Prospective evaluation of prior image constrained compressed sensing (PICCS) algorithm in abdominal CT: a comparison of reduced dose with standard dose imaging

Authors: Meghan G. Lubner, Perry J. Pickhardt, David H. Kim, Jie Tang, Alejandro Munoz del Rio, Guang-Hong Chen

Published in: Abdominal Radiology | Issue 1/2015

Login to get access

Abstract

Purpose

To prospectively study CT dose reduction using the “prior image constrained compressed sensing” (PICCS) reconstruction technique.

Methods

Immediately following routine standard dose (SD) abdominal MDCT, 50 patients (mean age, 57.7 years; mean BMI, 28.8) underwent a second reduced dose (RD) scan (targeted dose reduction, 70%–90%). DLP, CTDIvol, and SSDE were compared. Several reconstruction algorithms (FBP, ASIR, and PICCS) were applied to the RD series. SD images with FBP served as reference standard. Two blinded readers evaluated each series for subjective image quality and focal lesion detection.

Results

Mean DLP, CTDIvol, and SSDE for RD series were 140.3 mGy cm (median 79.4), 3.7 mGy (median 1.8), and 4.2 mGy (median 2.3) compared with 493.7 mGy cm (median 345.8), 12.9 mGy (median 7.9 mGy), and 14.6 mGy (median 10.1) for SD series, respectively. Mean effective patient diameter was 30.1 cm (median 30), which translates to a mean SSDE reduction of 72% (P < 0.001). RD-PICCS image quality score was 2.8 ± 0.5, improved over the RD-FBP (1.7 ± 0.7) and RD-ASIR (1.9 ± 0.8) (P < 0.001), but lower than SD (3.5 ± 0.5) (P < 0.001). Readers detected 81% (184/228) of focal lesions on RD-PICCS series, vs. 67% (153/228) and 65% (149/228) for RD-FBP and RD-ASIR, respectively. Mean image noise was significantly reduced on RD-PICCS series (13.9 HU) compared with RD-FBP (57.2) and RD-ASIR (44.1) (P < 0.001).

