Published in: Abdominal Radiology 3/2017

01-03-2017

Benefit of iodine density images to reduce out-of-field image artifacts at rapid kVp switching dual-energy CT

Authors: Brandan Dotson, Jack W. Lambert, Zhen J. Wang, Yuxin Sun, Michael A. Ohliger, Sebastian Winklhofer, Benjamin M. Yeh

Published in: Abdominal Radiology | Issue 3/2017

Login to get access

Abstract

Purpose

To evaluate the reduction of out-of-field artifacts caused by body parts outside the field of view (FOV) at rapid kVp switching dual-energy CT (rsDECT).

Materials and methods

This retrospective study was approved by our institutional review board. Informed consent was not required. We viewed 246 consecutive rsDECT thoracoabdominal scans to identify those with body parts outside the maximal FOV of 50 cm. The maximal length, thickness, and subjective severity of the out-of-field artifacts were recorded for the 40, 65, and 140 keV virtual monochromatic and iodine and water density images. Artifact severity was rated on a 6-point scale from 0 = absent to 5 = obscures intraabdominal/intrathoracic anatomic detail. Artifact thickness and severity scores were compared by t-test and Wilcoxon tests, respectively.

Results

In 20 of 246 scans (8.1%), body parts extended past the maximum FOV of 50 cm. The mean BMI of these 20 patients was 40.2 kg/m2 (range, 26.83–61.69 kg/m2), and out-of-field artifacts occurred for all 20. The mean out-of-field artifact maximal length was 16.6 cm. The mean artifact thickness was significantly less for iodine density (0.6 mm) than for the 65 keV and water density images (8.4 and 13.5 mm, respectively, p < 0.001 each comparison). The mean artifact severity score was lower for iodine density (0.2) than for the 65 keV and water density images (2.5 and 2.6, respectively, p < 0.001 each).

