Published in: Abdominal Radiology 12/2020

Open Access 01-12-2020 | Prostate Cancer | Pelvis

Quality Comparison of 3 Tesla multiparametric MRI of the prostate using a flexible surface receiver coil versus conventional surface coil plus endorectal coil setup

Authors: T. Ullrich, M. D. Kohli, M. A. Ohliger, K. Magudia, S. S. Arora, T. Barrett, L. K. Bittencourt, D. J. Margolis, L. Schimmöller, B. Turkbey, A. C. Westphalen

Published in: Abdominal Radiology | Issue 12/2020

Login to get access

Abstract

Purpose

To subjectively and quantitatively compare the quality of 3 Tesla magnetic resonance imaging of the prostate acquired with a novel flexible surface coil (FSC) and with a conventional endorectal coil (ERC).

Methods

Six radiologists independently reviewed 200 pairs of axial, high-resolution T2-weighted and diffusion-weighted image data sets, each containing one examination acquired with the FSC and one with the ERC, respectively. Readers selected their preferred examination from each pair and assessed every single examination using six quality criteria on 4-point scales. Signal-to-noise ratios were measured and compared.

Results

Two readers preferred FSC acquisition (36.5–45%) over ERC acquisition (13.5–15%) for both sequences combined, and four readers preferred ERC acquisition (41–46%). Analysis of pooled responses for both sequences from all readers shows no significant preference for FSC or ERC. Analysis of the individual sequences revealed a pooled preference for the FSC in T2WI (38.7% vs 17.8%) and for the ERC in DWI (50.9% vs 19.6%). Patients’ weight was the only weak predictor of a preference for the ERC acquisition (p = 0.04). SNR and CNR were significantly higher in the ERC acquisitions (p<0.001) except CNR differentiating tumor lesions from benign prostate (p=0.1).

