01-10-2018 | Cardiac

Cardiac implanted electronic devices and MRI safety in 2018—the state of play

Authors: Ryan Mark Shulman, Ben Hunt

Published in: European Radiology | Issue 10/2018

Login to get access

Abstract

Traditionally, the presence of cardiac implanted electronic devices (CIEDs) was a contra-indication to magnetic resonance (MR) imaging. Professional groups from around the world are releasing updated guidelines for the imaging of MR-conditional and legacy CIEDs, reflecting increasing evidence that this can be performed safely when strict protocols are followed.

Key Points

The presence of a pacemaker or automatic implanted cardioverter defibrillator is no longer an absolute contraindication to magnetic resonance imaging.
Strict protocols enable diagnostic quality images to be obtained with minimal risk.
Close collaboration among radiologists, cardiologists and device manufacturer representatives is required.
Literature
1.
go back to reference Sommer T, Bauer W, Fischbach K et al (2017) MR imaging in patients with cardiac pacemakers and implantable cardioverter defibrillators. Rofo 189:204–217CrossRef Sommer T, Bauer W, Fischbach K et al (2017) MR imaging in patients with cardiac pacemakers and implantable cardioverter defibrillators. Rofo 189:204–217CrossRef
2.
go back to reference Lowe MD, Plummer CJ, Manisty CH, Linker NJ, British Heart Rhythm Society (2015) Safe use of MRI in people with cardiac implantable electronic devices. Heart 101:1950–1953CrossRef Lowe MD, Plummer CJ, Manisty CH, Linker NJ, British Heart Rhythm Society (2015) Safe use of MRI in people with cardiac implantable electronic devices. Heart 101:1950–1953CrossRef
3.
go back to reference Indik JH, Gimbel JR, Abe H et al (2017) 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm 14:e97–e153CrossRef Indik JH, Gimbel JR, Abe H et al (2017) 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm 14:e97–e153CrossRef
4.
go back to reference Brignole M, Auricchio A, Baron-Esquivias G et al (2013) 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J 34:2281–2329CrossRef Brignole M, Auricchio A, Baron-Esquivias G et al (2013) 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J 34:2281–2329CrossRef
5.
go back to reference Irnich W, Irnich B, Bartsch C, Stertmann WA, Gufler H, Weiler G (2005) Do we need pacemakers resistant to magnetic resonance imaging? Europace 7:353–365CrossRef Irnich W, Irnich B, Bartsch C, Stertmann WA, Gufler H, Weiler G (2005) Do we need pacemakers resistant to magnetic resonance imaging? Europace 7:353–365CrossRef
6.
go back to reference Di Leo G, Zanardo M, Schiaffino S, Secchi F, Sardanelli F (2017) Safety of MRI in patients with conditional or non-conditional pacemaker or other implantable cardiac electronic devices (ICED): a systematic review. Paper presented at: 103rd Scientific Assembly and Annual Meeting RSNA; 2017 Nov 26 - Dec 1; Chicago, IL, USA Di Leo G, Zanardo M, Schiaffino S, Secchi F, Sardanelli F (2017) Safety of MRI in patients with conditional or non-conditional pacemaker or other implantable cardiac electronic devices (ICED): a systematic review. Paper presented at: 103rd Scientific Assembly and Annual Meeting RSNA; 2017 Nov 26 - Dec 1; Chicago, IL, USA
7.
go back to reference Expert Panel on MR Safety, Kanal E, Barkovich AJ et al (2013) ACR guidance document on MR safe practices: 2013. J Magn Reson Imaging 37:501–530CrossRef Expert Panel on MR Safety, Kanal E, Barkovich AJ et al (2013) ACR guidance document on MR safe practices: 2013. J Magn Reson Imaging 37:501–530CrossRef
8.
go back to reference Russo RJ, Costa HS, Silva PD et al (2017) Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med 376:755–764CrossRef Russo RJ, Costa HS, Silva PD et al (2017) Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med 376:755–764CrossRef
9.
go back to reference Friedman HL, Acker N, Dalzell C et al (2013) Magnetic resonance imaging in patients with recently implanted pacemakers. Pacing Clin Electrophysiol 36:1090–1095CrossRef Friedman HL, Acker N, Dalzell C et al (2013) Magnetic resonance imaging in patients with recently implanted pacemakers. Pacing Clin Electrophysiol 36:1090–1095CrossRef
10.
go back to reference Baker KB, Tkach JA, Nyenhuis JA et al (2004) Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating. J Magn Reson Imaging 20:315–320CrossRef Baker KB, Tkach JA, Nyenhuis JA et al (2004) Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating. J Magn Reson Imaging 20:315–320CrossRef
