PET/MRI system design

Authors: Gaspar Delso, Sibylle Ziegler

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Special Issue 1/2009

Login to get access

Abstract

Introduction

The combination of clinical MRI and PET systems has received increased attention in recent years. In contrast to currently used PET/CT systems, PET/MRI offers not only improved soft-tissue contrast and reduced levels of ionizing radiation, but also a wealth of MRI-specific information such as functional, spectroscopic and diffusion tensor imaging. Combining PET and MRI, however, has proven to be very challenging, due to the detrimental cross-talk effects between the two systems.

Objective

Significant progress has been made in the recent years to overcome these difficulties, with several groups reporting PET/MRI prototypes for animal imaging and a clinical insert for neurological applications being demonstrated at the 2007 Annual Meeting of the Society of Nuclear Medicine.

Discussion

In this paper we review different architectures for clinical PET/MRI systems, and their possibilities, limitations and technological obstacles.
Literature
1.
go back to reference Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000;41:1369–79.PubMed Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000;41:1369–79.PubMed
2.
go back to reference European Commission. Referral guidelines for imaging radiation protection 118. Technical Report. ISBN: 92-828-9454-1; 2001. European Commission. Referral guidelines for imaging radiation protection 118. Technical Report. ISBN: 92-828-9454-1; 2001.
3.
go back to reference Zaidi H, Mawlawi O, Orton CG. Point/counterpoint. Simultaneous PET/MR will replace PET/CT as the molecular multimodality imaging platform of choice. Med Phys 2007;34:1525–8.PubMedCrossRef Zaidi H, Mawlawi O, Orton CG. Point/counterpoint. Simultaneous PET/MR will replace PET/CT as the molecular multimodality imaging platform of choice. Med Phys 2007;34:1525–8.PubMedCrossRef
4.
go back to reference Christensen NL, Hammer BE, Heil BG, Fetterly K. Positron emission tomography within a magnetic field using photomultiplier tubes and lightguides. Phys Med Biol 1995;40:691–7.PubMedCrossRef Christensen NL, Hammer BE, Heil BG, Fetterly K. Positron emission tomography within a magnetic field using photomultiplier tubes and lightguides. Phys Med Biol 1995;40:691–7.PubMedCrossRef
5.
go back to reference Cumalat JP, Cheung HWK, Hassed J, Smith BD, Bross AD. Effects of magnetic fields on the light yield of scintillators. Nucl Instrum Methods Phys Res Sect A 1990;293:606–14.CrossRef Cumalat JP, Cheung HWK, Hassed J, Smith BD, Bross AD. Effects of magnetic fields on the light yield of scintillators. Nucl Instrum Methods Phys Res Sect A 1990;293:606–14.CrossRef
6.
go back to reference Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR-compatible gamma detectors. IEEE Trans Nucl Sci 2003;50:1683–5.CrossRef Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR-compatible gamma detectors. IEEE Trans Nucl Sci 2003;50:1683–5.CrossRef
7.
go back to reference Lucas AJ, Hawkes RC, Ansorge RE, Williams GB, Nutt RE, Clark JC, et al. Development of a combined microPET-MR system. Technol Cancer Res Treat 2006;5:337–41.PubMed Lucas AJ, Hawkes RC, Ansorge RE, Williams GB, Nutt RE, Clark JC, et al. Development of a combined microPET-MR system. Technol Cancer Res Treat 2006;5:337–41.PubMed
8.
go back to reference Marsden PK, Strul D, Keevil SF, Williams SCR, Cash D. Simultaneous PET and NMR. Br J Radiol 2002;75:53–9. Marsden PK, Strul D, Keevil SF, Williams SCR, Cash D. Simultaneous PET and NMR. Br J Radiol 2002;75:53–9.
9.
