Percutaneous Radiofrequency Ablation of Painful Spinal Tumors Adjacent to the Spinal Cord with Real-Time Monitoring of Spinal Canal Temperature: A Prospective Study

Authors: Atsuhiro Nakatsuka, Koichiro Yamakado, Haruyuki Takaki, Junji Uraki, Masashi Makita, Fumiyoshi Oshima, Kan Takeda

Published in: CardioVascular and Interventional Radiology | Issue 1/2009

Login to get access

Abstract

Purpose

To prospectively evaluate the feasibility, safety, and clinical utility of bone radiofrequency (RF) ablation with real-time monitoring of the spinal canal temperature for the treatment of spinal tumors adjacent to the spinal cord.

Materials and Methods

Our Institutional Review Board approved this study. Patients gave informed consent. The inclusion criteria were (a) a painful spinal metastasis and (b) a distance of 1 cm or less between the metastasis and the spinal cord. The thermocouple was placed in the spinal canal under CT fluoroscopic guidance. When the spinal canal temperature reached 45°C, RF application was immediately stopped. RF ablation was considered technically successful when the procedure was performed without major complications. Clinical success was defined as a fall in the visual analogue scale score of at least 2 points.

Results

Ten patients with spinal tumors measuring 3–8 cm (mean, 4.9 ± 1.5 cm) were enrolled. The distance between the tumor and the spinal cord was 1–6 mm (mean, 2.4 ± 1.6 mm). All procedures were judged technically successful (100%). The spinal canal temperature did not exceed 45°C in 9 of the 10 patients (90%). In the remaining patient, the temperature rose to 48°C, resulting in transient neural damage, although RF application was immediately stopped when the temperature reached 45°C. Clinical success was achieved within 1 week in all patients (100%).

