Open Access 01-01-2020 | Heart Surgery | Original Article

Indirect Calorimetry Overestimates Oxygen Consumption in Young Children: Caution is Advised Using Direct Fick Method as a Reference Method in Cardiac Output Comparison Studies

Authors: Theodor S. Sigurdsson, Lars Lindberg

Published in: Pediatric Cardiology | Issue 1/2020

Login to get access

Abstract

Direct Fick method is considered a standard reference method for estimation of cardiac output. It relies on indirect calorimetry to measure oxygen consumption. This is important as only a minor measurement error in oxygen consumption can result in false estimation of cardiac output. A number of studies have shown that indirect calorimetry overestimates oxygen consumption in adults. The aim of this prospective single center observational method comparison study was to compare the determination of oxygen consumption by indirect calorimetry and reverse Fick method in pediatric patients. Forty-two children mean age 352 days (range 30 to 1303 days) and mean weight 7.1 kg (range 2.7–13.6 kg) undergoing corrective cardiac surgery were included in the study. The mean (standard deviation) oxygen consumption by reverse Fick method was 43.5 (16.2) ml/min and by indirect calorimetry 49.9 (18.8) ml/min (p < 0.001). Indirect calorimetry overestimated the reverse Fick oxygen consumption by 14.7%. Bias between methods was 6.5 (11.3) ml/min, limits of agreement (LOA) − 15.7 and 28.7 ml/min and percentage error of 47.7%. A significant bias and large percentage error indicates that the methods are not interchangeable. Indirect calorimetry and the direct Fick method should be used with caution as a reference method in cardiac output comparison studies in young children.
Appendix
Available only for authorised users
Literature
1.
go back to reference Fick A (1870) Über die messung des blutquantums in den Herzventrikeln. Sitz der der Physiol-Med Ges zu Würzbg 2:16 Fick A (1870) Über die messung des blutquantums in den Herzventrikeln. Sitz der der Physiol-Med Ges zu Würzbg 2:16
2.
go back to reference Guyton AC, Hall JE (1996) Textbook of medical physiology, 9th edn. Saunders Company, Philadelphia, W.B, p 251 Guyton AC, Hall JE (1996) Textbook of medical physiology, 9th edn. Saunders Company, Philadelphia, W.B, p 251
3.
go back to reference Laszlo G (2004) Respiratory measurements of cardiac output: from elegant idea to useful test. J Appl Physiol 96:428–437CrossRef Laszlo G (2004) Respiratory measurements of cardiac output: from elegant idea to useful test. J Appl Physiol 96:428–437CrossRef
4.
go back to reference Selzer A, Sudrann RB (1958) Reliability of the determination of cardiac output in man by means of the Fick principle. Circ Res 6:485–490CrossRef Selzer A, Sudrann RB (1958) Reliability of the determination of cardiac output in man by means of the Fick principle. Circ Res 6:485–490CrossRef
5.
go back to reference de Boode WP (2010) Cardiac output monitoring in newborns. Early Hum Dev 86(3):143–148CrossRef de Boode WP (2010) Cardiac output monitoring in newborns. Early Hum Dev 86(3):143–148CrossRef
6.
go back to reference Lemson J, Nusmeier A, van der Hoeven JG (2011) Advanced hemodynamic monitoring in critically ill children. Pediatrics 128(3):560–567PubMed Lemson J, Nusmeier A, van der Hoeven JG (2011) Advanced hemodynamic monitoring in critically ill children. Pediatrics 128(3):560–567PubMed
7.
go back to reference Bizouarn P, Blanloeil Y, Pinaud M (1995) Comparison between oxygen consumption calculated by Fick’s principle using a continuous thermodilution technique and measured by indirect calorimetry. Br J Anaesth 75:719–723CrossRef Bizouarn P, Blanloeil Y, Pinaud M (1995) Comparison between oxygen consumption calculated by Fick’s principle using a continuous thermodilution technique and measured by indirect calorimetry. Br J Anaesth 75:719–723CrossRef
8.