go back to reference Raylman RR, Majewski S, Velan SS, Lemieux S, Kross B, Popov V, et al. Simultaneous acquisition of magnetic resonance spectroscopy (MRS) data and positron emission tomography (PET) images with a prototype MR-compatible, small animal PET imager. J Magn Reson 2007;186:305–10.PubMedCrossRef Raylman RR, Majewski S, Velan SS, Lemieux S, Kross B, Popov V, et al. Simultaneous acquisition of magnetic resonance spectroscopy (MRS) data and positron emission tomography (PET) images with a prototype MR-compatible, small animal PET imager. J Magn Reson 2007;186:305–10.PubMedCrossRef
10.
go back to reference Shao Y, Cherry S, Farahani K, Slates R, Silverman R, Meadors K, et al. Development of a PET detector system compatible with MRI/NMR systems. IEEE Trans Nucl Sci 1997;44:1167–71.CrossRef Shao Y, Cherry S, Farahani K, Slates R, Silverman R, Meadors K, et al. Development of a PET detector system compatible with MRI/NMR systems. IEEE Trans Nucl Sci 1997;44:1167–71.CrossRef
11.
go back to reference Pichler B, Lorenz E, Mirzoyan R, Pimpl W, Roder F, Schwaiger M, et al. Performance test of a LSO-APD PET module in a 9.4 Tesla magnet. Proc IEEE Nucl Sci Symp 1997;2:1237–9. Pichler B, Lorenz E, Mirzoyan R, Pimpl W, Roder F, Schwaiger M, et al. Performance test of a LSO-APD PET module in a 9.4 Tesla magnet. Proc IEEE Nucl Sci Symp 1997;2:1237–9.
12.
go back to reference Catana C, Wu Y, Judenhofer MS, Qi J, Pichler BJ, Cherry SR. Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 2006;47:1968–76.PubMed Catana C, Wu Y, Judenhofer MS, Qi J, Pichler BJ, Cherry SR. Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 2006;47:1968–76.PubMed
13.
go back to reference Catana C, Procissi D, Wu Y, Judenhofer MS, Qi J, Pichler BJ, et al. Simultaneous in vivo positron emission tomography and magnetic resonance imaging. Proc Natl Acad Sci U S A 2008;105:3705–10.PubMedCrossRef Catana C, Procissi D, Wu Y, Judenhofer MS, Qi J, Pichler BJ, et al. Simultaneous in vivo positron emission tomography and magnetic resonance imaging. Proc Natl Acad Sci U S A 2008;105:3705–10.PubMedCrossRef
14.
go back to reference Grazioso R, Zhanga N, Corbeila J, Schmanda M, Ladebeck R, Vester M, et al. APD-based PET detector for simultaneous PET/MR imaging. Nucl Instrum Methods Phys Res Sect A 2006;569:301–5.CrossRef Grazioso R, Zhanga N, Corbeila J, Schmanda M, Ladebeck R, Vester M, et al. APD-based PET detector for simultaneous PET/MR imaging. Nucl Instrum Methods Phys Res Sect A 2006;569:301–5.CrossRef
15.
go back to reference Pichler BJ, Judenhofer MS, Catana C, Walton JH, Kneilling M, Nutt RE, et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med 2006;47:639–47.PubMed Pichler BJ, Judenhofer MS, Catana C, Walton JH, Kneilling M, Nutt RE, et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med 2006;47:639–47.PubMed
16.
go back to reference Woody C, Schlyer D, Vaska P, Tomasi D, Solis-Najera S, Rooney W. Preliminary studies of a simultaneous PET/MRI scanner based on the RatCAP small animal tomograph. Nucl Instrum Methods Phys Res Sect A 2007;571:102–5.CrossRef Woody C, Schlyer D, Vaska P, Tomasi D, Solis-Najera S, Rooney W. Preliminary studies of a simultaneous PET/MRI scanner based on the RatCAP small animal tomograph. Nucl Instrum Methods Phys Res Sect A 2007;571:102–5.CrossRef
17.
go back to reference Junnarkar S, Fried J, O’Connor P, Radeka V, Vaska P, Purschke M. MRI compatible G-link and PCI based data acquisition hardware for the RatCAP scanner. IEEE Nucl Sci Symp 2006;1:380–3. Junnarkar S, Fried J, O’Connor P, Radeka V, Vaska P, Purschke M. MRI compatible G-link and PCI based data acquisition hardware for the RatCAP scanner. IEEE Nucl Sci Symp 2006;1:380–3.
