Step and shoot coronary CT angiography using 256-slice CT: effect of heart rate and heart rate variability on image quality

Authors: D. Muenzel, P. B. Noel, F. Dorn, M. Dobritz, E. J. Rummeny, A. Huber

Published in: European Radiology | Issue 11/2011

Login to get access

Abstract

Objective

To evaluate the effect of heart rate variability (HRV) and heart rate (HR) on intra-image “motion” and inter-image “stairstep” artefacts in step-and-shoot coronary CT angiography (CCTA) using a wide detector CT scanner.

Methods

66 patients underwent step-and-shoot CCTA using 256-slice CT. Patients were divided into two groups (Group 1: HR <65 bpm, Group 2 ≥65bpm). Motion artefacts were quantified using a 5-point-scale. Stairstep artefacts were defined by measurements of misalignment. Image noise, contrast-to-noise-ratio (CNR), signal-to-noise-ratio (SNR), and radiation dose were assessed.

Results

Mean HR was 66 ± 16.7 bpm (range: 45–125 bpm) and mean HRV was 10.7 ± 17.5 bpm. A significant correlation between HR and stairstep artefacts (r = 0.46, p < 0.001) and motion artefacts (r = 0.63, p < 0.001) was found. Group 2 showed significantly increased step artefacts with a mean misalignment of 1.4 mm compared to 0.4 mm in Group 1 (p < 0.001). There was no significant effect of HRV on stairstep artefacts (r = 0.15, p = 0.416) and motion artefacts (r = 0.13, p = 0.311). No significant differences in image noise, CNR, SNR, and radiation dose were seen.