go back to reference Peyton PJ, Robinson GJB (2005) Measured pulmonary oxygen consumption: difference between systemic oxygen uptake measured by the reverse Fick method and indirect calorimetry in cardiac surgery. Anaesthesia 60:146–150CrossRef Peyton PJ, Robinson GJB (2005) Measured pulmonary oxygen consumption: difference between systemic oxygen uptake measured by the reverse Fick method and indirect calorimetry in cardiac surgery. Anaesthesia 60:146–150CrossRef
9.
go back to reference Bizouarn P, Soulard D, Blanloeil Y et al (1992) Oxygen consumption after cardiac surgery—a comparison between calculation by Fick’s principle and measurement by indirect calorimetry. Intensive Care Med 18:206–209CrossRef Bizouarn P, Soulard D, Blanloeil Y et al (1992) Oxygen consumption after cardiac surgery—a comparison between calculation by Fick’s principle and measurement by indirect calorimetry. Intensive Care Med 18:206–209CrossRef
10.
go back to reference Cheong KF, Lee TL (1997) Oxygen consumption—a comparison between calculation by Fick’s principle and measurement by indirect calorimetry. Med J Malays 52(1):70–75 Cheong KF, Lee TL (1997) Oxygen consumption—a comparison between calculation by Fick’s principle and measurement by indirect calorimetry. Med J Malays 52(1):70–75
11.
go back to reference Sigurdsson TS, Aronsson A, Lindberg L (2019) Extracorporeal arteriovenous ultrasound measurement of cardiac output in small children. Anesthesiology 130:712–718CrossRef Sigurdsson TS, Aronsson A, Lindberg L (2019) Extracorporeal arteriovenous ultrasound measurement of cardiac output in small children. Anesthesiology 130:712–718CrossRef
12.
go back to reference Healthcare GE (2006) Datex-Ohmeda S-5 Anaesthetic Monitor-technical reference manual. General Electric Company, Helsinki Healthcare GE (2006) Datex-Ohmeda S-5 Anaesthetic Monitor-technical reference manual. General Electric Company, Helsinki
13.
go back to reference Eisenkraft JB (2011) Respiratory gas monitoring. In: Reich DL (ed) Monitoring in anesthesia and perioperative care. Cambridge University Press, New York, pp 164–165 Eisenkraft JB (2011) Respiratory gas monitoring. In: Reich DL (ed) Monitoring in anesthesia and perioperative care. Cambridge University Press, New York, pp 164–165
14.
go back to reference Meyer R, Briassouli E, Briassoulis G et al (2008) Evaluation of the M-COVX metabolic monitor in mechanically ventilated adult patients. Eur e-J Clin Nutr Metab 3:e232–e239CrossRef Meyer R, Briassouli E, Briassoulis G et al (2008) Evaluation of the M-COVX metabolic monitor in mechanically ventilated adult patients. Eur e-J Clin Nutr Metab 3:e232–e239CrossRef
15.
go back to reference Krivitski NM, Kislukhin VV, Thuramalla NV (2008) Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients. Pediatr Crit Care Med 9(4):423–428CrossRef Krivitski NM, Kislukhin VV, Thuramalla NV (2008) Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients. Pediatr Crit Care Med 9(4):423–428CrossRef
16.
go back to reference Stewart GN (1897) Researches on the circulation time and on the influences which affect it. J Physiol 22(3):159–183CrossRef Stewart GN (1897) Researches on the circulation time and on the influences which affect it. J Physiol 22(3):159–183CrossRef
17.
go back to reference Hamilton WF, Moore JW, Kinsman JM et al (1932) Studies on the circulation. Am J Physiol 99:534–551CrossRef Hamilton WF, Moore JW, Kinsman JM et al (1932) Studies on the circulation. Am J Physiol 99:534–551CrossRef
18.
go back to reference Reuter DA, Huang C, Edrich T et al (2010) Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 110(3):799–811CrossRef Reuter DA, Huang C, Edrich T et al (2010) Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 110(3):799–811CrossRef
19.
go back to reference Crittendon I III, Dreyer WJ, Decker JA et al (2012) Ultrasound dilution: an accurate means of determining cardiac output in children. Pediatr Crit Care Med 13(1):42–46CrossRef Crittendon I III, Dreyer WJ, Decker JA et al (2012) Ultrasound dilution: an accurate means of determining cardiac output in children. Pediatr Crit Care Med 13(1):42–46CrossRef
20.
