01-09-2010 | Computed Tomography

Quantification of radiation dose savings in cardiac computed tomography using prospectively triggered mode and ECG pulsing: a phantom study

Authors: Lukas Lehmkuhl, Dieter Gosch, H. D. Nagel, Patrick Stumpp, Thomas Kahn, Matthias Gutberlet

Published in: European Radiology | Issue 9/2010

Login to get access

Abstract

Objective

To quantify radiation dose reduction in cardiac computed tomography (CT) using a prospectively triggered mode compared with a retrospective ECG-gated helical mode.

Methods

Absorbed organ doses in cardiac 64-row multidetector CT were quantified using an anthropomorphic male Alderson phantom with 74 thermoluminescence dosimeters. Three different imaging protocols were applied: retrospective ECG-gating, retrospective ECG-gating with additional ECG-pulsing, and a prospectively triggered mode. The measured organ doses were compared with dose estimation by a mathematical phantom.

Results

Compared with the retrospective ECG-gating mode, the mean relative organ doses were reduced by 44% using ECG pulsing and by 76% using the prospectively triggered mode. The range of dose savings varied from 34% to 49% using ECG pulsing and from 65% to 87% using the prospectively triggered mode. The effective dose was 16.5 mSv using retrospective gating, 9.2 mSv using retrospective gating with ECG pulsing and 4.0 mSv using the prospectively triggered mode.