Conclusion

Iodine density images reduce out-of-field image artifact at rsDECT and assists in the evaluation of peripheral tissues that extend beyond the maximal CT FOV.
Literature
1.
go back to reference Bamberg F, Dierks A, Nikolaou K, et al. (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur Radiol 21:1424–1429CrossRefPubMed Bamberg F, Dierks A, Nikolaou K, et al. (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur Radiol 21:1424–1429CrossRefPubMed
2.
go back to reference Kelly D, Hasegawa I, Borders R, Hatabu H, Boiselle P (2004) High-resolution CT using MDCT: comparison of degree of motion artifact between volumetric and axial methods. Am J Roentgenol 182:757–759CrossRef Kelly D, Hasegawa I, Borders R, Hatabu H, Boiselle P (2004) High-resolution CT using MDCT: comparison of degree of motion artifact between volumetric and axial methods. Am J Roentgenol 182:757–759CrossRef
3.
go back to reference Brooks RA, Di Chiro G (1976) Beam hardening in X-ray reconstructive tomography. Phys Med Biol 21:390–398CrossRefPubMed Brooks RA, Di Chiro G (1976) Beam hardening in X-ray reconstructive tomography. Phys Med Biol 21:390–398CrossRefPubMed
5.
go back to reference Ng M, Fleming T, Robinson M, et al. (2014) Global, regional, and national prevalence of overweight and obesity in children and adults during 1980–2013: a systematic analysis for the Global Burden of Disease Study 2013. Lancet 384:766–781CrossRefPubMedPubMedCentral Ng M, Fleming T, Robinson M, et al. (2014) Global, regional, and national prevalence of overweight and obesity in children and adults during 1980–2013: a systematic analysis for the Global Burden of Disease Study 2013. Lancet 384:766–781CrossRefPubMedPubMedCentral
6.
go back to reference Ohnesorge B, Flohr T, Schwarz K, Heiken JP, Bae KT (2000) Efficient correction for CT image artifacts caused by objects extending outside the scan field of view. Med Phys 27:39–46. doi:10.1118/1.598855 CrossRefPubMed Ohnesorge B, Flohr T, Schwarz K, Heiken JP, Bae KT (2000) Efficient correction for CT image artifacts caused by objects extending outside the scan field of view. Med Phys 27:39–46. doi:10.​1118/​1.​598855 CrossRefPubMed
7.
go back to reference Maltz JS, Bose S, Shukla HP, Bani-Hashemi AR (2007) CT truncation artifact removal using water-equivalent thicknesses derived from truncated projection data. Conf Proc IEEE Eng Med Biol Soc 27:2907–2911. doi:10.1109/iembs.2007.4352937 Maltz JS, Bose S, Shukla HP, Bani-Hashemi AR (2007) CT truncation artifact removal using water-equivalent thicknesses derived from truncated projection data. Conf Proc IEEE Eng Med Biol Soc 27:2907–2911. doi:10.​1109/​iembs.​2007.​4352937
12.
go back to reference Pessis E, Campagna R, Sverzut J-M, et al. (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: Reduction of metal artifacts at CT. Radiographics 33:573–583CrossRefPubMed Pessis E, Campagna R, Sverzut J-M, et al. (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: Reduction of metal artifacts at CT. Radiographics 33:573–583CrossRefPubMed
13.
go back to reference Lee YH, Park KK, Song H-T, Kim S, Suh J-S (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340CrossRefPubMed Lee YH, Park KK, Song H-T, Kim S, Suh J-S (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340CrossRefPubMed
14.
go back to reference Coursey CA, Nelson RC, Boll DT, et al. (2010) Dual-energy multidetector CT: how does it work, what can it tell us, and when can we use it in abdominopelvic imaging? Radiographics 30:1037–1055. doi:10.1148/rg.304095175 CrossRefPubMed Coursey CA, Nelson RC, Boll DT, et al. (2010) Dual-energy multidetector CT: how does it work, what can it tell us, and when can we use it in abdominopelvic imaging? Radiographics 30:1037–1055. doi:10.​1148/​rg.​304095175 CrossRefPubMed
16.
go back to reference Stiller W, Skornitzke S, Fritz F, et al. (2015) Correlation of quantitative dual-energy computed tomography iodine maps and abdominal computed tomography perfusion measurements: are single-acquisition dual-energy computed tomography iodine maps more than a reduced-dose surrogate of conventional computed tomography perfusion? Invest Radiol 50:703–708. doi:10.1097/rli.0000000000000176 CrossRefPubMed Stiller W, Skornitzke S, Fritz F, et al. (2015) Correlation of quantitative dual-energy computed tomography iodine maps and abdominal computed tomography perfusion measurements: are single-acquisition dual-energy computed tomography iodine maps more than a reduced-dose surrogate of conventional computed tomography perfusion? Invest Radiol 50:703–708. doi:10.​1097/​rli.​0000000000000176​ CrossRefPubMed
18.
go back to reference Gupta S, Wagner-Bartak N, Jensen CT, et al. (2016) Dual-energy CT of pancreatic adenocarcinoma: reproducibility of primary tumor measurements and assessment of tumor conspicuity and margin sharpness. Abdom Radiol 41(7):1317–1324. doi:10.1007/s00261-016-0689-8 CrossRef Gupta S, Wagner-Bartak N, Jensen CT, et al. (2016) Dual-energy CT of pancreatic adenocarcinoma: reproducibility of primary tumor measurements and assessment of tumor conspicuity and margin sharpness. Abdom Radiol 41(7):1317–1324. doi:10.​1007/​s00261-016-0689-8 CrossRef
19.
go back to reference Winklhofer S, Lambert JW, Wang ZJ, et al. (2016) Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study. Abdom Radiol 41(8):1456–1465. doi:10.1007/s00261-016-0702-2 CrossRef Winklhofer S, Lambert JW, Wang ZJ, et al. (2016) Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study. Abdom Radiol 41(8):1456–1465. doi:10.​1007/​s00261-016-0702-2 CrossRef
21.
go back to reference Reynolds A (2011) Obesity and medical imaging challenges. Radiol Technol 82:219–239PubMed Reynolds A (2011) Obesity and medical imaging challenges. Radiol Technol 82:219–239PubMed
24.
go back to reference Guimaraes LS, Fletcher JG, Harmsen WS, et al. (2010) Appropriate patient selection at abdominal dual-energy CT using 80 kV: relationship between patient size, image noise, and image quality. Radiology 257:732–742. doi:10.1148/radiol.10092016 CrossRefPubMed Guimaraes LS, Fletcher JG, Harmsen WS, et al. (2010) Appropriate patient selection at abdominal dual-energy CT using 80 kV: relationship between patient size, image noise, and image quality. Radiology 257:732–742. doi:10.​1148/​radiol.​10092016 CrossRefPubMed
27.
go back to reference Wu X, Langan DA, Xu D, et al. (2009) Monochromatic CT image representation via fast switching dual kVp. In: Proceedings of SPIE 7258, Medical Imaging: Physics of Medical Imaging, 725845, 13 Mar 2009 Wu X, Langan DA, Xu D, et al. (2009) Monochromatic CT image representation via fast switching dual kVp. In: Proceedings of SPIE 7258, Medical Imaging: Physics of Medical Imaging, 725845, 13 Mar 2009
28.
go back to reference Winklhofer S, Lambert JW, Wang ZJ, et al. (2016) Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study. Abdom Radiol 41:1456–1465. doi:10.1007/s00261-016-0702-2 CrossRef Winklhofer S, Lambert JW, Wang ZJ, et al. (2016) Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study. Abdom Radiol 41:1456–1465. doi:10.​1007/​s00261-016-0702-2 CrossRef
Metadata
Title
Benefit of iodine density images to reduce out-of-field image artifacts at rapid kVp switching dual-energy CT
Authors
Brandan Dotson
Jack W. Lambert
Zhen J. Wang
Yuxin Sun
Michael A. Ohliger
Sebastian Winklhofer
Benjamin M. Yeh
Publication date
01-03-2017
Publisher
Springer US
Published in
Abdominal Radiology / Issue 3/2017
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-016-0978-2

Other articles of this Issue 3/2017

Abdominal Radiology 3/2017 Go to the issue

Classics in Abdominal Imaging

The northern exposure sign

Classics in Abdominal Imaging

Corkscrew esophagus