Conclusion

PICCS allows for marked dose reduction at abdominal CT with improved image quality and diagnostic performance over reduced dose FBP and ASIR. Further study is needed to determine indication-specific dose reduction levels that preserve acceptable diagnostic accuracy relative to higher dose protocols.
Appendix
Available only for authorised users
Literature
1.
go back to reference Brenner DJ, Hall EJ (2007) Computed tomography—an increasing source of radiation exposure. N Engl J Med 357(22):2277–2284PubMedCrossRef Brenner DJ, Hall EJ (2007) Computed tomography—an increasing source of radiation exposure. N Engl J Med 357(22):2277–2284PubMedCrossRef
2.
go back to reference Mahesh M (2009) NCRP report number 160: its significance to medical imaging. J Am Coll Radiol 6(12):890–892PubMedCrossRef Mahesh M (2009) NCRP report number 160: its significance to medical imaging. J Am Coll Radiol 6(12):890–892PubMedCrossRef
3.
go back to reference Measurements NCoRPa (1987) Ionizing radiation exposure of the population of the United States. Bethesda: NCoRPa Measurements NCoRPa (1987) Ionizing radiation exposure of the population of the United States. Bethesda: NCoRPa
4.
go back to reference Pearce MS, Salotti JA, Little MP, et al. (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380(9840):499–505PubMedCentralPubMedCrossRef Pearce MS, Salotti JA, Little MP, et al. (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380(9840):499–505PubMedCentralPubMedCrossRef
6.
go back to reference Mulkens TH, Bellinck P, Baeyaert M, et al. (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237(1):213–223PubMedCrossRef Mulkens TH, Bellinck P, Baeyaert M, et al. (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237(1):213–223PubMedCrossRef
7.
go back to reference McCollough CH, Bruesewitz MR, Kofler JM Jr (2006) CT dose reduction and dose management tools: overview of available options. Radiographics 26(2):503–512PubMedCrossRef McCollough CH, Bruesewitz MR, Kofler JM Jr (2006) CT dose reduction and dose management tools: overview of available options. Radiographics 26(2):503–512PubMedCrossRef
8.
go back to reference Graser A, Wintersperger BJ, Suess C, Reiser MF, Becker CR (2006) Dose reduction and image quality in MDCT colonography using tube current modulation. AJR Am J Roentgenol 187(3):695–701PubMedCrossRef Graser A, Wintersperger BJ, Suess C, Reiser MF, Becker CR (2006) Dose reduction and image quality in MDCT colonography using tube current modulation. AJR Am J Roentgenol 187(3):695–701PubMedCrossRef
9.
go back to reference Greess H, Wolf H, Baum U, et al. (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10(2):391–394PubMedCrossRef Greess H, Wolf H, Baum U, et al. (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10(2):391–394PubMedCrossRef
10.
11.
go back to reference Gies M, Kalender WA, Wolf H, Suess C (1999) Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. Med Phys 26(11):2235–2247PubMedCrossRef Gies M, Kalender WA, Wolf H, Suess C (1999) Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. Med Phys 26(11):2235–2247PubMedCrossRef
12.
go back to reference Kalender WA, Wolf H, Suess C (1999) Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements. Med Phys 26(11):2248–2253PubMedCrossRef Kalender WA, Wolf H, Suess C (1999) Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements. Med Phys 26(11):2248–2253PubMedCrossRef
13.
go back to reference Haaga JR, Miraldi F, MacIntyre W, et al. (1981) The effect of mAs variation upon computed tomography image quality as evaluated by in vivo and in vitro studies. Radiology 138(2):449–454PubMedCrossRef Haaga JR, Miraldi F, MacIntyre W, et al. (1981) The effect of mAs variation upon computed tomography image quality as evaluated by in vivo and in vitro studies. Radiology 138(2):449–454PubMedCrossRef
14.
go back to reference Goetti R, Winklehner A, Gordic S, et al. (2012) Automated attenuation-based kilovoltage selection: preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta. AJR Am J Roentgenol 199(3):W380–W385PubMedCrossRef Goetti R, Winklehner A, Gordic S, et al. (2012) Automated attenuation-based kilovoltage selection: preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta. AJR Am J Roentgenol 199(3):W380–W385PubMedCrossRef
15.
go back to reference Supanich M, Tao Y, Nett B, et al. (2009) Radiation dose reduction in time-resolved CT angiography using highly constrained back projection reconstruction. Phys Med Biol 54(14):4575–4593PubMedCentralPubMedCrossRef Supanich M, Tao Y, Nett B, et al. (2009) Radiation dose reduction in time-resolved CT angiography using highly constrained back projection reconstruction. Phys Med Biol 54(14):4575–4593PubMedCentralPubMedCrossRef
16.
