Published in: Intensive Care Medicine 6/2007

01-06-2007 | Original

Plethysmographic dynamic indices predict fluid responsiveness in septic ventilated patients

Authors: Marc Feissel, Jean-Louis Teboul, Paolo Merlani, Julio Badie, Jean-Pierre Faller, Karim Bendjelid

Published in: Intensive Care Medicine | Issue 6/2007

Login to get access

Abstract

Objectives

In septic patients, reliable non-invasive predictors of fluid responsiveness are needed. We hypothesised that the respiratory changes in the amplitude of the plethysmographic pulse wave (ΔPPLET) would allow the prediction of changes in cardiac index following volume administration in mechanically ventilated septic patients.

Design

Prospective clinical investigation.

Setting

An 11-bed hospital medical intensive care unit.

Patients

Twenty-three deeply sedated septic patients mechanically ventilated with tidal volume ≥ 8 ml/kg and equipped with an arterial catheter and a pulse oximetry plethysmographic sensor.

Interventions

Respiratory changes in pulse pressure (ΔPP), ΔPPLET and cardiac index (transthoracic Doppler echocardiography) were determined before and after volume infusion of colloids (8 ml/kg).

Measurements and main results

Twenty-eight volume challenges were performed in 23 patients. Before volume expansion, ΔPP correlated with ΔPPLET (r 2 = 0.71, p < 0.001). Changes in cardiac index after volume expansion significantly (p < 0.001) correlated with baseline ΔPP (r 2 = 0.76) and ΔPPLET (r 2 = 0.50). The patients were defined as responders to fluid challenge when cardiac index increased by at least 15% after the fluid challenge. Such an event occurred 18 times. Before volume challenge, a ΔPP value of 12% and a ΔPPLET value of 14% allowed discrimination between responders and non-responders with sensitivity of 100% and 94% respectively and specificity of 70% and 80% respectively. Comparison of areas under the receiver operator characteristic curves showed that ΔPP and ΔPPLET predicted similarly fluid responsiveness.

