High wear resistance of femoral components coated with titanium nitride: a retrieval analysis

Authors: Christian Fabry, Carmen Zietz, Axel Baumann, Reinhard Ehall, Rainer Bader

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 9/2018

Login to get access

Abstract

Purpose

The objective of this study was to evaluate the in vivo wear resistance of cobalt-chromium femoral components coated with titanium nitride (TiN). Our null hypothesis was that the surface damage and the thickness of the TiN coating do not correlate with the time in vivo.

Methods

Twenty-five TiN-coated bicondylar femoral retrievals with a mean implantation period of 30.7 ± 11.7 months were subjected to an objective surface damage analysis with a semi-quantitative assessment method. A visual examination of scratches, indentations, notches and coating breakthroughs of the surfaces was performed. The roughness and the coating thickness of the TiN coating were evaluated in the main articulation regions.

Results

Narrow scratches and indentations in the range of low flexion angles on the retrieval surfaces were the most common modes of damage. There was no evidence of delamination on the articulation surface but rather at the bottom of isolated severe indentations or notches. An analysis of three retrievals revealed a coating breakthrough in the patellofemoral joint region, resulting from patella maltracking and a dislocation. The arithmetical mean roughness of the TiN surface slightly increased with the implantation period. In contrast, the maximum peak height of the roughness profile was reduced at the condyles of the retrieved components in comparison with new, unused surfaces. No significant association between the coating thickness and implantation period was determined. Moreover, the measured values were retained in the range of the initial coating thickness even after several years of in vivo service.

