Published in: Knee Surgery, Sports Traumatology, Arthroscopy 4/2019

Open Access 01-04-2019 | KNEE

Patient-specific instrumentation improved axial alignment of the femoral component, operative time and perioperative blood loss after total knee arthroplasty

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 4/2019

Login to get access

Abstract

Purpose

The purpose of the present study was to compare patient-specific instrumentation (PSI) with standard instrumentation (SI) in patients undergoing total knee arthroplasty (TKA). PSI is hypothesized to have advantages with respect to component alignment; number of outliers (defined as alignment > 3° from the target alignment); operative time; perioperative blood loss; and length of hospital stay. This new surgical technique is expected to exhibit superior performance.

Methods

A total of 23 randomized controlled trials (RCTs) involving 2058 knees that compared the clinical outcomes of TKA between PSI and SI were included in the present analysis; these RCTs were identified via a literature search of the PubMed, Embase, and Cochrane Library databases through March 1, 2018. The outcomes of interest included coronal, sagittal and axial component alignment (presented as the angle of deviation from the transcondylar line); number of outliers; operative time; perioperative blood loss; and length of hospital stay.

Results

There was a significant difference in postoperative femoral axial alignment between PSI and SI patients (95% CI − 0.71 to − 0.21, p = 0.0004, I2 = 48%). PSI resulted in approximately 0.4° less deviation from the transcondylar line than SI. Based on our results, PSI reduced operative time by a mean of 7 min compared with SI (95% CI − 10.95 to − 3.75, p < 0.0001, I2 = 78%). According to the included literature, PSI reduced perioperative blood loss by approximately 90 ml compared with SI (95% CI − 146.65 to − 20.18, p = 0.01, I2 = 74%). We did not find any differences between PSI and SI with respect to any other parameters.

Conclusions

PSI has advantages in axial alignment of the femoral component, operative time, and perioperative blood loss relative to SI. No significant differences were found between PSI and SI with respect to alignment of the remaining components, number of outliers, or length of hospital stay.

