Published in: European Radiology 10/2011

01-10-2011 | Cardiac

Radiation dose threshold for coronary artery calcium score with MDCT: How low can you go?

Authors: Tom D. Newton, Hatem Mehrez, Kelly Wong, Ravi Menezes, Bernd J. Wintersperger, Andrew Crean, Elsie Nguyen, Narinder Paul

Published in: European Radiology | Issue 10/2011

Login to get access

Abstract

Objectives

To evaluate the lowest radiation exposure threshold at which coronary calcium scoring (CCS) remains accurate.

Methods

A prospective study of 43 consecutive eligible patients referred for CCS underwent imaging in accordance with the manufacturer’s recommended protocol. Dedicated software was used to generate 8 series of images simulating tube currents ranging from 20 to 300 mA. These images were randomised and read in blinded fashion to determine the lowest tube current at which the CCS remained accurate. The minimum mA was correlated with 6 different patients’ biometric parameters: bodyweight, body mass index, AP and lateral thoracic diameters, average thoracic diameter and the scout attenuation coefficient (SAC). The 95% confidence interval for each parameter was used to calculate tube current threshold levels and hence stratified CCS protocols were derived.

Results

Spearman’s correlation coefficients of the minimum tube current for the 6 parameters were: 0.66, 0.63, 0.65, 0.74, 0.77 and 0.86 respectively (p < 0.001). SAC offered the largest potential reduction in mean effective dose from 1.86 mSv to 0.88 mSv.

