Impact of [18F]FDG PET imaging parameters on automatic tumour delineation: need for improved tumour delineation methodology

Authors: Patsuree Cheebsumon, Maqsood Yaqub, Floris H. P. van Velden, Otto S. Hoekstra, Adriaan A. Lammertsma, Ronald Boellaard

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 12/2011

Login to get access

Abstract

Purpose

Delineation of tumour boundaries is important for quantification of [18F]fluoro-2-deoxy-D-glucose (FDG) positron emission tomography (PET) studies and for definition of biological target volumes in radiotherapy. Several (semi-)automatic tumour delineation methods have been proposed, but these methods differ substantially in estimating tumour volume and their performance may be affected by imaging parameters. The main purpose of this study was to explore the performance dependence of various (semi-)automatic tumour delineation methods on different imaging parameters, i.e. reconstruction parameters, noise levels and tumour characteristics, and thereby the need for standardization or inter-institute calibration.

Methods

Six different types of delineation methods were evaluated by assessing accuracy and precision in estimating tumour volume from simulations and phantom experiments. The evaluated conditions were various tumour sizes, iterative reconstruction algorithm settings and image filtering, tumour to background ratios (TBR), noise levels and region growing initializations.

Results

The accuracy of all automatic delineation methods was influenced when imaging parameters were varied. The performance of all tumour delineation methods depends on variation of TBR, image resolution and image noise level, and to a lesser extent on number of iterations during image reconstruction or the initialization method of the region generation. For sphere sizes larger than 20 mm diameter a contrast-oriented method provided the most accurate results, on average, over all simulated conditions. For threshold-based methods the accuracy of tumour delineation improved after image denoising/filtering.

