Published in: European Radiology 9/2007

01-09-2007 | Physics

Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system

Authors: Paula Toroi, Federica Zanca, Kenneth C. Young, Chantal van Ongeval, Guy Marchal, Hilde Bosmans

Published in: European Radiology | Issue 9/2007

Login to get access

Abstract

The signal difference-to-noise ratio (SDNR) between aluminium sheets and a homogeneous background was measured for various radiation qualities and breast thicknesses to determine the optimal radiation quality when using a Novation DR mammography system. Breast simulating phantoms, with a thickness from 2 cm to 7 cm, and aluminium sheet, with a thickness of 0.2 mm, were used. Three different combinations of anode/filter material and a wide range of tube voltages were employed for each phantom thickness. Each radiation quality was studied using three different dose levels. The tungsten (W) anode and rhodium (Rh) filter combination achieved the specified SDNR at the lowest mean glandular dose for all the phantom thicknesses and X-ray tube voltages. The difference between the doses for different anode/filter combinations increased with the phantom thickness. For a 5-cm phantom, with a peak tube voltage of 27 kV and a SDNR of 5, the mean glandular dose associated with the use of W/Rh was reduced by 49% when compared to the molybdenum/molybdenum (Mo/Mo) anode/filter combination and by 33% when compared to Mo/Rh. Based on these measurements, the use of the W/Rh anode/filter can be recommended. It remains important, however, to select the appropriate dose level.
Literature
1.
go back to reference van Ongeval C, Bosmans H, Van Steen A, Joossens K, Celis V, Van Goethem M, Verslegers I, Nijs K, Rogge F, Marchal G (2006) Evaluation of the diagnostic value of a computed radiography system by comparison of digital hard copy images with screen-film mammography: results of a prospective clinical trial. Eur Radiol 16:1360–1366. DOI 10.1007/s00330-005-0134-9 PubMedCrossRef van Ongeval C, Bosmans H, Van Steen A, Joossens K, Celis V, Van Goethem M, Verslegers I, Nijs K, Rogge F, Marchal G (2006) Evaluation of the diagnostic value of a computed radiography system by comparison of digital hard copy images with screen-film mammography: results of a prospective clinical trial. Eur Radiol 16:1360–1366. DOI 10.​1007/​s00330-005-0134-9 PubMedCrossRef
2.
go back to reference Saunders RS Jr, Samei E, Jesneck JL, Lo JL (2005) Physical characterization of a prototype selenium-based full field digital mammography detector. Med Phys 32:588–599PubMedCrossRef Saunders RS Jr, Samei E, Jesneck JL, Lo JL (2005) Physical characterization of a prototype selenium-based full field digital mammography detector. Med Phys 32:588–599PubMedCrossRef
3.
go back to reference Doyle P, Martin CJ, Gentle D (2006) Application of contrast-to-noise ration in optimizing beam quality for digital chest radiography: comparison of experimental measurements and theoretical simulations. Phys Med Biol 51:2953–2970PubMedCrossRef Doyle P, Martin CJ, Gentle D (2006) Application of contrast-to-noise ration in optimizing beam quality for digital chest radiography: comparison of experimental measurements and theoretical simulations. Phys Med Biol 51:2953–2970PubMedCrossRef
4.
go back to reference Gingold EL, Xizeng W, Barnes GT (1995) Contrast and dose with Mo-Mo, Mo-Rh and Rh-Rh target-filter combinations in mammography. Radiology 195:639–644PubMed Gingold EL, Xizeng W, Barnes GT (1995) Contrast and dose with Mo-Mo, Mo-Rh and Rh-Rh target-filter combinations in mammography. Radiology 195:639–644PubMed
5.
go back to reference Desponds L, Depeursinge C, Grecescu M, Hessler C, Samiri A, Valley JF (1991) Influence of anode and filter material on image quality and glandular dose for screen-film mammography. Phys Med Biol 36:1165–1182PubMedCrossRef Desponds L, Depeursinge C, Grecescu M, Hessler C, Samiri A, Valley JF (1991) Influence of anode and filter material on image quality and glandular dose for screen-film mammography. Phys Med Biol 36:1165–1182PubMedCrossRef
6.
go back to reference Jennings RJ, Eastgate RJ, Siedband MP, Ergun DL (1981) Optimal X-ray spectra for screen-film mammography. Med Phys 8:629–639PubMedCrossRef Jennings RJ, Eastgate RJ, Siedband MP, Ergun DL (1981) Optimal X-ray spectra for screen-film mammography. Med Phys 8:629–639PubMedCrossRef
