Published in: Pediatric Radiology 1/2020

01-01-2020 | Pneumothorax | Original Article

Modification of chest radiography exposure parameters using a neonatal chest phantom

Authors: Stefan B. Schäfer, Sabine Papst, Martin Fiebich, Claudia Rudolph, Jan de Laffolie, Gabriele A. Krombach

Published in: Pediatric Radiology | Issue 1/2020

Login to get access

Abstract

Background

The acquisition of chest radiographs in neonates is of critical importance in diagnostics because of the risk of respiratory distress syndrome and pneumothorax in preterm infants.

Objective

To achieve a dose reduction while preserving a diagnostic image quality for chest radiographs of neonates.

Materials and methods

All radiographs, generated on a fully digital X-ray unit by using a neonatal chest phantom, were evaluated under variation of the tube voltage (40–70 kV) and mAs levels (1–10.2 mAs) with and without an additional 0.1-mm copper (Cu) filtration. Noise, contrast and contrast-to-noise ratio for bronchus, heart, lungs and vessels were determined. Visual assessment of the image quality was carried out by three radiologists using a Likert scale. To evaluate a maximally possible dose reduction, the dose of the radiographs with still acceptable image quality at a minimal dose was compared to the dose of the radiographs with the standard settings used in clinical routine.

Results

The noise showed decreasing values with increasing dose, while the contrast values were increased. For the contrast-to-noise ratio, a digressive course of the values as a function of the tube voltage was found. The visual evaluation of image quality showed the best evaluation of the structures at the lowest possible dose in the settings (44 kV, 3.36 mAs) with copper filtration and in the settings (44 kV, 1.56 mAs) without copper filtration. A maximum dose reduction from 8.29 μSv to 2.21 μSv (about 73%) was obtained.

