A synthetic bone insert may protect the lateral cortex and fixation plate following a high tibial osteotomy by reducing the tensile strains
Authors:
Rachel MacDonald, Kyla Huebner, Jack Farr, Cynthia E. Dunning, Alan Getgood, Timothy A. Burkhart
Published in:
Knee Surgery, Sports Traumatology, Arthroscopy
|
Issue 6/2020
Login to get access
Abstract
Purpose
To determine the effectiveness of a synthetic bone insert on improving medial opening wedge high tibial osteotomy integrity in response to post-surgical cyclical loading.
Materials and methods
A medial opening wedge high tibial osteotomy, secured with a compression fixation plate, was performed on 12 cadaveric knee specimens that were randomised to either: (1) a synthetic insert condition (n = 6), in which a 9 mm bio-absorbable wedge was inserted into the gap space; or (2) a plate-only condition (n = 6). Uniaxial strain gauges, placed on the lateral cortex and fixation plate, measured the strain response as the specimens were subjected to a staircase cyclical loading protocol; a sinusoidal waveform between 100 and 800 N was applied and increased by increments of 200 N every 5000 cycles until failure. Peak strains at failure were compared between conditions using a one-tailed independent samples t test.
Results
The strains from the fixation plate were significantly different between the insert and plate only conditions (p = 0.02), transitioning from a compressive strain with the wedge (mean [SD] = − 8.6 [− 3.6] µε) to a tensile strain without the wedge (mean [SD] = 12.9 [23] µε). The strains measured at the lateral cortex were also significantly affected by the inclusion of a synthetic bone insert (p = 0.016), increasing from − 55.6 (− 54.3) µε when the insert was utilised to 23.7 (55.7) µε when only the plate was used.
Conclusions
The addition of a synthetic insert limited the tensile strains at the plate and lateral cortex, suggesting that this may protect these regions from fracture during prolonged loading.