Conclusion

The accuracy and precision of all studied tumour delineation methods was affected by physiological and imaging parameters. The latter illustrates the need for optimizing imaging parameters and/or for careful calibration and optimization of delineation methods.
Appendix
Available only for authorised users
Literature
1.
go back to reference Schaefer A, Kremp S, Hellwig D, Rübe C, Kirsch CM, Nestle U. A contrast-oriented algorithm for FDG-PET-based delineation of tumour volumes for the radiotherapy of lung cancer: derivation from phantom measurements and validation in patient data. Eur J Nucl Med Mol Imaging 2008;35:1989–99.PubMedCrossRef Schaefer A, Kremp S, Hellwig D, Rübe C, Kirsch CM, Nestle U. A contrast-oriented algorithm for FDG-PET-based delineation of tumour volumes for the radiotherapy of lung cancer: derivation from phantom measurements and validation in patient data. Eur J Nucl Med Mol Imaging 2008;35:1989–99.PubMedCrossRef
2.
go back to reference van Dalen JA, Hoffmann AL, Dicken V, Vogel WV, Wiering B, Ruers TJ, et al. A novel iterative method for lesion delineation and volumetric quantification with FDG PET. Nucl Med Commun 2007;28:485–93.PubMedCrossRef van Dalen JA, Hoffmann AL, Dicken V, Vogel WV, Wiering B, Ruers TJ, et al. A novel iterative method for lesion delineation and volumetric quantification with FDG PET. Nucl Med Commun 2007;28:485–93.PubMedCrossRef
3.
go back to reference Geets X, Lee JA, Bol A, Lonneux M, Grégoire V. A gradient-based method for segmenting FDG-PET images: methodology and validation. Eur J Nucl Med Mol Imaging 2007;34:1427–38.PubMedCrossRef Geets X, Lee JA, Bol A, Lonneux M, Grégoire V. A gradient-based method for segmenting FDG-PET images: methodology and validation. Eur J Nucl Med Mol Imaging 2007;34:1427–38.PubMedCrossRef
4.
go back to reference Zaidi H, El Naqa I. PET-guided delineation of radiation therapy treatment volumes: a survey of image segmentation techniques. Eur J Nucl Med Mol Imaging 2010;37:2165–87.PubMedCrossRef Zaidi H, El Naqa I. PET-guided delineation of radiation therapy treatment volumes: a survey of image segmentation techniques. Eur J Nucl Med Mol Imaging 2010;37:2165–87.PubMedCrossRef
5.
go back to reference Nestle U, Weber W, Hentschel M, Grosu AL. Biological imaging in radiation therapy: role of positron emission tomography. Phys Med Biol 2009;54:R1–R25.PubMedCrossRef Nestle U, Weber W, Hentschel M, Grosu AL. Biological imaging in radiation therapy: role of positron emission tomography. Phys Med Biol 2009;54:R1–R25.PubMedCrossRef
6.
go back to reference Fletcher JW, Djulbegovic B, Soares HP, Siegel BA, Lowe VJ, Lyman GH, et al. Recommendations on the use of 18F-FDG PET in oncology. J Nucl Med 2008;49:480–508.PubMedCrossRef Fletcher JW, Djulbegovic B, Soares HP, Siegel BA, Lowe VJ, Lyman GH, et al. Recommendations on the use of 18F-FDG PET in oncology. J Nucl Med 2008;49:480–508.PubMedCrossRef
7.
go back to reference de Geus-Oei LF, van der Heijden HF, Corstens FH, Oyen WJ. Predictive and prognostic value of FDG-PET in nonsmall-cell lung cancer: a systematic review. Cancer 2007;110:1654–64.PubMedCrossRef de Geus-Oei LF, van der Heijden HF, Corstens FH, Oyen WJ. Predictive and prognostic value of FDG-PET in nonsmall-cell lung cancer: a systematic review. Cancer 2007;110:1654–64.PubMedCrossRef
8.
go back to reference Erdi YE. The use of PET for radiotherapy. Curr Med Imaging Rev 2007;3:3–16.CrossRef Erdi YE. The use of PET for radiotherapy. Curr Med Imaging Rev 2007;3:3–16.CrossRef
9.
go back to reference Juweid ME, Cheson BD. Positron-emission tomography and assessment of cancer therapy. N Engl J Med 2006;354:496–507.PubMedCrossRef Juweid ME, Cheson BD. Positron-emission tomography and assessment of cancer therapy. N Engl J Med 2006;354:496–507.PubMedCrossRef
10.
go back to reference Roels S, Slagmolen P, Nuyts J, Lee JA, Loeckx D, Maes F, et al. Biological image-guided radiotherapy in rectal cancer: is there a role for FMISO or FLT, next to FDG? Acta Oncol 2008;47:1237–48.PubMedCrossRef Roels S, Slagmolen P, Nuyts J, Lee JA, Loeckx D, Maes F, et al. Biological image-guided radiotherapy in rectal cancer: is there a role for FMISO or FLT, next to FDG? Acta Oncol 2008;47:1237–48.PubMedCrossRef
11.
go back to reference MacManus M, Nestle U, Rosenzweig KE, Carrio I, Messa C, Belohlavek O, et al. Use of PET and PET/CT for radiation therapy planning: IAEA expert report 2006–2007. Radiother Oncol 2009;91:85–94.PubMedCrossRef MacManus M, Nestle U, Rosenzweig KE, Carrio I, Messa C, Belohlavek O, et al. Use of PET and PET/CT for radiation therapy planning: IAEA expert report 2006–2007. Radiother Oncol 2009;91:85–94.PubMedCrossRef