Conclusion

Although readers have strong individual preferences, comparable subjective image quality can be obtained for prostate MRI with an ERC and the novel FSC. ERC imaging might be particularly valuable for sequences with inherently lower SNR as DWI and larger patients whereas the FSC is generally preferred in T2WI. FSC imaging generates a lower SNR than with an ERC.
Appendix
Available only for authorised users
Literature
1.
go back to reference Hoeks CM, Barentsz JO, Hambrock T et al (2011) Prostate cancer: multiparametric MR imaging for detection, localization, and staging. Radiology 261:46-66CrossRefPubMed Hoeks CM, Barentsz JO, Hambrock T et al (2011) Prostate cancer: multiparametric MR imaging for detection, localization, and staging. Radiology 261:46-66CrossRefPubMed
2.
go back to reference Ahmed HU, El-Shater Bosaily A, Brown LC et al (2017) Diagnostic accuracy of multi-parametric MRI and TRUS biopsy in prostate cancer (PROMIS): a paired validating confirmatory study. Lancet 389:815–822CrossRefPubMed Ahmed HU, El-Shater Bosaily A, Brown LC et al (2017) Diagnostic accuracy of multi-parametric MRI and TRUS biopsy in prostate cancer (PROMIS): a paired validating confirmatory study. Lancet 389:815–822CrossRefPubMed
3.
go back to reference Haider MA, Krieger A, Elliott C et al (2014) Prostate imaging: evaluation of a reusable two-channel endorectal receiver coil for MR imaging at 1.5 T. Radiology 270:556-65CrossRefPubMed Haider MA, Krieger A, Elliott C et al (2014) Prostate imaging: evaluation of a reusable two-channel endorectal receiver coil for MR imaging at 1.5 T. Radiology 270:556-65CrossRefPubMed
4.
go back to reference Turkbey B, Rosenkrantz AB, Haider MA, et al (2019) Prostate Imaging Reporting and Data System Version 2.1: 2019 Update of Prostate Imaging Reporting and Data System Version 2. Eur Urol 76:340–51CrossRefPubMed Turkbey B, Rosenkrantz AB, Haider MA, et al (2019) Prostate Imaging Reporting and Data System Version 2.1: 2019 Update of Prostate Imaging Reporting and Data System Version 2. Eur Urol 76:340–51CrossRefPubMed
5.
go back to reference Mazaheri Y, Vargas HA, Nyman G et al (2013) Diffusion-weighted MRI of the prostate at 3.0 T: comparison of endorectal coil (ERC) MRI and phased-array coil (PAC) MRI The impact of SNR on ADC measurement. Eur J Radiol 82:515-20CrossRef Mazaheri Y, Vargas HA, Nyman G et al (2013) Diffusion-weighted MRI of the prostate at 3.0 T: comparison of endorectal coil (ERC) MRI and phased-array coil (PAC) MRI The impact of SNR on ADC measurement. Eur J Radiol 82:515-20CrossRef
6.
go back to reference Turkbey B, Merino MJ, Gallardo EC et al (2014) Comparison of endorectal coil and nonendorectal coil T2W and diffusion-weighted MRI at 3 Tesla for localizing prostate cancer: correlation with whole-mount histopathology. J Magn Reson Imaging 39:1443-8CrossRefPubMed Turkbey B, Merino MJ, Gallardo EC et al (2014) Comparison of endorectal coil and nonendorectal coil T2W and diffusion-weighted MRI at 3 Tesla for localizing prostate cancer: correlation with whole-mount histopathology. J Magn Reson Imaging 39:1443-8CrossRefPubMed
7.
go back to reference Costa DN, Yuan Q, Rofsky NM et al (2016) Comparison of prostate cancer detection at 3-T MRI with and without an endorectal coil: A prospective, paired-patient study. Urol Oncol 34:255-255CrossRefPubMed Costa DN, Yuan Q, Rofsky NM et al (2016) Comparison of prostate cancer detection at 3-T MRI with and without an endorectal coil: A prospective, paired-patient study. Urol Oncol 34:255-255CrossRefPubMed
8.
go back to reference Gawlitza J, Reiss-Zimmermann M, Thörmer G, et al (2017) Impact of the use of an endorectal coil for 3 T prostate MRI on image quality and cancer detection rate. Sci Rep 7:40640CrossRefPubMedPubMedCentral Gawlitza J, Reiss-Zimmermann M, Thörmer G, et al (2017) Impact of the use of an endorectal coil for 3 T prostate MRI on image quality and cancer detection rate. Sci Rep 7:40640CrossRefPubMedPubMedCentral
9.
go back to reference Barth BK, Cornelius A, Nanz D et al (2016) Comparison of image quality and patient discomfort in prostate MRI: pelvic phased array coil vs. endorectal coil. Abdom Radiol 41:2218-2226CrossRef Barth BK, Cornelius A, Nanz D et al (2016) Comparison of image quality and patient discomfort in prostate MRI: pelvic phased array coil vs. endorectal coil. Abdom Radiol 41:2218-2226CrossRef
10.
go back to reference Baur AD, Daqqaq T, Wagner M et al (2016) T2- and diffusion-weighted magnetic resonance imaging at 3T for the detection of prostate cancer with and without endorectal coil: An intraindividual comparison of image quality and diagnostic performance. Eur J Radiol 85:1075-84CrossRefPubMed Baur AD, Daqqaq T, Wagner M et al (2016) T2- and diffusion-weighted magnetic resonance imaging at 3T for the detection of prostate cancer with and without endorectal coil: An intraindividual comparison of image quality and diagnostic performance. Eur J Radiol 85:1075-84CrossRefPubMed
