CT-based attenuation correction in 82Rb-myocardial perfusion PET–CT: incidence of misalignment and effect on regional tracer distribution

Authors: Riikka Lautamäki, Tracy L. Y. Brown, Jennifer Merrill, Frank M. Bengel

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 2/2008

Login to get access

Abstract

Purpose

Misalignment of low-dose-CT used for attenuation correction (AC) may cause artifacts in cardiac-PET–CT. The aim was to evaluate incidence and severity of misalignment and its quantitative effects on regional myocardial 82Rb-distribution.

Methods

Rest/dipyridamole 82Rb-perfusion-PET–CT studies of 92 consecutive patients were analyzed for misalignment. Two different scanning protocols were employed: the first 57 patients had separate CTs for rest and stress PET. The following 35 patients had one CT at rest, used for AC of rest and stress PET. Misalignment was visually scored on a five-point scale (0 = no, 1 = minimal, 2 = mild, 3 = moderate, and 4 = severe). In five representative patients with normal perfusion and low probability of disease, 95 polarmaps were created by shifting CT vs PET prior to reconstruction of attenuation-corrected data sets using dedicated software (three dimensions of space; magnitude of shifts, 5, 10, 14 mm).

Results

PET/CT -misalignment was detected in 60% of rest and 67% of stress studies. Alignment for rest was better than that for stress (0.7 ± 0.7 vs 1.0 ± 0.9, P = 0.03). Comparison of the two protocols revealed no effect on the alignment of the stress study (1.0 ± 0.9 vs 1.0 ± 0.9, P = 0.9). Quantitatively, the largest individual effect of any artificial misalignment was a 25% reduction of relative 82Rb uptake. With a shift of 1 cm, the largest effect in an individual was a 19% decrease. Anterior wall was most frequently influenced by misalignment, but changes of uptake also occurred in all other segments.

