Published in: Neuroradiology 2/2014

01-02-2014 | Interventional Neuroradiology

Evaluation of a newly designed flow diverter for the treatment of intracranial aneurysms in an elastase-induced aneurysm model, in New Zealand white rabbits

Authors: Andreas Simgen, Desiree Ley, Christian Roth, Umut Yilmaz, Heiko Körner, Ruben Mühl-Benninghaus, Yoo-Jin Kim, Bruno Scheller, Wolfgang Reith

Published in: Neuroradiology | Issue 2/2014

Login to get access

Abstract

Introduction

In this study, we analyzed angiographic and histologic aneurysm occlusion of a newly designed flow-diverting device. Visibility and flexibility, as well as occlusions of side branches and neointimal proliferation were also evaluated.

Methods

Aneurysms were induced in 18 New Zealand white rabbits and treated with a braided, “closed-loop-designed” device of nitinol. Additional devices were implanted in the abdominal aorta to cover the origin of branch arteries. Angiographic follow-ups were performed immediately after placement of the device, after 3 months (n = 9) and 6 months (n = 9). The status of aneurysm occlusion (using a five-point scale) and the patency of branch arteries were assessed.

Results

Aneurysm occlusion rates were noted as grade 0 in 2 (11 %), grade I in 1 (6 %), grade II in 1 (6 %), grade III in 9 (50 %), and grade IV in 5 (28 %) of 18 aneurysms, respectively, indicating a complete or near-complete occlusion of 78 % under double antiplatelet therapy. Aneurysm occlusion was significantly higher at 6 months follow-up (P = 0.025). Radiopaque markers provided excellent visibility. Limited device flexibility led to incomplete aneurysm neck coverage and grade 0 occlusion rates in two cases. Distal device occlusions were found in three cases, most likely due to an extremely undersized vessel diameter in the subclavian artery. No case of branch artery occlusion was seen. Intimal proliferation and diameter stenosis were moderate.