11.
go back to reference Lui D, Modhafar A, Haider MA, Wong A (2015) Monte Carlo-based noise compensation in coil intensity corrected endorectal MRI. BMC Med Imaging 15:15:43CrossRefPubMedPubMedCentral Lui D, Modhafar A, Haider MA, Wong A (2015) Monte Carlo-based noise compensation in coil intensity corrected endorectal MRI. BMC Med Imaging 15:15:43CrossRefPubMedPubMedCentral
12.
go back to reference Murer S, Scheidler J, Mueller-Lisse UL et al (2019) Two-centre comparative experimental study of biparametric MRI at 3.0 T with and without endorectal coil using kiwifruit (Actinidia deliciosa) as a phantom for human prostate. Eur Radiol Exp 3:3 Murer S, Scheidler J, Mueller-Lisse UL et al (2019) Two-centre comparative experimental study of biparametric MRI at 3.0 T with and without endorectal coil using kiwifruit (Actinidia deliciosa) as a phantom for human prostate. Eur Radiol Exp 3:3
13.
go back to reference O’Donohoe RL, Dunne RM, Kimbrell V et al (2019) Prostate MRI using an external phased array wearable pelvic coil at 3T: comparison with an endorectal coil. Abdom Radiol (NY) 44:1062-1069CrossRef O’Donohoe RL, Dunne RM, Kimbrell V et al (2019) Prostate MRI using an external phased array wearable pelvic coil at 3T: comparison with an endorectal coil. Abdom Radiol (NY) 44:1062-1069CrossRef
14.
go back to reference Barth BK, Rupp NJ, Cornelius A et al (2019) Diagnostic Accuracy of a MR Protocol Acquired with and without Endorectal Coil for Detection of Prostate Cancer: A Multicenter Study. Curr Urol 12:88-96CrossRefPubMedPubMedCentral Barth BK, Rupp NJ, Cornelius A et al (2019) Diagnostic Accuracy of a MR Protocol Acquired with and without Endorectal Coil for Detection of Prostate Cancer: A Multicenter Study. Curr Urol 12:88-96CrossRefPubMedPubMedCentral
15.
go back to reference Woo S, Suh CH, Kim SY et al (2017) Diagnostic Performance of Prostate Imaging Reporting and Data System Version 2 for Detection of Prostate Cancer: A Systematic Review and Diagnostic Meta-analysis. Eur Urol 72:177-188CrossRefPubMed Woo S, Suh CH, Kim SY et al (2017) Diagnostic Performance of Prostate Imaging Reporting and Data System Version 2 for Detection of Prostate Cancer: A Systematic Review and Diagnostic Meta-analysis. Eur Urol 72:177-188CrossRefPubMed
16.
go back to reference Kim BS, Kim TH, Kwon TG et al (2012) Comparison of pelvic phased array versus endorectal coil magnetic resonance imaging at 3 Tesla for local staging of prostate cancer. Yonsei Med J 53:550-6CrossRefPubMedPubMedCentral Kim BS, Kim TH, Kwon TG et al (2012) Comparison of pelvic phased array versus endorectal coil magnetic resonance imaging at 3 Tesla for local staging of prostate cancer. Yonsei Med J 53:550-6CrossRefPubMedPubMedCentral
17.
go back to reference Mirak SA, Shakeri S, Bajgiran AM et al (2019) Three Tesla Multiparametric Magnetic Resonance Imaging: Comparison of Performance with and without Endorectal Coil for Prostate Cancer Detection, PI-RADS™ version 2 Category and Staging with Whole Mount Histopathology Correlation. J Urol 201:496-502CrossRefPubMed Mirak SA, Shakeri S, Bajgiran AM et al (2019) Three Tesla Multiparametric Magnetic Resonance Imaging: Comparison of Performance with and without Endorectal Coil for Prostate Cancer Detection, PI-RADS™ version 2 Category and Staging with Whole Mount Histopathology Correlation. J Urol 201:496-502CrossRefPubMed
18.
go back to reference Noworolski SM, Reed GD, Kurhanewicz J et al (2014) Post-processing correction of the endorectal coil reception effects in MR spectroscopic imaging of the prostate. J Magn Reson Imaging 32:654–662CrossRef Noworolski SM, Reed GD, Kurhanewicz J et al (2014) Post-processing correction of the endorectal coil reception effects in MR spectroscopic imaging of the prostate. J Magn Reson Imaging 32:654–662CrossRef
19.
go back to reference M. Ramalho, V. Heredia, M. Tsurusaki et al (2009) Quantitative and qualitative comparison of 1.5 T and 3.0 T MRI in patients with chronic liver diseases, J. Magn. Reson. Imaging 29: 869–879CrossRefPubMed M. Ramalho, V. Heredia, M. Tsurusaki et al (2009) Quantitative and qualitative comparison of 1.5 T and 3.0 T MRI in patients with chronic liver diseases, J. Magn. Reson. Imaging 29: 869–879CrossRefPubMed
20.
go back to reference Ullrich T, Quentin M, Oelers C et al (2017) Magnetic resonance imaging of the prostate at 1.5 versus 3.0T: A prospective comparison study of image quality. Eur J Radiol 90:192-197CrossRefPubMed Ullrich T, Quentin M, Oelers C et al (2017) Magnetic resonance imaging of the prostate at 1.5 versus 3.0T: A prospective comparison study of image quality. Eur J Radiol 90:192-197CrossRefPubMed
21.
