Variability of left ventricular ejection fraction and volumes with quantitative gated SPECT: influence of algorithm, pixel size and reconstruction parameters in small and normal-sized hearts

Authors: Anne-Sophie Hambye, Ann Vervaet, André Dobbeleir

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 12/2004

Login to get access

Abstract

Purpose

Several software packages are commercially available for quantification of left ventricular ejection fraction (LVEF) and volumes from myocardial gated single-photon emission computed tomography (SPECT), all of which display a high reproducibility. However, their accuracy has been questioned in patients with a small heart. This study aimed to evaluate the performances of different software and the influence of modifications in acquisition or reconstruction parameters on LVEF and volume measurements, depending on the heart size.

Methods

In 31 patients referred for gated SPECT, 642 and 1282 matrix acquisitions were consecutively obtained. After reconstruction by filtered back-projection (Butterworth, 0.4, 0.5 or 0.6 cycles/cm cut-off, order 6), LVEF and volumes were computed with different software [three versions of Quantitative Gated SPECT (QGS), the Emory Cardiac Toolbox (ECT) and the Stanford University (SU-Segami) Medical School algorithm] and processing workstations. Depending upon their end-systolic volume (ESV), patients were classified into two groups: group I (ESV>30 ml, n=14) and group II (ESV<30 ml, n=17). Agreement between the different software packages and the influence of matrix size and sharpness of the filter on LVEF and volumes were evaluated in both groups.

Results

In group I, the correlation coefficients between the different methods ranged from 0.82 to 0.94 except for SU-Segami (r=0.77), and were slightly lower for volumes than for LVEF. Mean differences between the methods were not significant, except for ECT, with which LVEF values were systematically higher by more than 10%. Changes in matrix size had no significant influence on LVEF or volumes. On the other hand, a sharper filter was associated with significantly larger volume values though this did not usually result in significant changes in LVEF. In group II, many patients had an LVEF in the higher range. The correlation coefficients between the different methods ranged between 0.80 and 0.96 except for SU-Segami (r=0.49), and were slightly worse for volumes than for LVEF values. In contrast to group I, however, inter-method variability was quite large and most mean LVEF differences were significant. LVEF was systematically highest with ECT and lowest with SU-Segami. With QGS, changes in matrix size from 642 to 1282 were associated with significantly larger volumes as well as lower LVEF values. Increasing the filter cut-off frequency had the same effect. With SU-Segami, a larger matrix was associated with larger end-diastolic volumes and smaller ESVs, resulting in a highly significant increase in LVEF. Increasing the filter sharpness, on the other hand, had no influence on LVEF though the measured volumes were significantly larger.