Conclusions

Our measurements confirm the high dose-saving potential of the prospectively triggered technique in cardiac CT. The reduction in the organ doses measured corresponds to estimates determined by the mathematical phantom. The effective dose calculated by the mathematical phantom was, in some cases, significantly lower than that calculated using the anthropomorphic phantom.
Literature
1.
go back to reference Rosamond W, Flegal K, Furie K, Go A, Greenlund K, Haase N et al (2008) Heart disease and stroke statistics–2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 117:e25–146CrossRefPubMed Rosamond W, Flegal K, Furie K, Go A, Greenlund K, Haase N et al (2008) Heart disease and stroke statistics–2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 117:e25–146CrossRefPubMed
2.
go back to reference Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E et al (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732CrossRefPubMed Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E et al (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732CrossRefPubMed
3.
go back to reference Meijboom WB, Meijs MF, Schuijf JD, Cramer MJ, Mollet NR, van Mieghem CA et al (2008) Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. J Am Coll Cardiol 52:2135–2144CrossRefPubMed Meijboom WB, Meijs MF, Schuijf JD, Cramer MJ, Mollet NR, van Mieghem CA et al (2008) Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. J Am Coll Cardiol 52:2135–2144CrossRefPubMed
4.
go back to reference Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I et al (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359:2324–2336CrossRefPubMed Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I et al (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359:2324–2336CrossRefPubMed
5.
go back to reference Coles DR, Smail MA, Negus IS, Wilde P, Oberhoff M, Karsch KR et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47:1840–1845CrossRefPubMed Coles DR, Smail MA, Negus IS, Wilde P, Oberhoff M, Karsch KR et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47:1840–1845CrossRefPubMed
6.
go back to reference Hausleiter J, Meyer T, Hadamitzky M, Huber E, Zankl M, Martinoff S et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310CrossRefPubMed Hausleiter J, Meyer T, Hadamitzky M, Huber E, Zankl M, Martinoff S et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310CrossRefPubMed
7.
go back to reference Hurwitz LM, Reiman RE, Yoshizumi TT, Goodman PC, Toncheva G, Nguyen G et al (2007) Radiation dose from contemporary cardiothoracic multidetector CT protocols with an anthropomorphic female phantom: implications for cancer induction. Radiology 245:742–750CrossRefPubMed Hurwitz LM, Reiman RE, Yoshizumi TT, Goodman PC, Toncheva G, Nguyen G et al (2007) Radiation dose from contemporary cardiothoracic multidetector CT protocols with an anthropomorphic female phantom: implications for cancer induction. Radiology 245:742–750CrossRefPubMed
8.
go back to reference Nickoloff EL, Alderson PO (2007) A comparative study of thoracic radiation doses from 64-slice cardiac CT. Br J Radiol 80:537–544CrossRefPubMed Nickoloff EL, Alderson PO (2007) A comparative study of thoracic radiation doses from 64-slice cardiac CT. Br J Radiol 80:537–544CrossRefPubMed
9.
go back to reference Gopal A, Mao SS, Karlsberg D, Young E, Waggoner J, Ahmadi N et al (2008) Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography. Int J Cardiovasc Imaging 25:405–416CrossRefPubMed Gopal A, Mao SS, Karlsberg D, Young E, Waggoner J, Ahmadi N et al (2008) Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography. Int J Cardiovasc Imaging 25:405–416CrossRefPubMed
10.
go back to reference Klass O, Jeltsch M, Feuerlein S, Brunner H, Nagel HD, Walker MJ et al (2008) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836CrossRefPubMed Klass O, Jeltsch M, Feuerlein S, Brunner H, Nagel HD, Walker MJ et al (2008) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836CrossRefPubMed
11.
go back to reference Shuman WP, Branch KR, May JM, Mitsumori LM, Lockhart DW, Dubinsky TJ et al (2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 248:431–437CrossRefPubMed Shuman WP, Branch KR, May JM, Mitsumori LM, Lockhart DW, Dubinsky TJ et al (2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 248:431–437CrossRefPubMed
12.
go back to reference Geleijns J, Van Unnik JG, Zoetelief J, Zweers D, Broerse JJ (1994) Comparison of two methods for assessing patient dose from computed tomography. Br J Radiol 67:360–365CrossRefPubMed Geleijns J, Van Unnik JG, Zoetelief J, Zweers D, Broerse JJ (1994) Comparison of two methods for assessing patient dose from computed tomography. Br J Radiol 67:360–365CrossRefPubMed
13.
go back to reference Shrimpton PC (2004) Assessment of patient dose in CT. NRPB-PE/1/2004 Shrimpton PC (2004) Assessment of patient dose in CT. NRPB-PE/1/2004
14.
go back to reference Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248CrossRefPubMed Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248CrossRefPubMed
15.
go back to reference Huda W, Sandison GA (1984) Estimation of mean organ doses in diagnostic radiology from Rando phantom measurements. Health Phys 47:463–467PubMed Huda W, Sandison GA (1984) Estimation of mean organ doses in diagnostic radiology from Rando phantom measurements. Health Phys 47:463–467PubMed
16.
go back to reference Golikov VY, Nikitin VV (1989) Estimation of the mean organ doses and the effective dose equivalent from Rando phantom measurements. Health Phys 56:111–115PubMed Golikov VY, Nikitin VV (1989) Estimation of the mean organ doses and the effective dose equivalent from Rando phantom measurements. Health Phys 56:111–115PubMed
17.
go back to reference Kramer R, Zankl M, Williams G, Drexler G (1982) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part I. The male (Adam) and female (Eva) adult mathematical phantoms. GSF-Report S-885. Neuherberg Kramer R, Zankl M, Williams G, Drexler G (1982) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part I. The male (Adam) and female (Eva) adult mathematical phantoms. GSF-Report S-885. Neuherberg
18.
go back to reference International Commission on Radiological Protection (ICRP) 60 (1991) Recommendations of the International Commission on Radiological Protection. Pergamon Press, Oxford International Commission on Radiological Protection (ICRP) 60 (1991) Recommendations of the International Commission on Radiological Protection. Pergamon Press, Oxford
19.
go back to reference Stamm G, Nagel HD (2002) CT-expo–a novel program for dose evaluation in CT. Rofo 174:1570–1576PubMed Stamm G, Nagel HD (2002) CT-expo–a novel program for dose evaluation in CT. Rofo 174:1570–1576PubMed
20.
go back to reference Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part VI. Organ doses from computed tomographic examinations. GSF-Report 30/91. Institute for Radiation Biology, Neuherberg Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part VI. Organ doses from computed tomographic examinations. GSF-Report 30/91. Institute for Radiation Biology, Neuherberg