Conclusions

As was demonstrated by the results of this study, the surface damage to the TiN coating did not deteriorate with the implantation period. The calculated damage scores and the measured coating thickness in particular both confirmed that the TiN coating provides low wear rates. Our findings support the use of wear-resistant TiN-coated components in total knee arthroplasty with the objective of reducing the risk of aseptic loosening. However, in terms of TiN-coated femoral components, particular attention should be paid to a correct patellar tracking in order to avoid wear propagation at the implant.
Literature
2.
go back to reference Arnholt CM, MacDonald DW, Malkani AL et al (2016) Corrosion damage and wear mechanisms in long-term retrieved CoCr femoral components for total knee arthroplasty. J Arthroplasty 31(12):2900–2906CrossRefPubMedPubMedCentral Arnholt CM, MacDonald DW, Malkani AL et al (2016) Corrosion damage and wear mechanisms in long-term retrieved CoCr femoral components for total knee arthroplasty. J Arthroplasty 31(12):2900–2906CrossRefPubMedPubMedCentral
3.
go back to reference Brandt JM, Guenther L, O’Brien S et al (2013) Performance assessment of femoral knee components made from cobalt-chromium alloy and oxidized zirconium. Knee 20(6):388–396CrossRefPubMed Brandt JM, Guenther L, O’Brien S et al (2013) Performance assessment of femoral knee components made from cobalt-chromium alloy and oxidized zirconium. Knee 20(6):388–396CrossRefPubMed
4.
go back to reference Burnell CD, Brandt JM, Petrak MJ et al (2011) Posterior condyle surface damage on retrieved femoral knee components. J Arthroplasty 26(8):1460–1467CrossRefPubMed Burnell CD, Brandt JM, Petrak MJ et al (2011) Posterior condyle surface damage on retrieved femoral knee components. J Arthroplasty 26(8):1460–1467CrossRefPubMed
5.
go back to reference Cooper J, Dowson D, Fisher J (1993) The effect of transfer film and surface roughness on the wear of lubricated ultra-high molecular weight polyethylene wear. Clin Mater 14:295–302CrossRef Cooper J, Dowson D, Fisher J (1993) The effect of transfer film and surface roughness on the wear of lubricated ultra-high molecular weight polyethylene wear. Clin Mater 14:295–302CrossRef
6.
go back to reference DeBaets T, Waelput W, Bellemans J (2008) Analysis of third body particles generated during total knee arthroplasty: is metal debris an issue? Knee 15:95–97CrossRef DeBaets T, Waelput W, Bellemans J (2008) Analysis of third body particles generated during total knee arthroplasty: is metal debris an issue? Knee 15:95–97CrossRef
7.
go back to reference DIN EN 1071-2:2003-02 (2003) Advanced technical ceramics—methods of test for ceramic coatings—part 2: determination of coating thickness by the crater grinding method DIN EN 1071-2:2003-02 (2003) Advanced technical ceramics—methods of test for ceramic coatings—part 2: determination of coating thickness by the crater grinding method
9.
go back to reference Eisenhuth SA, Saleh KJ, Cui Q et al (2006) Patellofemoral instability after total knee arthroplasty. Clin Orthop Relat Res 446:149–160CrossRefPubMed Eisenhuth SA, Saleh KJ, Cui Q et al (2006) Patellofemoral instability after total knee arthroplasty. Clin Orthop Relat Res 446:149–160CrossRefPubMed
10.
go back to reference Fabry C, Zietz C, Dammer R et al (2015) Patterns of wear in total knee arthroplasty. In: Hirschmann MT, Becker R (eds) The unhappy total knee arthroplasty: a comprehensive review and management guide. Springer, Heidelberg, pp 135–146CrossRef Fabry C, Zietz C, Dammer R et al (2015) Patterns of wear in total knee arthroplasty. In: Hirschmann MT, Becker R (eds) The unhappy total knee arthroplasty: a comprehensive review and management guide. Springer, Heidelberg, pp 135–146CrossRef
11.
go back to reference Halder A, Kutzner I, Graichen F et al (2012) Influence of limb alignment on mediolateral loading in total knee replacement: in vivo measurements in five patients. J Bone Joint Surg Am 94:1023–1029CrossRefPubMed Halder A, Kutzner I, Graichen F et al (2012) Influence of limb alignment on mediolateral loading in total knee replacement: in vivo measurements in five patients. J Bone Joint Surg Am 94:1023–1029CrossRefPubMed
12.
go back to reference Harman MK, Banks SA, Hodge WA (1997) Wear analysis of a retrieved hip implant with titanium nitride coating. J Arthroplasty 12:938–945CrossRefPubMed Harman MK, Banks SA, Hodge WA (1997) Wear analysis of a retrieved hip implant with titanium nitride coating. J Arthroplasty 12:938–945CrossRefPubMed
13.
go back to reference Heyse TJ, Davis J, Haas SB et al (2011) Retrieval analysis of femoral zirconium components in total knee arthroplasty: preliminary results. J Arthroplasty 26(3):445–450CrossRefPubMed Heyse TJ, Davis J, Haas SB et al (2011) Retrieval analysis of femoral zirconium components in total knee arthroplasty: preliminary results. J Arthroplasty 26(3):445–450CrossRefPubMed
14.
go back to reference Heyse TJ, Elpers ME, Nawabi DH et al (2014) Oxidized zirconium versus cobalt-chromium in TKA: profilometry of retrieved femoral components. Clin Orthop Relat Res 472(1):277–283CrossRefPubMed Heyse TJ, Elpers ME, Nawabi DH et al (2014) Oxidized zirconium versus cobalt-chromium in TKA: profilometry of retrieved femoral components. Clin Orthop Relat Res 472(1):277–283CrossRefPubMed
15.
go back to reference Holleyman RJ, Scholes SC, Weir D et al (2015) Changes in surface topography at the TKA backside articulation following in vivo service: a retrieval analysis. Knee Surg Sports Traumatol Arthrosc 23(12):3523–3531CrossRefPubMed Holleyman RJ, Scholes SC, Weir D et al (2015) Changes in surface topography at the TKA backside articulation following in vivo service: a retrieval analysis. Knee Surg Sports Traumatol Arthrosc 23(12):3523–3531CrossRefPubMed