Conclusion

The tested flow diverter achieved near-complete and complete aneurysm occlusion under double antiplatelet therapy of elastase-induced aneurysms in 78 %, while preserving branch arteries.
Literature
1.
go back to reference Molyneux AJ, Kerr RSC, Yu L-M et al (2005) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef Molyneux AJ, Kerr RSC, Yu L-M et al (2005) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef
2.
go back to reference McDougall CG, Spetzler RF, Zabramski JM et al (2012) The Barrow Ruptured Aneurysm Trial. J Neurosurg 116:135–144PubMedCrossRef McDougall CG, Spetzler RF, Zabramski JM et al (2012) The Barrow Ruptured Aneurysm Trial. J Neurosurg 116:135–144PubMedCrossRef
3.
go back to reference Grunwald IQ, Papanagiotou P, Struffert T et al (2007) Recanalization after endovascular treatment of intracerebral aneurysms. Neuroradiology 49:41–47PubMedCrossRef Grunwald IQ, Papanagiotou P, Struffert T et al (2007) Recanalization after endovascular treatment of intracerebral aneurysms. Neuroradiology 49:41–47PubMedCrossRef
4.
go back to reference Gallas S, Januel AC, Pasco A et al (2009) Long-term follow-up of 1036 cerebral aneurysms treated by bare coils: a multicentric cohort treated between 1998 and 2003. AJNR Am J Neuroradiol 30:1986–1992PubMedCrossRef Gallas S, Januel AC, Pasco A et al (2009) Long-term follow-up of 1036 cerebral aneurysms treated by bare coils: a multicentric cohort treated between 1998 and 2003. AJNR Am J Neuroradiol 30:1986–1992PubMedCrossRef
5.
go back to reference Kallmes DF, Ding YH, Dai D et al (2007) A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38:2346–2352PubMedCrossRef Kallmes DF, Ding YH, Dai D et al (2007) A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38:2346–2352PubMedCrossRef
6.
go back to reference Fiorella D, Woo HH, Albuquerque FC, Nelson PK (2008) Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device. Neurosurgery 62:1115–1120PubMedCrossRef Fiorella D, Woo HH, Albuquerque FC, Nelson PK (2008) Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device. Neurosurgery 62:1115–1120PubMedCrossRef
7.
go back to reference Lylyk P, Miranda C, Ceratto R et al (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64:632–642PubMedCrossRef Lylyk P, Miranda C, Ceratto R et al (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64:632–642PubMedCrossRef
8.
go back to reference Kallmes DF, Ding YH, Dai D et al (2009) A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms. AJNR Am J Neuroradiol 30:1153–1158PubMedCrossRef Kallmes DF, Ding YH, Dai D et al (2009) A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms. AJNR Am J Neuroradiol 30:1153–1158PubMedCrossRef
9.
go back to reference Sadasivan C, Cesar L, Seong J et al (2009) An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model. Stroke 40:952–958PubMedCentralPubMedCrossRef Sadasivan C, Cesar L, Seong J et al (2009) An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model. Stroke 40:952–958PubMedCentralPubMedCrossRef
10.
go back to reference Kulcsar Z, Houdart E, Bonafé A et al (2011) Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow-diversion treatment. AJNR Am J Neuroradiol 32:20–25PubMedCrossRef Kulcsar Z, Houdart E, Bonafé A et al (2011) Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow-diversion treatment. AJNR Am J Neuroradiol 32:20–25PubMedCrossRef
11.
go back to reference Turowski B, Macht S, Kulcsár Z et al (2011) Early fatal hemorrhage after endovascular cerebral aneurysm treatment with a flow diverter (SILK-Stent): do we need to rethink our concepts? Neuroradiology 53:37–41PubMedCrossRef Turowski B, Macht S, Kulcsár Z et al (2011) Early fatal hemorrhage after endovascular cerebral aneurysm treatment with a flow diverter (SILK-Stent): do we need to rethink our concepts? Neuroradiology 53:37–41PubMedCrossRef
12.
go back to reference Cebral JR, Mut F, Raschi M et al (2011) Aneurysm rupture following treatment with flow-diverting stents: computational hemodynamics analysis of treatment. AJNR Am J Neuroradiol 32:27–33PubMedCrossRef Cebral JR, Mut F, Raschi M et al (2011) Aneurysm rupture following treatment with flow-diverting stents: computational hemodynamics analysis of treatment. AJNR Am J Neuroradiol 32:27–33PubMedCrossRef
13.
go back to reference Berge J, Biondi A, Machi P et al (2012) Flow-diverter silk stent for the treatment of intracranial aneurysms: 1-year follow-up in a multicenter study. Am J Neuroradiol 33:1150–1155PubMedCrossRef Berge J, Biondi A, Machi P et al (2012) Flow-diverter silk stent for the treatment of intracranial aneurysms: 1-year follow-up in a multicenter study. Am J Neuroradiol 33:1150–1155PubMedCrossRef
14.
go back to reference Fischer S, Vajda Z, Aguilar Perez M et al (2012) Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections. Neuroradiology 54:369–382PubMedCentralPubMedCrossRef Fischer S, Vajda Z, Aguilar Perez M et al (2012) Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections. Neuroradiology 54:369–382PubMedCentralPubMedCrossRef
15.
go back to reference Maimon S, Gonen L, Nossek E et al (2012) Treatment of intracranial aneurysms with the SILK flow diverter: 2 years' experience with 28 patients at a single center. Acta Neurochir 154:979–987PubMedCrossRef Maimon S, Gonen L, Nossek E et al (2012) Treatment of intracranial aneurysms with the SILK flow diverter: 2 years' experience with 28 patients at a single center. Acta Neurochir 154:979–987PubMedCrossRef