Conclusion

A dose reduction while preserving diagnostic image quality in a digital X-ray system is generally possible by reducing the tube voltage and simultaneous adaptation of the mAs settings.
Literature
1.
go back to reference Speer CP (1999) lung diseases of the neonatal. in Harnack GA, Koletzko B (ed) Kinderheilkunde (Pediatrics). 11th edn. Springer Verlag, Berlin, pp 78-86 Speer CP (1999) lung diseases of the neonatal. in Harnack GA, Koletzko B (ed) Kinderheilkunde (Pediatrics). 11th edn. Springer Verlag, Berlin, pp 78-86
2.
go back to reference Hahn G, Glutig K (2009) medical imaging. in: Pädiatrie. 3rd edn. Springer, Berlin, Heidelberg, pp 1025-1041CrossRef Hahn G, Glutig K (2009) medical imaging. in: Pädiatrie. 3rd edn. Springer, Berlin, Heidelberg, pp 1025-1041CrossRef
4.
go back to reference Hall EJ (2002) Lessons we have learned from our children: cancer risks from diagnostic radiology. Pediatr Radiol 32:700–706CrossRef Hall EJ (2002) Lessons we have learned from our children: cancer risks from diagnostic radiology. Pediatr Radiol 32:700–706CrossRef
5.
go back to reference Spahn M, Heer V, Freytag R (2003) Flat-panel detectors in X-ray systems. Radiologe 43:340–350CrossRef Spahn M, Heer V, Freytag R (2003) Flat-panel detectors in X-ray systems. Radiologe 43:340–350CrossRef
6.
go back to reference Hamers S, Freyschmidt J, Neitzel U (2001) Digital radiography with large-scale electronic flat-panel detector vs screen-film radiography: observer preference in clinical skeletal diagnostics. Eur Radiol 11:1753–1759CrossRef Hamers S, Freyschmidt J, Neitzel U (2001) Digital radiography with large-scale electronic flat-panel detector vs screen-film radiography: observer preference in clinical skeletal diagnostics. Eur Radiol 11:1753–1759CrossRef
7.
go back to reference Busch HP (1999) Digial projection radiography. Technical bases. Imaging quality and usefulness. Radiologe 39:710–724CrossRef Busch HP (1999) Digial projection radiography. Technical bases. Imaging quality and usefulness. Radiologe 39:710–724CrossRef
9.
go back to reference Becht S, Bittner R, Ohmstede A et al (2008) Zimmer-Brossy M (ed) Textbook of the X-ray diagnostic adjustment technique, 6th edn. Springer-Verlag, Berlin Becht S, Bittner R, Ohmstede A et al (2008) Zimmer-Brossy M (ed) Textbook of the X-ray diagnostic adjustment technique, 6th edn. Springer-Verlag, Berlin
10.
go back to reference Bach M, Wesemann W, Kolling et al (2008) Photopic contrst sensitivity. Local contrast perception. Ophthalmologe 105:46–59CrossRef Bach M, Wesemann W, Kolling et al (2008) Photopic contrst sensitivity. Local contrast perception. Ophthalmologe 105:46–59CrossRef
11.
go back to reference DIN Deutsches Institut für Normung e.V (1997) terminology of physiological optics. (DIN 5340). Beuth, Berlin DIN Deutsches Institut für Normung e.V (1997) terminology of physiological optics. (DIN 5340). Beuth, Berlin
12.
go back to reference Hess R, Neitzel U (2012) Optimizing image qualitiy and dose for digital radiography of distal pediatric extremeities using the contrast-to-noise ratio. Rofo 184:643–649CrossRef Hess R, Neitzel U (2012) Optimizing image qualitiy and dose for digital radiography of distal pediatric extremeities using the contrast-to-noise ratio. Rofo 184:643–649CrossRef
13.
go back to reference Mohrs OK, Pettersen SE, Schulze T (2010) High-resolution 3D unenhanced ECG-gated respiratory-navigated MR angiography of the renal arteries: comparison with contrast-enhanced MR angiography. AJR Am J Roentgenol 195:1423–1428CrossRef Mohrs OK, Pettersen SE, Schulze T (2010) High-resolution 3D unenhanced ECG-gated respiratory-navigated MR angiography of the renal arteries: comparison with contrast-enhanced MR angiography. AJR Am J Roentgenol 195:1423–1428CrossRef
14.
go back to reference Körner M, Wirth S, Treitl M et al (2005) Initial clinical results with a new needle screen storage phosphor system in chest radiograms. Rofo 177:1491–1496CrossRef Körner M, Wirth S, Treitl M et al (2005) Initial clinical results with a new needle screen storage phosphor system in chest radiograms. Rofo 177:1491–1496CrossRef
15.
go back to reference Bulla S, Pache G, Blanke P et al (2011) Reduction of radiation dose in low-dose CT of the paranasal sinuses using iterative image reconstruction: feasibility and image quality. Rofo 183:VO409_1 Bulla S, Pache G, Blanke P et al (2011) Reduction of radiation dose in low-dose CT of the paranasal sinuses using iterative image reconstruction: feasibility and image quality. Rofo 183:VO409_1
16.
go back to reference Gosch G, Gosch K, Kahn T (2007) Conversion coefficients for estimation of effective dose to patients from dose area product during fluoroscopy x-ray examinations. Rofo 179:1035–1042CrossRef Gosch G, Gosch K, Kahn T (2007) Conversion coefficients for estimation of effective dose to patients from dose area product during fluoroscopy x-ray examinations. Rofo 179:1035–1042CrossRef
17.
go back to reference Gislason AJ, Davies AG, Cowen AR (2010) Dose optimization in pediatric cardiac x-ray imaging. Med Phys 37:5258–5269CrossRef Gislason AJ, Davies AG, Cowen AR (2010) Dose optimization in pediatric cardiac x-ray imaging. Med Phys 37:5258–5269CrossRef
18.
