Published in: European Radiology 6/2011

01-06-2011 | Physics

A novel approach to digital breast tomosynthesis for simultaneous acquisition of 2D and 3D images

Authors: Sara Vecchio, Achille Albanese, Paolo Vignoli, Angelo Taibi

Published in: European Radiology | Issue 6/2011

Login to get access

Abstract

Objective

The complexity of anatomical structure within the breast represents the ultimate limit to signal detection on a mammogram. To increase lesion conspicuity Digital Breast Tomosynthesis (DBT) has been recently proposed and several manufacturers are currently performing clinical trials. In this study we investigate the potential of DBT with variable dose distribution by using a phantom in which details of interest are within a heterogeneous background.

Methods

To compare the performance of a commercial digital mammography unit and a DBT prototype, 2D and 3D images of the breast phantom were obtained at similar dose levels.

Results

As expected, DBT showed superior performance over digital mammography. Although certain details of interest are not detectable with digital mammography, DBT can reveal their signal by reducing the complexity of tissue structures. Additionally, the potential of the central projection in variable dose DBT is similar to the standard projection obtained with digital mammography. Finally, the uniform and variable dose approaches provided almost identical reconstructed slices.

Conclusion

This preliminary investigation demonstrates that breast tomosynthesis acquired with variable dose distribution exhibits inherent 3D reconstruction advantages for structure noise removal and provides a 2D projection with a physical image quality close to that of standard mammography.
Literature
1.
go back to reference Dobbins JT, Godfrey DJ (2003) Digital x-ray tomosynthesis: current state of the art and clinical potential. Phys Med Biol 48:R65–R106PubMedCrossRef Dobbins JT, Godfrey DJ (2003) Digital x-ray tomosynthesis: current state of the art and clinical potential. Phys Med Biol 48:R65–R106PubMedCrossRef
2.
go back to reference Poplack SP, Tosteson TD, Kogel CA, Nagy HM (2007) Digital breast tomosynthesis: initial experience in 98 women with abnormal digital screening mammography. Am J Roentgenol 189:616–623CrossRef Poplack SP, Tosteson TD, Kogel CA, Nagy HM (2007) Digital breast tomosynthesis: initial experience in 98 women with abnormal digital screening mammography. Am J Roentgenol 189:616–623CrossRef
3.
go back to reference Andersson I, Ikeda DM, Zackrisson S, Ruschin M, Svahn T, Timberg P, Timberg A (2008) Breast tomosynthesis and digital mammography: a comparison of breast cancer visibility and BIRADS classification in a population of cancers with subtle mammographic findings. Eur Radiol 18:2817–2825PubMedCrossRef Andersson I, Ikeda DM, Zackrisson S, Ruschin M, Svahn T, Timberg P, Timberg A (2008) Breast tomosynthesis and digital mammography: a comparison of breast cancer visibility and BIRADS classification in a population of cancers with subtle mammographic findings. Eur Radiol 18:2817–2825PubMedCrossRef
4.
go back to reference Gennaro G, Toledano A, di Maggio C, Baldan E, Bezzon E, La Grassa M, Pescarini L, Polico I, Proietti A, Toffoli A, Muzzio PC (2009) Digital breast tomosynthesis versus digital mammography: a clinical performance study. Eur Radiol 20:1545–1553PubMedCrossRef Gennaro G, Toledano A, di Maggio C, Baldan E, Bezzon E, La Grassa M, Pescarini L, Polico I, Proietti A, Toffoli A, Muzzio PC (2009) Digital breast tomosynthesis versus digital mammography: a clinical performance study. Eur Radiol 20:1545–1553PubMedCrossRef
5.