go back to reference Marin D, Nelson RC, Schindera ST, et al. (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm—initial clinical experience. Radiology 254(1):145–153PubMedCrossRef Marin D, Nelson RC, Schindera ST, et al. (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm—initial clinical experience. Radiology 254(1):145–153PubMedCrossRef
17.
go back to reference Prakash P, Kalra MK, Kambadakone AK, et al. (2010) Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol 45(4):202–210PubMedCrossRef Prakash P, Kalra MK, Kambadakone AK, et al. (2010) Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol 45(4):202–210PubMedCrossRef
18.
go back to reference Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol 194(1):191–199PubMedCrossRef Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol 194(1):191–199PubMedCrossRef
19.
go back to reference Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34(11):4526–4544PubMedCrossRef Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34(11):4526–4544PubMedCrossRef
20.
go back to reference Chen GH, Tang J, Leng S (2008) Prior image constrained compressed sensing (PICCS): a method to accurately reconstruct dynamic CT images from highly undersampled projection data sets. Med Phys 35(2):660–663PubMedCentralPubMedCrossRef Chen GH, Tang J, Leng S (2008) Prior image constrained compressed sensing (PICCS): a method to accurately reconstruct dynamic CT images from highly undersampled projection data sets. Med Phys 35(2):660–663PubMedCentralPubMedCrossRef
21.
go back to reference Pickhardt PJ, Lubner MG, Kim DH, et al. (2012) Abdominal CT with model-based iterative reconstruction (MBIR): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging. AJR Am J Roentgenol 199(6):1266–1274PubMedCentralPubMedCrossRef Pickhardt PJ, Lubner MG, Kim DH, et al. (2012) Abdominal CT with model-based iterative reconstruction (MBIR): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging. AJR Am J Roentgenol 199(6):1266–1274PubMedCentralPubMedCrossRef
22.
go back to reference Chang W, Lee JM, Lee K, et al. (2013) Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography. Invest Radiol 48:598–606PubMedCrossRef Chang W, Lee JM, Lee K, et al. (2013) Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography. Invest Radiol 48:598–606PubMedCrossRef
23.
go back to reference Hara AK, Paden RG, Silva AC, et al. (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193(3):764–771PubMedCrossRef Hara AK, Paden RG, Silva AC, et al. (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193(3):764–771PubMedCrossRef
24.
go back to reference Sagara Y, Hara AK, Pavlicek W, et al. (2010) Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol 195(3):713–719PubMedCrossRef Sagara Y, Hara AK, Pavlicek W, et al. (2010) Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol 195(3):713–719PubMedCrossRef
25.
go back to reference Kambadakone AR, Chaudhary NA, Desai GS, et al. (2011) Low-dose MDCT and CT enterography of patients with Crohn disease: feasibility of adaptive statistical iterative reconstruction. AJR Am J Roentgenol 196(6):W743–W752PubMedCrossRef Kambadakone AR, Chaudhary NA, Desai GS, et al. (2011) Low-dose MDCT and CT enterography of patients with Crohn disease: feasibility of adaptive statistical iterative reconstruction. AJR Am J Roentgenol 196(6):W743–W752PubMedCrossRef
26.
go back to reference Mitsumori LM, Shuman WP, Busey JM, Kolokythas O, Koprowicz KM (2012) Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose. Eur Radiol 22(1):138–143PubMedCrossRef Mitsumori LM, Shuman WP, Busey JM, Kolokythas O, Koprowicz KM (2012) Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose. Eur Radiol 22(1):138–143PubMedCrossRef
27.
go back to reference McCollough CH, Chen GH, Kalender W, et al. (2012) Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. Radiology 264(2):567–580PubMedCentralPubMedCrossRef McCollough CH, Chen GH, Kalender W, et al. (2012) Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. Radiology 264(2):567–580PubMedCentralPubMedCrossRef
28.
go back to reference Lubner MG, Pickhardt PJ, Tang J, Chen GH (2011) Reduced image noise at low-dose multidetector CT of the abdomen with prior image constrained compressed sensing algorithm. Radiology 260(1):248–256PubMedCrossRef Lubner MG, Pickhardt PJ, Tang J, Chen GH (2011) Reduced image noise at low-dose multidetector CT of the abdomen with prior image constrained compressed sensing algorithm. Radiology 260(1):248–256PubMedCrossRef
29.
go back to reference McCollough C, Cody D, Edyvean S, et al. (2008) The measurement, reporting and management of radiation dose in CT. College Park: American Association of Physicists in Medicine McCollough C, Cody D, Edyvean S, et al. (2008) The measurement, reporting and management of radiation dose in CT. College Park: American Association of Physicists in Medicine
