Comparison of radiation dose estimates, image noise, and scan duration in pediatric body imaging for volumetric and helical modes on 320-detector CT and helical mode on 64-detector CT

Authors: Jennifer H. Johnston, Daniel J. Podberesky, Terry T. Yoshizumi, Erin Angel, Greta Toncheva, David B. Larson, John C. Egelhoff, Colin Anderson-Evans, Giao B. Nguyen, Alessandra Barelli, Christopher Alsip, Shelia R. Salisbury, Donald P. Frush

Published in: Pediatric Radiology | Issue 9/2013

Login to get access

Abstract

Background

Advanced multidetector CT systems facilitate volumetric image acquisition, which offers theoretic dose savings over helical acquisition with shorter scan times.

Objective

Compare effective dose (ED), scan duration and image noise using 320- and 64-detector CT scanners in various acquisition modes for clinical chest, abdomen and pelvis protocols.

Materials and methods

ED and scan durations were determined for 64-detector helical, 160-detector helical and volume modes under chest, abdomen and pelvis protocols on 320-detector CT with adaptive collimation and 64-detector helical mode on 64-detector CT without adaptive collimation in a phantom representing a 5-year-old child. Noise was measured as standard deviation of Hounsfield units.

Results

Compared to 64-detector helical CT, all acquisition modes on 320-detector CT resulted in lower ED and scan durations. Dose savings were greater for chest (27–46%) than abdomen/pelvis (18–28%) and chest/abdomen/pelvis imaging (8–14%). Noise was similar across scanning modes, although some protocols on 320-detector CT produced slightly higher noise.