Conclusion

In patients with a normal-sized heart, LVEF and volume estimates computed from different commercially available software packages for quantitative gated SPECT are well correlated. LVEF and volumes are only slightly sensitive to changes in matrix size. Smoothing, by contrast, is associated with significant changes in volumes but usually not in LVEF values. However, owing to the specific characteristics of each algorithm, software should not be interchanged for follow-up in an individual patient. In small hearts, on the other hand, both the used software and the matrix size or smoothing significantly influence the results of quantitative gated SPECT. LVEF values in the higher range are frequently observed with all the studied software except for SU-Segami. A larger matrix or a sharper filter could be suggested to enhance the accuracy of most commercial software, more particularly in patients with a small heart.
Literature
1.
go back to reference Germano G, Kiat H, Kavanagh PB, Moriel M, Mazzanti M, Su H, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995;36:2138–47.PubMed Germano G, Kiat H, Kavanagh PB, Moriel M, Mazzanti M, Su H, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995;36:2138–47.PubMed
2.
go back to reference Germano G, Kavanagh PB, Kavanagh JT, Wishner SH, Berman DS, Kavanagh GJ. Repeatability of automatic left ventricular cavity volume measurements from myocardial perfusion SPECT. J Nucl Cardiol 1998;5:477–83.CrossRefPubMed Germano G, Kavanagh PB, Kavanagh JT, Wishner SH, Berman DS, Kavanagh GJ. Repeatability of automatic left ventricular cavity volume measurements from myocardial perfusion SPECT. J Nucl Cardiol 1998;5:477–83.CrossRefPubMed
3.
go back to reference Goris ML, Thompson C, Malone L, Franken PR. Modeling the integration of myocardial regional perfusion and function. Nucl Med Commun 1994;15:9–20.PubMed Goris ML, Thompson C, Malone L, Franken PR. Modeling the integration of myocardial regional perfusion and function. Nucl Med Commun 1994;15:9–20.PubMed
4.
go back to reference Faber TL, Akers MS, Peshock RM, Corbett JR. Three dimensional motion and perfusion quantification in gated single-photon emission computed tomograms. J Nucl Med 1991;32:2311–7.PubMed Faber TL, Akers MS, Peshock RM, Corbett JR. Three dimensional motion and perfusion quantification in gated single-photon emission computed tomograms. J Nucl Med 1991;32:2311–7.PubMed
5.
go back to reference Faber Tl, Cooke CD, Folks RD, Vansant JP, Nichols KJ, DePuey EG, et al. Left ventricular function and perfusion from gated perfusion images: an integrated method. J Nucl Med 1999;40:650–9PubMed Faber Tl, Cooke CD, Folks RD, Vansant JP, Nichols KJ, DePuey EG, et al. Left ventricular function and perfusion from gated perfusion images: an integrated method. J Nucl Med 1999;40:650–9PubMed
6.
go back to reference Nichols K, DePuey RG, Rozanski A. Automation of gated tomography left ventricular ejection fraction. J Nucl Cardiol 1996;3:475–82CrossRefPubMed Nichols K, DePuey RG, Rozanski A. Automation of gated tomography left ventricular ejection fraction. J Nucl Cardiol 1996;3:475–82CrossRefPubMed
7.
go back to reference Everaert H, Bossuyt A, Franken P. Left ventricular ejection fraction and volumes from gated single photon emission tomographic myocardial perfusion images: comparison between two algorithms working in three-dimensional space. J Nucl Cardiol 1997;4:472–6CrossRefPubMed Everaert H, Bossuyt A, Franken P. Left ventricular ejection fraction and volumes from gated single photon emission tomographic myocardial perfusion images: comparison between two algorithms working in three-dimensional space. J Nucl Cardiol 1997;4:472–6CrossRefPubMed
8.
go back to reference Nichols K, Lefkowitz D, Faber T, Folks R, Cooke D, Garcia EV, et al. Echocardiographic validation of gated SPECT ventricular function measurements. J Nucl Med 2000;41:1308–14PubMed Nichols K, Lefkowitz D, Faber T, Folks R, Cooke D, Garcia EV, et al. Echocardiographic validation of gated SPECT ventricular function measurements. J Nucl Med 2000;41:1308–14PubMed
9.
go back to reference Nakajima J, Higuchi T, Taki J, Kawano M, Tonami N. Accuracy of ventricular volume and ejection fraction measured by gated myocardial SPECT: comparison of 4 software programs. J Nucl Med 2001;42:1571–8PubMed Nakajima J, Higuchi T, Taki J, Kawano M, Tonami N. Accuracy of ventricular volume and ejection fraction measured by gated myocardial SPECT: comparison of 4 software programs. J Nucl Med 2001;42:1571–8PubMed
10.