12.
go back to reference Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med 2009;50 Suppl 1:11S–20S.PubMedCrossRef Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med 2009;50 Suppl 1:11S–20S.PubMedCrossRef
13.
go back to reference Hatt M, Visvikis D, Le Rest CC. Autocontouring versus manual contouring. J Nucl Med 2011;52:658–9.PubMedCrossRef Hatt M, Visvikis D, Le Rest CC. Autocontouring versus manual contouring. J Nucl Med 2011;52:658–9.PubMedCrossRef
14.
go back to reference Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med 2004;45:1519–27.PubMed Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med 2004;45:1519–27.PubMed
15.
go back to reference Brix G, Zaers J, Adam LE, Bellemann ME, Ostertag H, Trojan H, et al. Performance evaluation of a whole-body PET scanner using the NEMA protocol. National Electrical Manufacturers Association. J Nucl Med 1997;38:1614–23.PubMed Brix G, Zaers J, Adam LE, Bellemann ME, Ostertag H, Trojan H, et al. Performance evaluation of a whole-body PET scanner using the NEMA protocol. National Electrical Manufacturers Association. J Nucl Med 1997;38:1614–23.PubMed
16.
go back to reference Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007;48:471–80.PubMed Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007;48:471–80.PubMed
17.
go back to reference Paulino AC, Koshy M, Howell R, Schuster D, Davis LW. Comparison of CT- and FDG-PET-defined gross tumor volume in intensity-modulated radiotherapy for head-and-neck cancer. Int J Radiat Oncol Biol Phys 2005;61:1385–92.PubMedCrossRef Paulino AC, Koshy M, Howell R, Schuster D, Davis LW. Comparison of CT- and FDG-PET-defined gross tumor volume in intensity-modulated radiotherapy for head-and-neck cancer. Int J Radiat Oncol Biol Phys 2005;61:1385–92.PubMedCrossRef
18.
go back to reference Nestle U, Kremp S, Schaefer-Schuler A, Sebastian-Welsch C, Hellwig D, Rübe C, et al. Comparison of different methods for delineation of 18F-FDG PET-positive tissue for target volume definition in radiotherapy of patients with non-small cell lung cancer. J Nucl Med 2005;46:1342–8.PubMed Nestle U, Kremp S, Schaefer-Schuler A, Sebastian-Welsch C, Hellwig D, Rübe C, et al. Comparison of different methods for delineation of 18F-FDG PET-positive tissue for target volume definition in radiotherapy of patients with non-small cell lung cancer. J Nucl Med 2005;46:1342–8.PubMed
19.
go back to reference Boellaard R, O’Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, et al. FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010;37:181–200.PubMedCrossRef Boellaard R, O’Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, et al. FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010;37:181–200.PubMedCrossRef
20.
go back to reference Daisne JF, Sibomana M, Bol A, Doumont T, Lonneux M, Grégoire V. Tri-dimensional automatic segmentation of PET volumes based on measured source-to-background ratios: influence of reconstruction algorithms. Radiother Oncol 2003;69:247–50.PubMedCrossRef Daisne JF, Sibomana M, Bol A, Doumont T, Lonneux M, Grégoire V. Tri-dimensional automatic segmentation of PET volumes based on measured source-to-background ratios: influence of reconstruction algorithms. Radiother Oncol 2003;69:247–50.PubMedCrossRef
21.
go back to reference Hatt M, Cheze Le Rest C, Albarghach N, Pradier O, Visvikis D. PET functional volume delineation: a robustness and repeatability study. Eur J Nucl Med Mol Imaging 2011;38:663–72.PubMedCrossRef Hatt M, Cheze Le Rest C, Albarghach N, Pradier O, Visvikis D. PET functional volume delineation: a robustness and repeatability study. Eur J Nucl Med Mol Imaging 2011;38:663–72.PubMedCrossRef
22.
go back to reference Hofheinz F, Dittrich S, Pötzsch C, Hoff J. Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold. Phys Med Biol 2010;55:1099–113.PubMedCrossRef Hofheinz F, Dittrich S, Pötzsch C, Hoff J. Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold. Phys Med Biol 2010;55:1099–113.PubMedCrossRef
Metadata
Title
Impact of [18F]FDG PET imaging parameters on automatic tumour delineation: need for improved tumour delineation methodology
Authors
Patsuree Cheebsumon
Maqsood Yaqub
Floris H. P. van Velden
Otto S. Hoekstra
Adriaan A. Lammertsma
Ronald Boellaard
Publication date
01-12-2011
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 12/2011
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-011-1899-5

Other articles of this Issue 12/2011

European Journal of Nuclear Medicine and Molecular Imaging 12/2011 Go to the issue

Editorial Commentary

Imaging of neuroinflammation