Conclusion

Dose savings can be achieved for chest, abdomen/pelvis and chest/abdomen/pelvis examinations on 320-detector CT compared to helical acquisition on 64-detector CT, with shorter scan durations. Although noise differences between some modes reached statistical significance, this is of doubtful diagnostic significance and will be studied further in a clinical setting.
Literature
1.
go back to reference Larson DB, Johnson LW, Schnell BM et al (2011) Rising use of CT in child visits to the emergency department in the United States, 1995–2008. Radiology 259:793–801PubMedCrossRef Larson DB, Johnson LW, Schnell BM et al (2011) Rising use of CT in child visits to the emergency department in the United States, 1995–2008. Radiology 259:793–801PubMedCrossRef
2.
go back to reference Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289–296PubMedCrossRef Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289–296PubMedCrossRef
3.
go back to reference Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505PubMedCrossRef Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505PubMedCrossRef
5.
go back to reference Berrington de Gonzalez A, Mahesh M, Kim KP et al (2009) Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med 169:2071–2077PubMedCrossRef Berrington de Gonzalez A, Mahesh M, Kim KP et al (2009) Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med 169:2071–2077PubMedCrossRef
6.
go back to reference Choi SI, George RT, Schuleri KH et al (2009) Recent developments in wide-detector cardiac computed tomography. Int J Cardiovasc Imaging 25:23–29PubMedCrossRef Choi SI, George RT, Schuleri KH et al (2009) Recent developments in wide-detector cardiac computed tomography. Int J Cardiovasc Imaging 25:23–29PubMedCrossRef
7.
go back to reference Kroft LJ, Roelofs JJ, Geleijns J (2009) Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions. Pediatr Radiol 40:294–300PubMedCrossRef Kroft LJ, Roelofs JJ, Geleijns J (2009) Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions. Pediatr Radiol 40:294–300PubMedCrossRef
9.
go back to reference Greenberg SB, Bhutta S, Braswell L et al (2011) Computed tomography angiography in children with cardiovascular disease: low dose techniques and image quality. Int J Cardiovasc Imaging 28:163–170PubMedCrossRef Greenberg SB, Bhutta S, Braswell L et al (2011) Computed tomography angiography in children with cardiovascular disease: low dose techniques and image quality. Int J Cardiovasc Imaging 28:163–170PubMedCrossRef
10.
go back to reference Al-Mousily F, Shifrin RY, Fricker FJ et al (2011) Use of 320-detector computed tomographic angiography for infants and young children with congenital heart disease. Pediatr Cardiol 32:426–432PubMedCrossRef Al-Mousily F, Shifrin RY, Fricker FJ et al (2011) Use of 320-detector computed tomographic angiography for infants and young children with congenital heart disease. Pediatr Cardiol 32:426–432PubMedCrossRef
11.
go back to reference Bardo DM, Asamato J, Mackay CS et al (2009) Low-dose coronary artery computed tomography angiogram of an infant with tetralogy of fallot using a 256-slice multidetector computed tomography scanner. Pediatr Cardiol 30:824–826PubMedCrossRef Bardo DM, Asamato J, Mackay CS et al (2009) Low-dose coronary artery computed tomography angiogram of an infant with tetralogy of fallot using a 256-slice multidetector computed tomography scanner. Pediatr Cardiol 30:824–826PubMedCrossRef
12.
go back to reference Zhang T, Wang W, Luo Z et al (2011) Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart disease. Int J Cardiovasc Imaging. doi:10.1007/s10554-011-0005-1 Zhang T, Wang W, Luo Z et al (2011) Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart disease. Int J Cardiovasc Imaging. doi:10.​1007/​s10554-011-0005-1
13.
go back to reference Podberesky DJ, Angel E, Yoshizumi TT et al (2012) Radiation dose estimation for prospective and retrospective ECG-gated cardiac CT angiography in infants and small children using a 320-MDCT volume scanner. AJR Am J Roentgenol 199:1129–1135PubMedCrossRef Podberesky DJ, Angel E, Yoshizumi TT et al (2012) Radiation dose estimation for prospective and retrospective ECG-gated cardiac CT angiography in infants and small children using a 320-MDCT volume scanner. AJR Am J Roentgenol 199:1129–1135PubMedCrossRef
14.
go back to reference ICRP (2007) The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37:1–332 ICRP (2007) The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37:1–332
15.
go back to reference Yoshizumi TT, Goodman PC, Frush DP et al (2007) Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: comparison with thermoluminescent dosimetry. AJR Am J Roentgenol 188:1332–1336PubMedCrossRef Yoshizumi TT, Goodman PC, Frush DP et al (2007) Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: comparison with thermoluminescent dosimetry. AJR Am J Roentgenol 188:1332–1336PubMedCrossRef
16.
go back to reference Cristy M, Eckerman KF (1987) Specific absorbed fractions of energy at various ages from internal photon sources. ORNL/TM-8381. Oak Ridge National Laboratory, Oak Ridge Cristy M, Eckerman KF (1987) Specific absorbed fractions of energy at various ages from internal photon sources. ORNL/TM-8381. Oak Ridge National Laboratory, Oak Ridge
17.
go back to reference Zhang D, Savandi AS, Demarco JJ et al (2009) Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms. Med Phys 36:1025–1038PubMedCrossRef Zhang D, Savandi AS, Demarco JJ et al (2009) Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms. Med Phys 36:1025–1038PubMedCrossRef
18.
go back to reference American Association of Physicists in Medicine (2011) Report 204: Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. American Association of Physicists in Medicine, College Park American Association of Physicists in Medicine (2011) Report 204: Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. American Association of Physicists in Medicine, College Park
19.
go back to reference O'Daniel JC, Stevens DM, Cody DD (2005) Reducing radiation exposure from survey CT scans. AJR Am J Roentgenol 185:509–515PubMedCrossRef O'Daniel JC, Stevens DM, Cody DD (2005) Reducing radiation exposure from survey CT scans. AJR Am J Roentgenol 185:509–515PubMedCrossRef
20.
go back to reference Kitajima K, Maeda T, Ohno Y et al (2010) Capability of abdominal 320-detector row CT for small vasculature assessment compared with that of 64-detector row CT. Eur J Radiol 80:219–223PubMedCrossRef Kitajima K, Maeda T, Ohno Y et al (2010) Capability of abdominal 320-detector row CT for small vasculature assessment compared with that of 64-detector row CT. Eur J Radiol 80:219–223PubMedCrossRef
21.
go back to reference Yamashiro T, Miyara T, Takahashi M et al (2012) Lung image quality with 320-row wide-volume CT scans: the effect of prospective ECG-gating and comparisons with 64-row helical CT scans. Acad Radiol 19:380–388PubMedCrossRef Yamashiro T, Miyara T, Takahashi M et al (2012) Lung image quality with 320-row wide-volume CT scans: the effect of prospective ECG-gating and comparisons with 64-row helical CT scans. Acad Radiol 19:380–388PubMedCrossRef
22.
go back to reference Motosugi U, Ichikawa T, Sou H et al (2012) Multi-organ perfusion CT in the abdomen using a 320-detector row CT scanner: preliminary results of perfusion changes in the liver, spleen, and pancreas of cirrhotic patients. Eur J Radiol 81:2533–2537PubMedCrossRef Motosugi U, Ichikawa T, Sou H et al (2012) Multi-organ perfusion CT in the abdomen using a 320-detector row CT scanner: preliminary results of perfusion changes in the liver, spleen, and pancreas of cirrhotic patients. Eur J Radiol 81:2533–2537PubMedCrossRef
23.
go back to reference Schilham A, van der Molen AJ, Prokop M et al (2010) Overranging at multisection CT: an underestimated source of excess radiation exposure. Radiographics 30:1057–1067PubMedCrossRef Schilham A, van der Molen AJ, Prokop M et al (2010) Overranging at multisection CT: an underestimated source of excess radiation exposure. Radiographics 30:1057–1067PubMedCrossRef
24.
go back to reference Tzedakis A, Damilakis J, Perisinakis K et al (2005) The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. Med Phys 32:1621–1629PubMedCrossRef Tzedakis A, Damilakis J, Perisinakis K et al (2005) The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. Med Phys 32:1621–1629PubMedCrossRef
25.
go back to reference Massoumzadeh P, Don S, Hildebolt CF et al (2009) Validation of CT dose-reduction simulation. Med Phys 36:174–189PubMedCrossRef Massoumzadeh P, Don S, Hildebolt CF et al (2009) Validation of CT dose-reduction simulation. Med Phys 36:174–189PubMedCrossRef
26.
go back to reference Martin CJ (2007) Effective dose: how should it be applied to medical exposures? Br J Radiol 80:639–647PubMedCrossRef Martin CJ (2007) Effective dose: how should it be applied to medical exposures? Br J Radiol 80:639–647PubMedCrossRef
Metadata
Title
Comparison of radiation dose estimates, image noise, and scan duration in pediatric body imaging for volumetric and helical modes on 320-detector CT and helical mode on 64-detector CT
Authors
Jennifer H. Johnston
Daniel J. Podberesky
Terry T. Yoshizumi
Erin Angel
Greta Toncheva
David B. Larson
John C. Egelhoff
Colin Anderson-Evans
Giao B. Nguyen
Alessandra Barelli
Christopher Alsip
Shelia R. Salisbury
Donald P. Frush
Publication date
01-09-2013
Publisher
Springer Berlin Heidelberg
Published in
Pediatric Radiology / Issue 9/2013
Print ISSN: 0301-0449
Electronic ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-013-2690-5

Other articles of this Issue 9/2013

Pediatric Radiology 9/2013 Go to the issue