16.
go back to reference Becske T, Kallmes DF, Saatci I et al (2013) Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology 267:858–868PubMedCrossRef Becske T, Kallmes DF, Saatci I et al (2013) Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology 267:858–868PubMedCrossRef
17.
go back to reference Cloft HJ, Altes TA, Marx WF et al (1999) Endovascular creation of an in vivo bifurcation aneurysm model in rabbits. Radiology 213:223–228PubMedCrossRef Cloft HJ, Altes TA, Marx WF et al (1999) Endovascular creation of an in vivo bifurcation aneurysm model in rabbits. Radiology 213:223–228PubMedCrossRef
18.
go back to reference Aurboonyawat T, Blanc R, Schmidt P et al (2011) An in vitro study of silk stent morphology. Neuroradiology 53:659–667PubMedCrossRef Aurboonyawat T, Blanc R, Schmidt P et al (2011) An in vitro study of silk stent morphology. Neuroradiology 53:659–667PubMedCrossRef
19.
go back to reference Gross BA, Frerichs KU (2013) Stent usage in the treatment of intracranial aneurysms: past, present and future. J Neurol Neurosurg Psychiatry 84:244–253PubMedCrossRef Gross BA, Frerichs KU (2013) Stent usage in the treatment of intracranial aneurysms: past, present and future. J Neurol Neurosurg Psychiatry 84:244–253PubMedCrossRef
20.
go back to reference Cremers B, Kelsch B, Clever YP et al (2012) Inhibition of neointimal proliferation after bare metal stent implantation with low-pressure drug delivery using a paclitaxel-coated balloon in porcine coronary arteries. Clin Res Cardiol 101:385–391PubMedCrossRef Cremers B, Kelsch B, Clever YP et al (2012) Inhibition of neointimal proliferation after bare metal stent implantation with low-pressure drug delivery using a paclitaxel-coated balloon in porcine coronary arteries. Clin Res Cardiol 101:385–391PubMedCrossRef
21.
go back to reference Kamran M, Yarnold J, Grunwald IQ, Byrne JV (2011) Assessment of angiographic outcomes after flow diversion treatment of intracranial aneurysms: a new grading schema. Neuroradiology 53:501–508PubMedCrossRef Kamran M, Yarnold J, Grunwald IQ, Byrne JV (2011) Assessment of angiographic outcomes after flow diversion treatment of intracranial aneurysms: a new grading schema. Neuroradiology 53:501–508PubMedCrossRef
22.
go back to reference Seong J, Wakhloo AK, Lieber BB (2007) In vitro evaluation of flow divertors in an elastase-induced saccular aneurysm model in rabbit. J Biomech Eng 129:863–872PubMedCrossRef Seong J, Wakhloo AK, Lieber BB (2007) In vitro evaluation of flow divertors in an elastase-induced saccular aneurysm model in rabbit. J Biomech Eng 129:863–872PubMedCrossRef
23.
go back to reference Kulcsár Z, Ernemann U, Wetzel SG et al (2010) High-profile flow diverter (Silk) implantation in the basilar artery. Stroke 41:1690–1696PubMedCrossRef Kulcsár Z, Ernemann U, Wetzel SG et al (2010) High-profile flow diverter (Silk) implantation in the basilar artery. Stroke 41:1690–1696PubMedCrossRef
24.
go back to reference van Rooij WJ, Sluzewski M (2010) Perforator infarction after placement of a pipeline flow-diverting stent for an unruptured A1 aneurysm. AJNR Am J Neuroradiol 31:E43–44PubMedCrossRef van Rooij WJ, Sluzewski M (2010) Perforator infarction after placement of a pipeline flow-diverting stent for an unruptured A1 aneurysm. AJNR Am J Neuroradiol 31:E43–44PubMedCrossRef
25.
go back to reference Dai D, Ding YH, Kadirvel R et al (2012) Patency of branches after coverage with multiple telescoping flow-diverter devices: an in vivo study in rabbits. AJNR Am J Neuroradiol 33:171–174PubMedCrossRef Dai D, Ding YH, Kadirvel R et al (2012) Patency of branches after coverage with multiple telescoping flow-diverter devices: an in vivo study in rabbits. AJNR Am J Neuroradiol 33:171–174PubMedCrossRef
26.
go back to reference Asakura M, Ueda Y, Nanto S et al (1998) Remodeling of in-stent neointima, which became thinner and transparent over 3 years: serial angiographic and angioscopic follow-up. Circulation 97:2003–2006PubMedCrossRef Asakura M, Ueda Y, Nanto S et al (1998) Remodeling of in-stent neointima, which became thinner and transparent over 3 years: serial angiographic and angioscopic follow-up. Circulation 97:2003–2006PubMedCrossRef
27.
go back to reference Dai D, Ding YH, Kadirvel R et al (2007) Bone formation in elastase-induced rabbit aneurysms embolized with platinum coils: report of 2 cases. AJNR Am J Neuroradiol 28:1176–1178PubMedCrossRef Dai D, Ding YH, Kadirvel R et al (2007) Bone formation in elastase-induced rabbit aneurysms embolized with platinum coils: report of 2 cases. AJNR Am J Neuroradiol 28:1176–1178PubMedCrossRef
Metadata
Title
Evaluation of a newly designed flow diverter for the treatment of intracranial aneurysms in an elastase-induced aneurysm model, in New Zealand white rabbits
Authors
Andreas Simgen
Desiree Ley
Christian Roth
Umut Yilmaz
Heiko Körner
Ruben Mühl-Benninghaus
Yoo-Jin Kim
Bruno Scheller
Wolfgang Reith
Publication date
01-02-2014
Publisher
Springer Berlin Heidelberg
Published in
Neuroradiology / Issue 2/2014
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-013-1296-9

Other articles of this Issue 2/2014

Neuroradiology 2/2014 Go to the issue