go back to reference Hansson B, Finnbogason T, Schuwert P, Persliden J (1997) Added copper filtration in digital paediatric double-contrast colon examinations: effects on radiation dose and image quality. Eur Radiol:1117–1122CrossRef Hansson B, Finnbogason T, Schuwert P, Persliden J (1997) Added copper filtration in digital paediatric double-contrast colon examinations: effects on radiation dose and image quality. Eur Radiol:1117–1122CrossRef
19.
go back to reference Redlich U, Hoeschen C, Doehring W (2005) Assessment and optimisation of the image quality of chest-radiography systems. Radiat Prot Dosimetry 114:264–268CrossRef Redlich U, Hoeschen C, Doehring W (2005) Assessment and optimisation of the image quality of chest-radiography systems. Radiat Prot Dosimetry 114:264–268CrossRef
20.
go back to reference Smet MH, Breysem L, Mussen E et al (2018) Visual grading analysis of digital neonatal chest phantom X-ray images: impact of detector type, dose and image processing on image quality. Eur Radiol 28:2951–2959CrossRef Smet MH, Breysem L, Mussen E et al (2018) Visual grading analysis of digital neonatal chest phantom X-ray images: impact of detector type, dose and image processing on image quality. Eur Radiol 28:2951–2959CrossRef
21.
go back to reference Rattan AS, Cohen MD (2013) Removal of comfort pads underneath babies. a method of reducing Radiation exposure to neonates Acad Radiol 20:1297–1300PubMed Rattan AS, Cohen MD (2013) Removal of comfort pads underneath babies. a method of reducing Radiation exposure to neonates Acad Radiol 20:1297–1300PubMed
22.
go back to reference Stenzel M, Mentzel HJ (2011) Digitale Thorax-Radiographie in der Neonatologie - Dose optimization using a neonatal phantom. Monatsschr Kinderheilkd 159(Suppl 3):230–240 Stenzel M, Mentzel HJ (2011) Digitale Thorax-Radiographie in der Neonatologie - Dose optimization using a neonatal phantom. Monatsschr Kinderheilkd 159(Suppl 3):230–240
23.
go back to reference Smans K, Struelens L, Smet M et al (2010) Cu filtration for dose reduction in neonatal chest imaging. Radiat Prot Dosim 139:281–286CrossRef Smans K, Struelens L, Smet M et al (2010) Cu filtration for dose reduction in neonatal chest imaging. Radiat Prot Dosim 139:281–286CrossRef
24.
go back to reference Launders JH, Cowen AR, Bury RF, Hawkridge P (2001) Towards image quality, beam energy and effective dose optimisation in digital thoracic radiography. Eur Radiol 11:870–875CrossRef Launders JH, Cowen AR, Bury RF, Hawkridge P (2001) Towards image quality, beam energy and effective dose optimisation in digital thoracic radiography. Eur Radiol 11:870–875CrossRef
25.
go back to reference Wiltz HJ, Petersen U, Axelsson B (2005) Reduction of absorbed dose in storage phosphor urography by significant lowering of tube voltage and adjustment of image display parameters. Acta Radiol 46:391–395CrossRef Wiltz HJ, Petersen U, Axelsson B (2005) Reduction of absorbed dose in storage phosphor urography by significant lowering of tube voltage and adjustment of image display parameters. Acta Radiol 46:391–395CrossRef
26.
go back to reference Steidle G (2004) Two-dimensional parallel acquisition technique in 3D MR colonography. Fortschr Röntgenstr 176(08)CrossRef Steidle G (2004) Two-dimensional parallel acquisition technique in 3D MR colonography. Fortschr Röntgenstr 176(08)CrossRef
27.
go back to reference Rühm SG, Debatin JF (1999) Contrast-enhanced 3D MR-angiography of the thorax, abdomen and lower extremities. Radiologe 39:100–109CrossRef Rühm SG, Debatin JF (1999) Contrast-enhanced 3D MR-angiography of the thorax, abdomen and lower extremities. Radiologe 39:100–109CrossRef
28.
go back to reference Schaefer-Prokop CM, Prokop M (1996) Digital radiography of the thorax - the selenium detector compared to other imaging systems. Kontraste 09: 14-23 Schaefer-Prokop CM, Prokop M (1996) Digital radiography of the thorax - the selenium detector compared to other imaging systems. Kontraste 09: 14-23
29.
go back to reference Schaefer-Prokop CM (1997) Digital Radiography of the Thorax - A system comparison, taking into account dose requirements and diagnostic performance in comparison to conventional film-screen radiography [Habilschr med.]. Hannover: Medical College Schaefer-Prokop CM (1997) Digital Radiography of the Thorax - A system comparison, taking into account dose requirements and diagnostic performance in comparison to conventional film-screen radiography [Habilschr med.]. Hannover: Medical College
30.
go back to reference Dougeni ED, Delis HB, Karatza AA et al (2007) Dose and image quality optimization in neonatal radiography. Br J Radiol 80:807–815CrossRef Dougeni ED, Delis HB, Karatza AA et al (2007) Dose and image quality optimization in neonatal radiography. Br J Radiol 80:807–815CrossRef
Metadata
Title
Modification of chest radiography exposure parameters using a neonatal chest phantom
Authors
Stefan B. Schäfer
Sabine Papst
Martin Fiebich
Claudia Rudolph
Jan de Laffolie
Gabriele A. Krombach
Publication date
01-01-2020
Publisher
Springer Berlin Heidelberg
Published in
Pediatric Radiology / Issue 1/2020
Print ISSN: 0301-0449
Electronic ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-019-04522-1

Other articles of this Issue 1/2020

Pediatric Radiology 1/2020 Go to the issue