Conclusions

Misalignment between CT and PET in cardiac-PET–CT influences regional tracer distribution in multiple segments. Repeated CT imaging after dipyridamole does not improve alignment. These results emphasize the need for strategies to improve coregistration in clinical imaging protocols.
Literature
1.
go back to reference Yoshinaga K, Chow BJ, Williams K, Chen L, deKemp RA, Garrard L, et al. What is the prognostic value of myocardial perfusion imaging using rubidium-82 positron emission tomography? J Am Coll Cardiol 2006;48 5:1029–39.PubMedCrossRef Yoshinaga K, Chow BJ, Williams K, Chen L, deKemp RA, Garrard L, et al. What is the prognostic value of myocardial perfusion imaging using rubidium-82 positron emission tomography? J Am Coll Cardiol 2006;48 5:1029–39.PubMedCrossRef
2.
go back to reference Bateman TM, Heller GV, McGhie AI, Friedman JD, Case JA, Bryngelson JR, et al. Diagnostic accuracy of rest/stress ECG-gated Rb-82 myocardial perfusion PET: comparison with ECG-gated Tc-99m sestamibi SPECT. J Nucl Cardiol 2006;13 1:24–33.PubMedCrossRef Bateman TM, Heller GV, McGhie AI, Friedman JD, Case JA, Bryngelson JR, et al. Diagnostic accuracy of rest/stress ECG-gated Rb-82 myocardial perfusion PET: comparison with ECG-gated Tc-99m sestamibi SPECT. J Nucl Cardiol 2006;13 1:24–33.PubMedCrossRef
3.
go back to reference Tarakji KG, Brunken R, McCarthy PM, Al Chekakie MO, Abdel-Latif A, Pothier CE, et al. Myocardial viability testing and the effect of early intervention in patients with advanced left ventricular systolic dysfunction. Circulation 2006;113 2:230–7.PubMedCrossRef Tarakji KG, Brunken R, McCarthy PM, Al Chekakie MO, Abdel-Latif A, Pothier CE, et al. Myocardial viability testing and the effect of early intervention in patients with advanced left ventricular systolic dysfunction. Circulation 2006;113 2:230–7.PubMedCrossRef
4.
go back to reference Sampson UK, Dorbala S, Limaye A, Kwong R, Di Carli MF. Diagnostic accuracy of rubidium-82 myocardial perfusion imaging with hybrid positron emission tomography/computed tomography in the detection of coronary artery disease. J Am Coll Cardiol 2007;49 10:1052–8.PubMedCrossRef Sampson UK, Dorbala S, Limaye A, Kwong R, Di Carli MF. Diagnostic accuracy of rubidium-82 myocardial perfusion imaging with hybrid positron emission tomography/computed tomography in the detection of coronary artery disease. J Am Coll Cardiol 2007;49 10:1052–8.PubMedCrossRef
5.
go back to reference Dorbala S, Hassan A, Heinonen T, Schelbert HR, Di Carli MF. Coronary vasodilator reserve and Framingham risk scores in subjects at risk for coronary artery disease. J Nucl Cardiol 2006;13 6:761–7.PubMedCrossRef Dorbala S, Hassan A, Heinonen T, Schelbert HR, Di Carli MF. Coronary vasodilator reserve and Framingham risk scores in subjects at risk for coronary artery disease. J Nucl Cardiol 2006;13 6:761–7.PubMedCrossRef
6.
go back to reference Chow BJ, Wong JW, Yoshinaga K, Ruddy TD, Williams K, deKemp RA, et al. Prognostic significance of dipyridamole-induced ST depression in patients with normal 82Rb PET myocardial perfusion imaging. J Nucl Med 2005;46 7:1095–101.PubMed Chow BJ, Wong JW, Yoshinaga K, Ruddy TD, Williams K, deKemp RA, et al. Prognostic significance of dipyridamole-induced ST depression in patients with normal 82Rb PET myocardial perfusion imaging. J Nucl Med 2005;46 7:1095–101.PubMed
7.
go back to reference McCord ME, Bacharach SL, Bonow RO, Dilsizian V, Cuocolo A, Freedman N. Misalignment between PET transmission and emission scans: its effect on myocardial imaging. J Nucl Med 1992;33 6:1209–14.PubMed McCord ME, Bacharach SL, Bonow RO, Dilsizian V, Cuocolo A, Freedman N. Misalignment between PET transmission and emission scans: its effect on myocardial imaging. J Nucl Med 1992;33 6:1209–14.PubMed
8.
go back to reference Loghin C, Sdringola S, Gould KL. Common artifacts in PET myocardial perfusion images due to attenuation-emission misregistration: clinical significance, causes, and solutions. J Nucl Med 2004;45 6:1029–39.PubMed Loghin C, Sdringola S, Gould KL. Common artifacts in PET myocardial perfusion images due to attenuation-emission misregistration: clinical significance, causes, and solutions. J Nucl Med 2004;45 6:1029–39.PubMed
9.
go back to reference Koepfli P, Hany TF, Wyss CA, Namdar M, Burger C, Konstantinidis AV, et al. CT attenuation correction for myocardial perfusion quantification using a PET/CT hybrid scanner. J Nucl Med 2004;45 4:537–42.PubMed Koepfli P, Hany TF, Wyss CA, Namdar M, Burger C, Konstantinidis AV, et al. CT attenuation correction for myocardial perfusion quantification using a PET/CT hybrid scanner. J Nucl Med 2004;45 4:537–42.PubMed
10.
go back to reference Martinez-Moller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. J Nucl Med 2007;48 2:188–93.PubMed Martinez-Moller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. J Nucl Med 2007;48 2:188–93.PubMed
11.