Level of evidence

Therapeutic study (systematic review and meta-analysis), Level I.
Literature
1.
go back to reference Abane L, Anract P, Boisgard S, Descamps S, Courpied JP, Hamadouche M (2015) A comparison of patient-specific and conventional instrumentation for total knee arthroplasty: a multicentre randomized controlled trial. Bone Jt J 97-B:56–63CrossRef Abane L, Anract P, Boisgard S, Descamps S, Courpied JP, Hamadouche M (2015) A comparison of patient-specific and conventional instrumentation for total knee arthroplasty: a multicentre randomized controlled trial. Bone Jt J 97-B:56–63CrossRef
2.
go back to reference Alcelik I, Blomfield M, Öztürk C, Soni A, Charity R, Acornley A (2017) A comparison of short term radiological alignment outcomes of the patient specific and standard instrumentation for primary total knee arthroplasty: a systematic review and meta-analysis. Acta Orthop Traumatol Turc 51(3):215–222CrossRefPubMedPubMedCentral Alcelik I, Blomfield M, Öztürk C, Soni A, Charity R, Acornley A (2017) A comparison of short term radiological alignment outcomes of the patient specific and standard instrumentation for primary total knee arthroplasty: a systematic review and meta-analysis. Acta Orthop Traumatol Turc 51(3):215–222CrossRefPubMedPubMedCentral
3.
go back to reference Anderl W, Pauzenberger L, Kölblinger R, Kiesselbach G, Brandl G, Laky B, Kriegleder B, Heuberer P, Schwameis E (2016) Patient-specific instrumentation improved mechanical alignment, while early clinical outcome was comparable to conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc 24:102–111. 183CrossRefPubMed Anderl W, Pauzenberger L, Kölblinger R, Kiesselbach G, Brandl G, Laky B, Kriegleder B, Heuberer P, Schwameis E (2016) Patient-specific instrumentation improved mechanical alignment, while early clinical outcome was comparable to conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc 24:102–111. 183CrossRefPubMed
4.
go back to reference Boonen B, Schotanus MG, Kerens B, Van der Weegen W, van Drumpt RA, Kort NP (2013) Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee Surg Sports Traumatol Arthrosc 21:2206–2212CrossRefPubMed Boonen B, Schotanus MG, Kerens B, Van der Weegen W, van Drumpt RA, Kort NP (2013) Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee Surg Sports Traumatol Arthrosc 21:2206–2212CrossRefPubMed
5.
go back to reference Boonen B, Schotanus MG, Kerens B, Van der Weegen W, Hoekstra HJ, Kort NP (2016) No difference in clinical outcome between patient-matched positioning guides and conventional instrumented total knee arthroplasty two years post-operatively: a multicentre, double-blind, randomised controlled trial. Bone Jt J 98-B:939–944. 191CrossRef Boonen B, Schotanus MG, Kerens B, Van der Weegen W, Hoekstra HJ, Kort NP (2016) No difference in clinical outcome between patient-matched positioning guides and conventional instrumented total knee arthroplasty two years post-operatively: a multicentre, double-blind, randomised controlled trial. Bone Jt J 98-B:939–944. 191CrossRef
6.
go back to reference Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C (2013) A prospective randomized controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Jt J 95-B:354–359CrossRef Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C (2013) A prospective randomized controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Jt J 95-B:354–359CrossRef
7.
go back to reference Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322CrossRefPubMed Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322CrossRefPubMed
8.
go back to reference Camarda L, D’Arienzo A, Morello S, Peri G, Valentino B, D’Arienzo M (2015) Patient-specific instrumentation for total knee arthroplasty: a literature review. Musculoskelet Surg 99:11–18CrossRefPubMed Camarda L, D’Arienzo A, Morello S, Peri G, Valentino B, D’Arienzo M (2015) Patient-specific instrumentation for total knee arthroplasty: a literature review. Musculoskelet Surg 99:11–18CrossRefPubMed
9.