18.
go back to reference Zang-Hee C, Young-Don S, Hang-Keun K, Kyoung-Nam K, Se-Hong O, Jae-Yong H, et al. A hybrid PET-MRI: an integrated molecular-genetic imaging system with HRRT-PET and 7.0-T MRI. Int J Imaging Sys Tech 2007;17:252–65.CrossRef Zang-Hee C, Young-Don S, Hang-Keun K, Kyoung-Nam K, Se-Hong O, Jae-Yong H, et al. A hybrid PET-MRI: an integrated molecular-genetic imaging system with HRRT-PET and 7.0-T MRI. Int J Imaging Sys Tech 2007;17:252–65.CrossRef
19.
go back to reference Shao Y, Cherry SR, Farahani K, Meadors K, Siegel S, Silverman RW, et al. Simultaneous PET and MR imaging. Phys Med Biol 1997;42:1965–70.PubMedCrossRef Shao Y, Cherry SR, Farahani K, Meadors K, Siegel S, Silverman RW, et al. Simultaneous PET and MR imaging. Phys Med Biol 1997;42:1965–70.PubMedCrossRef
20.
go back to reference Slates RB, Farahani K, Shao Y, Marsden PK, Taylor J, Summers PE. A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner. Phys Med Biol 1999;44:2015–27.PubMedCrossRef Slates RB, Farahani K, Shao Y, Marsden PK, Taylor J, Summers PE. A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner. Phys Med Biol 1999;44:2015–27.PubMedCrossRef
21.
go back to reference Pichler BJ, Swann BK, Rochelle J, Nutt RE, Cherry SR, Siegel SB. Lutetium oxyorthosilicate block detector readout by avalanche photodiode arrays for high resolution animal PET. Phys Med Biol 2004;49:4305–19.PubMedCrossRef Pichler BJ, Swann BK, Rochelle J, Nutt RE, Cherry SR, Siegel SB. Lutetium oxyorthosilicate block detector readout by avalanche photodiode arrays for high resolution animal PET. Phys Med Biol 2004;49:4305–19.PubMedCrossRef
22.
go back to reference Spanoudaki VC, McElroy DP, Torres-Espallardo I, Ziegler SI. Effect of temperature on the performance of proportional APD-based modules for gamma ray detection in positron emission tomography. IEEE Trans Nucl Sci 2008;55:469–80.CrossRef Spanoudaki VC, McElroy DP, Torres-Espallardo I, Ziegler SI. Effect of temperature on the performance of proportional APD-based modules for gamma ray detection in positron emission tomography. IEEE Trans Nucl Sci 2008;55:469–80.CrossRef
23.
go back to reference Carson RE, Herscovitch P, Daube-Witherspoon ME, editors. Quantitative functional brain imaging with positron emission tomography. 1st ed. New York: Academic Press; 1998. Carson RE, Herscovitch P, Daube-Witherspoon ME, editors. Quantitative functional brain imaging with positron emission tomography. 1st ed. New York: Academic Press; 1998.
24.
go back to reference Raylman RR, Majewski S, Lemieux SK, Velan SS, Kross B, Popov V. Simultaneous MRI and PET imaging of a rat brain. Phys Med Biol 2006;51:6371–9.PubMedCrossRef Raylman RR, Majewski S, Lemieux SK, Velan SS, Kross B, Popov V. Simultaneous MRI and PET imaging of a rat brain. Phys Med Biol 2006;51:6371–9.PubMedCrossRef
25.
go back to reference Schlyer D, Vaska P, Tomasi D, Woody C, Maramraju S-H, Southekal S. A simultaneous PET/MRI scanner based on RatCAP in small animals. IEEE Nucl Sci Symp 2007;5:3256–9. Schlyer D, Vaska P, Tomasi D, Woody C, Maramraju S-H, Southekal S. A simultaneous PET/MRI scanner based on RatCAP in small animals. IEEE Nucl Sci Symp 2007;5:3256–9.
26.
go back to reference Judenhofer MS, Wehrl HF, Newport DF, Catana C, Siegel SB, Becker M, et al. Simultaneous PET-MRI: a new approach for functional and morphological imaging. Nat Med 2008;14:459–65.PubMedCrossRef Judenhofer MS, Wehrl HF, Newport DF, Catana C, Siegel SB, Becker M, et al. Simultaneous PET-MRI: a new approach for functional and morphological imaging. Nat Med 2008;14:459–65.PubMedCrossRef
27.
