3D patient imaging and retrieval analysis help understand the clinical importance of rotation in knee replacements

Authors: Arianna Cerquiglini, Johann Henckel, Harry Hothi, Niccoló Rotigliano, Michael T. Hirschmann, Alister J. Hart

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 11/2018

Login to get access

Abstract

Purpose

The purpose of the present study was to correlate highly accurate CT measurements of pre-revision total knee arthroplasty (TKA) implant position with findings of retrieval analysis post-revision, to understand the clinical relevance of TKA orientation.

Methods

This study involved 53 retrieved TKA implants with pre-revision 3D-CT scans used to determine coronal (varus–valgus), sagittal (tibial slope) and rotational (internal rotation–external rotation) TKA orientation as well as tibiofemoral leg axis. Differences between femoral and tibial angles to describe the "relative rotational mismatch" were also calculated. All tibial inserts were forensically analyzed using the Hood score. Statistical analysis was performed to investigate correlations between TKA component orientation and surface damage (p < 0.05).

Results

Femoral components were found to have axial rotations mainly within ± 3° (68%), whilst 45% of the tibial components and 66% of the relative rotational mismatches were > 3° and < − 3°, respectively. The majority of femoral and tibial components (87% in both cases), as well as the femorotibial angle (70%), showed coronal orientations within ± 3°. The 64% of the tibial components showed posterior tibial slopes out of both the 0°–3° and 5°–7° ranges. There was a significant correlation between tibial slope and damage score on polyethylene tibial inserts (r = 0.2856; p = 0.0382) as well as a significant correlation between implants’ position in the axial plane and damage score on polyethylene tibial inserts (r = 0.6537, p = 0.0240).