Conclusion

Bone RF ablation with real-time monitoring of the spinal canal temperature is feasible, safe, and clinically useful for the treatment of painful spinal metastases adjacent to the spinal cord.
Literature
1.
go back to reference Livraghi T, Goldberg SN, Lazzaroni S et al (1999) Small hepatocellular carcinoma: treatment with radio-frequency ablation versus ethanol injection. Radiology 210:655–661PubMed Livraghi T, Goldberg SN, Lazzaroni S et al (1999) Small hepatocellular carcinoma: treatment with radio-frequency ablation versus ethanol injection. Radiology 210:655–661PubMed
2.
go back to reference Yamakado K, Nakatsuka A, Kobayashi S et al (2006) Radiofrequency ablation combined with renal arterial embolization for the treatment of unresectable renal cell carcinoma larger than 3.5 cm: initial experience. Cardiovasc Intervent Radiol 29:389–394PubMedCrossRef Yamakado K, Nakatsuka A, Kobayashi S et al (2006) Radiofrequency ablation combined with renal arterial embolization for the treatment of unresectable renal cell carcinoma larger than 3.5 cm: initial experience. Cardiovasc Intervent Radiol 29:389–394PubMedCrossRef
3.
go back to reference Yamakado K, Hase H, Matsuoka T et al (2007) Radiofrequency ablation for the treatment of unresectable lung metastases in patients with colorectal cancer: a multicenter study in Japan. J Vasc Interv Radiol 18:303–398CrossRef Yamakado K, Hase H, Matsuoka T et al (2007) Radiofrequency ablation for the treatment of unresectable lung metastases in patients with colorectal cancer: a multicenter study in Japan. J Vasc Interv Radiol 18:303–398CrossRef
4.
go back to reference Akeboshi M, Yamakado K, Nakatsuka A et al (2004) Percutaneous radiofrequency ablation of lung neoplasms: initial therapeutic response. J Vasc Interv Radiol 15:463–470PubMed Akeboshi M, Yamakado K, Nakatsuka A et al (2004) Percutaneous radiofrequency ablation of lung neoplasms: initial therapeutic response. J Vasc Interv Radiol 15:463–470PubMed
5.
go back to reference Gronemeyer DH, Schirp S, Gevargez A (2002) Image-guided radiofrequency ablation of spinal tumors: preliminary experience with an expandable array electrode. Cancer J 8:33–39PubMedCrossRef Gronemeyer DH, Schirp S, Gevargez A (2002) Image-guided radiofrequency ablation of spinal tumors: preliminary experience with an expandable array electrode. Cancer J 8:33–39PubMedCrossRef
6.
go back to reference Callstrom MR, Charboneau JW, Goetz MP et al (2002) Painful metastases involving bone: feasibility of percutaneous CT- and US-guided radio-frequency ablation. Radiology 224:87–97PubMedCrossRef Callstrom MR, Charboneau JW, Goetz MP et al (2002) Painful metastases involving bone: feasibility of percutaneous CT- and US-guided radio-frequency ablation. Radiology 224:87–97PubMedCrossRef
7.
go back to reference Toyota N, Naito A, Kakizawa H et al (2005) Radiofrequency ablation therapy combined with cementoplasty for painful bone metastases: initial experience. Cardiovasc Intervent Radiol 28:578–583PubMedCrossRef Toyota N, Naito A, Kakizawa H et al (2005) Radiofrequency ablation therapy combined with cementoplasty for painful bone metastases: initial experience. Cardiovasc Intervent Radiol 28:578–583PubMedCrossRef
8.
go back to reference Goetz MP, Callstrom MR, Charboneau JW et al (2004) Percutaneous image-guided radiofrequency ablation of painful metastases involving bone: a multicenter study. J Clin Oncol 22:300–306PubMedCrossRef Goetz MP, Callstrom MR, Charboneau JW et al (2004) Percutaneous image-guided radiofrequency ablation of painful metastases involving bone: a multicenter study. J Clin Oncol 22:300–306PubMedCrossRef
9.
go back to reference Nakatsuka A, Yamakado K, Maeda M et al (2004) Radiofrequency ablation combined with bone cement injection for the treatment of bone malignancies. J Vasc Interv Radiol 15:707–712PubMed Nakatsuka A, Yamakado K, Maeda M et al (2004) Radiofrequency ablation combined with bone cement injection for the treatment of bone malignancies. J Vasc Interv Radiol 15:707–712PubMed
10.
go back to reference Kojima H, Tanigawa N, Kariya S, Komemushi A et al (2006) Clinical assessment of percutaneous radiofrequency ablation for painful metastatic bone tumors. Cardiovasc Intervent Radiol 29:1022–1026PubMedCrossRef Kojima H, Tanigawa N, Kariya S, Komemushi A et al (2006) Clinical assessment of percutaneous radiofrequency ablation for painful metastatic bone tumors. Cardiovasc Intervent Radiol 29:1022–1026PubMedCrossRef
11.
go back to reference Rhim H, Goldberg SN, Dodd GD III et al (2001) Essential techniques for successful radiofrequency thermal ablation of malignant hepatic tumors. Radiographics 21:S17–S39PubMed Rhim H, Goldberg SN, Dodd GD III et al (2001) Essential techniques for successful radiofrequency thermal ablation of malignant hepatic tumors. Radiographics 21:S17–S39PubMed
12.
go back to reference Varghese T, Zagzebski JA, Chen Q et al (2002) Ultrasound monitoring of temperature change during radiofrequency ablation: preliminary in-vivo results. Ultrasound Med Biol 28:321–329PubMedCrossRef Varghese T, Zagzebski JA, Chen Q et al (2002) Ultrasound monitoring of temperature change during radiofrequency ablation: preliminary in-vivo results. Ultrasound Med Biol 28:321–329PubMedCrossRef
13.
go back to reference Froese G, Das RM, Dunscombe PB (1991) The sensitivity of the thoracolumbar spinal cord of the mouse to hyperthermia. Radiat Res 125:173–180PubMedCrossRef Froese G, Das RM, Dunscombe PB (1991) The sensitivity of the thoracolumbar spinal cord of the mouse to hyperthermia. Radiat Res 125:173–180PubMedCrossRef
14.