go back to reference Vourvouri E, Poldermans D, Bax JJ, Sianos G, Sozzi FB, Schinkel AF, et al. Evaluation of left ventricular function and volumes in patients with ischaemic cardiomyopathy: gated single-photon emission computed tomography versus two-dimensional echocardiography. Eur J Nucl Med 2001;28:1610–5CrossRefPubMed Vourvouri E, Poldermans D, Bax JJ, Sianos G, Sozzi FB, Schinkel AF, et al. Evaluation of left ventricular function and volumes in patients with ischaemic cardiomyopathy: gated single-photon emission computed tomography versus two-dimensional echocardiography. Eur J Nucl Med 2001;28:1610–5CrossRefPubMed
11.
go back to reference Lum DP, Coel MN. Comparison of automatic quantification software for the measurement of ventricular volume and ejection fraction in gated myocardial perfusion SPECT. Nucl Med Commun 2003;24:259–66CrossRefPubMed Lum DP, Coel MN. Comparison of automatic quantification software for the measurement of ventricular volume and ejection fraction in gated myocardial perfusion SPECT. Nucl Med Commun 2003;24:259–66CrossRefPubMed
12.
go back to reference Vallejo E, Dione DP, Bruni WL, Constable RT, Borek PP, Soares JP, et al. Reproducibility and accuracy of gated SPECT for determination of left ventricular volume and ejection fraction: experimental validation using MRI. J Nucl Med 2000;41:874–82PubMed Vallejo E, Dione DP, Bruni WL, Constable RT, Borek PP, Soares JP, et al. Reproducibility and accuracy of gated SPECT for determination of left ventricular volume and ejection fraction: experimental validation using MRI. J Nucl Med 2000;41:874–82PubMed
13.
go back to reference Nakajima K, Taki J, Higuchi T, Kawano M, Taniguchi M, Maruhashi K, et al. Gated SPET quantification of small hearts: mathematical simulation and clinical application. Eur J Nucl Med 2000;27:1372–9CrossRefPubMed Nakajima K, Taki J, Higuchi T, Kawano M, Taniguchi M, Maruhashi K, et al. Gated SPET quantification of small hearts: mathematical simulation and clinical application. Eur J Nucl Med 2000;27:1372–9CrossRefPubMed
14.
go back to reference Ford P, Chatziioannou S, Moore H, Dhekne R. Overestimation of the LVEF by quantitative gated SPECT in simulated left ventricles. J Nucl Med 2001;42:454–9PubMed Ford P, Chatziioannou S, Moore H, Dhekne R. Overestimation of the LVEF by quantitative gated SPECT in simulated left ventricles. J Nucl Med 2001;42:454–9PubMed
15.
go back to reference Manrique A, Hitzel A, Gardin I, Dacher JN, Vera P. Influence of Wiener filter in determining the left ventricle volume and ejection fraction using thallium-201 gated SPECT. Nucl Med Commun 2003;24:907–14CrossRefPubMed Manrique A, Hitzel A, Gardin I, Dacher JN, Vera P. Influence of Wiener filter in determining the left ventricle volume and ejection fraction using thallium-201 gated SPECT. Nucl Med Commun 2003;24:907–14CrossRefPubMed
16.
go back to reference De Bondt P, Van de Wiele C, De Sutter J, De Winter F, De Backer G, Dierckx RA. Age- and gender-specific differences in left ventricular cardiac function and volumes determined by gated SPET. Eur J Nucl Med 2001;28:620–4CrossRefPubMed De Bondt P, Van de Wiele C, De Sutter J, De Winter F, De Backer G, Dierckx RA. Age- and gender-specific differences in left ventricular cardiac function and volumes determined by gated SPET. Eur J Nucl Med 2001;28:620–4CrossRefPubMed
17.
go back to reference Achtert AD, King MA, Dahlberg ST, Pretorius PH, LaCroix KJ, Tsui BM. An investigation of the estimation of ejection fractions and cardiac volumes by a quantitative gated SPECT software package in simulated gated SPECT images. J Nucl Cardiol 1998;5:144–52CrossRefPubMed Achtert AD, King MA, Dahlberg ST, Pretorius PH, LaCroix KJ, Tsui BM. An investigation of the estimation of ejection fractions and cardiac volumes by a quantitative gated SPECT software package in simulated gated SPECT images. J Nucl Cardiol 1998;5:144–52CrossRefPubMed
18.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurements. Lancet 1986;1:307–10PubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurements. Lancet 1986;1:307–10PubMed
19.
go back to reference Nichols K, Santana CA, Folks R, Krawczynska E, Cooke CD, Faber TL, et al. Comparison between ECT and QGS for assessment of left ventricular function from gated myocardial perfusion SPECT. J Nucl Cardiol 2002;9:285–93CrossRefPubMed Nichols K, Santana CA, Folks R, Krawczynska E, Cooke CD, Faber TL, et al. Comparison between ECT and QGS for assessment of left ventricular function from gated myocardial perfusion SPECT. J Nucl Cardiol 2002;9:285–93CrossRefPubMed
Metadata
Title
Variability of left ventricular ejection fraction and volumes with quantitative gated SPECT: influence of algorithm, pixel size and reconstruction parameters in small and normal-sized hearts
Authors
Anne-Sophie Hambye
Ann Vervaet
André Dobbeleir
Publication date
01-12-2004
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 12/2004
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-004-1601-2

Other articles of this Issue 12/2004

European Journal of Nuclear Medicine and Molecular Imaging 12/2004 Go to the issue