16.
go back to reference Lützner C, Kirschner S, Lützner J (2014) Patient activity after TKA depends on patient-specific parameters. Clin Orthop Relat Res 472(12):3933–3940CrossRefPubMedPubMedCentral Lützner C, Kirschner S, Lützner J (2014) Patient activity after TKA depends on patient-specific parameters. Clin Orthop Relat Res 472(12):3933–3940CrossRefPubMedPubMedCentral
17.
go back to reference Malikian R, Maruthainar K, Stammers J et al (2014) In vivo roughening of retrieved total knee arthroplasty femoral components. Knee 21(1):278–282CrossRefPubMed Malikian R, Maruthainar K, Stammers J et al (2014) In vivo roughening of retrieved total knee arthroplasty femoral components. Knee 21(1):278–282CrossRefPubMed
18.
go back to reference Malo M, Vince KG (2003) The unstable patella after total knee arthroplasty: etiology, prevention, and management. J Am Acad Orthop Surg 11(5):364–371CrossRefPubMed Malo M, Vince KG (2003) The unstable patella after total knee arthroplasty: etiology, prevention, and management. J Am Acad Orthop Surg 11(5):364–371CrossRefPubMed
19.
go back to reference Mohammed A, Metcalfe A, Woodnutt D (2014) Medium-term outcome of titanium nitride, mobile bearing total knee replacement. Acta Orthop Belg 80(2):269–275PubMed Mohammed A, Metcalfe A, Woodnutt D (2014) Medium-term outcome of titanium nitride, mobile bearing total knee replacement. Acta Orthop Belg 80(2):269–275PubMed
20.
go back to reference Morrison ML, Jani S, Parikh A (2015) Design of an advanced bearing system for total knee arthroplasty. Lubricants 3:475–492CrossRef Morrison ML, Jani S, Parikh A (2015) Design of an advanced bearing system for total knee arthroplasty. Lubricants 3:475–492CrossRef
21.
go back to reference Muratoglu OK, Burroughs BR, Bragdon CR et al (2004) Knee simulator wear of polyethylene tibias articulating against explanted rough femoral components. Clin Orthop Relat Res 428:108–113CrossRef Muratoglu OK, Burroughs BR, Bragdon CR et al (2004) Knee simulator wear of polyethylene tibias articulating against explanted rough femoral components. Clin Orthop Relat Res 428:108–113CrossRef
22.
go back to reference Raimondi MT, Pietrabissa R (2000) The in vivo wear performance of prosthetic femoral heads with titanium nitride coating. Biomaterials 21:907–913CrossRefPubMed Raimondi MT, Pietrabissa R (2000) The in vivo wear performance of prosthetic femoral heads with titanium nitride coating. Biomaterials 21:907–913CrossRefPubMed
24.
go back to reference Scholes SC, Kennard E, Gangadharan R et al (2013) Topographical analysis of the femoral components of ex vivo total knee replacements. J Mater Sci Mater Med 24(2):547–554CrossRefPubMed Scholes SC, Kennard E, Gangadharan R et al (2013) Topographical analysis of the femoral components of ex vivo total knee replacements. J Mater Sci Mater Med 24(2):547–554CrossRefPubMed
25.
go back to reference Steinbrück A, Schröder C, Woiczinski M et al (2013) Patellofemoral contact patterns before and after total knee arthroplasty: an in vitro measurement. Biomed Eng Online 12(58):1–13. doi:10.1186/1475-925X-12-58 CrossRef Steinbrück A, Schröder C, Woiczinski M et al (2013) Patellofemoral contact patterns before and after total knee arthroplasty: an in vitro measurement. Biomed Eng Online 12(58):1–13. doi:10.​1186/​1475-925X-12-58 CrossRef
26.
go back to reference Thomas P, Weik T, Roider G et al (2016) Influence of surface coating on metal ion release: evaluation in patients with metal allergy. Orthopedics 39(3):24–30CrossRef Thomas P, Weik T, Roider G et al (2016) Influence of surface coating on metal ion release: evaluation in patients with metal allergy. Orthopedics 39(3):24–30CrossRef
27.
go back to reference Tuerkan U, Oeztuerk O, Eroglu AE (2006) Metal ion release from TiN coated CoCrMo orthopedic implant material. Surf Coat Technol 200(16–17):5020–5027CrossRef Tuerkan U, Oeztuerk O, Eroglu AE (2006) Metal ion release from TiN coated CoCrMo orthopedic implant material. Surf Coat Technol 200(16–17):5020–5027CrossRef
28.
go back to reference van Hove RP, Sierevelt IN, van Royen BJ et al (2015) Titanium-nitride coating of orthopaedic implants: a review of the literature. Biomed Res Int. Article ID 485975. doi: 10.1155/2015/485975 van Hove RP, Sierevelt IN, van Royen BJ et al (2015) Titanium-nitride coating of orthopaedic implants: a review of the literature. Biomed Res Int. Article ID 485975. doi: 10.​1155/​2015/​485975
29.
go back to reference van Hove RP, Brohet RM, van Royen BJ et al (2015) No clinical benefit of titanium nitride coating in cementless mobile-bearing total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 23(6):1833–1840CrossRefPubMed van Hove RP, Brohet RM, van Royen BJ et al (2015) No clinical benefit of titanium nitride coating in cementless mobile-bearing total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 23(6):1833–1840CrossRefPubMed
30.
go back to reference Zietz C, Reinders J, Schwiesau J et al (2015) Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions. J Mater Sci Mater Med 26(3):134–146CrossRefPubMed Zietz C, Reinders J, Schwiesau J et al (2015) Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions. J Mater Sci Mater Med 26(3):134–146CrossRefPubMed
Metadata
Title
High wear resistance of femoral components coated with titanium nitride: a retrieval analysis
Authors
Christian Fabry
Carmen Zietz
Axel Baumann
Reinhard Ehall
Rainer Bader
Publication date
01-09-2018
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 9/2018
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-017-4578-7

Other articles of this Issue 9/2018

Knee Surgery, Sports Traumatology, Arthroscopy 9/2018 Go to the issue