go back to reference Osman MM, Cohade C, Nakamoto Y, Marshall LT, Leal JP, Wahl RL. Clinically significant inaccurate localization of lesions with PET/CT: frequency in 300 patients. J Nucl Med 2003;44 2:240–3.PubMed Osman MM, Cohade C, Nakamoto Y, Marshall LT, Leal JP, Wahl RL. Clinically significant inaccurate localization of lesions with PET/CT: frequency in 300 patients. J Nucl Med 2003;44 2:240–3.PubMed
12.
go back to reference Osman MM, Cohade C, Nakamoto Y, Wahl RL. Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET–CT. Eur J Nucl Med Mol Imaging 2003;30 4:603–6.PubMedCrossRef Osman MM, Cohade C, Nakamoto Y, Wahl RL. Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET–CT. Eur J Nucl Med Mol Imaging 2003;30 4:603–6.PubMedCrossRef
13.
go back to reference Chin BB, Nakamoto Y, Kraitchman DL, Marshall L, Wahl R. PET–CT evaluation of 2-deoxy-2-[18F]fluoro-D-glucose myocardial uptake: effect of respiratory motion. Mol Imaging Biol 2003;5 2:57–64.PubMedCrossRef Chin BB, Nakamoto Y, Kraitchman DL, Marshall L, Wahl R. PET–CT evaluation of 2-deoxy-2-[18F]fluoro-D-glucose myocardial uptake: effect of respiratory motion. Mol Imaging Biol 2003;5 2:57–64.PubMedCrossRef
14.
go back to reference Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. J Nucl Med 2007;48 7:1112–21.PubMedCrossRef Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. J Nucl Med 2007;48 7:1112–21.PubMedCrossRef
15.
go back to reference Goetze S, Wahl RL. Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT. J Nucl Cardiol 2007;14 2:200–6.PubMedCrossRef Goetze S, Wahl RL. Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT. J Nucl Cardiol 2007;14 2:200–6.PubMedCrossRef
16.
go back to reference Matsunari I, Boning G, Ziegler SI, Kosa I, Nekolla SG, Ficaro EP, et al. Effects of misalignment between transmission and emission scans on attenuation-corrected cardiac SPECT. J Nucl Med 1998;39 3:411–6.PubMed Matsunari I, Boning G, Ziegler SI, Kosa I, Nekolla SG, Ficaro EP, et al. Effects of misalignment between transmission and emission scans on attenuation-corrected cardiac SPECT. J Nucl Med 1998;39 3:411–6.PubMed
17.
go back to reference Beyer T, Antoch G, Blodgett T, Freudenberg LF, Akhurst T, Mueller S. Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology. Eur J Nucl Med Mol Imaging 2003;30 4:588–96.PubMedCrossRef Beyer T, Antoch G, Blodgett T, Freudenberg LF, Akhurst T, Mueller S. Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology. Eur J Nucl Med Mol Imaging 2003;30 4:588–96.PubMedCrossRef
18.
go back to reference Goerres GW, Kamel E, Heidelberg TN, Schwitter MR, Burger C, von Schulthess GK. PET–CT image co-registration in the thorax: influence of respiration. Eur J Nucl Med Mol Imaging 2002;29 3:351–60.PubMedCrossRef Goerres GW, Kamel E, Heidelberg TN, Schwitter MR, Burger C, von Schulthess GK. PET–CT image co-registration in the thorax: influence of respiration. Eur J Nucl Med Mol Imaging 2002;29 3:351–60.PubMedCrossRef
19.
go back to reference Le Meunier L, Maass-Moreno R, Carrasquillo JA, Dieckmann W, Bacharach SL. PET/CT imaging: effect of respiratory motion on apparent myocardial uptake. J Nucl Cardiol 2006;13 6:821–30.PubMedCrossRef Le Meunier L, Maass-Moreno R, Carrasquillo JA, Dieckmann W, Bacharach SL. PET/CT imaging: effect of respiratory motion on apparent myocardial uptake. J Nucl Cardiol 2006;13 6:821–30.PubMedCrossRef
20.
go back to reference Nehmeh SA, Erdi YE, Pan T, Pevsner A, Rosenzweig KE, Yorke E, et al. Four-dimensional (4D) PET/CT imaging of the thorax. Med Phys 2004;31 12:3179–86.PubMedCrossRef Nehmeh SA, Erdi YE, Pan T, Pevsner A, Rosenzweig KE, Yorke E, et al. Four-dimensional (4D) PET/CT imaging of the thorax. Med Phys 2004;31 12:3179–86.PubMedCrossRef
21.
go back to reference Pan T, Mawlawi O, Nehmeh SA, Erdi YE, Luo D, Liu HH, et al. Attenuation correction of PET images with respiration-averaged CT images in PET/CT. J Nucl Med 2005;46 9:1481–7.PubMed Pan T, Mawlawi O, Nehmeh SA, Erdi YE, Luo D, Liu HH, et al. Attenuation correction of PET images with respiration-averaged CT images in PET/CT. J Nucl Med 2005;46 9:1481–7.PubMed
22.
go back to reference Goetze S, Brown TL, Lavely WC, Zhang Z, Bengel FM. Attenuation correction in myocardial perfusion SPECT/CT: effects of misregistration and value of reregistration. J Nucl Med 2007;48 7:1090–5.PubMedCrossRef Goetze S, Brown TL, Lavely WC, Zhang Z, Bengel FM. Attenuation correction in myocardial perfusion SPECT/CT: effects of misregistration and value of reregistration. J Nucl Med 2007;48 7:1090–5.PubMedCrossRef
Metadata
Title
CT-based attenuation correction in 82Rb-myocardial perfusion PET–CT: incidence of misalignment and effect on regional tracer distribution
Authors
Riikka Lautamäki
Tracy L. Y. Brown
Jennifer Merrill
Frank M. Bengel
Publication date
01-02-2008
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 2/2008
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-007-0607-y

Other articles of this Issue 2/2008

European Journal of Nuclear Medicine and Molecular Imaging 2/2008 Go to the issue

Society communications

Society communications