go back to reference Cavaignac E, Pailhé R, Laumond G, Murgier J, Reina N, Laffosse JM (2015) Evaluation of the accuracy of patient—specific cutting blocks for total knee arthroplasty: a meta-analysis. Int Orthop 39(8):1541–1552CrossRefPubMed Cavaignac E, Pailhé R, Laumond G, Murgier J, Reina N, Laffosse JM (2015) Evaluation of the accuracy of patient—specific cutting blocks for total knee arthroplasty: a meta-analysis. Int Orthop 39(8):1541–1552CrossRefPubMed
10.
go back to reference De RV, Pellikaan P, Dhollander A, Vander JS (2017) Three-dimensional analysis of accuracy of component positioning in total knee arthroplasty with patient specific and conventional instruments: a randomized controlled trial. Knee 24:1469–1477CrossRef De RV, Pellikaan P, Dhollander A, Vander JS (2017) Three-dimensional analysis of accuracy of component positioning in total knee arthroplasty with patient specific and conventional instruments: a randomized controlled trial. Knee 24:1469–1477CrossRef
11.
go back to reference Daniilidis K, Tibesku CO (2014) A comparison of conventional and patient-specific instruments in total knee arthroplasty. Int Orthop 38:503–508CrossRefPubMed Daniilidis K, Tibesku CO (2014) A comparison of conventional and patient-specific instruments in total knee arthroplasty. Int Orthop 38:503–508CrossRefPubMed
12.
go back to reference Fu H, Wang J, Zhou S, Cheng T, Zhang W, Wang Q (2015) No difference in mechanical alignment and femoral component placement between patient-specific instrumentation and conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc 23:3288–3295CrossRefPubMed Fu H, Wang J, Zhou S, Cheng T, Zhang W, Wang Q (2015) No difference in mechanical alignment and femoral component placement between patient-specific instrumentation and conventional instrumentation in TKA. Knee Surg Sports Traumatol Arthrosc 23:3288–3295CrossRefPubMed
13.
go back to reference Gan Y, Ding J, Xu Y, Hou C (2015) Accuracy and efficacy of osteotomy in total knee arthroplasty with patient-specific navigational template. Int J Clin Exp Med 8:12192–12201PubMedPubMedCentral Gan Y, Ding J, Xu Y, Hou C (2015) Accuracy and efficacy of osteotomy in total knee arthroplasty with patient-specific navigational template. Int J Clin Exp Med 8:12192–12201PubMedPubMedCentral
14.
go back to reference Gill GS, Joshi AB (2001) Long-term results of cemented, posterior cruciate ligament-retaining total knee arthroplasty in osteoarthritis. Am J Knee Surg 14:209–214PubMed Gill GS, Joshi AB (2001) Long-term results of cemented, posterior cruciate ligament-retaining total knee arthroplasty in osteoarthritis. Am J Knee Surg 14:209–214PubMed
15.
go back to reference Hamilton WG, Parks NL, Saxena A (2013) Patient-specific instrumentation does not shorten surgical time: a prospective, randomized trial. J Arthroplasty 28(8 Suppl):96–100CrossRefPubMed Hamilton WG, Parks NL, Saxena A (2013) Patient-specific instrumentation does not shorten surgical time: a prospective, randomized trial. J Arthroplasty 28(8 Suppl):96–100CrossRefPubMed
16.
go back to reference Huijbregts HJ, Khan RJ, Fick DP, Hall MJ, Punwar SA, Sorensen E (2016) Component alignment and clinical outcome following total knee arthroplasty: a randomised controlled trial comparing an intramedullary alignment system with patient-specific instrumentation. Bone Jt J 98-B:1043–1049CrossRef Huijbregts HJ, Khan RJ, Fick DP, Hall MJ, Punwar SA, Sorensen E (2016) Component alignment and clinical outcome following total knee arthroplasty: a randomised controlled trial comparing an intramedullary alignment system with patient-specific instrumentation. Bone Jt J 98-B:1043–1049CrossRef
17.
go back to reference Huijbregts HJTAM, Khan RJK, Sorensen E, Fick DP, Haebich S (2016) Patient-specific instrumentation does not improve radiographic alignment or clinical outcomes after total knee arthroplasty. Acta Orthop 87(4):386–394CrossRefPubMedPubMedCentral Huijbregts HJTAM, Khan RJK, Sorensen E, Fick DP, Haebich S (2016) Patient-specific instrumentation does not improve radiographic alignment or clinical outcomes after total knee arthroplasty. Acta Orthop 87(4):386–394CrossRefPubMedPubMedCentral
18.
