Sequential and Simultaneous 4-Antenna Microwave Ablation in an Ex Vivo Bovine Liver Model
Authors:
Tian-qi Zhang, Sen-miao Huang, Yang-kui Gu, Xiong-ying Jiang, Zhi-mei Huang, Han-xia Deng, Jin-hua Huang
Published in:
CardioVascular and Interventional Radiology
|
Issue 10/2019
Login to get access
Abstract
Aims
To determine the sizes and shapes of ablation zones in 4-antenna microwave ablation in ex vivo bovine liver model under different conditions of power delivery patterns, antenna spacings, and ablation durations, for further using of multi-antenna MWA strategies in the treatment of large hepatocellular carcinoma.
Methods
We tested protocols of eight ablations each on ex vivo bovine livers, involving simultaneous or sequential activation of four microwave antennas, spaced either 3 cm, 4 cm, or 5 cm apart, for either 10 or 15 min, at 60-W power. We determined the diameters, shapes, and temperatures of the ablation zones.
Results
Compared to sequential power delivery, simultaneous power delivery resulted in significantly larger ablation zone diameters (P < .001). The temperatures in ablation zones were significantly higher for simultaneous than for sequential power delivery. The largest ablation diameter (7.45 ± 0.06 cm) resulted from simultaneous delivery for 15 min using 4-cm antenna spacing.
Conclusions
Simultaneous 4-antenna microwave ablation results in larger ablation zones than sequential ablation, and 4-cm antenna spacing with a 15-minute ablation duration creates the largest ablation zone. This information may provide multi-antenna MWA strategies for large HCC in the further clinical practice.