30.
go back to reference Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257(1):158–166PubMedCrossRef Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257(1):158–166PubMedCrossRef
31.
go back to reference Boone JM, Strauss KJ, Cody DD, et al. (2011) Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. College Park: American Association of Physicists in Medicine Boone JM, Strauss KJ, Cody DD, et al. (2011) Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. College Park: American Association of Physicists in Medicine
32.
go back to reference Bankier AA, Kressel HY (2012) Through the looking glass revisited: the need for more meaning and less drama in the reporting of dose and dose reduction in CT. Radiology 265(1):4–8PubMedCrossRef Bankier AA, Kressel HY (2012) Through the looking glass revisited: the need for more meaning and less drama in the reporting of dose and dose reduction in CT. Radiology 265(1):4–8PubMedCrossRef
33.
go back to reference Flicek KT, Hara AK, Silva AC, et al. (2010) Reducing the radiation dose for CT colonography using adaptive statistical iterative reconstruction: a pilot study. AJR Am J Roentgenol 195(1):126–131PubMedCrossRef Flicek KT, Hara AK, Silva AC, et al. (2010) Reducing the radiation dose for CT colonography using adaptive statistical iterative reconstruction: a pilot study. AJR Am J Roentgenol 195(1):126–131PubMedCrossRef
34.
go back to reference Tang JT-LP, Chen G-H (2011) In: Proceedings of SPIE, Medical Imaging. Physics of Medical Imaging, vol 7961 Tang JT-LP, Chen G-H (2011) In: Proceedings of SPIE, Medical Imaging. Physics of Medical Imaging, vol 7961
35.
go back to reference Agresti ACB (1998) Approximate is better than “Exact” for interval estimation of binomial proportions. Am Stat 52(2):119–126 Agresti ACB (1998) Approximate is better than “Exact” for interval estimation of binomial proportions. Am Stat 52(2):119–126
36.
go back to reference Pinheiro JCBD (2000) Mixed-effects models in S and S-PLUS. New York: SpringerCrossRef Pinheiro JCBD (2000) Mixed-effects models in S and S-PLUS. New York: SpringerCrossRef
37.
go back to reference R Development Core Team (2009) R: a language and environment for statistical computing. Vienna: R Foundation for Statistical Computing R Development Core Team (2009) R: a language and environment for statistical computing. Vienna: R Foundation for Statistical Computing
38.
go back to reference Singh S, Kalra MK, Hsieh J, et al. (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257(2):373–383PubMedCrossRef Singh S, Kalra MK, Hsieh J, et al. (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257(2):373–383PubMedCrossRef
39.
go back to reference Noel PB, Fingerle AA, Renger B, et al. (2011) Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol 197(6):1404–1409PubMedCrossRef Noel PB, Fingerle AA, Renger B, et al. (2011) Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol 197(6):1404–1409PubMedCrossRef
40.
go back to reference Lee SJ, Park SH, Kim AY, et al. (2011) A prospective comparison of standard-dose CT enterography and 50% reduced-dose CT enterography with and without noise reduction for evaluating Crohn disease. AJR Am J Roentgenol 197(1):50–57PubMedCrossRef Lee SJ, Park SH, Kim AY, et al. (2011) A prospective comparison of standard-dose CT enterography and 50% reduced-dose CT enterography with and without noise reduction for evaluating Crohn disease. AJR Am J Roentgenol 197(1):50–57PubMedCrossRef
41.
go back to reference Gervaise A, Osemont B, Lecocq S, et al. (2012) CT image quality improvement using Adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22(2):295–301PubMedCrossRef Gervaise A, Osemont B, Lecocq S, et al. (2012) CT image quality improvement using Adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22(2):295–301PubMedCrossRef
42.
go back to reference Baker ME, Dong F, Primak A, et al. (2012) Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. AJR Am J Roentgenol 199(1):8–18PubMedCrossRef Baker ME, Dong F, Primak A, et al. (2012) Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. AJR Am J Roentgenol 199(1):8–18PubMedCrossRef
Metadata
Title
Prospective evaluation of prior image constrained compressed sensing (PICCS) algorithm in abdominal CT: a comparison of reduced dose with standard dose imaging
Authors
Meghan G. Lubner
Perry J. Pickhardt
David H. Kim
Jie Tang
Alejandro Munoz del Rio
Guang-Hong Chen
Publication date
01-01-2015
Publisher
Springer US
Published in
Abdominal Radiology / Issue 1/2015
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-014-0178-x

Other articles of this Issue 1/2015

Abdominal Radiology 1/2015 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.