go back to reference Wu T, Stewart A, Stanton M, McCauley T, Phillips W, Kopans DB, Moore RH, Eberhard JW, Opsahl-Ong B, Niklason L, Williams MB (2003) Tomographic mammography using a limited number of low-dose cone-beam projection images. Med Phys 30:365–380PubMedCrossRef Wu T, Stewart A, Stanton M, McCauley T, Phillips W, Kopans DB, Moore RH, Eberhard JW, Opsahl-Ong B, Niklason L, Williams MB (2003) Tomographic mammography using a limited number of low-dose cone-beam projection images. Med Phys 30:365–380PubMedCrossRef
6.
go back to reference Stanton M, Stewart A, Phillips W (2004) Method and system for low-dose three-dimensional imaging of a scene, United States Patent 6744848 Stanton M, Stewart A, Phillips W (2004) Method and system for low-dose three-dimensional imaging of a scene, United States Patent 6744848
7.
go back to reference Das M, Gifford HC, O’Connor JM, Glick SJ (2009) Evaluation of a variable dose acquisition technique for microcalcification and mass detection in digital breast tomosynthesis. Med Phys 36:1976–1984PubMedCrossRef Das M, Gifford HC, O’Connor JM, Glick SJ (2009) Evaluation of a variable dose acquisition technique for microcalcification and mass detection in digital breast tomosynthesis. Med Phys 36:1976–1984PubMedCrossRef
10.
go back to reference Zhao B, Zhao W (2008) Imaging performance of an amorphous selenium digital mammography detector in a breast tomosynthesis system. Med Phys 35:1978–1987PubMedCrossRef Zhao B, Zhao W (2008) Imaging performance of an amorphous selenium digital mammography detector in a breast tomosynthesis system. Med Phys 35:1978–1987PubMedCrossRef
11.
go back to reference Chawla AS, Lo JY, Baker JA, Samei E (2009) Optimized image acquisition for breast tomosynthesis in projection and reconstruction space. Med Phys 36:4859–4869PubMedCrossRef Chawla AS, Lo JY, Baker JA, Samei E (2009) Optimized image acquisition for breast tomosynthesis in projection and reconstruction space. Med Phys 36:4859–4869PubMedCrossRef
12.
go back to reference Sechopoulos I, Ghetti C (2009) Optimization of the acquisition geometry in digital tomosynthesis of the breast. Med Phys 36:1199–1207PubMedCrossRef Sechopoulos I, Ghetti C (2009) Optimization of the acquisition geometry in digital tomosynthesis of the breast. Med Phys 36:1199–1207PubMedCrossRef
13.
go back to reference European Commission (2006) European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th edition European Commission (2006) European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th edition
16.
go back to reference Kastanis I, Arridge S, Stewart A, Gunn S, Ullberg C, Francke T (2010) 3D Digital Breast Tomosynthesis Using Total Variation Regularization, LNCS vol. 5116 Springer Heidelberg Kastanis I, Arridge S, Stewart A, Gunn S, Ullberg C, Francke T (2010) 3D Digital Breast Tomosynthesis Using Total Variation Regularization, LNCS vol. 5116 Springer Heidelberg
17.
go back to reference Sotthivirat S, Fessler JA (2002) Image recovery using partitioned-separable paraboloidal surrogate coordinate ascent algorithms. IEEE Trans Image Process 11:306–317PubMedCrossRef Sotthivirat S, Fessler JA (2002) Image recovery using partitioned-separable paraboloidal surrogate coordinate ascent algorithms. IEEE Trans Image Process 11:306–317PubMedCrossRef
18.
19.
go back to reference Bochud FO, Valley J-F, Verdun FR, Hessler C, Schnyder P (1999) Estimation of the noisy component of anatomical backgrounds. Med Phys 26:1365–1370PubMedCrossRef Bochud FO, Valley J-F, Verdun FR, Hessler C, Schnyder P (1999) Estimation of the noisy component of anatomical backgrounds. Med Phys 26:1365–1370PubMedCrossRef
Metadata
Title
A novel approach to digital breast tomosynthesis for simultaneous acquisition of 2D and 3D images
Authors
Sara Vecchio
Achille Albanese
Paolo Vignoli
Angelo Taibi
Publication date
01-06-2011
Publisher
Springer-Verlag
Published in
European Radiology / Issue 6/2011
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-010-2041-y

Other articles of this Issue 6/2011

European Radiology 6/2011 Go to the issue