Conclusions

This is the first study to use accurate measurements from pre-revision 3DCT to compare tibial and femoral orientation in all three planes with retrieval findings in total knee replacements. A significant correlation between implant position and polyethylene surface damage was found. These results showed the importance of optimizing component position to minimize polyethylene damage. Further analysis involving more accurate polyethylene wear measurements are fundamental to fully understand the role of components’ orientation in TKAs.
Literature
2.
go back to reference Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4):780–785 Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4):780–785
4.
go back to reference Robertsson O, Bizjajeva S, Fenstad AM et al (2010) Knee arthroplasty in Denmark, Norway and Sweden: a pilot study from the Nordic Arthroplasty Register Association. Acta Orthop 81(1):82–89CrossRef Robertsson O, Bizjajeva S, Fenstad AM et al (2010) Knee arthroplasty in Denmark, Norway and Sweden: a pilot study from the Nordic Arthroplasty Register Association. Acta Orthop 81(1):82–89CrossRef
5.
go back to reference Callaghan JJ, O’roukre MR, Saleh K (2004) Why knees fail: lessons learned. J Arthroplasty 19(4 Suppl 1):31–34CrossRef Callaghan JJ, O’roukre MR, Saleh K (2004) Why knees fail: lessons learned. J Arthroplasty 19(4 Suppl 1):31–34CrossRef
6.
go back to reference Dalury DF, Pomeroy DL, Gorab RS, Adams MJ (2013) Why are total knee arthroplasties being revised? J Arthroplasty 28(8 Suppl):120–121CrossRef Dalury DF, Pomeroy DL, Gorab RS, Adams MJ (2013) Why are total knee arthroplasties being revised? J Arthroplasty 28(8 Suppl):120–121CrossRef
7.
go back to reference Hirschmann MT, Becker R (2015) The unhappy knee replacement. A comprehensive review and management guide. Roland Becker Editors. Springer, New York Hirschmann MT, Becker R (2015) The unhappy knee replacement. A comprehensive review and management guide. Roland Becker Editors. Springer, New York
8.
go back to reference Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components. J Bone Joint Surg Br 93(5):629–633CrossRef Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components. J Bone Joint Surg Br 93(5):629–633CrossRef
9.
go back to reference Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME, Ritter MA, Nunley RM (2013) Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplasty 28(8 Suppl):116–119CrossRef Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME, Ritter MA, Nunley RM (2013) Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplasty 28(8 Suppl):116–119CrossRef
10.
go back to reference Armstrong AD, Brien HJ, Dunning CE, King GJ, Johnson JA, Chess DG (2003) Patellar position after total knee arthroplasty: influence of femoral component malposition. J Arthroplasty 18(4):458–465CrossRef Armstrong AD, Brien HJ, Dunning CE, King GJ, Johnson JA, Chess DG (2003) Patellar position after total knee arthroplasty: influence of femoral component malposition. J Arthroplasty 18(4):458–465CrossRef
11.
go back to reference Berger RA, Crossett LS, Jacobs JJ, Rubash HE (1998) Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. (356):144–153CrossRef Berger RA, Crossett LS, Jacobs JJ, Rubash HE (1998) Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. (356):144–153CrossRef
12.
go back to reference Nicoll D, Rowley DI (2010) Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg Br 92(9):1238–4124CrossRef Nicoll D, Rowley DI (2010) Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg Br 92(9):1238–4124CrossRef
13.
go back to reference Slevin O, Amsler F, Hirschmann MT (2016) No correlation between coronal alignment of total knee arthroplasty and clinical outcomes: a prospective clinical study using 3D-CT. Knee Surg Sports Traumatol Arthrosc 25(12):3892–3900CrossRef Slevin O, Amsler F, Hirschmann MT (2016) No correlation between coronal alignment of total knee arthroplasty and clinical outcomes: a prospective clinical study using 3D-CT. Knee Surg Sports Traumatol Arthrosc 25(12):3892–3900CrossRef
14.
15.
go back to reference Slevin O, Schmid FA, Schiapparelli FF, Rasch H, Amsler F, Hirschmann MT (2017) Coronal femoral TKA position significantly influences in vivo patellar loading in unresurfaced patellae after primary total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 25(11):3605–3610CrossRef Slevin O, Schmid FA, Schiapparelli FF, Rasch H, Amsler F, Hirschmann MT (2017) Coronal femoral TKA position significantly influences in vivo patellar loading in unresurfaced patellae after primary total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 25(11):3605–3610CrossRef
16.
go back to reference Chambers AW, Wood AR, Kosmopoulos V, Sanchez HB, Wagner RA (2016) Effect of posterior tibial slope on flexion and anterior-posterior tibial translation in posterior cruciate-retaining total knee arthroplasty. J Arthroplasty 31(1):103–106CrossRef Chambers AW, Wood AR, Kosmopoulos V, Sanchez HB, Wagner RA (2016) Effect of posterior tibial slope on flexion and anterior-posterior tibial translation in posterior cruciate-retaining total knee arthroplasty. J Arthroplasty 31(1):103–106CrossRef
17.
go back to reference Di Laura A, Whittaker R, Hothi H, Kwon YM, Skinner J, Hart A (2016) Dual mobility vs standard bearings: what is the effect on taper junction? Bone Joint J 98(sup 11):3 Di Laura A, Whittaker R, Hothi H, Kwon YM, Skinner J, Hart A (2016) Dual mobility vs standard bearings: what is the effect on taper junction? Bone Joint J 98(sup 11):3
18.
go back to reference Wasielewski RC, Galante JO, Leighty RM, Roseneberg AG (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical consideration during total knee arthroplasty. Clin Orthop Relat Res 299:31–43 Wasielewski RC, Galante JO, Leighty RM, Roseneberg AG (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical consideration during total knee arthroplasty. Clin Orthop Relat Res 299:31–43
19.
go back to reference Metthies AK, Henckel J, Skinner JA, Hart AJ (2011) A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties. Hip Int 21(06):724–731CrossRef Metthies AK, Henckel J, Skinner JA, Hart AJ (2011) A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties. Hip Int 21(06):724–731CrossRef