go back to reference Iorio R, Bolle G, Conteduca F, Valeo L, Conteduca J, Mazza D (2013) Accuracy of manual instrumentation of tibial cutting guide in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 21:2296–2300CrossRefPubMed Iorio R, Bolle G, Conteduca F, Valeo L, Conteduca J, Mazza D (2013) Accuracy of manual instrumentation of tibial cutting guide in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 21:2296–2300CrossRefPubMed
19.
go back to reference Jiang J, Kang X, Lin Q, Teng Y, An L, Ma J (2015) Accuracy of patient-specific instrumentation compared with conventional instrumentation in total knee arthroplasty. Orthopedics 38(4):e305CrossRefPubMed Jiang J, Kang X, Lin Q, Teng Y, An L, Ma J (2015) Accuracy of patient-specific instrumentation compared with conventional instrumentation in total knee arthroplasty. Orthopedics 38(4):e305CrossRefPubMed
20.
go back to reference Khuangsirikul S, Lertcharoenchoke T, Chotanaphuti T (2014) Rotational alignment of femoral component between custom cutting block and conventional technique in total knee arthroplasty. J Med Assoc Thai 97(Suppl 2):S47–S51PubMed Khuangsirikul S, Lertcharoenchoke T, Chotanaphuti T (2014) Rotational alignment of femoral component between custom cutting block and conventional technique in total knee arthroplasty. J Med Assoc Thai 97(Suppl 2):S47–S51PubMed
21.
go back to reference Kotela A, Kotela I (2014) Patient-specific computed tomography based instrumentation in total knee arthroplasty: a prospective randomized controlled study. Int Orthop 38:2099–2107CrossRefPubMed Kotela A, Kotela I (2014) Patient-specific computed tomography based instrumentation in total knee arthroplasty: a prospective randomized controlled study. Int Orthop 38:2099–2107CrossRefPubMed
22.
go back to reference Kotela A, Lorkowski J, Kucharzewski M, Wilkfrańczuk M, Śliwiński Z, Frańczuk B (2015) Patient-specific ct-based instrumentation versus conventional instrumentation in total knee arthroplasty: a prospective randomized controlled study on clinical outcomes and in-hospital data. Biomed Res Int 2015:165908CrossRefPubMedPubMedCentral Kotela A, Lorkowski J, Kucharzewski M, Wilkfrańczuk M, Śliwiński Z, Frańczuk B (2015) Patient-specific ct-based instrumentation versus conventional instrumentation in total knee arthroplasty: a prospective randomized controlled study on clinical outcomes and in-hospital data. Biomed Res Int 2015:165908CrossRefPubMedPubMedCentral
23.
go back to reference Kosse NM, Heesterbeek PJC, Schimmel JJP, Hellemondt GGV, Wymenga AB, Defoort KC (2018) Stability and alignment do not improve by using patient-specific instrumentation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 26(6):1–8CrossRef Kosse NM, Heesterbeek PJC, Schimmel JJP, Hellemondt GGV, Wymenga AB, Defoort KC (2018) Stability and alignment do not improve by using patient-specific instrumentation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 26(6):1–8CrossRef
24.
go back to reference León VJ, Lengua MA, Calvo V, Lisón AJ (2017) Use of patient-specific cutting blocks reduces blood loss after total knee arthroplasty. Eur J Orthop Surg Traumatol 27:273–277CrossRefPubMed León VJ, Lengua MA, Calvo V, Lisón AJ (2017) Use of patient-specific cutting blocks reduces blood loss after total knee arthroplasty. Eur J Orthop Surg Traumatol 27:273–277CrossRefPubMed
26.
go back to reference Marimuthu K, Chen DB, Harris IA, Wheatley E, Bryant CJ, Macdessi SJ (2014) A multi-planar ct-based comparative analysis of patient-specific cutting guides with conventional instrumentation in total knee arthroplasty. J Arthroplasty 29:1138–1142CrossRefPubMed Marimuthu K, Chen DB, Harris IA, Wheatley E, Bryant CJ, Macdessi SJ (2014) A multi-planar ct-based comparative analysis of patient-specific cutting guides with conventional instrumentation in total knee arthroplasty. J Arthroplasty 29:1138–1142CrossRefPubMed
27.
go back to reference Moubarak H, Brilhault J (2014) Contribution of patient-specific cutting guides to lower limb alignment for total knee arthroplasty. Orthop Traumatol Surg Res 100(4 Suppl):S239–S242CrossRefPubMed Moubarak H, Brilhault J (2014) Contribution of patient-specific cutting guides to lower limb alignment for total knee arthroplasty. Orthop Traumatol Surg Res 100(4 Suppl):S239–S242CrossRefPubMed
28.
