01-03-2012 | Paediatric Neuroradiology

Radiation dose optimization in pediatric temporal bone computed tomography: influence of tube tension on image contrast and image quality

Authors: Claude Bertrand Nauer, Christoph Zubler, Christian Weisstanner, Christof Stieger, Pascal Senn, Andreas Arnold

Published in: Neuroradiology | Issue 3/2012

Login to get access

Abstract

Introduction

The purpose of this experimental study was to investigate the effect of tube tension reduction on image contrast and image quality in pediatric temporal bone computed tomography (CT).

Methods

Seven lamb heads with infant-equivalent sizes were scanned repeatedly, using four tube tensions from 140 to 80 kV while the CT-Dose Index (CTDI) was held constant. Scanning was repeated with four CTDI values from 30 to 3 mGy. Image contrast was calculated for the middle ear as the Hounsfield unit (HU) difference between bone and air and for the inner ear as the HU difference between bone and fluid. The influence of tube tension on high-contrast detail delineation was evaluated using a phantom. The subjective image quality of eight middle and inner ear structures was assessed using a 4-point scale (scores 1–2 = insufficient; scores 3–4 = sufficient).

Results

Middle and inner ear contrast showed a near linear increase with tube tension reduction (r = −0.94/−0.88) and was highest at 80 kV. Tube tension had no influence on spatial resolution. Subjective image quality analysis showed significantly better scoring at lower tube tensions, with highest image quality at 80 kV. However, image quality improvement was most relevant for low-dose scans.