go back to reference Yamane T, Tateishi A, Cho S et al (1992) The effects of hyperthermia on the spinal cord. Spine 17:1386–1391PubMedCrossRef Yamane T, Tateishi A, Cho S et al (1992) The effects of hyperthermia on the spinal cord. Spine 17:1386–1391PubMedCrossRef
15.
go back to reference Dupuy DE, Hong R, Oliver B et al (2000) Radiofrequency ablation of spinal tumors: temperature distribution in the spinal canal. AJR Am J Roentgenol 175:1263–1266PubMed Dupuy DE, Hong R, Oliver B et al (2000) Radiofrequency ablation of spinal tumors: temperature distribution in the spinal canal. AJR Am J Roentgenol 175:1263–1266PubMed
16.
go back to reference Diehn FE, Neeman Z, Hvizda JL et al (2003) Remote thermometry to avoid complications in radiofrequency ablation. J Vasc Interv Radiol 14:1569–1576PubMed Diehn FE, Neeman Z, Hvizda JL et al (2003) Remote thermometry to avoid complications in radiofrequency ablation. J Vasc Interv Radiol 14:1569–1576PubMed
17.
go back to reference Zubrod CG, Schneiderman M, Frei E et al (1960) Appraisal of methods for the study of chemotherapy of cancer in man: comparative trial of nitrogen mustard and trimethylene triophoshoramide. J Chron Dis 11:7–33CrossRef Zubrod CG, Schneiderman M, Frei E et al (1960) Appraisal of methods for the study of chemotherapy of cancer in man: comparative trial of nitrogen mustard and trimethylene triophoshoramide. J Chron Dis 11:7–33CrossRef
18.
go back to reference Chen WT, Shih TTF, Chen RC et al (2002) Blood perfusion of vertebral lesions evaluated with gadolinium-enhanced dynamic MRI: in comparison with compression fracture and metastasis. J Magn Reson Imaging 15:308–314PubMedCrossRef Chen WT, Shih TTF, Chen RC et al (2002) Blood perfusion of vertebral lesions evaluated with gadolinium-enhanced dynamic MRI: in comparison with compression fracture and metastasis. J Magn Reson Imaging 15:308–314PubMedCrossRef
19.
go back to reference Maeda M, Sakuma H, Maier SE et al (2003) Quantitative assessment of diffusion abnormalities in benign and malignant vertebral compression fractures by line scan diffusion-weighted imaging. AJR Am J Roentgenol 181:1203–1209PubMed Maeda M, Sakuma H, Maier SE et al (2003) Quantitative assessment of diffusion abnormalities in benign and malignant vertebral compression fractures by line scan diffusion-weighted imaging. AJR Am J Roentgenol 181:1203–1209PubMed
20.
go back to reference Goldberg SN, Grassi CJ, Cardella JF et al (2005) Image-guided tumor ablation: standardization of terminology and reporting criteria. Radiology 235:728–739 Goldberg SN, Grassi CJ, Cardella JF et al (2005) Image-guided tumor ablation: standardization of terminology and reporting criteria. Radiology 235:728–739
21.
go back to reference Boon JM, Abrahams PH, Meiring JH et al (2004) Lumbar puncture: anatomical review of a clinical skill. Clin Anat 17:544–553 Boon JM, Abrahams PH, Meiring JH et al (2004) Lumbar puncture: anatomical review of a clinical skill. Clin Anat 17:544–553
22.
go back to reference Parnass SM, Schmidt KJ (1990) Adverse effects of spinal and epidural anesthesia. Drug Safety 5:179–194PubMedCrossRef Parnass SM, Schmidt KJ (1990) Adverse effects of spinal and epidural anesthesia. Drug Safety 5:179–194PubMedCrossRef
23.
go back to reference Goldberg SN, Stein MC, Gazelle GS et al (1999) Percutaneous radiofrequency tissue ablation: optimization of pulsed-radiofrequency technique to increase coagulation necrosis. J Vasc Interv Radiol 10:907–916PubMedCrossRef Goldberg SN, Stein MC, Gazelle GS et al (1999) Percutaneous radiofrequency tissue ablation: optimization of pulsed-radiofrequency technique to increase coagulation necrosis. J Vasc Interv Radiol 10:907–916PubMedCrossRef
24.
go back to reference Thiagalingam A, D’Avila A, McPherson C et al (2007) Impedance and temperature monitoring improve the safety of closed-loop irrigated-tip radiofrequency ablation. J Cardiovasc Electrophysiol 18:318–325PubMedCrossRef Thiagalingam A, D’Avila A, McPherson C et al (2007) Impedance and temperature monitoring improve the safety of closed-loop irrigated-tip radiofrequency ablation. J Cardiovasc Electrophysiol 18:318–325PubMedCrossRef
25.
go back to reference Antz M, Otomo K, Nakagawa H et al (2001) Radiofrequency catheter ablation with the split-tip electrode in the temperature-controlled mode. J Pacing Clin Electrophysiol 24:1765–1773CrossRef Antz M, Otomo K, Nakagawa H et al (2001) Radiofrequency catheter ablation with the split-tip electrode in the temperature-controlled mode. J Pacing Clin Electrophysiol 24:1765–1773CrossRef
26.
go back to reference Lepetit-Coiffe M, Quesson B, Seror O et al (2006) Real-time monitoring of radiofrequency ablation of rabbit liver by respiratory-gated quantitative temperature MRI. J Magn Reson Imaging 24:152–159PubMedCrossRef Lepetit-Coiffe M, Quesson B, Seror O et al (2006) Real-time monitoring of radiofrequency ablation of rabbit liver by respiratory-gated quantitative temperature MRI. J Magn Reson Imaging 24:152–159PubMedCrossRef
Metadata
Title
Percutaneous Radiofrequency Ablation of Painful Spinal Tumors Adjacent to the Spinal Cord with Real-Time Monitoring of Spinal Canal Temperature: A Prospective Study
Authors
Atsuhiro Nakatsuka
Koichiro Yamakado
Haruyuki Takaki
Junji Uraki
Masashi Makita
Fumiyoshi Oshima
Kan Takeda
Publication date
01-01-2009
Publisher
Springer-Verlag
Published in
CardioVascular and Interventional Radiology / Issue 1/2009
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-008-9390-9

Other articles of this Issue 1/2009

CardioVascular and Interventional Radiology 1/2009 Go to the issue