Conclusion

The present study found ΔPPLET to be as accurate as ΔPP for predicting fluid responsiveness in mechanically ventilated septic patients.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bendjelid K, Romand JA (2003) Fluid responsiveness in mechanically ventilated patients: a review of indices used in intensive care. Intensive Care Med 29:352–360PubMedCrossRef Bendjelid K, Romand JA (2003) Fluid responsiveness in mechanically ventilated patients: a review of indices used in intensive care. Intensive Care Med 29:352–360PubMedCrossRef
2.
go back to reference Michard F, Teboul JL (2002) Predicting fluid responsiveness in ICU patients: a critical analysis of the evidence. Chest 121:2000–2008PubMedCrossRef Michard F, Teboul JL (2002) Predicting fluid responsiveness in ICU patients: a critical analysis of the evidence. Chest 121:2000–2008PubMedCrossRef
3.
go back to reference Feissel M, Michard F, Mangin I, Ruyer O, Faller JP, Teboul JL (2001) Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock. Chest 119:867–873PubMedCrossRef Feissel M, Michard F, Mangin I, Ruyer O, Faller JP, Teboul JL (2001) Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock. Chest 119:867–873PubMedCrossRef
4.
go back to reference Michard F, Boussat S, Chemla D, Anguel N, Mercat A, Lecarpentier Y, Richard C, Pinsky MR, Teboul JL (2000) Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med 162:134–138PubMed Michard F, Boussat S, Chemla D, Anguel N, Mercat A, Lecarpentier Y, Richard C, Pinsky MR, Teboul JL (2000) Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med 162:134–138PubMed
5.
go back to reference Reuter DA, Felbinger TW, Schmidt C, Kilger E, Goedje O, Lamm P, Goetz AE (2002) Stroke volume variations for assessment of cardiac responsiveness to volume loading in mechanically ventilated patients after cardiac surgery. Intensive Care Med 28:392–398PubMedCrossRef Reuter DA, Felbinger TW, Schmidt C, Kilger E, Goedje O, Lamm P, Goetz AE (2002) Stroke volume variations for assessment of cardiac responsiveness to volume loading in mechanically ventilated patients after cardiac surgery. Intensive Care Med 28:392–398PubMedCrossRef
6.
go back to reference Awad AA, Ghobashy MA, Stout RG, Silverman DG, Shelley KH (2001) How does the plethysmogram derived from the pulse oximeter relate to arterial blood pressure in coronary artery bypass graft patients? Anesth Analg 93:1466–1471PubMedCrossRef Awad AA, Ghobashy MA, Stout RG, Silverman DG, Shelley KH (2001) How does the plethysmogram derived from the pulse oximeter relate to arterial blood pressure in coronary artery bypass graft patients? Anesth Analg 93:1466–1471PubMedCrossRef
7.
go back to reference Murray WB, Foster PA (1996) The peripheral pulse wave: information overlooked. J Clin Monit 12:365–377PubMedCrossRef Murray WB, Foster PA (1996) The peripheral pulse wave: information overlooked. J Clin Monit 12:365–377PubMedCrossRef
8.
go back to reference Shelley KH, Murray WB, Chang D (1997) Arterial-pulse oximetry loops: a new method of monitoring vascular tone. J Clin Monit 13:223–228PubMedCrossRef Shelley KH, Murray WB, Chang D (1997) Arterial-pulse oximetry loops: a new method of monitoring vascular tone. J Clin Monit 13:223–228PubMedCrossRef
9.
go back to reference Cannesson M, Besnard C, Durand PG, Bohe J, Jacques D (2005) Relation between respiratory variations in pulse oximetry plethysmographic waveform amplitude and arterial pulse pressure in ventilated patients. Crit Care 9:R562–568PubMedCrossRef Cannesson M, Besnard C, Durand PG, Bohe J, Jacques D (2005) Relation between respiratory variations in pulse oximetry plethysmographic waveform amplitude and arterial pulse pressure in ventilated patients. Crit Care 9:R562–568PubMedCrossRef
10.
go back to reference Alexander CM, Teller LE, Gross JB (1989) Principles of pulse oximetry: theoretical and practical considerations. Anesth Analg 68:368–376PubMedCrossRef Alexander CM, Teller LE, Gross JB (1989) Principles of pulse oximetry: theoretical and practical considerations. Anesth Analg 68:368–376PubMedCrossRef
11.
go back to reference Awad AA, Stout RG, Ghobashy MA, Rezkanna HA, Silverman DG, Shelley KH (2006) Analysis of the ear pulse oximeter waveform. J Clin Monit Comput 20:175–184PubMedCrossRef Awad AA, Stout RG, Ghobashy MA, Rezkanna HA, Silverman DG, Shelley KH (2006) Analysis of the ear pulse oximeter waveform. J Clin Monit Comput 20:175–184PubMedCrossRef
12.
go back to reference Monnet X, Lamia B, Teboul JL (2005) Pulse oximeter as a sensor of fluid responsiveness: do we have our finger on the best solution? Crit Care 9:429–430PubMedCrossRef Monnet X, Lamia B, Teboul JL (2005) Pulse oximeter as a sensor of fluid responsiveness: do we have our finger on the best solution? Crit Care 9:429–430PubMedCrossRef
13.
go back to reference Shamir M, Eidelman LA, Floman Y, Kaplan L, Pizov R (1999) Pulse oximetry plethysmographic waveform during changes in blood volume. Br J Anaesth 82:178–181PubMed Shamir M, Eidelman LA, Floman Y, Kaplan L, Pizov R (1999) Pulse oximetry plethysmographic waveform during changes in blood volume. Br J Anaesth 82:178–181PubMed
14.