Conclusions

Unlike CCTA using narrow CT detectors, HRV has no significant effect on motion and stairstep artefacts using a wide CT detector with high z-coverage. However, a higher HR still increases stairstep and motion artefacts.
Literature
1.
go back to reference Scheffel H, Alkadhi H, Leschka S et al (2008) Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance. Heart 94:1132–1137PubMedCrossRef Scheffel H, Alkadhi H, Leschka S et al (2008) Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance. Heart 94:1132–1137PubMedCrossRef
2.
go back to reference de Graaf FR, Schuijf JD, van Velzen JE et al (2010) Diagnostic accuracy of 320-row multidetector computed tomography coronary angiography in the non-invasive evaluation of significant coronary artery disease. Eur Heart J 31:1908–1915PubMedCrossRef de Graaf FR, Schuijf JD, van Velzen JE et al (2010) Diagnostic accuracy of 320-row multidetector computed tomography coronary angiography in the non-invasive evaluation of significant coronary artery disease. Eur Heart J 31:1908–1915PubMedCrossRef
3.
go back to reference Chao SP, Law WY, Kuo CJ et al (2010) The diagnostic accuracy of 256-row computed tomographic angiography compared with invasive coronary angiography in patients with suspected coronary artery disease. Eur Heart J 31:1916–1923PubMedCrossRef Chao SP, Law WY, Kuo CJ et al (2010) The diagnostic accuracy of 256-row computed tomographic angiography compared with invasive coronary angiography in patients with suspected coronary artery disease. Eur Heart J 31:1916–1923PubMedCrossRef
4.
go back to reference Budoff MJ, Dowe D, Jollis JG et al (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732PubMedCrossRef Budoff MJ, Dowe D, Jollis JG et al (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732PubMedCrossRef
5.
go back to reference Earls JP, Berman EL, Urban BA et al (2008) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 246:742–753PubMedCrossRef Earls JP, Berman EL, Urban BA et al (2008) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 246:742–753PubMedCrossRef
6.
go back to reference Efstathopoulos EP, Kelekis NL, Pantos I et al (2009) Reduction of the estimated radiation dose and associated patient risk with prospective ECG-gated 256-slice CT coronary angiography. Phys Med Biol 54:5209–5222PubMedCrossRef Efstathopoulos EP, Kelekis NL, Pantos I et al (2009) Reduction of the estimated radiation dose and associated patient risk with prospective ECG-gated 256-slice CT coronary angiography. Phys Med Biol 54:5209–5222PubMedCrossRef
7.
go back to reference Klass O, Jeltsch M, Feuerlein S et al (2009) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836PubMedCrossRef Klass O, Jeltsch M, Feuerlein S et al (2009) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836PubMedCrossRef
8.
go back to reference Hou Y, Yue Y, Guo W et al (2010) Prospectively versus retrospectively ECG-gated 256-slice coronary CT angiography: image quality and radiation dose over expanded heart rates. Int J Cardiovasc Imaging Hou Y, Yue Y, Guo W et al (2010) Prospectively versus retrospectively ECG-gated 256-slice coronary CT angiography: image quality and radiation dose over expanded heart rates. Int J Cardiovasc Imaging
10.
go back to reference Hirai N, Horiguchi J, Fujioka C et al (2008) Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology 248:424–430PubMedCrossRef Hirai N, Horiguchi J, Fujioka C et al (2008) Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology 248:424–430PubMedCrossRef
11.
go back to reference Husmann L, Herzog BA, Burkhard N et al (2009) Body physique and heart rate variability determine the occurrence of stair-step artefacts in 64-slice CT coronary angiography with prospective ECG-triggering. Eur Radiol 19:1698–1703PubMedCrossRef Husmann L, Herzog BA, Burkhard N et al (2009) Body physique and heart rate variability determine the occurrence of stair-step artefacts in 64-slice CT coronary angiography with prospective ECG-triggering. Eur Radiol 19:1698–1703PubMedCrossRef
12.
go back to reference Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249:71–80PubMedCrossRef Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249:71–80PubMedCrossRef
13.
go back to reference Choi SI, George RT, Schuleri KH et al (2009) Recent developments in wide-detector cardiac computed tomography. Int J Cardiovasc Imaging 25:23–29PubMedCrossRef Choi SI, George RT, Schuleri KH et al (2009) Recent developments in wide-detector cardiac computed tomography. Int J Cardiovasc Imaging 25:23–29PubMedCrossRef
14.
go back to reference Hsiao EM, Rybicki FJ, Steigner M (2010) CT coronary angiography: 256-slice and 320-detector row scanners. Curr Cardiol Rep 12:68–75PubMedCrossRef Hsiao EM, Rybicki FJ, Steigner M (2010) CT coronary angiography: 256-slice and 320-detector row scanners. Curr Cardiol Rep 12:68–75PubMedCrossRef
15.
