Published in: European Radiology 9/2006

01-09-2006 | Cardiac

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

Authors: Sebastian Leschka, Lars Husmann, Lotus M. Desbiolles, Oliver Gaemperli, Tiziano Schepis, Pascal Koepfli, Thomas Boehm, Borut Marincek, Philipp A. Kaufmann, Hatem Alkadhi

Published in: European Radiology | Issue 9/2006

Login to get access

Abstract

The reconstruction intervals providing best image quality for non-invasive coronary angiography with 64-slice computed tomography (CT) were evaluated. Contrast-enhanced, retrospectively electrocardiography (ECG)-gated 64-slice CT coronary angiography was performed in 80 patients (47 male, 33 female; mean age 62.1±10.6 years). Thirteen data sets were reconstructed in 5% increments from 20 to 80% of the R-R interval. Depending on the average heart rate during scanning, patients were grouped as <65 bpm (n=49) and ≥65 bpm (n=31). Two blinded and independent readers assessed the image quality of each coronary segment with a diameter ≥1.5 mm using the following scores: 1, no motion artifacts; 2, minor artifacts; 3, moderate artifacts; 4, severe artifacts; and 5, not evaluative. The average heart rate was 63.3±13.1 bpm (range 38–102). Acceptable image quality (scores 1–3) was achieved in 99.1% of all coronary segments (1,162/1,172; mean image quality score 1.55±0.77) in the best reconstruction interval. Best image quality was found at 60% and 65% of the R-R interval for all patients and for each heart rate subgroup, whereas motion artifacts occurred significantly more often (P<0.01) at other reconstruction intervals. At heart rates <65 bpm, acceptable image quality was found in all coronary segments at 60%. At heart rates ≥65 bpm, the whole coronary artery tree could be visualized with acceptable image quality in 87% (27/31) of the patients at 60%, while ten segments in four patients were rated as non-diagnostic (scores 4–5) at any reconstruction interval. In conclusion, 64-slice CT coronary angiography provides best overall image quality in mid-diastole. At heart rates <65 bpm, diagnostic image quality of all coronary segments can be obtained at a single reconstruction interval of 60%.
Literature
1.
go back to reference Leschka S et al (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:1482–1487PubMedCrossRef Leschka S et al (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:1482–1487PubMedCrossRef
2.
go back to reference Raff GL, Gallagher MJ, O'Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef Raff GL, Gallagher MJ, O'Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef
3.
go back to reference Leber AW et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef Leber AW et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef
4.
go back to reference Mollet NR et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef Mollet NR et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef
5.
go back to reference Pugliese F et al (2005) Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris. Eur Radiol:1–8 Pugliese F et al (2005) Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris. Eur Radiol:1–8
6.
go back to reference Flohr TG, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough CH (2005) Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. Med Phys 32:2536–2547PubMedCrossRef Flohr TG, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough CH (2005) Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. Med Phys 32:2536–2547PubMedCrossRef
7.
go back to reference Giesler T et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR Am J Roentgenol 179:911–916PubMed Giesler T et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR Am J Roentgenol 179:911–916PubMed
8.
go back to reference Hong C et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMedCrossRef Hong C et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMedCrossRef
9.
go back to reference Kopp AF et al (2001) Coronary arteries: retrospectively ECG-gated multi-detector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMedCrossRef Kopp AF et al (2001) Coronary arteries: retrospectively ECG-gated multi-detector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMedCrossRef
10.
go back to reference Hamoir XL et al (2005) Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time. Eur Radiol 15:296–304PubMedCrossRef Hamoir XL et al (2005) Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time. Eur Radiol 15:296–304PubMedCrossRef
11.
go back to reference Bley TA et al (2005) Computed tomography coronary angiography with 370-millisecond gantry rotation time: evaluation of the best image reconstruction interval. J Comput Assist Tomogr 29:1–5PubMedCrossRef Bley TA et al (2005) Computed tomography coronary angiography with 370-millisecond gantry rotation time: evaluation of the best image reconstruction interval. J Comput Assist Tomogr 29:1–5PubMedCrossRef
12.
go back to reference Sanz J et al (2005) The importance of end-systole for optimal reconstruction protocol of coronary angiography with 16-slice multidetector computed tomography. Invest Radiol 40:155–163PubMedCrossRef Sanz J et al (2005) The importance of end-systole for optimal reconstruction protocol of coronary angiography with 16-slice multidetector computed tomography. Invest Radiol 40:155–163PubMedCrossRef
13.
go back to reference Jakobs TF et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086PubMedCrossRef Jakobs TF et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086PubMedCrossRef
14.
go back to reference Kachelriess M, Ulzheimer S, Kalender WA (2000) ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Med Phys 27:1881–1902PubMedCrossRef Kachelriess M, Ulzheimer S, Kalender WA (2000) ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Med Phys 27:1881–1902PubMedCrossRef
15.
go back to reference Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart. J Comput Assist Tomogr 25:907–923PubMedCrossRef Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart. J Comput Assist Tomogr 25:907–923PubMedCrossRef
16.
go back to reference Austen WG et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc committee for grading of coronary artery disease, council on cardiovascular surgery, American heart association. Circulation 51:5–40PubMed Austen WG et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc committee for grading of coronary artery disease, council on cardiovascular surgery, American heart association. Circulation 51:5–40PubMed
17.
go back to reference Shim SS, Kim Y, Lim SM (2005) Improvement of image quality with beta-blocker premedication on ECG-gated 16-MDCT coronary angiography. AJR Am J Roentgenol 184:649–654PubMed Shim SS, Kim Y, Lim SM (2005) Improvement of image quality with beta-blocker premedication on ECG-gated 16-MDCT coronary angiography. AJR Am J Roentgenol 184:649–654PubMed
18.
go back to reference Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20:37–46CrossRef Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20:37–46CrossRef
19.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
20.
go back to reference Choi HS et al (2004) Pitfalls, artifacts, and remedies in multi- detector row CT coronary angiography. Radiographics 24:787–800PubMedCrossRef Choi HS et al (2004) Pitfalls, artifacts, and remedies in multi- detector row CT coronary angiography. Radiographics 24:787–800PubMedCrossRef
21.
go back to reference Hoffmann MH et al (2005) Noninvasive coronary angiography with 16-detector row CT: effect of heart rate. Radiology 234:86–97PubMedCrossRef Hoffmann MH et al (2005) Noninvasive coronary angiography with 16-detector row CT: effect of heart rate. Radiology 234:86–97PubMedCrossRef
22.
go back to reference Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054PubMedCrossRef Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054PubMedCrossRef
23.
go back to reference Cademartiri F et al (2005) Diagnostic accuracy of multislice computed tomography coronary angiography is improved at low heart rates. Int J Cardiovasc Imaging:1–5 Cademartiri F et al (2005) Diagnostic accuracy of multislice computed tomography coronary angiography is improved at low heart rates. Int J Cardiovasc Imaging:1–5
Metadata
Title
Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT
Authors
Sebastian Leschka
Lars Husmann
Lotus M. Desbiolles
Oliver Gaemperli
Tiziano Schepis
Pascal Koepfli
Thomas Boehm
Borut Marincek
Philipp A. Kaufmann
Hatem Alkadhi
Publication date
01-09-2006
Publisher
Springer-Verlag
Published in
European Radiology / Issue 9/2006
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-006-0262-x

Other articles of this Issue 9/2006

European Radiology 9/2006 Go to the issue