11.
go back to reference Gimbel JR, Bailey SM, Tchou PJ, Ruggieri PM, Wilkoff BL (2005) Strategies for the safe magnetic resonance imaging of pacemaker-dependent patients. Pacing Clin Electrophysiol 28:1041–1046CrossRef Gimbel JR, Bailey SM, Tchou PJ, Ruggieri PM, Wilkoff BL (2005) Strategies for the safe magnetic resonance imaging of pacemaker-dependent patients. Pacing Clin Electrophysiol 28:1041–1046CrossRef
12.
go back to reference Faulkner W (2016) New MRI safety labels & devices. Signals 5:1 Faulkner W (2016) New MRI safety labels & devices. Signals 5:1
13.
go back to reference Nazarian S, Hansford R, Rahsepar AA et al (2017) Safety of magnetic resonance imaging in patients with cardiac devices. N Engl J Med 377:2555–2564CrossRef Nazarian S, Hansford R, Rahsepar AA et al (2017) Safety of magnetic resonance imaging in patients with cardiac devices. N Engl J Med 377:2555–2564CrossRef
14.
go back to reference Mason S, Osborn JS, Dhar R et al (2017) Real world MRI experience with nonconditional and conditional cardiac rhythm devices after MagnaSafe. J Cardiovasc Electrophysiol 28:1468–1474CrossRef Mason S, Osborn JS, Dhar R et al (2017) Real world MRI experience with nonconditional and conditional cardiac rhythm devices after MagnaSafe. J Cardiovasc Electrophysiol 28:1468–1474CrossRef
15.
go back to reference Do DH, Eyvazian V, Bayoneta AJ et al (2018) Cardiac magnetic resonance imaging using wideband sequences in patients with nonconditional cardiac implanted electronic devices. Heart Rhythm 15:218–225CrossRef Do DH, Eyvazian V, Bayoneta AJ et al (2018) Cardiac magnetic resonance imaging using wideband sequences in patients with nonconditional cardiac implanted electronic devices. Heart Rhythm 15:218–225CrossRef
16.
go back to reference Hwang YM, Kim J, Lee JH et al (2016) Cardiac implantable electronic device safety during magnetic resonance imaging. Korean Circ J 46:804–810CrossRef Hwang YM, Kim J, Lee JH et al (2016) Cardiac implantable electronic device safety during magnetic resonance imaging. Korean Circ J 46:804–810CrossRef
17.
go back to reference Higgins JV, Gard JJ, Sheldon SH et al (2014) Safety and outcomes of magnetic resonance imaging in patients with abandoned pacemaker and defibrillator leads. Pacing Clin Electrophysiol 37:1284–1290CrossRef Higgins JV, Gard JJ, Sheldon SH et al (2014) Safety and outcomes of magnetic resonance imaging in patients with abandoned pacemaker and defibrillator leads. Pacing Clin Electrophysiol 37:1284–1290CrossRef
18.
go back to reference Miller JD, Nazarian S, Halperin HR (2016) Implantable electronic cardiac devices and compatibility with magnetic resonance imaging. J Am Coll Cardiol 68:1590–1598CrossRef Miller JD, Nazarian S, Halperin HR (2016) Implantable electronic cardiac devices and compatibility with magnetic resonance imaging. J Am Coll Cardiol 68:1590–1598CrossRef
19.
go back to reference Shulman RM, Moodley M, Vaishnav M, Wright J (2018) Cardiac magnetic resonance with conditional pacemaker at three Tesla (3T) field strength. Cardiovascular Imaging Asia 2:28–30CrossRef Shulman RM, Moodley M, Vaishnav M, Wright J (2018) Cardiac magnetic resonance with conditional pacemaker at three Tesla (3T) field strength. Cardiovascular Imaging Asia 2:28–30CrossRef
20.
go back to reference Naehle CP, Meyer C, Thomas D et al (2008) Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology 249:991–1001CrossRef Naehle CP, Meyer C, Thomas D et al (2008) Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology 249:991–1001CrossRef
21.
go back to reference Gimbel JR (2008) Magnetic resonance imaging of implantable cardiac rhythm devices at 3.0 tesla. Pacing Clin Electrophysiol 31:795–801CrossRef Gimbel JR (2008) Magnetic resonance imaging of implantable cardiac rhythm devices at 3.0 tesla. Pacing Clin Electrophysiol 31:795–801CrossRef
22.
go back to reference Delfino JG, Viohl I, Woods TO (2017) Risks of MRI in patients with a pacemaker or defibrillator. N Engl J Med 376:2495CrossRef Delfino JG, Viohl I, Woods TO (2017) Risks of MRI in patients with a pacemaker or defibrillator. N Engl J Med 376:2495CrossRef
Metadata
Title
Cardiac implanted electronic devices and MRI safety in 2018—the state of play
Authors
Ryan Mark Shulman
Ben Hunt
Publication date
01-10-2018
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 10/2018
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5396-0

Other articles of this Issue 10/2018

European Radiology 10/2018 Go to the issue