go back to reference Molicnik A, Naranda J, Dolinar D (2015) Patient-matched instruments versus standard instrumentation in total knee arthroplasty: a prospective randomized study. Wien Klin Wochenschr 127(Suppl 5):S235–S240CrossRefPubMed Molicnik A, Naranda J, Dolinar D (2015) Patient-matched instruments versus standard instrumentation in total knee arthroplasty: a prospective randomized study. Wien Klin Wochenschr 127(Suppl 5):S235–S240CrossRefPubMed
29.
go back to reference MacDessi SJ, Jang B, Harris IA, Wheatley E, Bryant C, Chen DB (2014) A comparison of alignment using patient specific guides, computer navigation and conventional instrumentation in total knee arthroplasty. Knee 21:406–409CrossRefPubMed MacDessi SJ, Jang B, Harris IA, Wheatley E, Bryant C, Chen DB (2014) A comparison of alignment using patient specific guides, computer navigation and conventional instrumentation in total knee arthroplasty. Knee 21:406–409CrossRefPubMed
30.
go back to reference Mannan A, Smith TO, Sagar C, London NJ, Molitor PJ (2015) No demonstrable benefit for coronal alignment outcomes in psi knee arthroplasty: a systematic review and meta-analysis. Orthop Traumatol Surg Res 101(4):461–468CrossRefPubMed Mannan A, Smith TO, Sagar C, London NJ, Molitor PJ (2015) No demonstrable benefit for coronal alignment outcomes in psi knee arthroplasty: a systematic review and meta-analysis. Orthop Traumatol Surg Res 101(4):461–468CrossRefPubMed
31.
go back to reference Nam D, Park A, Stambough JB, Johnson SR, Nunley RM, Barrack RL (2016) The mark coventry award: custom cutting guides do not improve total knee arthroplasty clinical outcomes at 2 years followup. Clin Orthop Relat Res 474:40–46CrossRefPubMed Nam D, Park A, Stambough JB, Johnson SR, Nunley RM, Barrack RL (2016) The mark coventry award: custom cutting guides do not improve total knee arthroplasty clinical outcomes at 2 years followup. Clin Orthop Relat Res 474:40–46CrossRefPubMed
32.
go back to reference Nunley RM, Ellison BS, Ruh EL, Williams BM, Foreman K, Ford AD (2012) Are patient-specific cutting blocks cost-effective for total knee arthroplasty? Clin Orthop Relat Res 470:889–894CrossRefPubMed Nunley RM, Ellison BS, Ruh EL, Williams BM, Foreman K, Ford AD (2012) Are patient-specific cutting blocks cost-effective for total knee arthroplasty? Clin Orthop Relat Res 470:889–894CrossRefPubMed
33.
go back to reference Ng VY, Declaire JH, Berend KR, Gulick BC Jr, Lombardi A (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107CrossRefPubMed Ng VY, Declaire JH, Berend KR, Gulick BC Jr, Lombardi A (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107CrossRefPubMed
34.
go back to reference Nabavi A. Olwill CM (2015) Early outcome after total knee replacement using computed tomography-based patient specific cutting blocks versus standard instrumentation. J Orthop Surg 23:182–184CrossRef Nabavi A. Olwill CM (2015) Early outcome after total knee replacement using computed tomography-based patient specific cutting blocks versus standard instrumentation. J Orthop Surg 23:182–184CrossRef
35.
go back to reference Nam D, Mcarthur BA, Cross MB, Pearle AD, Mayman DJ, Haas SB (2012) Patient-specific instrumentation in total knee arthroplasty: a review. J Knee Surg 25:213–219CrossRefPubMed Nam D, Mcarthur BA, Cross MB, Pearle AD, Mayman DJ, Haas SB (2012) Patient-specific instrumentation in total knee arthroplasty: a review. J Knee Surg 25:213–219CrossRefPubMed
36.
go back to reference Oremus M, Wolfson C, Perrault A, Demers L, Momoli F, Moride Y (2001) Interrater reliability of the modified jadad quality scale for systematic reviews of alzheimer’s disease drug trials. Dement Geriatr Cogn Disord 12(3):232–236CrossRefPubMed Oremus M, Wolfson C, Perrault A, Demers L, Momoli F, Moride Y (2001) Interrater reliability of the modified jadad quality scale for systematic reviews of alzheimer’s disease drug trials. Dement Geriatr Cogn Disord 12(3):232–236CrossRefPubMed
37.
go back to reference Olivo SA, Macedo LG, Gadotti IC, Fuentes J, Stanton T, Magee DJ (2008) Scales to assess the quality of randomized controlled trials: a systematic review. Phys Ther 88(2):156–175CrossRefPubMed Olivo SA, Macedo LG, Gadotti IC, Fuentes J, Stanton T, Magee DJ (2008) Scales to assess the quality of randomized controlled trials: a systematic review. Phys Ther 88(2):156–175CrossRefPubMed