20.
go back to reference Currier JH, Bill MA, Mayor MB (2005) Analysis of wear asymmetry in a series of 94 retrieved polyethylene tibial bearings. J Biomech 38(2):367–375CrossRef Currier JH, Bill MA, Mayor MB (2005) Analysis of wear asymmetry in a series of 94 retrieved polyethylene tibial bearings. J Biomech 38(2):367–375CrossRef
21.
go back to reference Flannery M, Jones E, Birkinshaw C (2010) Compliant layer knee bearings. Part II: preliminary wear assessment. Wear 269(5–6):331–338CrossRef Flannery M, Jones E, Birkinshaw C (2010) Compliant layer knee bearings. Part II: preliminary wear assessment. Wear 269(5–6):331–338CrossRef
22.
go back to reference Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol Avon) 17(2):140–146CrossRef Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol Avon) 17(2):140–146CrossRef
23.
go back to reference Lokte PA, Ecker ML (1977) Influence of positioning of Prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRef Lokte PA, Ecker ML (1977) Influence of positioning of Prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRef
24.
go back to reference Pang HN, Bin Abd Razak HR, Jamiecson P, Teeter MG, Naudie DD, MacDonald SJ (2016) Factors affecting wear of constrained polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty 31(6):1340–1345CrossRef Pang HN, Bin Abd Razak HR, Jamiecson P, Teeter MG, Naudie DD, MacDonald SJ (2016) Factors affecting wear of constrained polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty 31(6):1340–1345CrossRef
25.
go back to reference Srivastava A, Lee GY, Steklov N, Colwell CW Jr, Ezzet KA, D’Lima DD (2012) Effect of tibial component varus on wear in total knee arthroplasty. Knee 19(5):560–563CrossRef Srivastava A, Lee GY, Steklov N, Colwell CW Jr, Ezzet KA, D’Lima DD (2012) Effect of tibial component varus on wear in total knee arthroplasty. Knee 19(5):560–563CrossRef
26.
go back to reference Henckel J, Richards R, Lozhkin K, Harris S, Rodriguez y Baena FM, Barrett ARW, Cobb JP (2006) Very low-dose computed tomography for planning and outcome measurement in knee replacement. The imperial knee protocol. J Bone Joint Surg 88-B:1513–1518CrossRef Henckel J, Richards R, Lozhkin K, Harris S, Rodriguez y Baena FM, Barrett ARW, Cobb JP (2006) Very low-dose computed tomography for planning and outcome measurement in knee replacement. The imperial knee protocol. J Bone Joint Surg 88-B:1513–1518CrossRef
27.
go back to reference Harman MK, Banks SA, Kirschner S, Lützner J (2012) Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations. BMC Musculoskelet Disord 13:206CrossRef Harman MK, Banks SA, Kirschner S, Lützner J (2012) Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations. BMC Musculoskelet Disord 13:206CrossRef
28.
go back to reference Gromov K, Korchi M, Thomsen MG, Husted H, Troelsen A (2014) What is the optimal alignment of the tibial and femoral components in knee arthroplasty? An overview of the literature. Acta Orthop 85(5):480–487CrossRef Gromov K, Korchi M, Thomsen MG, Husted H, Troelsen A (2014) What is the optimal alignment of the tibial and femoral components in knee arthroplasty? An overview of the literature. Acta Orthop 85(5):480–487CrossRef
29.
go back to reference Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73(5):709–714CrossRef Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73(5):709–714CrossRef
30.
go back to reference Hood RW, Wright TM, Burstein AH (1983) Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res 17:829–842CrossRef Hood RW, Wright TM, Burstein AH (1983) Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res 17:829–842CrossRef
31.
go back to reference Ferrara F, Cipriani A, Rapisarda S, Iacobucci M, Magarelli N, Leone A, Bonomo L (2016) Assessment of implant position after total knee arthroplasty by dual-energy computed tomography. Acta Radiol 57(5):612–619CrossRef Ferrara F, Cipriani A, Rapisarda S, Iacobucci M, Magarelli N, Leone A, Bonomo L (2016) Assessment of implant position after total knee arthroplasty by dual-energy computed tomography. Acta Radiol 57(5):612–619CrossRef
32.
go back to reference Yau WP, Chiu KY, Tang WM (2007) How precise is the determination of rotational alignment of the femoral prosthesis in total knee arthroplasty: an in vivo study. J Arthroplasty 22(7):1042–1048CrossRef Yau WP, Chiu KY, Tang WM (2007) How precise is the determination of rotational alignment of the femoral prosthesis in total knee arthroplasty: an in vivo study. J Arthroplasty 22(7):1042–1048CrossRef
33.
go back to reference Knowlton CB, Bhutani P, Wimmer MA (2016) Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners. J Biomed Mater Res Part B 105(7):2053–2059CrossRef Knowlton CB, Bhutani P, Wimmer MA (2016) Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners. J Biomed Mater Res Part B 105(7):2053–2059CrossRef
34.
go back to reference Gilbert SL, Rana AJ, Lipman JD, Wright GH, Westrich GH (2014) Design changes improve contact patterns and articular surface damage in total knee arthroplasty. Knee 21(6):1129–1134CrossRef Gilbert SL, Rana AJ, Lipman JD, Wright GH, Westrich GH (2014) Design changes improve contact patterns and articular surface damage in total knee arthroplasty. Knee 21(6):1129–1134CrossRef
35.
go back to reference McClung CD, Zahiri CA, Higa JK, Amstutz HC, Schmalzried TP (2000) Relationship between body mass index and activity in hip or knee arthroplasty patients. J Orthop Res 18(1):35–39CrossRef McClung CD, Zahiri CA, Higa JK, Amstutz HC, Schmalzried TP (2000) Relationship between body mass index and activity in hip or knee arthroplasty patients. J Orthop Res 18(1):35–39CrossRef
Metadata
Title
3D patient imaging and retrieval analysis help understand the clinical importance of rotation in knee replacements
Authors
Arianna Cerquiglini
Johann Henckel
Harry Hothi
Niccoló Rotigliano
Michael T. Hirschmann
Alister J. Hart
Publication date
01-11-2018
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 11/2018
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-018-4891-9

Other articles of this Issue 11/2018

Knee Surgery, Sports Traumatology, Arthroscopy 11/2018 Go to the issue