go back to reference Weustink AC, Mollet NR, Pugliese F et al (2008) Optimal electrocardiographic pulsing windows and heart rate: effect on image quality and radiation exposure at dual-source coronary CT angiography. Radiology 248:792–798PubMedCrossRef Weustink AC, Mollet NR, Pugliese F et al (2008) Optimal electrocardiographic pulsing windows and heart rate: effect on image quality and radiation exposure at dual-source coronary CT angiography. Radiology 248:792–798PubMedCrossRef
16.
go back to reference Wintersperger BJ, Nikolaou K, von Ziegler F et al (2006) Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed. Invest Radiol 41:436–442PubMedCrossRef Wintersperger BJ, Nikolaou K, von Ziegler F et al (2006) Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed. Invest Radiol 41:436–442PubMedCrossRef
17.
go back to reference Austen WG et al (1975) A reporting systm on patients evaluated for coronary artery disease. Report of the Ad Hoc committee for grading of coronary artery disease, council on cardiovascular surgery, American heart association. Circulation 51:5–40PubMed Austen WG et al (1975) A reporting systm on patients evaluated for coronary artery disease. Report of the Ad Hoc committee for grading of coronary artery disease, council on cardiovascular surgery, American heart association. Circulation 51:5–40PubMed
18.
go back to reference Hausleiter J, Meyer T, Hadamitzky M et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310PubMedCrossRef Hausleiter J, Meyer T, Hadamitzky M et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310PubMedCrossRef
19.
go back to reference McCollough C, Cody D, Edyvean S et al (2008) The measurement, reporting, and management of radiation dose in CT. College Park, MD: American Association of Physicists in Medicine, AAPM report no. 96 McCollough C, Cody D, Edyvean S et al (2008) The measurement, reporting, and management of radiation dose in CT. College Park, MD: American Association of Physicists in Medicine, AAPM report no. 96
20.
go back to reference Leschka S, Husmann L, Desbiolles LM et al (2006) Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT. Eur Radiol 16:1964–1972PubMedCrossRef Leschka S, Husmann L, Desbiolles LM et al (2006) Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT. Eur Radiol 16:1964–1972PubMedCrossRef
21.
go back to reference Leschka S, Wildermuth S, Boehm T et al (2006) Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality. Radiology 241:378–385PubMedCrossRef Leschka S, Wildermuth S, Boehm T et al (2006) Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality. Radiology 241:378–385PubMedCrossRef
22.
go back to reference Leschka S, Scheffel H, Desbiolles L et al (2007) Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing windowing. Invest Radiol 42:543–549PubMedCrossRef Leschka S, Scheffel H, Desbiolles L et al (2007) Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing windowing. Invest Radiol 42:543–549PubMedCrossRef
23.
go back to reference Adler G, Meille L, Rohnean A et al (2010) Robustness of end-systolic reconstructions in coronary dual-source CT angiography for high heart rate patients. Eur Radiol 20:1118–1123PubMedCrossRef Adler G, Meille L, Rohnean A et al (2010) Robustness of end-systolic reconstructions in coronary dual-source CT angiography for high heart rate patients. Eur Radiol 20:1118–1123PubMedCrossRef
24.
go back to reference Brodoefel H, Burgstahler C, Tsiflikas I et al (2008) Dual-source CT: effect of heart rate, heart rate variability, and calcification on image quality and diagnostic accuracy. Radiology 247:346–355PubMedCrossRef Brodoefel H, Burgstahler C, Tsiflikas I et al (2008) Dual-source CT: effect of heart rate, heart rate variability, and calcification on image quality and diagnostic accuracy. Radiology 247:346–355PubMedCrossRef
25.
go back to reference Feuchtner G, Goetti R, Plass A et al (2010) Dual-step prospective ECG-triggered 128-slice dual-source CT for evaluation of coronary arteries and cardiac function without heart rate control: a technical note. Eur Radiol 20:2092–2099PubMedCrossRef Feuchtner G, Goetti R, Plass A et al (2010) Dual-step prospective ECG-triggered 128-slice dual-source CT for evaluation of coronary arteries and cardiac function without heart rate control: a technical note. Eur Radiol 20:2092–2099PubMedCrossRef
26.
go back to reference Herzog BA, Husmann L, Burkhard N et al (2009) Low-dose CT coronary angiography using prospective ECG-triggering: impact of mean heart rate and heart rate variability on image quality. Acad Radiol 16:15–21PubMedCrossRef Herzog BA, Husmann L, Burkhard N et al (2009) Low-dose CT coronary angiography using prospective ECG-triggering: impact of mean heart rate and heart rate variability on image quality. Acad Radiol 16:15–21PubMedCrossRef
27.
go back to reference Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef
Metadata
Title
Step and shoot coronary CT angiography using 256-slice CT: effect of heart rate and heart rate variability on image quality
Authors
D. Muenzel
P. B. Noel
F. Dorn
M. Dobritz
E. J. Rummeny
A. Huber
Publication date
01-11-2011
Publisher
Springer-Verlag
Published in
European Radiology / Issue 11/2011
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-011-2185-4

Other articles of this Issue 11/2011

European Radiology 11/2011 Go to the issue