go back to reference Schlemmer H, Pichler B, Wienhard K, Schmand M, Nahmias C, Townsend D, et al. Simultaneous MR/PET for brain imaging: first patient scans (abstract). J Nucl Med 2007;48(Suppl 2):45P. Schlemmer H, Pichler B, Wienhard K, Schmand M, Nahmias C, Townsend D, et al. Simultaneous MR/PET for brain imaging: first patient scans (abstract). J Nucl Med 2007;48(Suppl 2):45P.
28.
go back to reference Schlemmer H, Pichler BJ, Schmand M, Burbar Z, Michel C, Ladebeck R, et al. Simultaneous MR/PET imaging of the human brain: feasibility study. Radiology 2008;248:1028–35.PubMedCrossRef Schlemmer H, Pichler BJ, Schmand M, Burbar Z, Michel C, Ladebeck R, et al. Simultaneous MR/PET imaging of the human brain: feasibility study. Radiology 2008;248:1028–35.PubMedCrossRef
29.
go back to reference Handler W, Chronik B, Scholl T, Gilbert K, Peng H. Combining field-cycled magnetic resonance imaging with positron emission tomography (abstract). J Nucl Med Meeting Abstracts 2007;48(Suppl 2):89P. Handler W, Chronik B, Scholl T, Gilbert K, Peng H. Combining field-cycled magnetic resonance imaging with positron emission tomography (abstract). J Nucl Med Meeting Abstracts 2007;48(Suppl 2):89P.
30.
go back to reference Martinez M, Torres I, Ladebeck R, Nekolla S, Ziegler S, Schwaiger M. Whole-body MR-PET: characterization of PET performance by Monte Carlo simulations (abstract). J Nucl Med 2007;48(Suppl 2):46P. Martinez M, Torres I, Ladebeck R, Nekolla S, Ziegler S, Schwaiger M. Whole-body MR-PET: characterization of PET performance by Monte Carlo simulations (abstract). J Nucl Med 2007;48(Suppl 2):46P.
31.
go back to reference Otte A, Barral J, Dolgoshein B, Hose J, Klemin S, Lorenz E, et al. A test of silicon photomultipliers as readout for PET. Nucl Instrum Methods Phys Res Sect A 2005;545:705–15.CrossRef Otte A, Barral J, Dolgoshein B, Hose J, Klemin S, Lorenz E, et al. A test of silicon photomultipliers as readout for PET. Nucl Instrum Methods Phys Res Sect A 2005;545:705–15.CrossRef
32.
go back to reference Spanoudaki VC, Mann AB, Otte AN, Konorov I, Torres-Espallardo I, Paul S, et al. Use of single photon counting detector arrays in combined PET/MR: characterization of LYSO-SiPM detector modules and comparison with a LSO-APD detector. J Instrum 2007;2:P12002.CrossRef Spanoudaki VC, Mann AB, Otte AN, Konorov I, Torres-Espallardo I, Paul S, et al. Use of single photon counting detector arrays in combined PET/MR: characterization of LYSO-SiPM detector modules and comparison with a LSO-APD detector. J Instrum 2007;2:P12002.CrossRef
33.
go back to reference Strul D, Cash D, Keevil S, Halsted P, Williams S, Marsden P. Gamma shielding materials for MR-compatible PET. IEEE Trans Nucl Sci 2003;50:60–9.CrossRef Strul D, Cash D, Keevil S, Halsted P, Williams S, Marsden P. Gamma shielding materials for MR-compatible PET. IEEE Trans Nucl Sci 2003;50:60–9.CrossRef
34.
go back to reference Lee H, Li Z, Chen K, Hsu A, Xu C, Xie J, et al. PET/MRI dual-modality tumor imaging using arginine-glycine-aspartic (RGD)-conjugated radiolabeled iron oxide nanoparticles. J Nucl Med 2008;49:1371–9.PubMedCrossRef Lee H, Li Z, Chen K, Hsu A, Xu C, Xie J, et al. PET/MRI dual-modality tumor imaging using arginine-glycine-aspartic (RGD)-conjugated radiolabeled iron oxide nanoparticles. J Nucl Med 2008;49:1371–9.PubMedCrossRef
Metadata
Title
PET/MRI system design
Authors
Gaspar Delso
Sibylle Ziegler
Publication date
01-03-2009
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue Special Issue 1/2009
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-008-1008-6

Other articles of this Special Issue 1/2009

European Journal of Nuclear Medicine and Molecular Imaging 1/2009 Go to the issue