7.
go back to reference Dance DR, Thilander Klang A, Sandborg M, Skinner CL, Castellano S, Alm Carlsson G (2000) Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study. Br J Radiol 73:1056–1067PubMed Dance DR, Thilander Klang A, Sandborg M, Skinner CL, Castellano S, Alm Carlsson G (2000) Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study. Br J Radiol 73:1056–1067PubMed
8.
go back to reference Fahrig R, Yaffe MJ (1994) Optimization of spectral shape in digital mammography: Dependence on anode material, breast thickness, and lesion type. Med Phys 21:1473–1481PubMedCrossRef Fahrig R, Yaffe MJ (1994) Optimization of spectral shape in digital mammography: Dependence on anode material, breast thickness, and lesion type. Med Phys 21:1473–1481PubMedCrossRef
9.
go back to reference Fahrig R, Yaffe MJ (1994) A model for optimization of spectral shape in digital mammography. Med Phys 21:1463–1471PubMedCrossRef Fahrig R, Yaffe MJ (1994) A model for optimization of spectral shape in digital mammography. Med Phys 21:1463–1471PubMedCrossRef
10.
go back to reference Fahrig R, Rowlands JA, Yaffe MJ (1996) X-ray imaging with amorphous selenium: Optimal spectra for digital mammography. Med Phys 23:557–567PubMedCrossRef Fahrig R, Rowlands JA, Yaffe MJ (1996) X-ray imaging with amorphous selenium: Optimal spectra for digital mammography. Med Phys 23:557–567PubMedCrossRef
11.
go back to reference Berns EA, Hendrick RE, Cutter GR (2003) Optimization of technique factors for silicon diode array full-field digital mammography system and comparison to screen-film mammography with matched average glandular dose. Med Phys 30:334–340PubMedCrossRef Berns EA, Hendrick RE, Cutter GR (2003) Optimization of technique factors for silicon diode array full-field digital mammography system and comparison to screen-film mammography with matched average glandular dose. Med Phys 30:334–340PubMedCrossRef
12.
go back to reference Obenauer S, Hermann K-P, Grabbe E (2003) Dose reduction in full-field digital mammography: an anthromorphic breast phantom study. Br J Radiol 76:478–482PubMedCrossRef Obenauer S, Hermann K-P, Grabbe E (2003) Dose reduction in full-field digital mammography: an anthromorphic breast phantom study. Br J Radiol 76:478–482PubMedCrossRef
13.
go back to reference Venkatakrishnan V, Yavuz M, Niklason LT, Opsahl-Ong B, Han S, Landberg C, Nevin R, Hamberg L, Kopans D (1999) Experimental and theoretical spectral optimization for digital mammography. Proceedings of the SPIE 3659:142–149CrossRef Venkatakrishnan V, Yavuz M, Niklason LT, Opsahl-Ong B, Han S, Landberg C, Nevin R, Hamberg L, Kopans D (1999) Experimental and theoretical spectral optimization for digital mammography. Proceedings of the SPIE 3659:142–149CrossRef
14.
go back to reference Young KC, Oduko JM, Bosmans H (2005) Optimal beam quality selection in digital mammography, In: Proceedings of the 7th International Workshop on Digital Mammography:105–109 Young KC, Oduko JM, Bosmans H (2005) Optimal beam quality selection in digital mammography, In: Proceedings of the 7th International Workshop on Digital Mammography:105–109
15.
go back to reference Young KC (2004) Breast dose surveys in the NHSBSP: software and instruction manual, Version 2.0. NHSBSP Report 04/05, Guildford Young KC (2004) Breast dose surveys in the NHSBSP: software and instruction manual, Version 2.0. NHSBSP Report 04/05, Guildford
16.
go back to reference Dance DR, Skinner CL, Young KC, Beckett JR, Korte CJ (2000) Additional factors for estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys Med Biol 45:3225–3240PubMedCrossRef Dance DR, Skinner CL, Young KC, Beckett JR, Korte CJ (2000) Additional factors for estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys Med Biol 45:3225–3240PubMedCrossRef
17.
go back to reference Moore AC, Dance DR, Evans DS, Lawinski CP, Pitcher EM, Rust A, Young KC (2005) Commissioning and routine testing of mammographic X-ray systems, Institute of Physics and Engineering in Medicine. IPEM Report No. 89 Moore AC, Dance DR, Evans DS, Lawinski CP, Pitcher EM, Rust A, Young KC (2005) Commissioning and routine testing of mammographic X-ray systems, Institute of Physics and Engineering in Medicine. IPEM Report No. 89