38.
go back to reference Parratte S, Blanc G, Boussemart T, Ollivier M, Le Corroller T, Argenson JN (2013) Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation. Knee Surg Sports Traumatol Arthrosc 21:2213–2219CrossRefPubMed Parratte S, Blanc G, Boussemart T, Ollivier M, Le Corroller T, Argenson JN (2013) Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation. Knee Surg Sports Traumatol Arthrosc 21:2213–2219CrossRefPubMed
39.
go back to reference Pietsch M, Djahani O, Zweiger C, Plattner F, Radl R, Tschauner C, Hofmann S (2013) Custom-fit minimally invasive total knee arthroplasty: effect on blood loss and early clinical outcomes. Knee Surg Sports Traumatol Arthrosc 21:2234–2240CrossRefPubMed Pietsch M, Djahani O, Zweiger C, Plattner F, Radl R, Tschauner C, Hofmann S (2013) Custom-fit minimally invasive total knee arthroplasty: effect on blood loss and early clinical outcomes. Knee Surg Sports Traumatol Arthrosc 21:2234–2240CrossRefPubMed
40.
go back to reference Pourgiezis N, Reddy SP, Nankivell M, Morrison G, VanEssen J (2016) Alignment and component position after patient-matched instrumentation versus conventional total knee arthroplasty. J Orthop Surg 24:170–174CrossRef Pourgiezis N, Reddy SP, Nankivell M, Morrison G, VanEssen J (2016) Alignment and component position after patient-matched instrumentation versus conventional total knee arthroplasty. J Orthop Surg 24:170–174CrossRef
41.
go back to reference Predescu V, Prescura C, Olaru R, Savin L, Botez P, Deleanu B (2017) Patient specific instrumentation versus conventional knee arthroplasty: comparative study. Int Orthop 41:1361–1367CrossRefPubMed Predescu V, Prescura C, Olaru R, Savin L, Botez P, Deleanu B (2017) Patient specific instrumentation versus conventional knee arthroplasty: comparative study. Int Orthop 41:1361–1367CrossRefPubMed
42.
go back to reference Qing-Meng Z, Ji-Ying C, Heng L, Wei C, Ming N, Zhen-Dong Z (2015) No evidence of superiority in reducing outliers of component alignment for patient-specific instrumentation for total knee arthroplasty: a systematic review. Orthop Surg 7(1):19–25CrossRef Qing-Meng Z, Ji-Ying C, Heng L, Wei C, Ming N, Zhen-Dong Z (2015) No evidence of superiority in reducing outliers of component alignment for patient-specific instrumentation for total knee arthroplasty: a systematic review. Orthop Surg 7(1):19–25CrossRef
43.
go back to reference Roh YW, Kim TW, Lee S, Seong SC, Lee MC (2013) Is TKA using patient-specific instruments comparable to conventional TKA? A randomized controlled study of one system. Clin Orthop Relat Res 471:3988–3995CrossRefPubMedPubMedCentral Roh YW, Kim TW, Lee S, Seong SC, Lee MC (2013) Is TKA using patient-specific instruments comparable to conventional TKA? A randomized controlled study of one system. Clin Orthop Relat Res 471:3988–3995CrossRefPubMedPubMedCentral
44.
go back to reference Renson L, Poilvache P, Wyngaert HVD (2014) Improved alignment and operating room efficiency with patient-specific instrumentation for TKA. Knee 21:1216–1220CrossRefPubMed Renson L, Poilvache P, Wyngaert HVD (2014) Improved alignment and operating room efficiency with patient-specific instrumentation for TKA. Knee 21:1216–1220CrossRefPubMed
45.
go back to reference Russell R, Brown T, Huo M, Jones R (2014) Patient-specific instrumentation does not improve alignment in total knee arthroplasty. J Knee Surg 27(06):501–504CrossRefPubMed Russell R, Brown T, Huo M, Jones R (2014) Patient-specific instrumentation does not improve alignment in total knee arthroplasty. J Knee Surg 27(06):501–504CrossRefPubMed
46.
go back to reference Ritter MA, Herbst SA, Keating EM, Faris PM, Meding JB (1994) Long-term survival analysis of a posterior cruciate-retaining total condylar total knee arthroplasty. Clin Orthop Relat Res 309:136–145 Ritter MA, Herbst SA, Keating EM, Faris PM, Meding JB (1994) Long-term survival analysis of a posterior cruciate-retaining total condylar total knee arthroplasty. Clin Orthop Relat Res 309:136–145