Conclusion

CCS with at least 50% reduction in radiation exposure and below 1 mSv is feasible if CT scout projections are utilised effectively.
Appendix
Available only for authorised users
Literature
1.
go back to reference Genders TS, Pugliese F, Mollet NR et al (2010) Incremental value of the CT coronary calcium score for the prediction of coronary artery disease. Eur Radiol 20:2331–2340PubMedCrossRef Genders TS, Pugliese F, Mollet NR et al (2010) Incremental value of the CT coronary calcium score for the prediction of coronary artery disease. Eur Radiol 20:2331–2340PubMedCrossRef
2.
go back to reference Greenland P, LaBree L, Azen SP, Doherty TM, Detrano RC (2004) Coronary artery calcium score combined with Framingham score for risk prediction in asymptomatic individuals. JAMA 291:210–215PubMedCrossRef Greenland P, LaBree L, Azen SP, Doherty TM, Detrano RC (2004) Coronary artery calcium score combined with Framingham score for risk prediction in asymptomatic individuals. JAMA 291:210–215PubMedCrossRef
3.
go back to reference Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832PubMedCrossRef Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832PubMedCrossRef
4.
go back to reference Hunold P, Vogt FM, Schmermund A et al (2003) Radiation exposure during cardiac CT: effective doses at multi-detector row CT and electron-beam CT. Radiology 226:145–152PubMedCrossRef Hunold P, Vogt FM, Schmermund A et al (2003) Radiation exposure during cardiac CT: effective doses at multi-detector row CT and electron-beam CT. Radiology 226:145–152PubMedCrossRef
5.
go back to reference Kim KP, Einstein AJ, Berrington de Gonzalez A (2009) Coronary artery calcification screening: estimated radiation dose and cancer risk. Arch Intern Med 169:1188–1194PubMedCrossRef Kim KP, Einstein AJ, Berrington de Gonzalez A (2009) Coronary artery calcification screening: estimated radiation dose and cancer risk. Arch Intern Med 169:1188–1194PubMedCrossRef
6.
go back to reference Klass O, Jeltsch M, Feuerlein S et al (2009) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836PubMedCrossRef Klass O, Jeltsch M, Feuerlein S et al (2009) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836PubMedCrossRef
7.
go back to reference Klass O, Walker M, Siebach A et al (2010) Prospectively gated axial CT coronary angiography: comparison of image quality and effective radiation dose between 64- and 256-slice CT. Eur Radiol 20:1124–1131PubMedCrossRef Klass O, Walker M, Siebach A et al (2010) Prospectively gated axial CT coronary angiography: comparison of image quality and effective radiation dose between 64- and 256-slice CT. Eur Radiol 20:1124–1131PubMedCrossRef
8.
go back to reference Paul JF, Abada HT (2007) Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol 17:2028–2037PubMedCrossRef Paul JF, Abada HT (2007) Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol 17:2028–2037PubMedCrossRef
9.
go back to reference Carr JJ, Nelson JC, Wong ND et al (2005) Calcified coronary artery plaque measurement with cardiac CT in population-based studies: standardized protocol of Multi-Ethnic Study of Atherosclerosis (MESA) and Coronary Artery Risk Development in Young Adults (CARDIA) study. Radiology 234:35–43PubMedCrossRef Carr JJ, Nelson JC, Wong ND et al (2005) Calcified coronary artery plaque measurement with cardiac CT in population-based studies: standardized protocol of Multi-Ethnic Study of Atherosclerosis (MESA) and Coronary Artery Risk Development in Young Adults (CARDIA) study. Radiology 234:35–43PubMedCrossRef
10.
go back to reference McCollough CH, Ulzheimer S, Halliburton SS, Shanneik K, White RD, Kalender WA (2007) Coronary artery calcium: a multi-institutional, multimanufacturer international standard for quantification at cardiac CT. Radiology 243:527–538PubMedCrossRef McCollough CH, Ulzheimer S, Halliburton SS, Shanneik K, White RD, Kalender WA (2007) Coronary artery calcium: a multi-institutional, multimanufacturer international standard for quantification at cardiac CT. Radiology 243:527–538PubMedCrossRef
11.
go back to reference McNitt-Gray MF (2002) AAPM/RSNA Physics Tutorial for Residents: Topics in CT. Radiation dose in CT. Radiographics 22:1541–1553PubMedCrossRef McNitt-Gray MF (2002) AAPM/RSNA Physics Tutorial for Residents: Topics in CT. Radiation dose in CT. Radiographics 22:1541–1553PubMedCrossRef
12.
go back to reference Hong C, Bae KT, Pilgram TK, Suh J, Bradley D (2002) Coronary artery calcium measurement with multi-detector row CT: in vitro assessment of effect of radiation dose. Radiology 225:901–906PubMedCrossRef Hong C, Bae KT, Pilgram TK, Suh J, Bradley D (2002) Coronary artery calcium measurement with multi-detector row CT: in vitro assessment of effect of radiation dose. Radiology 225:901–906PubMedCrossRef
13.
go back to reference Shemesh J, Evron R, Koren-Morag N et al (2005) Coronary artery calcium measurement with multi-detector row CT and low radiation dose: comparison between 55 and 165 mAs. Radiology 236:810–814PubMedCrossRef Shemesh J, Evron R, Koren-Morag N et al (2005) Coronary artery calcium measurement with multi-detector row CT and low radiation dose: comparison between 55 and 165 mAs. Radiology 236:810–814PubMedCrossRef
14.
go back to reference Horiguchi J, Matsuura N, Yamamoto H et al (2009) Coronary artery calcium scoring on low-dose prospective electrocardiographically-triggered 64-slice CT. Acad Radiol 16:187–193PubMedCrossRef Horiguchi J, Matsuura N, Yamamoto H et al (2009) Coronary artery calcium scoring on low-dose prospective electrocardiographically-triggered 64-slice CT. Acad Radiol 16:187–193PubMedCrossRef
15.
