, Josephs W (2016) Numerically simulated exposure of children and adults to pulsed gradient fields in MRI. J Magn Reson Imaging 44:1360–1367CrossRefPubMed
Metadata
Title
Risk assessment of copper-containing contraceptives: the impact for women with implanted intrauterine devices during clinical MRI and CT examinations
Authors
Wiebke Neumann
Tanja Uhrig
Matthias Malzacher
Verena Kossmann
Lothar R. Schad
Frank G. Zoellner
Publication date
01-06-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 6/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5864-6

Other articles of this Issue 6/2019

European Radiology 6/2019 Go to the issue