18.
go back to reference Abramoff MD, Magelhaes PJ, Ram SJ (2004) Image processing with Image. J Biophoton Int 11:36–42 Abramoff MD, Magelhaes PJ, Ram SJ (2004) Image processing with Image. J Biophoton Int 11:36–42
19.
go back to reference Carton AK, Bosmans H, Van Ongeval C (2003) Contrast visibility of simulated microcalcifications in full field mammography systems. Proceedings of the SPIE 5034:412–423CrossRef Carton AK, Bosmans H, Van Ongeval C (2003) Contrast visibility of simulated microcalcifications in full field mammography systems. Proceedings of the SPIE 5034:412–423CrossRef
20.
go back to reference Perry N, Broeders M, de Wolf C, Törnberg S, Holland R, von Karsa L (2006) European guidelines for quality assurance in breast cancer screening and diagnosis, 4th edition. European Communities, Luxembourg: chapter 2 Perry N, Broeders M, de Wolf C, Törnberg S, Holland R, von Karsa L (2006) European guidelines for quality assurance in breast cancer screening and diagnosis, 4th edition. European Communities, Luxembourg: chapter 2
21.
go back to reference Pachoud M, Lepori D, Valley J-F, Verdun FR (2004) A new test phantom with different breast tissue compositions for image quality assessment in conventional and digital mammography. Phys Med Biol 49:5267–5281PubMedCrossRef Pachoud M, Lepori D, Valley J-F, Verdun FR (2004) A new test phantom with different breast tissue compositions for image quality assessment in conventional and digital mammography. Phys Med Biol 49:5267–5281PubMedCrossRef
22.
go back to reference Pöyry P, Zanca F, Bosmans H (2006) Experimental investigation of the necessity for extra flat field corrections in quality control of digital mammography. In: Proceedings of the 8th International Workshop on Digital Mammography:475–481 Pöyry P, Zanca F, Bosmans H (2006) Experimental investigation of the necessity for extra flat field corrections in quality control of digital mammography. In: Proceedings of the 8th International Workshop on Digital Mammography:475–481
23.
go back to reference Young KC, Cook JJH, Oduko JM (2006) Use of the European Protocol to Optimise a Digital Mammography System. In: Proceedings of the 8th International Workshop on Digital Mammography:362–369 Young KC, Cook JJH, Oduko JM (2006) Use of the European Protocol to Optimise a Digital Mammography System. In: Proceedings of the 8th International Workshop on Digital Mammography:362–369
24.
go back to reference ICRP 60 (1991) 1990 Recommendations of the International Commission on Radiological Protection. Ann ICRP 21:1–3 ICRP 60 (1991) 1990 Recommendations of the International Commission on Radiological Protection. Ann ICRP 21:1–3
25.
go back to reference Smans K, Bosmans H, Xiao M, Carton AK, Marchal G (2006) Towards a proposition of diagnostic (dose) reference level for mammographic acquisitions in breast screening measurements in Belgium. Rad Prot Dosim 117(1–3):321–326CrossRef Smans K, Bosmans H, Xiao M, Carton AK, Marchal G (2006) Towards a proposition of diagnostic (dose) reference level for mammographic acquisitions in breast screening measurements in Belgium. Rad Prot Dosim 117(1–3):321–326CrossRef
26.
go back to reference Young KC, Burch A, Oduko JM (2005) Radiation doses received in the UK Breast Screening Programme in 2001 and 2002. Br J Radiol 78:207–218PubMedCrossRef Young KC, Burch A, Oduko JM (2005) Radiation doses received in the UK Breast Screening Programme in 2001 and 2002. Br J Radiol 78:207–218PubMedCrossRef
27.
go back to reference Burgess AE, Jacobson FL, Judy PF (2001) Human observer detection experiments with mammograms and power-law noise. Med Phys 28:419–437PubMedCrossRef Burgess AE, Jacobson FL, Judy PF (2001) Human observer detection experiments with mammograms and power-law noise. Med Phys 28:419–437PubMedCrossRef
Metadata
Title
Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system
Authors
Paula Toroi
Federica Zanca
Kenneth C. Young
Chantal van Ongeval
Guy Marchal
Hilde Bosmans
Publication date
01-09-2007
Publisher
Springer-Verlag
Published in
European Radiology / Issue 9/2007
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-006-0574-x

Other articles of this Issue 9/2007

European Radiology 9/2007 Go to the issue