48.
go back to reference Silva A, Sampaio R, Pinto E (2014) Patient-specific instrumentation improves tibial component rotation in TKA. Knee Surg Sports Traumatol Arthrosc 22:636–642CrossRefPubMed Silva A, Sampaio R, Pinto E (2014) Patient-specific instrumentation improves tibial component rotation in TKA. Knee Surg Sports Traumatol Arthrosc 22:636–642CrossRefPubMed
49.
go back to reference Stronach BM, Pelt CE, Erickson JA, Peters CL (2014) Patient-specific instrumentation in total knee arthroplasty provides no improvement in component alignment. J Arthroplasty 29:1705–1708CrossRefPubMed Stronach BM, Pelt CE, Erickson JA, Peters CL (2014) Patient-specific instrumentation in total knee arthroplasty provides no improvement in component alignment. J Arthroplasty 29:1705–1708CrossRefPubMed
50.
go back to reference Steinhaus ME, McLawhorn AS, Richardson SS, Maher P, Mayman DJ (2016) Handheld navigation device and patient-specific cutting guides result in similar coronal alignment for primary total knee arthroplasty: a retrospective matched cohort study. HSS J 12:224–234CrossRefPubMedPubMedCentral Steinhaus ME, McLawhorn AS, Richardson SS, Maher P, Mayman DJ (2016) Handheld navigation device and patient-specific cutting guides result in similar coronal alignment for primary total knee arthroplasty: a retrospective matched cohort study. HSS J 12:224–234CrossRefPubMedPubMedCentral
51.
go back to reference Sharareh B, Schwarzkopf R (2015) Review article: patient-specific versus standard instrumentation for total knee arthroplasty. J Orthop Surg 23(1):100CrossRef Sharareh B, Schwarzkopf R (2015) Review article: patient-specific versus standard instrumentation for total knee arthroplasty. J Orthop Surg 23(1):100CrossRef
52.
go back to reference Shen C, Tang ZH, Hu JZ, Zou GY, Xiao RC, Yan DX (2015) Patient-specific instrumentation does not improve accuracy in total knee arthroplasty. Orthopedics 38(3):178–188CrossRef Shen C, Tang ZH, Hu JZ, Zou GY, Xiao RC, Yan DX (2015) Patient-specific instrumentation does not improve accuracy in total knee arthroplasty. Orthopedics 38(3):178–188CrossRef
53.
go back to reference Thienpont E, Schwab PE, Fennema P (2017) Efficacy of patient-specific instruments in total knee arthroplasty: a systematic review and meta-analysis. J Bone Jt Surg Am 99(6):521CrossRef Thienpont E, Schwab PE, Fennema P (2017) Efficacy of patient-specific instruments in total knee arthroplasty: a systematic review and meta-analysis. J Bone Jt Surg Am 99(6):521CrossRef
54.
go back to reference Thienpont E, Schwab PE, Fennema P (2014) A systematic review and meta-analysis of patient-specific instrumentation for improving alignment of the components in total knee replacement. Bone Jt J 410(6825):1052–1061CrossRef Thienpont E, Schwab PE, Fennema P (2014) A systematic review and meta-analysis of patient-specific instrumentation for improving alignment of the components in total knee replacement. Bone Jt J 410(6825):1052–1061CrossRef
55.
go back to reference Vundelinckx BJ, Bruckers L, De Mulder K, De Schepper J, Van Esbroeck G (2013) Functional and radiographic short-term outcome evaluation of the visionaire system, a patient-matched instrumentation system for total knee arthroplasty. J Arthroplasty 28:964–970CrossRefPubMed Vundelinckx BJ, Bruckers L, De Mulder K, De Schepper J, Van Esbroeck G (2013) Functional and radiographic short-term outcome evaluation of the visionaire system, a patient-matched instrumentation system for total knee arthroplasty. J Arthroplasty 28:964–970CrossRefPubMed
56.
go back to reference Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty: a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty: a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed
57.
go back to reference Vide J, Freitas TP, Ramos A, Cruz H, Sousa JP (2017) Patient-specific instrumentation in total knee arthroplasty: simpler, faster and more accurate than standard instrumentation—a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 25:2616–2621CrossRefPubMed Vide J, Freitas TP, Ramos A, Cruz H, Sousa JP (2017) Patient-specific instrumentation in total knee arthroplasty: simpler, faster and more accurate than standard instrumentation—a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 25:2616–2621CrossRefPubMed