go back to reference Westphalen AC, Noworolski SM, Harisinghani M et al (2016) High-Resolution 3-T Endorectal Prostate MRI: A Multireader Study of Radiologist Preference and Perceived Interpretive Quality of 2D and 3D T2-Weighted Fast Spin-Echo MR Images AJR Am J Roentgenol 206:86–91 Westphalen AC, Noworolski SM, Harisinghani M et al (2016) High-Resolution 3-T Endorectal Prostate MRI: A Multireader Study of Radiologist Preference and Perceived Interpretive Quality of 2D and 3D T2-Weighted Fast Spin-Echo MR Images AJR Am J Roentgenol 206:86–91
22.
go back to reference Landis, J, & Koch, G (1977). The Measurement of Observer Agreement for Categorical Data. Biometrics 33:159-174CrossRefPubMed Landis, J, & Koch, G (1977). The Measurement of Observer Agreement for Categorical Data. Biometrics 33:159-174CrossRefPubMed
23.
go back to reference Barth BK, Cornelius A, Nanz D et al (2015) Diffusion-Weighted Imaging of the Prostate: Image Quality and Geometric Distortion of Readout-Segmented Versus Selective-Excitation Accelerated Acquisitions. Invest Radiol 50: 785-91CrossRefPubMed Barth BK, Cornelius A, Nanz D et al (2015) Diffusion-Weighted Imaging of the Prostate: Image Quality and Geometric Distortion of Readout-Segmented Versus Selective-Excitation Accelerated Acquisitions. Invest Radiol 50: 785-91CrossRefPubMed
24.
go back to reference Dikaios N, Punwani S, Hamy V et al (2014) Noise estimation from averaged diffusion weighted images: Can unbiased quantitative decay parameters assist cancer evaluation? Magn Reson Med 71:2105-17CrossRefPubMed Dikaios N, Punwani S, Hamy V et al (2014) Noise estimation from averaged diffusion weighted images: Can unbiased quantitative decay parameters assist cancer evaluation? Magn Reson Med 71:2105-17CrossRefPubMed
25.
go back to reference Ullrich T, Quentin M, Schmaltz AK et al (2018) Hyoscine butylbromide significantly decreases motion artefacts and allows better delineation of anatomic structures in mp-MRI of the prostate. Eur Radiol 28:17-23CrossRefPubMed Ullrich T, Quentin M, Schmaltz AK et al (2018) Hyoscine butylbromide significantly decreases motion artefacts and allows better delineation of anatomic structures in mp-MRI of the prostate. Eur Radiol 28:17-23CrossRefPubMed
26.
go back to reference Slough RA, Caglic I, Hansen NL et al (2018) Effect of hyoscine butylbromide on prostate multiparametric MRI anatomical and functional image quality. Clin Radiol 73:216.e9-216.e14CrossRef Slough RA, Caglic I, Hansen NL et al (2018) Effect of hyoscine butylbromide on prostate multiparametric MRI anatomical and functional image quality. Clin Radiol 73:216.e9-216.e14CrossRef
27.
go back to reference van Griethuysen JJM, Bus EM, Hauptmann M, et al (2018) Gas-induced sus- ceptibility artefacts on diffusion-weighted MRI of the rectum at 1.5 Tdeffect of applying a micro-enema to improve image quality. Eur J Radiol 99:131e7 van Griethuysen JJM, Bus EM, Hauptmann M, et al (2018) Gas-induced sus- ceptibility artefacts on diffusion-weighted MRI of the rectum at 1.5 Tdeffect of applying a micro-enema to improve image quality. Eur J Radiol 99:131e7
28.
go back to reference Martin GV, Kudchadker RJ, Bruno TL et al (2018) Comparison of prostate distortion by inflatable and rigid endorectal MRI coils in permanent prostate brachytherapy imaging. Brachytherapy 17:298-305CrossRefPubMed Martin GV, Kudchadker RJ, Bruno TL et al (2018) Comparison of prostate distortion by inflatable and rigid endorectal MRI coils in permanent prostate brachytherapy imaging. Brachytherapy 17:298-305CrossRefPubMed
29.
go back to reference Boroomand A, Shafiee MJ, Khalvati F et al (2016) Noise- compensated, bias-corrected diffusion weighted endorectal magnetic resonance imaging via a stochastically fully connected joint conditional random field model. IEEE Trans Med Imaging 35:2587–2597CrossRefPubMed Boroomand A, Shafiee MJ, Khalvati F et al (2016) Noise- compensated, bias-corrected diffusion weighted endorectal magnetic resonance imaging via a stochastically fully connected joint conditional random field model. IEEE Trans Med Imaging 35:2587–2597CrossRefPubMed
30.
go back to reference Heijmink SW, Futterer JJ, Hambrock T, et al (2007) Prostate cancer: body-array versus endorectal coil MR imaging at 3 T– comparison of image quality, localization, and staging performance. Radiology 244:184–195CrossRefPubMed Heijmink SW, Futterer JJ, Hambrock T, et al (2007) Prostate cancer: body-array versus endorectal coil MR imaging at 3 T– comparison of image quality, localization, and staging performance. Radiology 244:184–195CrossRefPubMed
Metadata
Title
Quality Comparison of 3 Tesla multiparametric MRI of the prostate using a flexible surface receiver coil versus conventional surface coil plus endorectal coil setup
Authors
T. Ullrich
M. D. Kohli
M. A. Ohliger
K. Magudia
S. S. Arora
T. Barrett
L. K. Bittencourt
D. J. Margolis
L. Schimmöller
B. Turkbey
A. C. Westphalen
Publication date
01-12-2020
Publisher
Springer US
Published in
Abdominal Radiology / Issue 12/2020
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-020-02641-0

Other articles of this Issue 12/2020

Abdominal Radiology 12/2020 Go to the issue
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.