Conclusions

Image contrast in the temporal bone is significantly higher at low tube tensions, leading to a better subjective image quality. Highest contrast and best quality were found at 80 kV. This image quality improvement might be utilized to further reduce the radiation dose in pediatric low-dose CT protocols.
Literature
1.
go back to reference Fazel R, Krumholz HM, Wang Y, Ross JS, Chen J, Ting HH, Shah ND, Nasir K, Einstein AJ, Nallamothu BK (2009) Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med 361(9):849–857. doi:10.1056/NEJMoa0901249 PubMedCrossRef Fazel R, Krumholz HM, Wang Y, Ross JS, Chen J, Ting HH, Shah ND, Nasir K, Einstein AJ, Nallamothu BK (2009) Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med 361(9):849–857. doi:10.​1056/​NEJMoa0901249 PubMedCrossRef
4.
go back to reference Brenner D, Elliston C, Hall E, Berdon W (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176(2):289–296PubMed Brenner D, Elliston C, Hall E, Berdon W (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176(2):289–296PubMed
5.
go back to reference Huda W, Lieberman KA, Chang J, Roskopf ML (2004) Patient size and X-ray technique factors in head computed tomography examinations. II. Image quality. Med Phys 31(3):595–601PubMedCrossRef Huda W, Lieberman KA, Chang J, Roskopf ML (2004) Patient size and X-ray technique factors in head computed tomography examinations. II. Image quality. Med Phys 31(3):595–601PubMedCrossRef
6.
go back to reference Khursheed A, Hillier MC, Shrimpton PC, Wall BF (2002) Influence of patient age on normalized effective doses calculated for CT examinations. Br J Radiol 75(898):819–830PubMed Khursheed A, Hillier MC, Shrimpton PC, Wall BF (2002) Influence of patient age on normalized effective doses calculated for CT examinations. Br J Radiol 75(898):819–830PubMed
7.
go back to reference Bettman R, Beek E, Van Olphen A, Zonneveld F, Huizing E (2004) MRI versus CT in assessment of cochlear patency in cochlear implant candidates. Acta Otolaryngol 124(5):577–581PubMedCrossRef Bettman R, Beek E, Van Olphen A, Zonneveld F, Huizing E (2004) MRI versus CT in assessment of cochlear patency in cochlear implant candidates. Acta Otolaryngol 124(5):577–581PubMedCrossRef
9.
go back to reference Vazquez E, Castellote A, Piqueras J, Mauleon S, Creixell S, Pumarola F, Figueras C, Carreno JC, Lucaya J (2003) Imaging of complications of acute mastoiditis in children. Radiographics 23(2):359–372PubMedCrossRef Vazquez E, Castellote A, Piqueras J, Mauleon S, Creixell S, Pumarola F, Figueras C, Carreno JC, Lucaya J (2003) Imaging of complications of acute mastoiditis in children. Radiographics 23(2):359–372PubMedCrossRef
11.
go back to reference Goske MJ, Applegate KE, Boylan J, Butler PF, Callahan MJ, Coley BD, Farley S, Frush DP, Hernanz-Schulman M, Jaramillo D, Johnson ND, Kaste SC, Morrison G, Strauss KJ, Tuggle N (2008) The Image Gently campaign: working together to change practice. AJR Am J Roentgenol 190(2):273–274. doi:10.2214/AJR.07.3526 PubMedCrossRef Goske MJ, Applegate KE, Boylan J, Butler PF, Callahan MJ, Coley BD, Farley S, Frush DP, Hernanz-Schulman M, Jaramillo D, Johnson ND, Kaste SC, Morrison G, Strauss KJ, Tuggle N (2008) The Image Gently campaign: working together to change practice. AJR Am J Roentgenol 190(2):273–274. doi:10.​2214/​AJR.​07.​3526 PubMedCrossRef
13.
go back to reference Szucs-Farkas Z, Semadeni M, Bensler S, Patak MA, von Allmen G, Vock P, Schindera ST (2009) Endoleak detection with CT angiography in an abdominal aortic aneurysm phantom: effect of tube energy, simulated patient size, and physical properties of endoleaks. Radiology 251(2):590–598. doi:10.1148/radiol.2512081687 PubMedCrossRef Szucs-Farkas Z, Semadeni M, Bensler S, Patak MA, von Allmen G, Vock P, Schindera ST (2009) Endoleak detection with CT angiography in an abdominal aortic aneurysm phantom: effect of tube energy, simulated patient size, and physical properties of endoleaks. Radiology 251(2):590–598. doi:10.​1148/​radiol.​2512081687 PubMedCrossRef
14.
go back to reference Wintermark M, Maeder P, Verdun FR, Thiran JP, Valley JF, Schnyder P, Meuli R (2000) Using 80 kVp versus 120 kVp in perfusion CT measurement of regional cerebral blood flow. AJNR Am J Neuroradiol 21(10):1881–1884PubMed Wintermark M, Maeder P, Verdun FR, Thiran JP, Valley JF, Schnyder P, Meuli R (2000) Using 80 kVp versus 120 kVp in perfusion CT measurement of regional cerebral blood flow. AJNR Am J Neuroradiol 21(10):1881–1884PubMed
15.
go back to reference Seibel VA, Lavinsky L, De Oliveira JA (2006) Morphometric study of the external and middle ear anatomy in sheep: a possible model for ear experiments. Clin Anat 19(6):503–509. doi:10.1002/ca.20218 PubMedCrossRef Seibel VA, Lavinsky L, De Oliveira JA (2006) Morphometric study of the external and middle ear anatomy in sheep: a possible model for ear experiments. Clin Anat 19(6):503–509. doi:10.​1002/​ca.​20218 PubMedCrossRef
16.
go back to reference Seibel VA, Lavinsky L, Irion K (2006) CT-Scan sheep and human inner ear morphometric comparison. Braz J Otorhinolaryngol 72(3):370–376PubMed Seibel VA, Lavinsky L, Irion K (2006) CT-Scan sheep and human inner ear morphometric comparison. Braz J Otorhinolaryngol 72(3):370–376PubMed
17.
go back to reference Suetens P (2009) Fundamentals of medical imaging, 2nd edn. Cambridge University Press, CambridgeCrossRef Suetens P (2009) Fundamentals of medical imaging, 2nd edn. Cambridge University Press, CambridgeCrossRef
18.
go back to reference Walsh C, Dowling A, Meade A, Malone J (2005) Subjective and objective measures of image quality in digital fluoroscopy. Radiat Prot Dosim 117(1–3):34–37. doi:10.1093/rpd/nci708 Walsh C, Dowling A, Meade A, Malone J (2005) Subjective and objective measures of image quality in digital fluoroscopy. Radiat Prot Dosim 117(1–3):34–37. doi:10.​1093/​rpd/​nci708
20.
go back to reference Huda W, Lieberman KA, Chang J, Roskopf ML (2004) Patient size and x-ray technique factors in head computed tomography examinations. I. Radiation doses. Med Phys 31(3):588–594PubMedCrossRef Huda W, Lieberman KA, Chang J, Roskopf ML (2004) Patient size and x-ray technique factors in head computed tomography examinations. I. Radiation doses. Med Phys 31(3):588–594PubMedCrossRef
Metadata
Title
Radiation dose optimization in pediatric temporal bone computed tomography: influence of tube tension on image contrast and image quality
Authors
Claude Bertrand Nauer
Christoph Zubler
Christian Weisstanner
Christof Stieger
Pascal Senn
Andreas Arnold
Publication date
01-03-2012
Publisher
Springer-Verlag
Published in
Neuroradiology / Issue 3/2012
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-011-0961-0

Other articles of this Issue 3/2012

Neuroradiology 3/2012 Go to the issue