go back to reference Cannesson M, Desebbe O, Hachemi M, Jacques D, Bastien O, Lehot JJ (2007) Respiratory variations in pulse oximeter waveform amplitude are influenced by venous return in mechanically ventilated patients under general anaesthesia. Eur J Anaesthesiol 24:245–251PubMedCrossRef Cannesson M, Desebbe O, Hachemi M, Jacques D, Bastien O, Lehot JJ (2007) Respiratory variations in pulse oximeter waveform amplitude are influenced by venous return in mechanically ventilated patients under general anaesthesia. Eur J Anaesthesiol 24:245–251PubMedCrossRef
15.
go back to reference Natalini G, Rosano A, Taranto M, Faggian B, Vittorielli E, Bernardini A (2006) Arterial versus plethysmographic dynamic indices to test responsiveness for testing fluid administration in hypotensive patients: a clinical trial. Anesth Analg 103:1478–1484PubMedCrossRef Natalini G, Rosano A, Taranto M, Faggian B, Vittorielli E, Bernardini A (2006) Arterial versus plethysmographic dynamic indices to test responsiveness for testing fluid administration in hypotensive patients: a clinical trial. Anesth Analg 103:1478–1484PubMedCrossRef
16.
go back to reference De Backer D, Heenen S, Piagnerelli M, Koch M, Vincent JL (2005) Pulse pressure variations to predict fluid responsiveness: influence of tidal volume. Intensive Care Med 31:517–523PubMedCrossRef De Backer D, Heenen S, Piagnerelli M, Koch M, Vincent JL (2005) Pulse pressure variations to predict fluid responsiveness: influence of tidal volume. Intensive Care Med 31:517–523PubMedCrossRef
17.
go back to reference Levy MM, Fink MP, Marshall JC, Abraham E, Angus D, Cook D, Cohen J, Opal SM, Vincent JL, Ramsay G (2003) 2001 SCCM/ESICM/ACCP/ATS/SIS International Sepsis Definitions Conference. Intensive Care Med 29:530–538PubMed Levy MM, Fink MP, Marshall JC, Abraham E, Angus D, Cook D, Cohen J, Opal SM, Vincent JL, Ramsay G (2003) 2001 SCCM/ESICM/ACCP/ATS/SIS International Sepsis Definitions Conference. Intensive Care Med 29:530–538PubMed
18.
go back to reference Ramsay MA, Savege TM, Simpson BR, Goodwin R (1974) Controlled sedation with alphaxalone–alphadolone. Br Med J 2:656–659PubMedCrossRef Ramsay MA, Savege TM, Simpson BR, Goodwin R (1974) Controlled sedation with alphaxalone–alphadolone. Br Med J 2:656–659PubMedCrossRef
19.
go back to reference Preisman S, Kogan S, Berkenstadt H, Perel A (2005) Predicting fluid responsiveness in patients undergoing cardiac surgery: functional haemodynamic parameters including the Respiratory Systolic Variation Test and static preload indicators. Br J Anaesth 95:746–755PubMedCrossRef Preisman S, Kogan S, Berkenstadt H, Perel A (2005) Predicting fluid responsiveness in patients undergoing cardiac surgery: functional haemodynamic parameters including the Respiratory Systolic Variation Test and static preload indicators. Br J Anaesth 95:746–755PubMedCrossRef
20.
go back to reference Stetz CW, Miller RG, Kelly GE, Raffin TA (1982) Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 126:1001–1004PubMed Stetz CW, Miller RG, Kelly GE, Raffin TA (1982) Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 126:1001–1004PubMed
21.
go back to reference Hanley JA, McNeil BJ (1983) A method of comparing the areas under receiver operating characteristic curves derived from the same cases. Radiology 148:839–843PubMed Hanley JA, McNeil BJ (1983) A method of comparing the areas under receiver operating characteristic curves derived from the same cases. Radiology 148:839–843PubMed
22.
go back to reference Feissel M, Badie J, Merlani PG, Faller JP, Bendjelid K (2005) Pre-ejection period variations predict the fluid responsiveness of septic ventilated patients. Crit Care Med 33:2534–2539PubMedCrossRef Feissel M, Badie J, Merlani PG, Faller JP, Bendjelid K (2005) Pre-ejection period variations predict the fluid responsiveness of septic ventilated patients. Crit Care Med 33:2534–2539PubMedCrossRef
23.
go back to reference Vieillard-Baron A, Chergui K, Rabiller A, Peyrouset O, Page B, Beauchet A, Jardin F (2004) Superior vena caval collapsibility as a gauge of volume status in ventilated septic patients. Intensive Care Med 30:1734–1739PubMed Vieillard-Baron A, Chergui K, Rabiller A, Peyrouset O, Page B, Beauchet A, Jardin F (2004) Superior vena caval collapsibility as a gauge of volume status in ventilated septic patients. Intensive Care Med 30:1734–1739PubMed
24.
go back to reference Duperret S, Lhuillier F, Piriou V, Vivier E, Metton O, Branche P, Annat G, Bendjelid K, Viale JP (2007) Increased intra-abdominal pressure affects respiratory variations in arterial pressure in normovolaemic and hypovolaemic mechanically ventilated healthy pigs. Intensive Care Med 33:163–171PubMedCrossRef Duperret S, Lhuillier F, Piriou V, Vivier E, Metton O, Branche P, Annat G, Bendjelid K, Viale JP (2007) Increased intra-abdominal pressure affects respiratory variations in arterial pressure in normovolaemic and hypovolaemic mechanically ventilated healthy pigs. Intensive Care Med 33:163–171PubMedCrossRef
Metadata
Title
Plethysmographic dynamic indices predict fluid responsiveness in septic ventilated patients
Authors
Marc Feissel
Jean-Louis Teboul
Paolo Merlani
Julio Badie
Jean-Pierre Faller
Karim Bendjelid
Publication date
01-06-2007
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 6/2007
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0602-6

Other articles of this Issue 6/2007

Intensive Care Medicine 6/2007 Go to the issue