go back to reference Horiguchi J, Matsuura N, Yamamoto H et al (2009) Evaluation of attenuation-based tube current control in coronary artery calcium scoring on prospective ECG-triggered 64-detector CT. Acad Radiol 16:1231–1240PubMedCrossRef Horiguchi J, Matsuura N, Yamamoto H et al (2009) Evaluation of attenuation-based tube current control in coronary artery calcium scoring on prospective ECG-triggered 64-detector CT. Acad Radiol 16:1231–1240PubMedCrossRef
16.
go back to reference Mahnken AH, Wildberger JE, Simon J et al (2003) Detection of coronary calcifications: feasibility of dose reduction with a body weight-adapted examination protocol. AJR Am J Roentgenol 181:533–538PubMed Mahnken AH, Wildberger JE, Simon J et al (2003) Detection of coronary calcifications: feasibility of dose reduction with a body weight-adapted examination protocol. AJR Am J Roentgenol 181:533–538PubMed
17.
go back to reference Nasir K, Budoff MJ, Post WS et al (2003) Electron beam CT versus helical CT scans for assessing coronary calcification: current utility and future directions. Am Heart J 146:969–977PubMedCrossRef Nasir K, Budoff MJ, Post WS et al (2003) Electron beam CT versus helical CT scans for assessing coronary calcification: current utility and future directions. Am Heart J 146:969–977PubMedCrossRef
18.
go back to reference Silverman JD, Paul NS, Siewerdsen JH (2009) Investigation of lung nodule detectability in low-dose 320-slice computed tomography. Med Phys 36:1700–1710PubMedCrossRef Silverman JD, Paul NS, Siewerdsen JH (2009) Investigation of lung nodule detectability in low-dose 320-slice computed tomography. Med Phys 36:1700–1710PubMedCrossRef
19.
go back to reference Primak AN, McCollough CH, Bruesewitz MR, Zhang J, Fletcher JG (2006) Relationship between noise, dose, and pitch in cardiac multi-detector row CT. Radiographics 26:1785–1794PubMedCrossRef Primak AN, McCollough CH, Bruesewitz MR, Zhang J, Fletcher JG (2006) Relationship between noise, dose, and pitch in cardiac multi-detector row CT. Radiographics 26:1785–1794PubMedCrossRef
20.
go back to reference Bielak LF, Kaufmann RB, Moll PP, McCollough CH, Schwartz RS, Sheedy PF 2nd (1994) Small lesions in the heart identified at electron beam CT: calcification or noise? Radiology 192:631–636PubMed Bielak LF, Kaufmann RB, Moll PP, McCollough CH, Schwartz RS, Sheedy PF 2nd (1994) Small lesions in the heart identified at electron beam CT: calcification or noise? Radiology 192:631–636PubMed
21.
go back to reference Callister TQ, Cooil B, Raya SP, Lippolis NJ, Russo DJ, Raggi P (1998) Coronary artery disease: improved reproducibility of calcium scoring with an electron-beam CT volumetric method. Radiology 208:807–814PubMed Callister TQ, Cooil B, Raya SP, Lippolis NJ, Russo DJ, Raggi P (1998) Coronary artery disease: improved reproducibility of calcium scoring with an electron-beam CT volumetric method. Radiology 208:807–814PubMed
22.
go back to reference Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507PubMedCrossRef Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507PubMedCrossRef
23.
go back to reference Menke J (2005) Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology 236:565–571PubMedCrossRef Menke J (2005) Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology 236:565–571PubMedCrossRef
24.
go back to reference Rutten A, Isgum I, Prokop M (2008) Coronary calcification: effect of small variation of scan starting position on Agatston, volume, and mass scores. Radiology 246:90–98PubMedCrossRef Rutten A, Isgum I, Prokop M (2008) Coronary calcification: effect of small variation of scan starting position on Agatston, volume, and mass scores. Radiology 246:90–98PubMedCrossRef
25.
go back to reference Yoon HC, Goldin JG, Greaser LE 3rd, Sayre J, Fonarow GC (2000) Interscan variation in coronary artery calcium quantification in a large asymptomatic patient population. AJR Am J Roentgenol 174:803–809PubMed Yoon HC, Goldin JG, Greaser LE 3rd, Sayre J, Fonarow GC (2000) Interscan variation in coronary artery calcium quantification in a large asymptomatic patient population. AJR Am J Roentgenol 174:803–809PubMed
26.
go back to reference Starck G, Lonn L, Cederblad A, Alpsten M, Sjostrom L, Ekholm S (1998) Radiation dose reduction in CT: application to tissue area and volume determination. Radiology 209:397–403PubMed Starck G, Lonn L, Cederblad A, Alpsten M, Sjostrom L, Ekholm S (1998) Radiation dose reduction in CT: application to tissue area and volume determination. Radiology 209:397–403PubMed
27.
go back to reference Takahashi N, Bae KT (2003) Quantification of coronary artery calcium with multi-detector row CT: assessing interscan variability with different tube currents pilot study. Radiology 228:101–106PubMedCrossRef Takahashi N, Bae KT (2003) Quantification of coronary artery calcium with multi-detector row CT: assessing interscan variability with different tube currents pilot study. Radiology 228:101–106PubMedCrossRef
28.
go back to reference Nakazato R, Dey D, Gutstein A et al (2009) Coronary artery calcium scoring using a reduced tube voltage and radiation dose protocol with dual-source computed tomography. J Cardiovasc Comput Tomogr 3:394–400PubMedCrossRef Nakazato R, Dey D, Gutstein A et al (2009) Coronary artery calcium scoring using a reduced tube voltage and radiation dose protocol with dual-source computed tomography. J Cardiovasc Comput Tomogr 3:394–400PubMedCrossRef
Metadata
Title
Radiation dose threshold for coronary artery calcium score with MDCT: How low can you go?
Authors
Tom D. Newton
Hatem Mehrez
Kelly Wong
Ravi Menezes
Bernd J. Wintersperger
Andrew Crean
Elsie Nguyen
Narinder Paul
Publication date
01-10-2011
Publisher
Springer-Verlag
Published in
European Radiology / Issue 10/2011
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-011-2159-6

Other articles of this Issue 10/2011

European Radiology 10/2011 Go to the issue