21.
go back to reference Shleien B (1973) Review of determinations of radiation dose to the active bone marrow from diagnostic X-ray examinations. DHEW publication—no. (FDA) 74–8007, Rockville Shleien B (1973) Review of determinations of radiation dose to the active bone marrow from diagnostic X-ray examinations. DHEW publication—no. (FDA) 74–8007, Rockville
22.
go back to reference DeMarco JJ, Cagnon CH, Cody DD, Stevens DM, McCollough CH, Zankl M et al (2007) Estimating radiation doses from multidetector CT using Monte Carlo simulations: effects of different size voxelized patient models on magnitudes of organ and effective dose. Phys Med Biol 52:2583–2597CrossRefPubMed DeMarco JJ, Cagnon CH, Cody DD, Stevens DM, McCollough CH, Zankl M et al (2007) Estimating radiation doses from multidetector CT using Monte Carlo simulations: effects of different size voxelized patient models on magnitudes of organ and effective dose. Phys Med Biol 52:2583–2597CrossRefPubMed
23.
go back to reference Hidajat N, Vogl T, Schroder RJ, Felix R (1996) Calculated organ doses and effective dosage for computerized tomography examination of the thorax and abdomen: are these doses realistic? Rofo 164:382–387PubMed Hidajat N, Vogl T, Schroder RJ, Felix R (1996) Calculated organ doses and effective dosage for computerized tomography examination of the thorax and abdomen: are these doses realistic? Rofo 164:382–387PubMed
24.
go back to reference Veit R, Panzer W, Zankl M, Scheurer C (1992) Vergleich berechneter und gemessener Dosen an einem anthropomorphen Phantom. Z Med Phys 2:123–126 Veit R, Panzer W, Zankl M, Scheurer C (1992) Vergleich berechneter und gemessener Dosen an einem anthropomorphen Phantom. Z Med Phys 2:123–126
25.
go back to reference Stolzmann P, Scheffel H, Schertler T, Frauenfelder T, Leschka S, Husmann L et al (2008) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18:592–599CrossRefPubMed Stolzmann P, Scheffel H, Schertler T, Frauenfelder T, Leschka S, Husmann L et al (2008) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18:592–599CrossRefPubMed
26.
go back to reference Wang M, Qi HT, Wang XM, Wang T, Chen JH, Liu C (2008) Dose performance and image quality: Dual source CT versus single source CT in cardiac CT angiography. Eur J Radiol 72:396–400CrossRefPubMed Wang M, Qi HT, Wang XM, Wang T, Chen JH, Liu C (2008) Dose performance and image quality: Dual source CT versus single source CT in cardiac CT angiography. Eur J Radiol 72:396–400CrossRefPubMed
27.
go back to reference Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507CrossRefPubMed Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507CrossRefPubMed
28.
go back to reference Hermann F, Martinoff S, Meyer T, Hadamitzky M, Jiang C, Hendrich E et al (2008) Reduction of radiation dose estimates in cardiac 64-slice CT angiography in patients after coronary artery bypass graft surgery. Invest Radiol 43:253–260CrossRefPubMed Hermann F, Martinoff S, Meyer T, Hadamitzky M, Jiang C, Hendrich E et al (2008) Reduction of radiation dose estimates in cardiac 64-slice CT angiography in patients after coronary artery bypass graft surgery. Invest Radiol 43:253–260CrossRefPubMed
29.
go back to reference Trabold T, Buchgeister M, Kuttner A, Heuschmid M, Kopp AF, Schroder S et al (2003) Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo 175:1051–1055PubMed Trabold T, Buchgeister M, Kuttner A, Heuschmid M, Kopp AF, Schroder S et al (2003) Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo 175:1051–1055PubMed
30.
go back to reference Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197CrossRefPubMed Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197CrossRefPubMed
31.
go back to reference Blankstein R, Shah A, Pale R, Abbara S, Bezerra H, Bolen M et al (2009) Radiation dose and image quality of prospective triggering with dual-source cardiac computed tomography. Am J Cardiol 103:1168–1173CrossRefPubMed Blankstein R, Shah A, Pale R, Abbara S, Bezerra H, Bolen M et al (2009) Radiation dose and image quality of prospective triggering with dual-source cardiac computed tomography. Am J Cardiol 103:1168–1173CrossRefPubMed
32.
go back to reference Hirai N, Horiguchi J, Fujioka C, Kiguchi M, Yamamoto H, Matsuura N et al (2008) Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology 248:424–430CrossRefPubMed Hirai N, Horiguchi J, Fujioka C, Kiguchi M, Yamamoto H, Matsuura N et al (2008) Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology 248:424–430CrossRefPubMed
33.
go back to reference Shuman WP, Branch KR, May JM, Mitsumori LM, Strote JN, Warren BH et al (2009) Whole-chest 64-MDCT of emergency department patients with nonspecific chest pain: radiation dose and coronary artery image quality with prospective ECG triggering versus retrospective ECG gating. AJR Am J Roentgenol 192:1662–1667CrossRefPubMed Shuman WP, Branch KR, May JM, Mitsumori LM, Strote JN, Warren BH et al (2009) Whole-chest 64-MDCT of emergency department patients with nonspecific chest pain: radiation dose and coronary artery image quality with prospective ECG triggering versus retrospective ECG gating. AJR Am J Roentgenol 192:1662–1667CrossRefPubMed
34.
go back to reference Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323CrossRefPubMed Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323CrossRefPubMed
35.
go back to reference The 2007 Recommendations of the International Commission on Radiological Protection (2007) ICRP publication 103, Ann ICRP 37:1–332 The 2007 Recommendations of the International Commission on Radiological Protection (2007) ICRP publication 103, Ann ICRP 37:1–332
36.
go back to reference Hart D, Jones DG, Wang M (1994) Normalised organ doses for medical X-ray examinations calculated using Monte-Carlo techniques (NRPB-SR262). National Radiological Protection Board, Chilton Hart D, Jones DG, Wang M (1994) Normalised organ doses for medical X-ray examinations calculated using Monte-Carlo techniques (NRPB-SR262). National Radiological Protection Board, Chilton
37.
go back to reference Schindera ST, Nelson RC, Lee ER, DeLong DM, Ngyen G, Toncheva G et al (2007) Abdominal multislice CT for obese patients: effect on image quality and radiation dose in a phantom study. Acad Radiol 14:486–494CrossRefPubMed Schindera ST, Nelson RC, Lee ER, DeLong DM, Ngyen G, Toncheva G et al (2007) Abdominal multislice CT for obese patients: effect on image quality and radiation dose in a phantom study. Acad Radiol 14:486–494CrossRefPubMed
Metadata
Title
Quantification of radiation dose savings in cardiac computed tomography using prospectively triggered mode and ECG pulsing: a phantom study
Authors
Lukas Lehmkuhl
Dieter Gosch
H. D. Nagel
Patrick Stumpp
Thomas Kahn
Matthias Gutberlet
Publication date
01-09-2010
Publisher
Springer-Verlag
Published in
European Radiology / Issue 9/2010
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-010-1767-x

Other articles of this Issue 9/2010

European Radiology 9/2010 Go to the issue