58.
go back to reference Van JL, Snorrason F, Röhrl SM (2018) No radiological and clinical advantages with patient-specific positioning guides in total knee replacement. Acta Orthop 89:89–94CrossRef Van JL, Snorrason F, Röhrl SM (2018) No radiological and clinical advantages with patient-specific positioning guides in total knee replacement. Acta Orthop 89:89–94CrossRef
59.
go back to reference Vaishya R, Vijay V, Birla VP, Agarwal AK (2016) Computerized tomography based “patient specific blocks” improve postoperative mechanical alignment in primary total knee arthroplasty. World J Orthop 7:426–433CrossRefPubMedPubMedCentral Vaishya R, Vijay V, Birla VP, Agarwal AK (2016) Computerized tomography based “patient specific blocks” improve postoperative mechanical alignment in primary total knee arthroplasty. World J Orthop 7:426–433CrossRefPubMedPubMedCentral
60.
go back to reference Voleti PB, Hamula MJ, Baldwin KD, Lee GC (2014) Current data do not support routine use of patient-specific instrumentation in total knee arthroplasty. J Arthroplasty 29(9):1709–1712CrossRefPubMed Voleti PB, Hamula MJ, Baldwin KD, Lee GC (2014) Current data do not support routine use of patient-specific instrumentation in total knee arthroplasty. J Arthroplasty 29(9):1709–1712CrossRefPubMed
61.
go back to reference Woolson ST, Harris AH, Wagner DW, Giori NJ (2014) Component alignment during total knee arthroplasty with use of standard or custom instrumentation: a randomized clinical trial using computed tomography for postoperative alignment measurement. J Bone Jt Surg Am 96:366–372CrossRef Woolson ST, Harris AH, Wagner DW, Giori NJ (2014) Component alignment during total knee arthroplasty with use of standard or custom instrumentation: a randomized clinical trial using computed tomography for postoperative alignment measurement. J Bone Jt Surg Am 96:366–372CrossRef
62.
go back to reference Wan X, Wang W, Liu J, Tong T (2014) Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol 14(1):135CrossRefPubMedPubMedCentral Wan X, Wang W, Liu J, Tong T (2014) Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol 14(1):135CrossRefPubMedPubMedCentral
63.
go back to reference Wu PL, Lee M, Huang TT (2017) Effectiveness of physical activity on patients with depression and parkinson’s disease: a systematic review. PLoS One 12(7):e0181515CrossRefPubMedPubMedCentral Wu PL, Lee M, Huang TT (2017) Effectiveness of physical activity on patients with depression and parkinson’s disease: a systematic review. PLoS One 12(7):e0181515CrossRefPubMedPubMedCentral
64.
go back to reference Yan CH, Chiu KY, Ng FY, Chan PK, Fang CX (2015) Comparison between patient-specific instruments and conventional instruments and computer navigation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 23:3637–3645CrossRefPubMed Yan CH, Chiu KY, Ng FY, Chan PK, Fang CX (2015) Comparison between patient-specific instruments and conventional instruments and computer navigation in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 23:3637–3645CrossRefPubMed
65.
go back to reference Zhang QM1, Chen JY, Li H, Chai W, Ni M, Zhang ZD, Yang F (2015) No evidence of superiority in reducing outliers of component alignment for patient-specific instrumentation for total knee arthroplasty: a systematic review. Orthop Surg 7:19–25CrossRefPubMed Zhang QM1, Chen JY, Li H, Chai W, Ni M, Zhang ZD, Yang F (2015) No evidence of superiority in reducing outliers of component alignment for patient-specific instrumentation for total knee arthroplasty: a systematic review. Orthop Surg 7:19–25CrossRefPubMed
Metadata
Title
Patient-specific instrumentation improved axial alignment of the femoral component, operative time and perioperative blood loss after total knee arthroplasty
Publication date
01-04-2019
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 4/2019
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-018-5256-0

Other articles of this Issue 4/2019

Knee Surgery, Sports Traumatology, Arthroscopy 4/2019 Go to the issue