go back to reference Lindberg L, Johansson S, Perez-de-Sa V (2014) Validation of an ultrasound dilution technology for cardiac output measurement and shunt detection in infants and children. Pediatr Crit Care Med 15(2):139–147CrossRef Lindberg L, Johansson S, Perez-de-Sa V (2014) Validation of an ultrasound dilution technology for cardiac output measurement and shunt detection in infants and children. Pediatr Crit Care Med 15(2):139–147CrossRef
21.
go back to reference Saxena R, Krivitski N, Peacock K et al (2015) Accuracy of the transpulmonary ultrasound dilution method for detection of small anatomic shunts. J Clin Monit Comput 29(3):407–414CrossRef Saxena R, Krivitski N, Peacock K et al (2015) Accuracy of the transpulmonary ultrasound dilution method for detection of small anatomic shunts. J Clin Monit Comput 29(3):407–414CrossRef
22.
go back to reference Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15:85–91CrossRef Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15:85–91CrossRef
23.
go back to reference Kelman GR, Nunn JF (1966) Nomograms for correction of blood PO2, PCO2, pH and base excess for time and temperature. J Appl Physiol 21:1484CrossRef Kelman GR, Nunn JF (1966) Nomograms for correction of blood PO2, PCO2, pH and base excess for time and temperature. J Appl Physiol 21:1484CrossRef
24.
go back to reference Oudemans-van Straaten HM, Scheffer GJ, Eysman L et al (1993) Oxygen consumption after cardiopulmonary bypass—implications of different measuring methods. Intensive Care Med 19:105–110CrossRef Oudemans-van Straaten HM, Scheffer GJ, Eysman L et al (1993) Oxygen consumption after cardiopulmonary bypass—implications of different measuring methods. Intensive Care Med 19:105–110CrossRef
25.
go back to reference de Boode WP, Vrancken SL, Lemson J et al (2013) Gold standard must be solid gold. Intensive Care Med 39:1330–1331CrossRef de Boode WP, Vrancken SL, Lemson J et al (2013) Gold standard must be solid gold. Intensive Care Med 39:1330–1331CrossRef
26.
go back to reference Floh AA, La Rotta G, Wermelt JZ et al (2013) Validation of a new method base on ultrasound velocity dilution to measure cardiac output in paediatric patients. Intensive Care Med 39(5):926–933CrossRef Floh AA, La Rotta G, Wermelt JZ et al (2013) Validation of a new method base on ultrasound velocity dilution to measure cardiac output in paediatric patients. Intensive Care Med 39(5):926–933CrossRef
27.
go back to reference Lemson J, Tibby SM (2016) Reliability of the ultrasound cardiac output monitor for pediatric patients. Pediatr Cardiol 37(3):618CrossRef Lemson J, Tibby SM (2016) Reliability of the ultrasound cardiac output monitor for pediatric patients. Pediatr Cardiol 37(3):618CrossRef
28.
go back to reference Le Manach Y, Collins GS (2016) Disagreement between cardiac output measurement devices: which device is the gold standard? Br J Anaesth 116(4):451–453CrossRef Le Manach Y, Collins GS (2016) Disagreement between cardiac output measurement devices: which device is the gold standard? Br J Anaesth 116(4):451–453CrossRef
29.
go back to reference Cecconi M, Rhodes A, Poloniecki J et al (2009) Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies with specific reference to the measurement of cardiac output. Crit Care 13(1):201CrossRef Cecconi M, Rhodes A, Poloniecki J et al (2009) Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies with specific reference to the measurement of cardiac output. Crit Care 13(1):201CrossRef
30.
go back to reference Suehiro K, Joosten A, Murphy LSL et al (2016) Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis. J Clin Monit Comput 30(5):603–620CrossRef Suehiro K, Joosten A, Murphy LSL et al (2016) Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis. J Clin Monit Comput 30(5):603–620CrossRef
Metadata
Title
Indirect Calorimetry Overestimates Oxygen Consumption in Young Children: Caution is Advised Using Direct Fick Method as a Reference Method in Cardiac Output Comparison Studies
Authors
Theodor S. Sigurdsson
Lars Lindberg
Publication date
01-01-2020
Publisher
Springer US
Published in
Pediatric Cardiology / Issue 1/2020
Print ISSN: 0172-0643
Electronic ISSN: 1432-1971
DOI
https://doi.org/10.1007/s00246-019-02238-5

Other articles of this Issue 1/2020

Pediatric Cardiology 1/2020 Go to the issue